Wear of polyethylene inserts plays an important role in failure
of total knee replacement and can be monitored Before revision, the minimum joint space width values and their
locations on the insert were measured in 15 fully weight-bearing
radiographs. These measurements were compared with the actual minimum
thickness values and locations of the retrieved tibial inserts after
revision. Introduction
Method
We are entering a new era with governmental bodies
taking an increasingly guiding role, gaining control of registries,
demanding direct access with release of open public information
for quality comparisons between hospitals. This review is written
by physicians and scientists who have worked with the Swedish Knee
Arthroplasty Register (SKAR) periodically since it began. It reviews
the history of the register and describes the methods used and lessons
learned. Cite this article:
The LockDown device (previously called Surgilig)
is a braided polyester mesh which is mostly used to reconstruct the
dislocated acromioclavicular joint. More than 11 000 have been implanted
worldwide. Little is known about the tissue reaction to the device
nor to its wear products when implanted in an extra-articular site
in humans. This is of importance as an adverse immunological reaction
could result in osteolysis or damage to the local tissues, thereby affecting
the longevity of the implant. We analysed the histology of five LockDown implants retrieved
from five patients over the last seven years by one of the senior
authors. Routine analysis was carried out in all five cases and
immunohistochemistry in one. The LockDown device acts as a scaffold for connective tissue
which forms an investing fibrous pseudoligament. The immunological
response at the histological level seems favourable with a limited
histiocytic and giant cell response to micron-sized wear particles.
The connective tissue envelope around the implant is less organised
than a native ligament. Cite this article:
Fresh-frozen allograft bone is frequently used
in orthopaedic surgery. We investigated the incidence of allograft-related
infection and analysed the outcomes of recipients of bacterial culture-positive
allografts from our single-institute bone bank during bone transplantation.
The fresh-frozen allografts were harvested in a strict sterile environment
during total joint arthroplasty surgery and immediately stored in
a freezer at -78º to -68º C after packing. Between January 2007
and December 2012, 2024 patients received 2083 allografts with a
minimum of 12 months of follow-up. The overall allograft-associated
infection rate was 1.2% (24/2024). Swab cultures of 2083 allografts
taken before implantation revealed 21 (1.0%) positive findings.
The 21 recipients were given various antibiotics at the individual
orthopaedic surgeon’s discretion. At the latest follow-up, none
of these 21 recipients displayed clinical signs of infection following
treatment. Based on these findings, we conclude that an incidental positive
culture finding for allografts does not correlate with subsequent
surgical site infection. Additional prolonged post-operative antibiotic
therapy may not be necessary for recipients of fresh-frozen bone
allograft with positive culture findings. Cite this article:
As the number of younger and more active patients
treated with total knee arthroplasty (TKA) continues to increase,
consideration of better fixation as a means of improving implant
longevity is required. Cemented TKA remains the reference standard
with the largest body of evidence and the longest follow-up to support
its use. However, cementless TKA, may offer the opportunity of a
more bone-sparing procedure with long lasting biological fixation
to the bone. We undertook a review of the literature examining advances
of cementless TKA and the reported results. Cite this article:
Modular junctions are ubiquitous in contemporary hip arthroplasty. The head-trunnion junction is implicated in the failure of large diameter metal-on-metal (MoM) hips which are the currently the topic of one the largest legal actions in the history of orthopaedics (estimated costs are stated to exceed $4 billion). Several factors are known to influence the strength of these press-fit modular connections. However, the influence of different head sizes has not previously been investigated. The aim of the study was to establish whether the choice of head size influences the initial strength of the trunnion-head connection. Ti-6Al-4V trunnions (n = 60) and two different sizes of cobalt-chromium (Co-Cr) heads (28 mm and 36 mm; 30 of each size) were used in the study. Three different levels of assembly force were considered: 4 kN; 5 kN; and 6 kN (n = 10 each). The strength of the press-fit connection was subsequently evaluated by measuring the pull-off force required to break the connection. The statistical differences in pull-off force were examined using a Kruskal–Wallis test and two-sample Mann–Whitney U test. Finite element and analytical models were developed to understand the reasons for the experimentally observed differences.Objectives
Materials and Methods
Following the recall of modular neck hip stems
in July 2012, research into femoral modularity will intensify over
the next few years. This review aims to provide surgeons with an
up-to-date summary of the clinically relevant evidence. The development
of femoral modularity, and a classification system, is described.
The theoretical rationale for modularity is summarised and the clinical
outcomes are explored. The review also examines the clinically relevant problems
reported following the use of femoral stems with a modular neck. Joint replacement registries in the United Kingdom and Australia
have provided data on the failure rates of modular devices but cannot
identify the mechanism of failure. This information is needed to
determine whether modular neck femoral stems will be used in the
future, and how we should monitor patients who already have them implanted. Cite this article:
We wished to investigate the influence of metal debris exposure
on the subsequent immune response and resulting soft-tissue injury
following metal-on-metal (MoM) hip arthroplasty. Some reports have
suggested that debris generated from the head-neck taper junction
is more destructive than equivalent doses from metal bearing surfaces. We investigated the influence of the source and volume of metal
debris on chromium (Cr) and cobalt (Co) concentrations in corresponding
blood and hip synovial fluid samples and the observed agglomerated
particle sizes in excised tissues using multiple regression analysis
of prospectively collected data. A total of 199 explanted MoM hips
(177 patients; 132 hips female) were analysed to determine rates
of volumetric wear at the bearing surfaces and taper junctions. Aims
Patients and Methods
The August 2013 Hip &
Pelvis Roundup360 looks at: are we getting it right first time?; tantalum augments in revision hip surgery; lower wear in dual mobility?; changing faces changes outcomes; synovial fluid aspiration in MOM hips; taper disease: the new epidemic of hip surgery; the super-obese and THR; and whether well fixed stems can remain in infected hips
Since the Oxford knee was first used unicompartmentally
in 1982, a small number of bearings have fractured. Of 14 retrieved
bearings, we examined ten samples with known durations
Pseudotumours (abnormal peri-prosthetic soft-tissue reactions)
following metal-on-metal hip resurfacing arthroplasty (MoMHRA) have
been associated with elevated metal ion levels, suggesting that
excessive wear may occur due to edge-loading of these MoM implants.
This study aimed to quantify The duration and magnitude of edge-loading Objectives
Methods
Initial stability of tibial trays is crucial for long-term success of total knee arthroplasty (TKA) in both primary and revision settings. Rotating platform (RP) designs reduce torque transfer at the tibiofemoral interface. We asked if this reduced torque transfer in RP designs resulted in subsequently reduced micromotion at the cemented fixation interface between the prosthesis component and the adjacent bone. Composite tibias were implanted with fixed and RP primary and revision tibial trays and biomechanically tested under up to 2.5 kN of axial compression and 10° of external femoral component rotation. Relative micromotion between the implanted tibial tray and the neighbouring bone was quantified using high-precision digital image correlation techniques.Objectives
Methods
This study reports the mid-term results of a large-bearing hybrid metal-on-metal total hip replacement in 199 hips (185 patients) with a mean follow-up of 62 months (32 to 83). Two patients died of unrelated causes and 13 were lost to follow-up. In all, 17 hips (8.5%) have undergone revision, and a further 14 are awaiting surgery. All revisions were symptomatic. Of the revision cases, 14 hips showed evidence of adverse reactions to metal debris. The patients revised or awaiting revision had significantly higher whole blood cobalt ion levels (p = 0.001), but no significant difference in acetabular component size or position compared with the unrevised patients. Wear analysis (n = 5) showed increased wear at the trunnion-head interface, normal levels of wear at the articulating surfaces and evidence of corrosion on the surface of the stem. The cumulative survival rate, with revision for any reason, was 92.4% (95% confidence interval 87.4 to 95.4) at five years. Including those awaiting surgery, the revision rate would be 15.1% with a cumulative survival at five years of 89.6% (95% confidence interval 83.9 to 93.4). This hybrid metal-on-metal total hip replacement series has shown an unacceptably high rate of failure, with evidence of high wear at the trunnion-head interface and passive corrosion of the stem surface. This raises concerns about the use of large heads on conventional 12/14 tapers.
This study was undertaken to evaluate the safety and efficacy of retrievable inferior vena cava filters in high-risk orthopaedic patients. A total of 58 patients had a retrievable inferior vena cava filter placed as an adjunct to chemical and mechanical prophylaxis, most commonly for a history of previous deep-vein thrombosis or pulmonary embolism, polytrauma, or expected prolonged immobilisation. In total 56 patients (96.6%) had an uncomplicated post-operative course. Two patients (3.4%) died in the peri-operative period for unrelated reasons. Of the 56 surviving patients, 50 (89%) were available for follow-up. A total of 32 filters (64%) were removed without complication at a mean of 37.8 days (4 to 238) after placement. There were four filters (8%) which were retained because of thrombosis at the filter site, and four (8%) were retained because of incorporation of the filter into the wall of the inferior vena cava. In ten cases (20%) the retrievable filter was left in place to continue as primary prophylaxis. No patient had post-removal thromboembolic complications. A retrievable inferior vena cava filter, as an adjunct to chemical and mechanical prophylaxis, was a safe and effective means of reducing the acute risk of pulmonary embolism in this high-risk group of patients. Although most filters were removed without complications, thereby avoiding the long-term complications that have plagued permanent indwelling filters, a relatively high percentage of filters had to be left
Oxidised zirconium was introduced as a material for femoral components
in total knee arthroplasty (TKA) as an attempt to reduce polyethylene
wear. However, the long-term survival of this component is not known. We performed a retrospective review of a prospectively collected
database to assess the ten year survival and clinical and radiological
outcomes of an oxidised zirconium total knee arthroplasty with the
Genesis II prosthesis. The Western Ontario and McMaster Universities Osteoarthritis
Index (WOMAC), Knee Injury and Osteoarthritis Outcome Score (KOOS)
and a patient satisfaction scale were used to assess outcome.Aims
Methods
The October 2015 Hip &
Pelvis Roundup360 looks at: Smoking and complications in arthroplasty; Smoking cessation beneficial in arthroplasty; Intermediate care and arthroplasty; Do we still need cell salvage?; Femoroacetabular impingement in the Japanese population; Trunnionosis or taperosis and geometry; Decontamination for staphylococcus aureus works!; Policeman or opportunity? Quality improvement with registries; Death rates higher in readmission to other hospitals
We report the five-year outcome of a randomised
controlled trial which used radiostereometric analysis (RSA) to assess
the influence of surface oxidised zirconium (OxZr, Oxinium) on polyethylene
wear A total of 120 patients, 85 women and 35 men with a mean age
of 70 years (59 to 80) who were scheduled for primary cemented total
hip arthroplasty were randomly allocated to four study groups. Patients
were blinded to their group assignment and received either a conventional
polyethylene (CPE) or a highly cross-linked (HXL) acetabular component
of identical design. On the femoral side patients received a 28
mm head made of either cobalt-chromium (CoCr) or OxZr. The proximal head penetration (wear) was measured with repeated
RSA examinations over five years. Clinical outcome was measured
using the Harris hip score. There was no difference in polyethylene wear between the two
head materials when used with either of the two types of acetabular
component (p = 0.3 to 0.6). When comparing the two types of polyethylene
there was a significant difference in favour of HXLPE, regardless
of the head material used (p <
0.001). In conclusion, we found no advantage of OxZr over CoCr in terms
of polyethylene wear after five years of follow-up. Our findings
do not support laboratory results which have shown a reduced rate
of wear with OxZr. They do however add to the evidence on the better
resistance to wear of HXLPE over CPE. Cite this article:
Noise generation has been reported with ceramic-on-ceramic
articulations in total hip replacement (THR). This study evaluated
208 consecutive Delta Motion THRs at a mean follow-up of 21 months
(12 to 35). There were 141 women and 67 men with a mean age of 59
years (22 to 84). Patients were reviewed clinically and radiologically,
and the incidence of noise was determined using a newly described
assessment method. Noise production was examined against range of
movement, ligamentous laxity, patient-reported outcome scores, activity
level and orientation of the acetabular component. There were 143
silent hips (69%), 22 (11%) with noises other than squeaking, 17
(8%) with unreproducible squeaking and 26 (13%) with reproducible
squeaking. Hips with reproducible squeaking had a greater mean range
of movement (p <
0.001) and mean ligament laxity (p = 0.004), smaller
median head size (p = 0.01) and decreased mean acetabular component
inclination (p = 0.02) and anteversion angle (p = 0.02) compared
with the other groups. There was no relationship between squeaking
and age (p = 0.13), height (p = 0.263), weight (p = 0.333), body
mass index (p = 0.643), gender (p = 0.07) or patient outcome score
(p = 0.422). There were no revisions during follow-up. Despite the
surprisingly high incidence of squeaking, all patients remain satisfied
with their hip replacement. Cite this article: