This is a 15-year follow-up observational study of 4390 patients with 4606 primary total knee replacements (TKRs) implanted in the Trent health region between 1990 and 1992. The operations were performed in 21 hospitals, including both district general and teaching hospitals, with 77 different surgeons as named consultant. The main objective was to analyse the survival of the patients and of the prostheses, and to evaluate what impact different variables have on survival. In addition, the 1480 patients (33.7%) (1556 TKRs) alive at 15 years following operation were sent a self-administered questionnaire which examined their level of satisfaction, of pain, and their quality of life at 15 years. Completed responses were received from 912 TKRs (58.6%). Three survival curves were constructed: a best-case scenario based on the patients entered into the life tables, another included failures not reported in the revision database, and a third worst-case scenario based on all patients lost to follow-up presumed to have had a failed primary TKR. In the best-case scenario survival at 15 years was 92.2%, and in the worst-case scenario was 81.1%. Survival was significantly increased in women and older patients (Mantel-Cox log-rank test, p <
0.005 and p <
0.001, respectively). Revision as a result of infection was required in 40 TKRs (18.8%) representing 0.87% of the original cohort. The limited information available from the questionnaire indicated that satisfaction was less frequent among men, patients with osteoarthritis and those who required revision (chi-squared test, p <
0.05, p <
0.05 and p <
0.0001, respectively). With regard to pain, older patients, females and patients who still had their primary replacement in place at 15 years, reported the least pain (chi-squared test for trends, p <
0.0005, p <
0.005 and p <
0.0001, respectively). The reported quality of life was not affected by any variable.
The Oxford medial unicompartmental knee replacement was designed to reproduce normal mobility and forces in the knee, but its detailed effect on the patellofemoral joint has not been studied previously. We have examined the effect on patellofemoral mechanics of the knee by simultaneously measuring patellofemoral kinematics and forces in 11 cadaver knee specimens in a supine leg-extension rig. Comparison was made between the intact normal knee and sequential unicompartmental and total knee replacement. Following medial mobile-bearing unicompartmental replacement in 11 knees, patellofemoral kinematics and forces did not change significantly from those in the intact knee across any measured parameter. In contrast, following posterior cruciate ligament retaining total knee replacement in eight knees, there were significant changes in patellofemoral movement and forces. The Oxford device appears to produce near-normal patellofemoral mechanics, which may partly explain the low incidence of complications with the extensor mechanism associated with clinical use.
We prospectively reviewed 1000 consecutive patients who underwent a cementless, hydroxyapatite-coated, stemless, total knee replacement over a period of nine years. Regular post-operative clinical follow-up was performed using the Knee Society score. The mean pre-operative score was 96, improving to 182 and 180 at five and ten years, respectively. To date, there have been seven (0.5%) cases which required revision, primarily for septic loosening (four cases), with low rates of other post-operative complications. The cumulative survival at ten years with revision as the end-point, was 99.14% (95% confidence interval 92.5 to 99.8). These results support the use of hydroxyapatite in a cementless total knee replacement since it can give reliable fixation with an excellent clinical and functional outcome.
We have carried out a radiostereometric study of 50 patients (54 knees) with osteoarthritis of the knee who were randomly allocated to receive a cemented or a hydroxyapatite-coated femoral component for total knee replacement. The patients were also stratified to receive one of three types of articulating surface (standard, rotating platform, Freeman-Samuelson (FS)1000) all based on the Freeman-Samuelson design. The tibial components were cemented in all cases. Radiostereometry was performed post-operatively and at 3, 12 and 24 months. The analysis was restricted to rotation of the femoral component over time. After two years, rotation of the femoral components in the transverse, longitudinal and sagittal planes did not differ between the cemented and the hydroxyapatite-coated implants (p = 0.2 to 0.9). In total knee replacements with a rotating platform, the femoral component tended to tilt more posteriorly than in the other two designs, regardless of the choice of fixation (cemented or hydroxyapatite-coated, p = 0.04). The standard version of the femoral component, whether cemented or hydroxyapatite-coated, rotated more into valgus than was observed with the rotating-platform and FS1000 designs (p = 0.005). The increased constraint provided by the FS1000 component did not appear to have any adverse effect on fixation of the femoral component.
Bilateral sequential total knee replacement was carried out under one anaesthetic in 100 patients. One knee was replaced using a CT-free computer-assisted navigation system and the other conventionally without navigation. The two methods were compared for accuracy of orientation and alignment of the components. There were 85 women and 15 men with a mean age of 67.6 years (54 to 83). Radiological and CT imaging was carried out to determine the alignment of the components. The mean follow-up was 2.3 years (2 to 3). The operating and tourniquet times were significantly longer in the navigation group (p <
0.001). There were no significant pre- or post-operative differences between the knee scores of the two groups (p = 0.288 and p = 0.429, respectively). The results of imaging and the number of outliers for all radiological parameters were not statistically different (p = 0.109 to p = 0.920). In this series computer-assisted navigated total knee replacement did not result in more accurate orientation and alignment of the components than that achieved by conventional total knee replacement.
We studied the influence of soft-tissue releases and soft-tissue balance on the outcome of 526 total knee replacements one year after operation. The surgery had been performed by seven surgeons in five centres in the United Kingdom between October 1999 and December 2002. Balancing was carried out by five surgeons using spacers and trials and by two surgeons using a ‘balancer’ instrument. All the surgeons assessed the adequacy of their releases by taking measurements with the balancer after soft-tissue release before implanting the components. Independent observers collected the Oxford knee scores and applied the American Knee Society functional and knee scores as well as recording the range of movement of the replaced knee. These were compared with the pre-operative scores and the extent of the releases. We found differences in outcomes between minimal and extensive releases and between balanced and imbalanced knees. Knees requiring extensive soft-tissue releases showed greater change in the short-term clinical outcome without increased complications and achieved similar results at one year compared with those with less deformity pre-operatively which had required less soft-tissue release. Balancing an imbalanced knee improved the short-term knee outcome.
Using the Trent regional arthroplasty register, we analysed the survival at ten years of 1198 consecutive Charnley total hip replacements carried out across a single health region of the United Kingdom in 1990. At ten years, information regarding outcome was available for 1001 hips (83.6%). The crude revision rate was 6.2% (62 of 1001) and the cumulative survival rate with revision of the components as an end-point was 93.1%. At five years, a review of this series of patients identified gross radiological failure in 25 total hip replacements which had previously been unrecognised. At ten years the outcome was known for 18 of these 25 patients (72%), of whom 13 had not undergone revision. This is the first study to assess the survival at ten years for the primary Charnley total hip replacement performed in a broad cross-section of hospitals in the United Kingdom, as opposed to specialist centres. Our results highlight the importance of the arthroplasty register in identifying the long-term outcome of hip prostheses.
We carried out a prospective study of 118 hydroxyapatite-coated, cementless total knee replacements in patients who were ≤ 55 years of age and who had primary (92; 78%) or post-traumatic (26; 22%) osteoarthritis. The mean period of follow-up was 7.9 years (5 to 12.5). The Knee Society clinical scores improved from a pre-operative mean of 98 (0 to 137) to a mean of 185 (135 to 200) at five years, and 173 (137 to 200) at ten years. There were two revisions of the tibial component because of aseptic loosening, and one case of polyethylene wear requiring further surgery. There was no osteolysis or progressive radiological loosening of any other component. At 12 years, the overall rate of implant survival was 97.5% (excluding exchange of spacer) and 92.1% (including exchange of spacer). Cementless total knee replacement can achieve excellent long-term results in young, active patients with osteoarthritis. In contrast to total hip replacement, polyethylene wear, osteolysis and loosening of the prosthesis were not major problems for these patients, although it is possible that this observation could change with longer periods of follow-up.
Iontophoresis is a novel technique which may be used to facilitate the movement of antibiotics into the substance of bone using an electrical potential applied externally. We have examined the rate of early infection in allografts following application of this technique in clinical practice. A total of 31 patients undergoing revision arthroplasty or surgery for limb salvage received 34 iontophoresed sequential allografts, of which 26 survived for a minimum of two years. The mean serum antibiotic levels after operation were low (gentamicin 0.37 mg/l (0.2 to 0.5); flucloxacillin 1 mg/l (0 to 1) and the levels in the drains were high (gentamicin 40 mg/l (2.5 to 131); flucloxacillin 17 mg/l (1 to 43). There were no early deep infections. Two late infections were presumed to be haemotogenous; 28 of the 34 allografts were retained. In 12 patients with pre-existing proven infection further infection has not occurred at a mean follow-up of 51 months (24 to 82).
Clinical experience of impaction bone grafting for revision knee arthroplasty is limited, with initial stability of the tibial tray emerging as a major concern. The length of the stem and its diameter have been altered to improve stability. Our aim was to investigate the effect of the type of stem, support of the rim and graft impaction on early stability of the tray. We developed a system for impaction grafting of trays which we used with morsellised bone in artificial tibiae. Trays with short, long thick or long thin stems were implanted, with or without support of the rim. They were cyclically loaded while measuring relative movement. Long-stemmed trays migrated 4.5 times less than short-stemmed trays, regardless of diameter. Those with support migrated 2.8 times less than those without. The migration of short-stemmed trays correlated inversely with the density of the impacted groups. That of impaction-grafted tibial trays was in the range reported for uncemented primary trays. Movements of short-stemmed trays without cortical support were largest and sensitive to the degree of compaction of the graft. If support of the rim was sufficient or a long stem was used, impacted morsellised bone graft achieved adequate initial stability.
Although the importance of sound statistical principles in the design and analysis of data has gained prominence in recent years, biostatistics, the application of statistics to the analysis of biological and medical data, is still a subject which is poorly understood and often mishandled. This review introduces, in the context of orthopaedic research, the terminology and the principles involved in simple data analysis, and outlines areas of medical statistics that have gained prominence in recent years. It also lists and provides an insight into some of the more common errors that occur in published orthopaedic journals and which are frequently encountered at the review stage in papers submitted to the