High failure rates of metal-on-metal hip arthroplasty implants have highlighted the need for more careful introduction and monitoring of new implants and for the evaluation of the safety of medical devices. The National Joint Registry and other regulatory services are unable to detect failing implants at an early enough stage. We aimed to identify validated surrogate markers of long-term outcome in patients undergoing primary total hip arthroplasty (THA). We conducted a systematic review of studies evaluating surrogate markers for predicting long-term outcome in primary THA. Long-term outcome was defined as revision rate of an implant at ten years according to National Institute of Health and Care Excellence guidelines. We conducted a search of Medline and Embase (OVID) databases. Separate search strategies were devised for the Cochrane database and Google Scholar. Each search was performed to include articles from the date of their inception to June 8, 2015.Objectives
Methods
Revision total hip replacement (THR) for young
patients is challenging because of technical complexity and the potential
need for subsequent further revisions. We have assessed the survivorship,
functional outcome and complications of this procedure in patients
aged <
50 years through a large longitudinal series with consistent treatment
algorithms. Of 132 consecutive patients (181 hips) who underwent
revision THR, 102 patients (151 hips) with a mean age of 43 years
(22 to 50) were reviewed at a mean follow-up of 11 years (2 to 26)
post-operatively. We attempted to restore bone stock with allograft
where indicated. Using further revision for any reason as an end point,
the survival of the acetabular component was 71% ( This overall perspective on the mid- to long-term results is
valuable when advising young patients on the prospects of revision
surgery at the time of primary replacement. Cite this article:
We report the long-term results of revision total
hip replacement using femoral impaction allografting with both uncemented
and cemented Freeman femoral components. A standard design of component
was used in both groups, with additional proximal hydroxyapatite
coating in the uncemented group. A total of 33 hips in 30 patients received
an uncemented component and 31 hips in 30 patients a cemented component.
The mean follow-up was 9.8 years (2 to 17) in the uncemented group
and 6.2 years (1 to 11) in the cemented group. Revision procedures
(for all causes) were required in four patients (four hips) in the
uncemented group and in five patients (five hips) in the cemented
group. Harris hip scores improved significantly in both groups and
were maintained independently of the extent of any migration of
the femoral component within the graft or graft–cement mantle.
We conducted a retrospective study to assess
the prevalence of adverse reactions to metal debris (ARMD) in patients
operated on at our institution with metal-on-metal (MoM) total hip
replacements with 36 mm heads using a Pinnacle acetabular shell.
A total of 326 patients (150 males, 175 hips; 176 females, 203 hips)
with a mean age of 62.7 years (28 to 85) and mean follow-up of 7.5
years (0.1 to 10.8) participating in our in-depth modern MoM follow-up
programme were included in the study, which involved recording whole
blood cobalt and chromium ion measurements, Oxford hip scores (OHS)
and plain radiographs of the hip and targeted cross-sectional imaging. Elevated
blood metal ion levels (>
5 parts per billion) were seen in 32 (16.1%)
of the 199 patients who underwent unilateral replacement. At 23
months after the start of our modern MoM follow-up programme, 29
new cases of ARMD had been revealed. Hence, the nine-year survival
of this cohort declined from 96% (95% CI 95 to 98) with the old
surveillance routine to 86% (95% CI 82 to 90) following the new
protocol. Although ARMD may not be as common in 36 mm MoM THRs as
in those with larger heads, these results support the Medicines
and Healthcare Products Regulatory Agency guidelines on regular
reviews and further investigations, and emphasise the need for specific
a follow-up programme for patients with MoM THRs. Cite this article:
Slipped capital femoral epiphysis (SCFE) may
lead to symptomatic femoroacetabular impingement (FAI). We report our
experience of arthroscopic treatment, including osteochondroplasty,
for the sequelae of SCFE. Data were prospectively collected on patients undergoing arthroscopy
of the hip for the sequelae of SCFE between March 2007 and February
2013, including demographic data, radiological assessment of the
deformity and other factors that may influence outcome, such as
the presence of established avascular necrosis. Patients completed
the modified Harris hip score (mHHS) and the non-arthritic hip score
(NAHS) before and after surgery. In total, 18 patients with a mean age of 19 years (13 to 42),
were included in the study. All patients presented with pain in
the hip and mechanical symptoms, and had evidence of FAI (cam or
mixed impingement) on plain radiographs. The patients underwent arthroscopic osteoplasty of the femoral
neck. The mean follow-up was 29 months (23 to 56). The mean mHHS and NAHS scores improved from 56.2 (27.5 to 100.1)
and 52.1 (12.5 to 97.5) pre-operatively to 75.1 (33.8 to 96.8, p
= 0.01) and 73.6 (18.8 to 100, p = 0.02) at final follow-up, respectively.
Linear regression analysis demonstrated a significant association
between poorer outcome scores and increased time to surgery following SCFE
(p <
0.05 for all parameters except baseline MHHS). Symptomatic FAI following (SCFE) may be addressed using arthroscopic
techniques, and should be treated promptly to minimise progressive
functional impairment and chondrolabral degeneration. Take home message: Arthroscopy of the hip can be used to treat
femoroacetabular impingement successfully following SCFE. However,
this should be performed promptly after presentation in order to
prevent irreversible progression and poorer clinical outcomes. Cite this article:
The clinical and radiological results of 50 consecutive acetabular reconstructions in 48 patients using impaction grafting have been retrospectively reviewed. A 1:1 mixture of frozen, ground irradiated bone graft and Apapore 60, a synthetic bone graft substitute, was used in all cases. There were 13 complex primary and 37 revision procedures with a mean follow-up of five years (3.4 to 7.6). The clinical survival rate was 100%, with improvements in the mean Harris Hip Scores for pain and function. Radiologically, 30 acetabular grafts showed evidence of incorporation, ten had radiolucent lines and two acetabular components migrated initially before stabilising. Acetabular reconstruction in both primary and revision surgery using a 1:1 mixture of frozen, ground, irriadiated bone and Apapore 60 appears to be a reliable method of managing acetabular defects. Longer follow-up will be required to establish whether this technique is as effective as using fresh-frozen allograft.
We present a series of 35 patients (19 men and
16 women) with a mean age of 64 years (36.7 to 75.9), who underwent
total hip replacement using the ESKA dual-modular short stem with
metal on-polyethylene bearing surfaces. This implant has a modular
neck section in addition to the modular head. Of these patients,
three presented with increasing post-operative pain due to pseudotumour
formation that resulted from corrosion at the modular neck-stem
junction. These patients underwent further surgery and aseptic lymphocytic
vaculitis associated lesions were demonstrated on histological analysis. Retrieval analysis of two modular necks showed corrosion at the
neck-stem taper. Blood cobalt and chromium levels were measured
at a mean of nine months (3 to 28) following surgery. These were
compared with the levels in seven control patients (three men and
four women) with a mean age of 53.4 years (32.1 to 64.1), who had
an identical prosthesis and articulation but with a prosthesis that
had no modularity at neck-stem junction. The mean blood levels of
cobalt in the study group were raised at 50.75 nmol/l (5 to 145)
compared with 5.6 nmol/l (2 to 13) in control patients. Corrosion at neck-stem tapers has been identified as an important
source of metal ion release and pseudotumour formation requiring
revision surgery. Finite element modelling of the dual modular stem
demonstrated high stresses at the modular stem-neck junction. Dual
modular cobalt-chrome hip prostheses should be used with caution
due to these concerns.
We evaluated the outcome of primary total hip replacement (THR) in 3290 patients with the primary diagnosis of osteoarthritis at a minimum follow-up of two years. They were stratified into categories of body mass index (BMI) based on the World Health Organisation classification of obesity. Statistical analysis was carried out to determine if there was a difference in the post-operative Western Ontario and McMaster Universities osteoarthritis index, the Harris hip score and the Short-Form-12 outcome based on the BMI. While the pre- and post-operative scores were lower for the group classified as morbidly obese, the overall change in outcome scores suggested an equal if not greater improvement compared with the non-morbidly obese patients. The overall survivorship and rate of complications were similar in the BMI groups although there was a slightly higher rate of revision for sepsis in the morbidly obese group. Morbid obesity does not affect the post-operative outcome after THR, with the possible exception of a marginally increased rate of infection. Therefore withholding surgery based on the BMI is not justified.
In this prospective study a total of 80 consecutive
Chinese patients with Crowe type I or II developmental dysplasia of
the hip were randomly assigned for hip resurfacing arthroplasty
(HRA) or total hip replacement (THR). Three patients assigned to HRA were converted to THR, and three
HRA patients and two THR patients were lost to follow-up. This left
a total of 34 patients (37 hips) who underwent HRA and 38 (39 hips)
who underwent THR. The mean follow-up was 59.4 months (52 to 70)
in the HRA group and 60.6 months (50 to 72) in the THR group. There was
no failure of the prosthesis in either group. Flexion of the hip
was significantly better after HRA, but there was no difference
in the mean post-operative Harris hip scores between the groups.
The mean size of the acetabular component in the HRA group was significantly
larger than in the THR group (49.5 mm vs 46.1 mm, p = 0.001). There was
no difference in the mean abduction angle of the acetabular component
between the two groups. Although the patients in this series had risk factors for failure
after HRA, such as low body weight, small femoral heads and dysplasia,
the clinical results of resurfacing in those with Crowe type I or
II hip dysplasia were satisfactory. Patients in the HRA group had
a better range of movement, although neck-cup impingement was observed.
However, more acetabular bone was sacrificed in HRA patients, and
it is unclear whether this will have an adverse effect in the long
term.
Fungal peri-prosthetic infections of the knee
and hip are rare but likely to result in devastating complications.
In this study we evaluated the results of their management using
a single-stage exchange technique. Between 2001 and 2011, 14 patients
(ten hips, four knees) were treated for a peri-prosthetic fungal
infection. One patient was excluded because revision surgery was
not possible owing to a large acetabular defect. One patient developed
a further infection two months post-operatively and was excluded
from the analysis. Two patients died of unrelated causes. After a mean of seven years (3 to 11) a total of ten patients
were available for follow-up. One patient, undergoing revision replacement
of the hip, had a post-operative dislocation. Another patient, undergoing
revision replacement of the knee, developed a wound infection and
required revision 29 months post-operatively following a peri-prosthetic femoral
fracture. The mean Harris hip score increased to 74 points (63 to 84; p
<
0.02) in those undergoing revision replacement of the hip,
and the mean Hospital for Special Surgery knee score increased to
75 points (70 to 80; p <
0.01) in those undergoing revision replacement
of the knee. A single-stage revision following fungal peri-prosthetic infection
is feasible, with an acceptable rate of a satisfactory outcome. Cite this article:
We have previously described the mid- to long-term
results of conventional simple varus intertrochanteric osteotomy
for osteonecrosis of the femoral head, showing that 19 of the 26
hips had good or excellent results. We extended the follow-up to
a mean of 18.1 years (10.5 to 26) including a total of 34 hips in
28 patients, with a mean age at surgery of 33 years (19 to 53).
There were 18 men and ten women and 25 hips (74%) had a satisfactory
result with a Harris hip score ≥ 80. In all, six hips needed total
hip replacement (THR) or hemiarthroplasty. The collapse of the femoral
head or narrowing of the joint space was found to have progressed
in nine hips (26%). Leg shortening after osteotomy was a mean of
19 mm (8 to 36). With conversion to THR or hemiarthroplasty as the
endpoint, the ten-year survival rate was 88.2% (95% confidence interval
(CI) 82.7 to 93.7) and the 20-year survival rate was 79.7% (95%
CI 72.1 to 87.3); four hips were converted at ten years and other
two hips were converted at 20 years. Shortening of the leg after osteotomy remains a concern; however,
the conventional varus half-wedge osteotomy provides favourable
long-term results in hips with less than two-thirds of the medial
part of the femoral head affected by necrotic bone and with normal
bone superolaterally.
The benefit of arthroscopy of the hip in the
treatment of femoroacetabular impingement (FAI) in terms of quality
of life (QoL) has not been reported. We prospectively collected
data on 612 patients (257 women (42%) and 355 men (58%)) with a
mean age at the time of surgery of 36.7 years (14 to 75) who underwent
arthroscopy of the hip for FAI under the care of a single surgeon.
The minimum follow-up was one year (mean 3.2 years (1 to 7)). The
responses to the modified Harris hip score were translated using
the Rosser Index Matrix in order to provide a QoL score. The mean
QoL score increased from 0.946 (-1.486 to 0.995) to 0.974 (0.7 to
1) at one year after surgery (p <
0.001). The mean QoL score
in men was significantly higher than in women, both before and one
year after surgery (both p <
0.001). However, the mean change
in the QoL score was not statistically different between men and
women (0.02 (-0.21 to 0.27) and 0.04 (-0.16 to 0.87), respectively;
p = 0.12). Linear regression analysis revealed that the significant predictors
of a change in QoL score were pre-operative QoL score (p <
0.001)
and gender (p = 0.04). The lower the pre-operative score, the higher
the gain in QoL post-operatively (ρ = -0.66; p <
0.001). One
year after surgery the QoL scores in the 612 patients had improved
in 469 (76.6%), remained unchanged in 88 (14.4%) and had deteriorated
in 55 (9.0%).
We compared the clinical and radiological outcomes
of two cementless femoral stems in the treatment of patients with
a Garden III or IV fracture of the femoral neck. A total of 70 patients At final follow-up there were no statistically significant differences
between the short anatomical and the conventional stems with regard
to the mean Harris hip score (85.7 (66 to 100) Our study demonstrated that despite the poor bone quality in
these elderly patients with a fracture of the femoral neck, osseo-integration
was obtained in all hips in both groups. However, the incidence
of thigh pain, pulmonary microemboli and peri-prosthetic fracture
was significantly higher in the conventional stem group than in
the short stem group.
Reconstructive acetabular osteotomy is a well established and effective procedure in the treatment of acetabular dysplasia. However, the dysplasia is frequently accompanied by intra-articular pathology such as labral tears. We intended to determine whether a concomitant hip arthroscopy with peri-acetabular rotational osteotomy could identify and treat intra-articular pathology associated with dysplasia and thereby produce a favourable outcome. We prospectively evaluated 43 consecutive hips treated by combined arthroscopy and acetabular osteotomy. Intra-operative arthroscopic examination revealed labral lesions in 38 hips. At a mean follow-up of 74 months (60 to 97) the mean Harris hip score improved from 72.4 to 94.0 (p < 0.001), as did all the radiological parameters (p < 0.001). Complications included penetration of the joint by the osteotome in one patient, a fracture of the posterior column in another and deep-vein thrombosis in one further patient. This combined surgical treatment gave good results in the medium term. We suggest that arthroscopy of the hip can be performed in conjunction with peri-acetabular osteotomy to provide good results in patients with symptomatic dysplasia of the hip, and the arthroscopic treatment of intra-articular pathology may alter the progression of osteoarthritis.
Pelvic discontinuity with associated bone loss is a complex challenge in acetabular revision surgery. Reconstruction using ilio-ischial cages combined with trabecular metal acetabular components and morsellised bone (the component-cage technique) is a relatively new method of treatment. We reviewed a consecutive series of 26 cases of acetabular revision reconstructions in 24 patients with pelvic discontinuity who had been treated by the component-cage technique. The mean follow-up was 44.6 months (24 to 68). Failure was defined as migration of a component of >
5 mm. In 23 hips (88.5%) there was no clinical or radiological evidence of loosening at the last follow-up. The mean Harris hip score improved significantly from 46.6 points (29.5 to 68.5) to 76.6 points (55.5 to 92.0) at two years (p <
0.001). In three hips (11.5%) the construct had migrated at one year after operation. The complications included two dislocations, one infection and one partial palsy of the peroneal nerve. Our findings indicate that treatment of pelvic discontinuity using the component-cage construct is a reliable option.
Peri-articular soft-tissue masses or ‘pseudotumours’
can occur after large-diameter metal-on-metal (MoM) resurfacing
of the hip and conventional total hip replacement (THR). Our aim
was to assess the incidence of pseudotumour formation and to identify
risk factors for their formation in a prospective cohort study. A total of 119 patients who underwent 120 MoM THRs with large-diameter
femoral heads between January 2005 and November 2007 were included
in the study. Outcome scores, serum metal ion levels, radiographs
and CT scans were obtained. Patients with symptoms or an identified
pseudotumour were offered MRI and an ultrasound-guided biopsy. There were 108 patients (109 hips) eligible for evaluation by
CT scan at a mean follow-up of 3.6 years (2.5 to 4.5); 42 patients
(39%) were diagnosed with a pseudotumour. The hips of 13 patients
(12%) were revised to a polyethylene acetabular component with small-diameter
metal head. Patients with elevated serum metal ion levels had a
four times increased risk of developing a pseudotumour. This study shows a substantially higher incidence of pseudotumour
formation and subsequent revisions in patients with MoM THRs than
previously reported. Because most revision cases were identified
only after an intensive screening protocol, we recommend close monitoring
of patients with MoM THR.
Revision after failed femoral components may
be technically demanding due to loss of peri-prosthetic bone. This retrospective
study evaluated the long-term results of femoral revision using
the cementless Wagner Self-Locking stem. Between 1992 and 1998,
68 consecutive hips in 66 patients underwent femoral revision using
this implant. A total of 25 patients died from unrelated causes
without further revision; the remaining 41 hips in 41 patients (12 men
and 29 women) with a mean age of 61 years (29 to 80) were reviewed
at a mean follow-up of 13.9 years (10.4 to 15.8). A transfemoral
approach was used in 32 hips. A total of five stems required further
revision because of infection in two, progressive subsidence in
two and recurrent dislocation in one. Four hips had dislocated and
eight stems had subsided ≥ 10 mm. The mean Harris hip score improved
from 33 points pre-operatively to 75 points at final follow-up (p
<
0.001). In all, 33 stems (91.7%) showed radiological signs
of stable bone fixation. The cumulative survival rates at 15.8 years
with femoral revision for any reason and for stem failure as the
endpoints were 92.0% (95% confidence interval (CI) 86.0% to 98.4%)
and 96.6% (95% CI 92.2% to 100%), respectively. The survivorship
with revision and ≥ 10 mm migration of the stem as the endpoint
was 83.6% (95% CI 76.6% to 91.4%). This study shows quite good survival and moderate clinical outcome
when using a monoblock tapered titanium stem for supporting the
regeneration of bone in complex revision hip surgery.
We investigated factors that were thought to be associated with an increased incidence of squeaking of ceramic-on-ceramic total hip replacements. Between June 1997 and December 2008 the three senior authors implanted 2406 primary total hip replacements with a ceramic-on-ceramic bearing surface. The mean follow-up was 10.6 years. The diagnosis was primary osteoarthritis in each case, and no patient had undergone previous surgery to the hip. We identified 74 squeaking hips (73 patients) giving an incidence of 3.1% at a mean follow-up of 9.5 years (4.1 to 13.3). Taller, heavier and younger patients were significantly more likely to have hips that squeaked. Squeaking hips had a significantly higher range of post-operative internal (p = 0.001) and external rotation (p = 0.003) compared with silent hips. Patients with squeaking hips had significantly higher activity levels (p = 0.009). A squeaking hip was not associated with a significant difference in patient satisfaction (p = 0.24) or Harris hip score (p = 0.34). Four implant position factors enabled good prediction of squeaking. These were high acetabular component inclination, high femoral offset, lateralisation of the hip centre and either high or low acetabular component anteversion. This is the largest study to date to examine patient factors and implant position factors that predispose to squeaking of a ceramic-on-ceramic hip. The results suggest that factors which increase the mechanical forces across the hip joint and factors which increase the risk of neck-to-rim impingement, and therefore edge-loading, are those that predispose to squeaking.
Femoral impaction bone allografting has been developed as a means of restoring bone stock in revision total hip replacement. We report the results of 75 consecutive patients (75 hips) with a mean age of 68 years (35 to 87) who underwent impaction grafting using the Exeter collarless, polished, tapered femoral stem between 1992 and 1998. The mean follow-up period was 10.5 years (6.3 to 14.1). The median pre-operative bone defect score was 3 (interquartile range (IQR) 2 to 3) using the Endo-Klinik classification. The median subsidence at one year post-operatively was 2 mm (IQR 1 to 3). At the final review the median Harris hip score was 80.6 (IQR 67.6 to 88.9) and the median subsidence 2 mm (IQR 1 to 4). Incorporation of the allograft into trabecular bone and secondary remodelling were noted radiologically at the final follow-up in 87% (393 of 452 zones) and 40% (181 of 452 zones), respectively. Subsidence of the Exeter stem correlated with the pre-operative Endo-Klinik bone loss score (p = 0.037). The degree of subsidence at one year had a strong association with long-term subsidence (p <
0.001). There was a significant correlation between previous revision surgery and a poor Harris Hip score (p = 0.028), and those who had undergone previous revision surgery for infection had a higher risk of complications (p = 0.048). Survivorship at 10.5 years with any further femoral operation as the end-point was 92% (95% confidence interval 82 to 97).
Implantation of a large-diameter femoral head prosthesis with a metal-on-metal bearing surface reduces the risk of dislocation, increases the range of movement, minimises the risk of impingement and, in theory, results in little wear. Between February 2004 and March 2007 we implanted 100 consecutive total hip replacements with a metal-on-metal bearing and a large femoral head into 92 patients. There were 51 men and 41 women with a mean age of 50 years (18 to 70) at the time of surgery. Outcome was assessed using the Western Ontario McMaster University osteoarthritis index and the Harris hip score as well as the Devane activity score. These all improved significantly (p <
0.0001). At the last follow-up there were no cases of dislocation, no impingement, a good range of movement and no osteolysis, but seven revisions, two for infection and five for aseptic loosening. The probability of groin pain increased if the other acetabular component inclination exceeded 50° (p = 0.0007). At 4.8 years of follow-up, the projected survival of the Durom acetabular component, with revision for any reason, was 92.4% ( The design of the component made it difficult both to orientate and seat, which when combined with a poor porous coating, produced unpredictable fixation and a low survival at five years.