We report on the outcome of the Exeter Contemporary flanged cemented
all-polyethylene acetabular component with a mean follow-up of 12
years (10 to 13.9). This study reviewed 203 hips in 194 patients.
129 hips in 122 patients are still A retrospective review was undertaken of a consecutive series
of 203 routine primary cemented total hip arthroplasties (THA) in
194 patients.Aims
Patients and Methods
The highly cross-linked polyethylene Exeter RimFit flanged cemented
acetabular component was introduced in the United Kingdom in 2010.
This study aimed to examine the rates of emergence of radiolucent
lines observed when the Rimfit acetabular component was implanted
at total hip arthroplasty (THA) using two different techniques: firstly,
the ‘rimcutter’ technique in which the flange sits on a pre-prepared
acetabular rim; and secondly, the ‘trimmed flange’ technique in
which the flange is trimmed and the acetabular component is seated
inside the rim of the acetabulum. The radiographs of 150 THAs (75 ‘rimcutter’, 75 ‘trimmed flange’)
involving this component were evaluated to assess for radiolucencies
at the cement/bone interface by three observers. Aims
Patients and Methods
Initial stability of tibial trays is crucial for long-term success of total knee arthroplasty (TKA) in both primary and revision settings. Rotating platform (RP) designs reduce torque transfer at the tibiofemoral interface. We asked if this reduced torque transfer in RP designs resulted in subsequently reduced micromotion at the cemented fixation interface between the prosthesis component and the adjacent bone. Composite tibias were implanted with fixed and RP primary and revision tibial trays and biomechanically tested under up to 2.5 kN of axial compression and 10° of external femoral component rotation. Relative micromotion between the implanted tibial tray and the neighbouring bone was quantified using high-precision digital image correlation techniques.Objectives
Methods
The aim of this study was to evaluate whether
coating titanium discs with selenium in the form of sodium selenite decreased
bacterial adhesion of In order to evaluate bacterial adhesion, sterile titanium discs
were coated with increasing concentrations of selenium and incubated
with bacterial solutions of The tested Selenium coating is a promising method to reduce bacterial attachment
on prosthetic material. Cite this article:
Sagittal alignment of the lumbosacral spine, and specifically pelvic incidence (PI), has been implicated in the development of spine pathology, but generally ignored with regards to diseases of the hip. We aimed to determine if increased PI is correlated with higher rates of hip osteoarthritis (HOA). The effect of PI on the development of knee osteoarthritis (KOA) was used as a negative control. We studied 400 well-preserved cadaveric skeletons ranging from 50 to 79 years of age at death. Each specimen’s OA of the hip and knee were graded using a previously described method. PI was measured from standardised lateral photographs of reconstructed pelvises. Multiple regression analysis was performed to determine the relationship between age and PI with HOA and KOA.Objectives
Methods
The aim of this study was to report a single surgeon series of
consecutive patients with moderate hallux valgus managed with a
percutaneous extra-articular reverse-L chevron (PERC) osteotomy. A total of 38 patients underwent 45 procedures. There were 35
women and three men. The mean age of the patients was 48 years (17
to 69). An additional percutaneous Akin osteotomy was performed
in 37 feet and percutaneous lateral capsular release was performed
in 22 feet. Clinical and radiological assessments included the type
of forefoot, range of movement, the American Orthopedic Foot and
Ankle (AOFAS) score, a subjective rating and radiological parameters. The mean follow-up was 59.1 months (45.9 to 75.2). No patients
were lost to follow-up.Aims
Patients and Methods
The primary aim of this study was to investigate the effect of
an enhanced recovery program (ERP) on the short-term functional
outcome after total hip arthroplasty (THA). Secondary outcomes included
its effect on rates of dislocation and mortality. Data were gathered on 1161 patients undergoing primary THA which
included 611 patients treated with traditional rehabilitation and
550 treated with an ERP. Aims
Patients and Methods
Revision of a cemented hemiarthroplasty of the
hip may be a hazardous procedure with high rates of intra-operative complications.
