The October 2013 Shoulder &
Elbow Roundup360 looks at: Deltoid impairment not necessarily a contra-indication for shoulder arthroplasty; The tricky radiograph; Not so asymptomatic cuff tears; Total shoulder arthroplasty: kinder on the glenoid; Barbotage for calcific tendonitis; What happens to the arthritic glenoid?; Two screws a screw too few?; Sloppy hinge best for elbow arthroplasty.
Patients with transfemoral amputation (TFA) often
experience problems related to the use of socket-suspended prostheses.
The clinical development of osseointegrated percutaneous prostheses
for patients with a TFA started in 1990, based on the long-term
successful results of osseointegrated dental implants. Between1999 and 2007, 51 patients with 55 TFAs were consecutively
enrolled in a prospective, single-centre non-randomised study and
followed for two years. The indication for amputation was trauma
in 33 patients (65%) and tumour in 12 (24%). A two-stage surgical
procedure was used to introduce a percutaneous implant to which
an external amputation prosthesis was attached. The assessment of
outcome included the use of two self-report questionnaires, the
Questionnaire for Persons with a Transfemoral Amputation (Q-TFA)
and the Short-Form (SF)-36. The cumulative survival at two years’ follow-up was 92%. The
Q-TFA showed improved prosthetic use, mobility, global situation
and fewer problems (all p <
0.001). The physical function SF-36
scores were also improved (p <
0.001). Superficial infection
was the most frequent complication, occurring 41 times in 28 patients
(rate of infection 54.9%). Most were treated effectively with oral
antibiotics. The implant was removed in four patients because of loosening
(three aseptic, one infection). Osseointegrated percutaneous implants constitute a novel form
of treatment for patients with TFA. The high cumulative survival
rate at two years (92%) combined with enhanced prosthetic use and
mobility, fewer problems and improved quality of life, supports
the ‘revolutionary change’ that patients with TFA have reported
following treatment with osseointegrated percutaneous prostheses. Cite this article:
In developmental dysplasia of the hip (DDH),
a bone defect is often observed superior to the acetabulum after
the reconstruction at the level of the true acetabulum during total
hip replacement (THR). However, the essential amount of uncemented
acetabular component coverage required for a satisfactory outcome
remains controversial. The purpose of this study was to assess the
stability and function of acetabular components with a lack of coverage >
30% (31% to 50%). A total of 760 DDH patients underwent THR with
acetabular reconstruction at the level of the true floor. Lack of
coverage above the acetabular component of >
30% occurred in 56
patients. Intra-operatively, autogenous morcellised bone grafts
were used to fill the uncovered portion. Other than two screws inserted through
the acetabular shell, no additional structural supports were used
in these hips. In all, four patients were lost to follow-up. Therefore,
52 patients (52 hips, 41 women and 11 men) with a mean age of 60.1
years (42 to 78) were available for this study at a mean of 4.8
years (3 to 7). There were no instances of prosthesis revision or
marked loosening during the follow-up. The Harris hip score improved
from a mean of 40.7 points ( Cite this article:
In a prospective study, 93 unselected consecutive uncemented hip arthroplasties were performed in 80 patients using the titanium-coated RM acetabular component and the CLS femoral component. The mean age of the patients at operation was 52 years (28 to 81). None were lost to follow-up. In the 23 patients who had died (26 hips) only one acetabular component had been revised. In the 57 living patients (67 hips), 13 such revisions had been performed. Of the 14 revisions, seven were for osteolysis, five for loosening and two for infection. Survival analysis of this implant showed a total probability of survival of 83% (95% confidence interval 73 to 90), with all revisions as the endpoint, and a probability of 94% (95% confidence interval 87 to 98) with revision for aseptic loosening as the endpoint, indicating reliable long-term fixation of the titanium-coated RM acetabular component.
We have managed 27 patients (16 women and 11 men) with a mean age of 68.4 years (50 to 84), with failed total hip replacement and severe proximal femoral bone loss by revision using a distal fix/proximal wrap prosthesis. The mean follow-up was for 55.3 months (25 to 126). The mean number of previous operations was 2.2 (1 to 4). The mean Oxford hip score decreased from 46.2 (38 to 60) to 28.5 (17 to 42) (paired t-test, p <
0.001) and the mean Harris Hip score increased from 30.4 (3 to 57.7) to 71.7 (44 to 99.7) (paired t-test, p <
0.001). There were two dislocations, and in three patients we failed to eradicate previous infection. None required revision of the femoral stem. This technique allows instant distal fixation while promoting biological integration and restoration of bone stock. In the short term, the functional outcome is encouraging and the complication rates acceptable in this difficult group of patients.
We implanted titanium and carbon fibre-reinforced plastic (CFRP) femoral prostheses of the same dimensions into five prosthetic femora. An abductor jig was attached and a 1 kN load applied. This was repeated with five control femora. Digital image correlation was used to give a detailed two-dimensional strain map of the medial cortex of the proximal femur. Both implants caused stress shielding around the calcar. Distally, the titanium implant showed stress shielding, whereas the CFRP prosthesis did not produce a strain pattern which was statistically different from the controls. There was a reduction in strain beyond the tip of both the implants. This investigation indicates that use of the CFRP stem should avoid stress shielding in total hip replacement.
The outcome at ten years of 100 Freeman hip stems (Finsbury Orthopaedics, Leatherhead, United Kingdom) retaining the neck with a proximal hydroxyapatite coating in a series of 52 men (six bilateral) and 40 women (two bilateral), has been described previously. None required revision for aseptic loosening. We have extended the follow-up to 20 years with a minimum of 17 years. The mean age of the patients at total hip replacement was 58.9 years (19 to 84). Six patients were lost to follow-up, but were included up to their last clinical review. A total of 22 patients (22 hips) had died, all from causes unrelated to their surgery. There have been 43 re-operations for failure of the acetabular component. However, in 38 of these the stem was not revised since it remained stable and there was no associated osteolysis. Two of the revisions were for damage to the trunnion after fracture of a modular ceramic head, and in another two, removal of the femoral component was because of the preference of the surgeon. In all cases the femoral component was well fixed, but could be extracted at the time of acetabular revision. In one case both components were revised for deep infection. There has been one case of aseptic loosening of the stem which occurred at 14 years. This stem had migrated distally by 7.6 mm in ten years and 8.4 mm at the time of revision at which stage it was found to be rotationally loose. With hindsight this component had been undersized at implantation. The survivorship for the stem at 17 years with aseptic loosening as the endpoint was 98.6% (95% confidence interval 95.9 to 100) when 62 hips were at risk. All remaining stems had a satisfactory clinical and radiological outcome. The Freeman proximally hydroxyapatite-coated femoral component is therefore a dependable implant and its continued use can be recommended.
The December 2012 Research Roundup360 looks at: whether the rheumatoid factor is just a ‘quick test’; osteonecrosis in smokers; pasteurisation effect on bone reconstruction; venous thromboembolism risk in rheumatoids; whether stem cells reverse age-related osteopenia; the effect of running on rat knees; rapid fracture healing in rats with ultrasound; magnetic stem cells; and the safety of surgery.
In this study we present our experience with
four generations of uncemented total knee arthroplasty (TKA) from Smith
&
Nephew: Tricon M, Tricon LS, Tricon II and Profix, focusing
on the failure rates correlating with each design change. Beginning
in 1984, 380 Tricon M, 435 Tricon LS, 305 Tricon 2 and 588 Profix
were implanted by the senior author. The rate of revision for loosening
was 1.1% for the Tricon M, 1.1% for the Tricon LS, 0.5% for the
Tricon 2 with a HA coated tibial component, and 1.3% for the Profix
TKA. No loosening of the femoral component was seen with the Tricon
M, Tricon LS or Tricon 2, with no loosening seen of the tibial component
with the Profix TKA. Regarding revision for wear, the incidence
was 13.1% for the Tricon M, 6.6% for the Tricon LS, 2.3% for the
Tricon 2, and 0% for the Profix. These results demonstrate that
improvements in the design of uncemented components, including increased
polyethylene thickness, improved polyethylene quality, and the introduction
of hydroxyapatite coating, has improved the outcomes of uncemented
TKA over time.
Peri-prosthetic bone loss caused by stress shielding may be associated with aseptic loosening of femoral components. In order to increase primary stability and to reduce stress shielding, a three-dimensional, cementless individual femoral (Evolution K) component was manufactured using pre-operative CT scans. Using dual energy x-ray absorptiometry, peri-prosthetic bone density was measured in 43 patients, three months, six months, 3.6 and 4.6 years after surgery. At final follow-up there was a significant reduction in mean bone density in the proximal Gruen zones of −30.3% (zone 7) and −22.8% (zone 1). The density in the other zones declined by a mean of between −4% and −16%. We conclude that the manufacture of a three-dimensional, custom-made femoral component could not prevent a reduction in peri-prosthetic bone density.
Eighteen hip fusions were converted to total
hip replacements. A constrained acetabular liner was used in three hips.
Mean follow up was five years (two to 15). Two (11%) hips failed,
requiring revision surgery and two patients (11%) had injury to
the peroneal nerve. Heterotopic ossification developed in seven
(39%) hips, in one case resulting in joint ankylosis. No hips dislocated. Conversion of hip fusion to hip replacement carries an increased
risk of heterotopic ossification and neurological injury. We advise
prophylaxis against heterotropic ossification. When there is concern
about hip stability we suggest that the use of a constrained acetabular
liner is considered. Despite the potential for complications, this procedure
had a high success rate and was effective in restoring hip function.
A clinical and radiological study was conducted on 97 total hip replacements performed for congenital hip dislocation in 79 patients between 1989 and 1998 using a three-dimensional custom-made cementless stem. The mean age at operation was 48 years (17 to 72) and the mean follow-up was for 123 months (83 to 182). According to the Crowe classification, there were 37 class I, 28 class II, 13 class III and 19 class IV hips. The mean leg lengthening was 25 mm (5 to 58), the mean pre-operative femoral anteversion was 38.6° (2° to 86°) and the mean correction in the prosthetic neck was −23.6° (−71° to 13°). The mean Harris hip score improved from 58 (15 to 84) to 93 (40 to 100) points. A revision was required in six hips (6.2%). The overall survival rate was 89.5% (95% confidence interval 89.2 to 89.8) at 13 years when two hips were at risk. This custom-made cementless femoral component, which can be accommodated in the abnormal proximal femur and will correct the anteversion and frontal offset, provided good results without recourse to proximal femoral corrective osteotomy.
The in-cement technique for revision hip arthroplasty involves retaining the original cement-bone interface. This has been proven to be a biomechanically stronger method than recementing after complete removal of the original cement mantle. This study reviewed a series of 54 consecutive revision hip arthroplasty procedures, using the in-cement technique, between November 1999 and November 2003. Clinical and radiological follow-up included functional assessment. There were 54 procedures performed in 51 patients, whose mean age at surgery was 70.3 years (45 to 85). A total of 42 were available at a mean follow-up of 29.2 months (6 to 51). There was no radiological evidence of loosening. Functional assessments were available for 40 patients who had a mean Harris hip score of 85.2 (51.9 to 98.5), a mean Oxford hip score of 19.6 (12 to 41), a mean UCLA activity profile score of 5.9 (3 to 8) and a mean SF-36 score of 78.0 (31.6 to 100). The in-cement technique provides consistent, high functional outcomes and should be considered in appropriately selected cases.
Systemic mastocytosis is a rare condition that often involves the bone marrow. We report the case of a patient with systemic mastocytosis who underwent total hip replacement. Technical difficulties encountered during the procedure included a narrow medullary canal and abnormally hard bone, later confirmed by laboratory measurements. Follow-up at five years showed a good clinical and radiological outcome.
This study reviewed the results of a cementless anatomical femoral component to give immediate post-operative stability, and with a narrow distal section in order not to contact the femoral cortex in the diaphysis, ensuring exclusively metaphyseal loading. A total of 471 patients (601 hips) who had a total hip replacement between March 1995 and February 2002 were included in the study. There were 297 men and 174 women. The mean age at the time of operation was 52.7 years (28 to 63). Clinical and radiological evaluation were performed at each follow-up. Bone densitometry was carried out on all patients two weeks after operation and at the final follow-up examination. The mean follow-up was 8.8 years (5 to 12). The mean pre-operative Harris hip score was 41 points (16 to 54), which improved to a mean of 96 (68 to 100) at the final follow-up. No patient complained of thigh pain at any stage. No acetabular or femoral osteolysis was observed and no hip required revision for aseptic loosening of either component. Deep infection occurred in two hips (0.3%) which required revision. One hip (0.2%) required revision of the acetabular component for recurrent dislocation. Bone mineral densitometry revealed a minimal bone loss in the proximal femur. This cementless anatomical femoral component with metaphyseal loading but without distal fixation produced satisfactory fixation and encourages proximal femoral loading.
An experimental sheep model was used for impaction allografting of 12 hemiarthroplasty femoral components placed into two equal-sized groups. In group 1, a 50:50 mixture of ApaPore hydroxyapatite bone-graft substitute and allograft was used. In group 2, ApaPore and allograft were mixed in a 90:10 ratio. Both groups were killed at six months. Ground reaction force results demonstrated no significant differences (p >
0.05) between the two groups at 8, 16 and 24 weeks post-operatively, and all animals remained active. The mean bone turnover rates were significantly greater in group 1, at 0.00206 mm/day, compared to group 2 at 0.0013 mm/day (p <
0.05). The results for the area of new bone formation demonstrated no significant differences (p >
0.05) between the two groups. No significant differences were found between the two groups in thickness of the cement mantle (p >
0.05) and percentage ApaPore-bone contact (p >
0.05). The results of this animal study demonstrated that a mixture of ApaPore allograft in a 90:10 ratio was comparable to using a 50:50 mixture.
We investigated whether improvements in design have altered the outcome for patients undergoing endoprosthetic replacement of the distal femur after resection of a tumour. Survival of the implant and ‘servicing’ procedures have been documented using a prospective database, review of the design of the implant and case records. In total, 335 patients underwent a distal femoral replacement, 162 having a fixed-hinge design and 173 a rotating-hinge. The median age of the patients was 24 years (interquartile range 17 to 48). A total of 192 patients remained alive with a mean follow-up of 12 years (5 to 30). The risk of revision for any reason was 17% at five years, 33% at ten years and 58% at 20 years. Aseptic loosening was the main reason for revision of the fixed-hinge knees while infection and fracture of the stem were the most common for the rotating-hinge implant. The risk of revision for aseptic loosening was 35% at ten years with the fixed-hinge knee, which has, however, been replaced by the rotating-hinge knee with a hydroxyapatite collar. The overall risk of revision for any reason fell by 52% when the rotating-hinge implant was used. Improvements in the design of distal femoral endoprostheses have significantly decreased the need for revision operations, but infection remains a serious problem. We believe that a cemented, rotating-hinge prosthesis with a hydroxyapatite collar offers the best chance of long-term survival of the prosthesis.
We prospectively reviewed 1000 consecutive patients who underwent a cementless, hydroxyapatite-coated, stemless, total knee replacement over a period of nine years. Regular post-operative clinical follow-up was performed using the Knee Society score. The mean pre-operative score was 96, improving to 182 and 180 at five and ten years, respectively. To date, there have been seven (0.5%) cases which required revision, primarily for septic loosening (four cases), with low rates of other post-operative complications. The cumulative survival at ten years with revision as the end-point, was 99.14% (95% confidence interval 92.5 to 99.8). These results support the use of hydroxyapatite in a cementless total knee replacement since it can give reliable fixation with an excellent clinical and functional outcome.
The outcome of tibial allograft reconstruction after resection of a tumour is inconsistent and has a high rate of failure. There are few reports on the use of tibial allografts in children with open growth plates. We performed 21 allograft reconstructions (16 osteoarticular, five intercalary) in 19 consecutive patients between seven and 17 years of age. Two had Ewing’s sarcoma, one an adamantinoma and 16 osteosarcoma, one with multifocal disease. Five patients have died; the other 14 were free from disease at the time of follow-up. Six surviving patients (eight allograft reconstructions) continue to have good or excellent function at a mean of 59 months (14 to 132). One patient has poor function at 31 months. The other seven patients have a good or excellent function after additional procedures including exchange of the allograft and resurfacing or revision to an endoprosthesis at a mean of 101 months (43 to 198). The additional operations were performed at a mean of 47 months (20 to 84) after the first reconstruction. With the use of allograft reconstruction in growing children, joints and growth plates may be preserved, at least partially. Although our results remain inconsistent, tibial allograft reconstruction in selected patients may restore complete and durable function of the limb.
Bioengineering reasons for increased wear and failure of metal-on-metal (MoM) bearings in hip prostheses have been described. Low wear occurs in MoM hips when the centre of the femoral head is concentric with the centre of the acetabular component and the implants are correctly positioned. Translational or rotational malpositioning of the components can lead to the contact-patch of the femoral component being displaced to the rim of the acetabular component, resulting in a ten- to 100-fold increase in wear and metal ion levels. This may cause adverse tissue reactions, loosening of components and failure of the prosthesis.