Removing well-fixed cement is time consuming and risks damaging
already weak bone or perforating the femoral shaft. The cement-in-cement
method avoids removal of intact cement and has shown good results
when used for revision total hip arthroplasty (THA). The use of
this technique for the revision of a hemiarthroplasty to THA has
not been previously reported. A total of 28 consecutive hemiarthroplasties (in 28 patients)
were revised to a THA using an Exeter stem and the cement-in-cement
technique. There were four men and 24 women; their mean age was
80 years (35 to 93). Clinical and radiographic data, as well as
operative notes, were collected prospectively and no patient was
lost to follow-up. Four patients died within two years of surgery. The mean follow
up of the remainder was 70 months (25 to 124). Intra-operatively
there was one proximal perforation, one crack of the
femoral calcar and one acetabular fracture. No femoral components
have required subsequent revision for aseptic loosening or are radiologically loose. Four patients with late complications (14%) have since undergone
surgery (two for a peri-prosthetic fracture, and one each for deep
infection and recurrent dislocation) resulting in an overall major
rate of complication of 35.7%. The cement-in-cement technique provides reliable femoral fixation
in this elderly population and may reduce operating time and rates
of complication. Cite this article:
The October 2015 Trauma Roundup360 looks at: PCA not the best in resuscitation; Impact of trauma centre care; Quality of life after a hip fracture; Recovery and severity of injury: open tibial fractures in the spotlight; Assessment of the triplane fractures; Signs of an unstable paediatric pelvis; Safe insertion of SI screws: are two views required?; Post-operative delirium under the spotlight; Psychological effects of fractures; K-wires cost effective in DRAFFT
Between 2002 and 2011, 81 patients with a traumatic total brachial
plexus injury underwent reconstruction by double free muscle transfer
(DFMT, 47 cases), single muscle transfer (SMT, 16 cases) or nerve
transfers (NT, 18 cases). They were evaluated for functional outcome and quality of life
(QoL) using the Disability of Arm, Shoulder and Hand questionnaire,
both pre- and post-operatively. The three groups were compared and
followed-up for at least 24 months.Aims
Methods
Orthopaedic surgeons use a variety of instruments to help correct, treat, and heal bone disease. The development of these instruments mirrors the history of orthopaedic surgery. The history of bonesetting, the treatment and replacement of joints, and of those who performed these techniques, appears to originate deep in antiquity. Changing ideas within medicine and surgery over the last 200 years have shaped the discovery and evolution of orthopaedic instruments and of the bonesetters themselves. Advances have led to the use of computers as instruments in the navigational guidance of arthroplasty surgery, the use of robotics, the development of cordless drills and improvements in the design of blades to cut bone. Yet some of the old instruments remain; plaster of Paris bandages, the Thomas Splint, Liston’s bonecutter, Gigli’s saw, bone nibblers and Macewan’s osteotomes are still in use. This paper presents a historical review of bonesetters and examines how orthopaedic instruments have evolved from antiquity to the 21st century.
This study compared the primary stability of two commercially
available acetabular components from the same manufacturer, which
differ only in geometry; a hemispherical and a peripherally enhanced
design (peripheral self-locking (PSL)). The objective was to determine
whether altered geometry resulted in better primary stability. Acetabular components were seated with 0.8 mm to 2 mm interference
fits in reamed polyethylene bone substrate of two different densities
(0.22 g/cm3 and 0.45 g/cm3). The primary stability
of each component design was investigated by measuring the peak
failure load during uniaxial pull-out and tangential lever-out tests.Objective
Methods
The early failure and revision of bimodular primary
total hip arthroplasty prostheses requires the identification of the
risk factors for material loss and wear at the taper junctions through
taper wear analysis. Deviations in taper geometries between revised
and pristine modular neck tapers were determined using high resolution
tactile measurements. A new algorithm was developed and validated
to allow the quantitative analysis of material loss, complementing
the standard visual inspection currently used. The algorithm was applied to a sample of 27 retrievals ( Cite this article:
Despite many claims of good wear properties following
total knee replacement (TKR) with an oxidised zirconium (OxZr) femoral
component, there are conflicting clinical results. We hypothesised
that there would be no difference in either the mid-term clinical
and radiological outcomes or the characteristics of the polyethylene
wear particles (weight, size and shape) in patients using an OxZr
or cobalt-chrome (CoCr) femoral component. In all 331 patients underwent
bilateral TKR, receiving an OxZr femoral component in one knee and
a CoCr femoral component in the other. The mean follow-up was 7.5
years (6 to 8). Following aspiration, polyethylene wear particles
were analysed using thermogravimetric methods and scanning electron
microscopy. At the most recent follow-up, the mean Knee Society
score, Western Ontario and McMaster Universities Osteoarthritis
Index score, range of movement and satisfaction score were not significantly
different in the two groups. The mean weight, size, aspect ratio
and roundness of the aspirated wear particles were similar for each
femoral component. Survivorship of the femoral, tibial and patellar
components was 100% in both groups. In the absence of evidence of an advantage in the medium term
we cannot justify the additional expense of an OxZr femoral component.
The purpose of this study was to evaluate the
expression of acid-sensing ion channels (ASICs) in the capsule and synovial
fluid of patients with frozen shoulder. Capsular tissue and synovial
fluid were obtained from 18 patients with idiopathic frozen shoulder
(FS group) and 18 patients with instability of the shoulder (control
group). The expressions of ASIC1, ASIC2, and ASIC3 in the capsule
were determined using the reverse transcriptase-polymerase chain
reaction, immunoblot analysis, and immunohistochemistry (IHC). The
concentrations in synovial fluid were evaluated using an enzyme-linked
immunosorbent assay. The mRNA expression of ASIC1, ASIC2 and ASIC3 in the capsule
were significantly increased in the FS group compared with the control
group. The protein levels of these three ASICs were also increased.
The increased expressions were confirmed by IHC. Of the ASICs, ASIC3
showed the greatest increase in both mRNA and levels of expression
compared with the control group. The levels of ASIC1 and ASIC3 in
synovial fluid were significantly increased in the FS group. This study suggests that ASICs may play a role as mediators of
inflammatory pain and be involved in the pathogenesis of frozen
shoulder. Cite this article:
This multicentre study analysed 12 alumina ceramic-on-ceramic
components retrieved from squeaking total hip replacements after
a mean of 23 months
A common situation presenting to the orthopaedic
surgeon today is a worn acetabular liner with substantial acetabular
and pelvic osteolysis. The surgeon has many options for dealing
with osteolytic defects. These include allograft, calcium based
substitutes, demineralised bone matrix, or combinations of these
options with or without addition of platelet rich plasma. To date
there are no clinical studies to determine the efficacy of using
bone-stimulating materials in osteolytic defects at the time of
revision surgery and there are surprisingly few studies demonstrating
the clinical efficacy of these treatment options. Even when radiographs
appear to demonstrate incorporation of graft material CT studies
have shown that incorporation is incomplete. The surgeon, in choosing
a graft material for a surgical procedure must take into account
the efficacy, safety, cost and convenience of that material. Cite this article:
The aim of this study was to assess the role
of synovial C-reactive protein (CRP) in the diagnosis of chronic periprosthetic
hip infection. We prospectively collected synovial fluid from 89
patients undergoing revision hip arthroplasty and measured synovial
CRP, serum CRP, erythrocyte sedimentation rate (ESR), synovial white
blood cell (WBC) count and synovial percentages of polymorphonuclear
neutrophils (PMN). Patients were classified as septic or aseptic
by means of clinical, microbiological, serum and synovial fluid
findings. The high viscosity of the synovial fluid precluded the
analyses in nine patients permitting the results in 80 patients
to be studied. There was a significant difference in synovial CRP
levels between the septic (n = 21) and the aseptic (n = 59) cohort.
According to the receiver operating characteristic curve, a synovial
CRP threshold of 2.5 mg/l had a sensitivity of 95.5% and specificity
of 93.3%. The area under the curve was 0.96. Compared with serum
CRP and ESR, synovial CRP showed a high diagnostic value. According
to these preliminary results, synovial CRP may be a useful parameter
in diagnosing chronic periprosthetic hip infection. Cite this article:
Graft-tunnel mismatch of the bone-patellar tendon-bone
(BPTB) graft is a major concern during anatomical anterior cruciate
ligament (ACL) reconstruction if the femoral tunnel is positioned
using a far medial portal technique, as the femoral tunnel tends
to be shorter compared with that positioned using a transtibial
portal technique. This study describes an accurate method of calculating
the ideal length of bone plugs of a BPTB graft required to avoid
graft–tunnel mismatch during anatomical ACL reconstruction using
a far medial portal technique of femoral tunnel positioning. Based on data obtained intra-operatively from 60 anatomical ACL
reconstruction procedures, we calculated the length of bone plugs
required in the BPTB graft to avoid graft–tunnel mismatch. When
this was prevented in all the 60 cases, we found that the mean length
of femoral bone plug that remained in contact with the interference
screw within the femoral tunnel was 14 mm (12 to 22) and the mean
length of tibial bone plug that remained in contact with the interference
screw within the tibial tunnel was 23 mm (18 to 28). These results
were used to validate theoretical formulae developed to predict
the required length of bone plugs in BPTB graft during anatomical
ACL reconstruction using a far medial portal technique. Cite this article:
The aim of this study was to assess the effect
of frictional torque and bending moment on fretting corrosion at
the taper interface of a modular femoral component and to investigate
whether different combinations of material also had an effect. The
combinations we examined were 1) cobalt–chromium (CoCr) heads on
CoCr stems 2) CoCr heads on titanium alloy (Ti) stems and 3) ceramic
heads on CoCr stems. In test 1 increasing torque was imposed by offsetting the stem
in the anteroposterior plane in increments of 0 mm, 4 mm, 6 mm and
8 mm when the torque generated was equivalent to 0 Nm, 9 Nm, 14
Nm and 18 Nm. In test 2 we investigated the effect of increasing the bending
moment by offsetting the application of axial load from the midline
in the mediolateral plane. Increments of offset equivalent to head
+ 0 mm, head + 7 mm and head + 14 mm were used. Significantly higher currents and amplitudes were seen with increasing
torque for all combinations of material. However, Ti stems showed
the highest corrosion currents. Increased bending moments associated
with using larger offset heads produced more corrosion: Ti stems
generally performed worse than CoCr stems. Using ceramic heads did
not prevent corrosion, but reduced it significantly in all loading
configurations. Cite this article: