Following the recall of modular neck hip stems
in July 2012, research into femoral modularity will intensify over
the next few years. This review aims to provide surgeons with an
up-to-date summary of the clinically relevant evidence. The development
of femoral modularity, and a classification system, is described.
The theoretical rationale for modularity is summarised and the clinical
outcomes are explored. The review also examines the clinically relevant problems
reported following the use of femoral stems with a modular neck. Joint replacement registries in the United Kingdom and Australia
have provided data on the failure rates of modular devices but cannot
identify the mechanism of failure. This information is needed to
determine whether modular neck femoral stems will be used in the
future, and how we should monitor patients who already have them implanted. Cite this article:
We report the clinical and radiological outcome at ten years of 104 primary total hip replacements (100 patients) using the Metasul metal-on-metal bearing. Of these, 52 had a cemented Stuehmer-Weber polyethylene acetabular component with a Metasul bearing and 52 had an uncemented Allofit acetabular component with a Metasul liner. A total of 15 patients (16 hips) died before their follow-up at ten years and three were lost to follow-up. The study group therefore comprised 82 patients (85 hips). The mean Oxford score at ten years was 20.7 (12 to 42). Six of 85 hips required revision surgery. One was performed because of infection, one for aseptic loosening of the acetabular component and four because of unexplained pain. Histological examination showed an aseptic lymphocytic vasculitis associated lesion-type tissue response in two of these. Continued follow-up is advocated in order to monitor the long-term performance of the Metasul bearing and tissue responses to metal debris.
We report the results of Vulpius transverse gastrocsoleus
recession for equinus gait in 26 children with cerebral palsy (CP),
using the Gait Profile Score (GPS), Gait Variable Scores (GVS) and
movement analysis profile. All children had an equinus deformity
on physical examination and equinus gait on three-dimensional gait
analysis prior to surgery. The pre-operative and post-operative
GPS and GVS were statistically analysed. There were 20 boys and
6 girls in the study cohort with a mean age at surgery of 9.2 years
(5.1 to 17.7) and 11.5 years (7.3 to 20.8) at follow-up. Of the
26 children, 14 had spastic diplegia and 12 spastic hemiplegia.
Gait function improved for the cohort, confirmed by a decrease in
mean GPS from 13.4° pre-operatively to 9.0° final review (p <
0.001). The change was 2.8 times the minimal clinically important
difference (MCID). Thus the improvements in gait were both clinically and
statistically significant. The transverse gastrocsoleus recession
described by Vulpius is an effective procedure for equinus gait
in selected children with CP, when there is a fixed contracture
of the gastrocnemius and soleus muscles. Cite this article:
Most proximal humeral fractures are stable injuries of the ageing population, and can be successfully treated non-operatively. The management of the smaller number of more complex displaced fractures is more controversial and new fixation techniques have greatly increased the range of fractures that may benefit from surgery. This article explores current concepts in the classification and clinical aspects of these injuries, reviewing the indications, innovations and outcomes for the most common methods of treatment.
The deformity index is a new radiological measurement of the degree of deformity of the femoral head in unilateral Perthes’ disease. Its values represent a continuous outcome measure of deformity incorporating changes in femoral epiphyseal height and width compared with the unaffected side. The sphericity of the femoral head in 30 radiographs (ten normal and 20 from patients with Perthes’ disease) were rated blindly as normal, mild, moderate or severe by three observers. Further blinded measurements of the deformity index were made on two further occasions with intervals of one month. There was good agreement between the deformity index score and the subjective grading of deformity. Intra- and interobserver agreement for the deformity index was high. The intraobserver intraclass correlation coefficient for each observer was 0.98, 0.99 and 0.97, respectively, while the interobserver intraclass correlation coefficient was 0.98 for the first and 0.97 for the second set of calculations. We also reviewed retrospectively 96 radiographs of children with Perthes’ disease, who were part of a multicentre trial which followed them to skeletal maturity. We found that the deformity index at two years correlated well with the Stulberg grading at skeletal maturity. A deformity index value above 0.3 was associated with the development of an aspherical femoral head. Using a deformity index value of 0.3 to divide groups for risk gives a sensitivity of 80% and specificity of 81% for predicting a Stulberg grade of III or IV. We conclude that the deformity index at two years is a valid and reliable radiological outcome measure in unilateral Perthes’ disease.
Cancellous allograft bone chips are commonly
used in the reconstruction of defects in bone after removal of benign tumours.
We investigated the MRI features of grafted bone chips and their
change over time, and compared them with those with recurrent tumour.
We retrospectively reviewed 66 post-operative MRIs from 34 patients
who had undergone curettage and grafting with cancellous bone chips
to fill the defect after excision of a tumour. All grafts showed
consistent features at least six months after grafting: homogeneous
intermediate or low signal intensities with or without scattered
hyperintense foci (speckled hyperintensities) on T1 images; high
signal intensities with scattered hypointense foci (speckled hypointensities)
on T2 images, and peripheral rim enhancement with or without central
heterogeneous enhancements on enhanced images. Incorporation of
the graft occurred from the periphery to the centre, and was completed
within three years. Recurrent lesions consistently showed the same signal
intensities as those of pre-operative MRIs of the primary lesions.
There were four misdiagnoses, three of which were chondroid tumours. We identified typical MRI features and clarified the incorporation
process of grafted cancellous allograft bone chips. The most important
characteristics of recurrent tumours were that they showed the same
signal intensities as the primary tumours. It might sometimes be
difficult to differentiate grafted cancellous allograft bone chips
from a recurrent chondroid tumour. Cite this article:
We describe the clinical and radiological results
of cementless primary total hip replacement (THR) in 25 patients
(18 women and seven men; 30 THRs) with severe developmental dysplasia
of the hip (DDH). Their mean age at surgery was 47 years (23 to
89). In all, 21 hips had Crowe type III dysplasia and nine had Crowe
type IV. Cementless acetabular components with standard polyethylene
liners were introduced as close to the level of the true acetabulum
as possible. The modular cementless S-ROM femoral component was
used with a low resection of the femoral neck. A total of 21 patients (25 THRs) were available for review at
a mean follow-up of 18.7 years (15.8 to 21.8). The mean modified
Harris hip score improved from 46 points pre-operatively to 90 at
final follow up (p <
0.001). A total of 15 patients (17 THRs; 57%) underwent revision of the
acetabular component at a mean of 14.6 years (7 to 20.8), all for
osteolysis. Two patients (two THRs) had symptomatic loosening. No
patient underwent femoral revision. Survival with revision of either
component for any indication was 81% at 15 years (95% CI 60.1 to
92.3), with 21 patients at risk. This technique may reduce the need for femoral osteotomy in severe
DDH, while providing a good long-term functional result. Cite this article:
Between June 1991 and January 1995, 42 hydroxyapatite-coated CAD-CAM femoral components were inserted in 25 patients with inflammatory polyarthropathy, 21 of whom had juvenile idiopathic arthritis. Their mean age was 21 years (11 to 35). All the patients were reviewed clinically and radiologically at one, three and five years. At the final review at a mean of 11.2 years (8 to 13) 37 hips in 23 patients were available for assessment. A total of four femoral components (9.5%) had failed, of which two were radiologically loose and two were revised. The four failed components were in patients aged 16 years or less at the time of surgery. Hydroxyapatite-coated customised femoral components give excellent medium- to long-term results in skeletally-mature young adults with inflammatory polyarthropathy. Patients aged less than 16 years at the time of surgery have a risk of 28.5% of failure of the femoral component at approximately ten years.
In 2004 we described the ten-year prospective results of 38 total hip replacements using the Furlong hydroxyapatite-ceramic-coated femoral component in 35 patients <
50 years old. We have now reviewed the surviving 35 arthroplasties in 33 patients at a mean of 16 years (10.3 to 19.9). The mean age of the surviving patients at the time of operation was 41.3 years (26.0 to 49.0). Of these, eight have undergone revision of their acetabular component for aseptic loosening. None of the femoral components has had revision for aseptic loosening giving a survival rate of 100% at 16 years (95% confidence interval 89% to 100%). The Furlong hydroxyapatite-ceramic-coated femoral component gives excellent long-term survival in young and active patients.
It is accepted that resurfacing hip replacement
preserves the bone mineral density (BMD) of the femur better than total
hip replacement (THR). However, no studies have investigated any
possible difference on the acetabular side. Between April 2007 and March 2009, 39 patients were randomised
into two groups to receive either a resurfacing or a THR and were
followed for two years. One patient’s resurfacing subsequently failed,
leaving 19 patients in each group. Resurfaced replacements maintained proximal femoral BMD and,
compared with THR, had an increased bone mineral density in Gruen
zones 2, 3, 6, and particularly zone 7, with a gain of 7.5% (95%
confidence interval (CI) 2.6 to 12.5) compared with a loss of 14.6%
(95% CI 7.6 to 21.6). Resurfacing replacements maintained the BMD
of the medial femoral neck and increased that in the lateral zones
between 12.8% (95% CI 4.3 to 21.4) and 25.9% (95% CI 7.1 to 44.6). On the acetabular side, BMD was similar in every zone at each
point in time. The mean BMD of all acetabular regions in the resurfaced
group was reduced to 96.2% (95% CI 93.7 to 98.6) and for the total
hip replacement group to 97.6% (95% CI 93.7 to 101.5) (p = 0.4863).
A mean total loss of 3.7% (95% CI 1.0 to 6.5) and 4.9% (95% CI 0.8
to 9.0) of BMD was found above the acetabular component in W1 and
10.2% (95% CI 0.9 to 19.4) and 9.1% (95% CI 3.8 to 14.4) medial
to the implant in W2 for resurfaced replacements and THRs respectively.
Resurfacing resulted in a mean loss of BMD of 6.7% (95% CI 0.7 to
12.7) in W3 but the BMD inferior to the acetabular component was
maintained in both groups. These results suggest that the ability of a resurfacing hip replacement
to preserve BMD only applies to the femoral side.
Different calcaneal plates with locked screws were compared in an experimental model of a calcaneal fracture. Four plate models were tested, three with uniaxially-locked screws (Synthes, Newdeal, Darco), and one with polyaxially-locked screws (90° ± 15°) (Rimbus). Synthetic calcanei were osteotomised to create a fracture model and then fixed with the plates and screws. Seven specimens for each plate model were subjected to cyclic loading (preload 20 N, 1000 cycles at 800 N, 0.75 mm/s), and load to failure (0.75 mm/s). During cyclic loading, the plate with polyaxially-locked screws (Rimbus) showed significantly lower displacement in the primary loading direction than the plates with uniaxially-locked screws (mean values of maximum displacement during cyclic loading: Rimbus, 3.13 mm ( The increased stability of a plate with polyaxially-locked screws demonstrated during cyclic loading compared with plates with uniaxially-locked screws may be beneficial for clinical use.
We reviewed 142 consecutive primary total hip replacements implanted into 123 patients between 1988 and 1993 using the Exeter Universal femoral stem. A total of 74 patients (88 hips) had survived for ten years or more and were reviewed at a mean of 12.7 years (10 to 17). There was no loss to follow-up. The rate of revision of the femoral component for aseptic loosening and osteolysis was 1.1% (1 stem), that for revision for any cause was 2.2% (2 stems), and for re-operation for any cause was 21.6% (19 hips). Re-operation was because of failure of the acetabular component in all but two hips. All but one femoral component subsided within the cement mantle to a mean of 1.52 mm (0 to 8.3) at the final follow-up. One further stem had subsided excessively (8 mm) and had lucent lines at the cement-stem and cement-bone interfaces. This was classified as a radiological failure and is awaiting revision. One stem was revised for deep infection and one for excessive peri-articular osteolysis. Defects of the cement mantle (Barrack grade C and D) were found in 28% of stems (25 hips), associated with increased subsidence (p = 0.01), but were not associated with endosteal lysis or failure. Peri-articular osteolysis was significantly related to the degree of polyethylene wear (p <
0.001), which was in turn associated with a younger age (p = 0.01) and male gender (p <
0.001). The use of the Exeter metal-backed acetabular component was a notable failure with 12 of 32 hips (37.5%) revised for loosening. The Harris-Galante components failed with excessive wear, osteolysis and dislocation with 15% revised (5 of 33 hips). Only one of 23 hips with a cemented Elite component (4%) was revised for loosening and osteolysis. Our findings show that the Exeter Universal stem implanted outside the originating centre has excellent medium-term results.
We describe the results at five years of a prospective study of a new tri-tapered polished, cannulated, cemented femoral stem implanted in 51 patients (54 hips) with osteoarthritis. The mean age and body mass index of the patients was 74 years and 27.9, respectively. Using the anterolateral approach, half of the stems were implanted by a consultant orthopaedic surgeon and half by six different registrars. There were three withdrawals from the study because of psychiatric illness, a deep infection and a recurrent dislocation. Five deaths occurred prior to five-year follow-up and one patient withdrew from clinical review. In the remaining 51 hips the mean pre-operative Oxford hip score was 47 points which decreased to 19 points at five years (45 hips). Of the stems 49 (98%) were implanted within 1° of neutral in the femoral canal. The mean migration of the stem at five years was 1.9 mm and the survivorship for aseptic loosening was 100%. There was no significant difference in outcome between the consultant and registrar groups. At five years, the results were comparable with those of other polished, tapered, cemented stems. Long-term surveillance continues.
We evaluated the outcome of 104 consecutive primary cemented Exeter femoral components in 78 patients (34 men, 44 women) under the age of 40 years who underwent total hip replacement between October 1993 and May 2004. The mean age at operation was 31 years (16 to 39). No hip was lost to follow-up, but three patients (four hips) died. None of the deaths were related to the surgery. At a mean follow-up of 6.2 years (2 to 13), three femoral components had been revised for septic loosening. Using Kaplan-Meier survival analysis, the seven-year survival of the component with revision for any reason as the endpoint was 95.8% (95% confidence interval 86.67 to 98.7). The seven-year survival with aseptic femoral loosening as the endpoint was 100% (95% confidence interval 100). The cemented Exeter femoral component in patients under the age of 40 shows promising medium-term results. As it is available in a wide range of sizes and offsets, we could address all types of anatomical variation in this series without the need for custom-made components.
We reviewed the long-term results at ten to 12 years of 118 total hip replacements in 109 patients using a second-generation hemispherical cementless acetabular component (Reflection) designed to address the problem of backside wear. Five patients (five hips) died and six patients (seven hips) were lost to follow-up. The remaining 98 patients (106 hips) had a mean age of 62.9 years (34.0 to 86.2) A rate of revision for aseptic loosening of 0.9%, and predictable results were found with respect to radiological evidence of fixation, lack of pain, walking ability, range of movement and function. One component was revised for aseptic loosening, and of the 101 hips (95.2%) that did not have a revision, minor osteolytic lesions of the pelvis were seen in six (5.9%). Kaplan-Meier survival analysis for the total cohort of 118 hips revealed a 96.4% survival at both ten (95% confidence interval 90 to 98) and 12 years (95% confidence interval 86 to 98).
The June 2014 Hip &
Pelvis Roundup360 looks at: Modular femoral necks: early signs are not good; is corrosion to blame for modular neck failures; metal-on-metal is not quite a closed book; no excess failures in fixation of displaced femoral neck fractures; noise no problem in hip replacement; heterotopic ossification after hip arthroscopy: are NSAIDs the answer?; thrombotic and bleeding events surprisingly low in total joint replacement; and the elephant in the room: complications and surgical volume.
This paper investigates whether cortical comminution
and intra-articular involvement can predict displacement in distal
radius fractures by using a classification that includes volar comminution
as a separate parameter. A prospective multicentre study involving non-operative treatment
of distal radius fractures in 387 patients aged between 15 and 74
years (398 fractures) was conducted. The presence of cortical comminution
and intra-articular involvement according to the Buttazzoni classification
is described. Minimally displaced fractures were treated with immobilisation
in a cast while displaced fractures underwent closed reduction with
subsequent immobilisation. Radiographs were obtained after reduction,
at 10 to 14 days and after union. The outcome measure was re-displacement
or union. In fractures with volar comminution (Buttazzoni type 4), 96%
(53 of 55) displaced. In intra-articular fractures without volar
comminution (Buttazzoni 3), 72% (84 of 117) displaced. In extra-articular
fractures with isolated dorsal comminution (Buttazzoni 2), 73% (106
of 145) displaced while in non-comminuted fractures (Buttazzoni
1), 16 % (13 of 81 ) displaced. A total of 32% (53 of 165) of initially minimally displaced fractures
later displaced. All of the initially displaced volarly comminuted
fractures re-displaced. Displacement occurred in 31% (63 of 205)
of fractures that were still in good alignment after 10 to 14 days. Regression analysis showed that volar and dorsal comminution
predicted later displacement, while intra-articular involvement
did not predict displacement. Volar comminution was the strongest
predictor of displacement. Cite this article:
This prospective study presents the ten-year (5 to 16) clinical and radiological results of 55 primary total hip replacements (THR) using a cementless modular femoral component (S-ROM). All patients had a significant anatomical abnormality which rendered the primary THR difficult. The mean Harris hip score was 36 (12 to 72) pre-operatively, 83 (44 to 100) at five years, and 85 (45 to 99) at ten years. The Western Ontario and McMaster Universities osteoarthritis index (WOMAC) and short-form (SF)-12 scores were recorded from the year 2000. The mean SF-12 score at five years after surgery was 45.24 (22.74 to 56.58) for the physical component and 54.14 (29.20 to 66.61) for the mental component. By ten years the SF-12 scores were 42.86 (21.59 to 58.95) and 51.03 (33.78 to 61.40), respectively. The mean WOMAC score at five years post-operatively was 25 (0 to 59), and at ten years was 27 (2 to 70). No femoral components were radiologically loose, although five had osteolysis in Gruen zone 1, three had osteolysis in zone 7, and two showed osteolysis in both zones 1 and 7. No osteolysis was observed around or distal to the prosthetic sleeve. No femoral components were revised, although three hips underwent an acetabular revision and two required a liner exchange. At a mean of ten years’ follow-up the S-ROM femoral component implanted for an anatomically difficult primary THR has excellent clinical and radiological results.
While frequently discussed as a standard treatment for the management of an infected shoulder replacement, there is little information on the outcome of two-stage re-implantation. We examined the outcome of 17 consecutive patients (19 shoulders) who were treated between 1995 and 2004 with a two-stage re-implantation for the treatment of a deep-infection after shoulder replacement. All 19 shoulders were followed for a minimum of two years or until the time of further revision surgery. The mean clinical follow-up was for 35 months (24 to 80). The mean radiological follow-up was 27 months (7 to 80). There were two excellent results, four satisfactory and 13 unsatisfactory. In 12 of the 19 shoulders (63%) infection was considered to be eradicated. The mean pain score improved from 4.2 (3 to 5 (out of 5)) to 1.8 (1 to 4). The mean elevation improved from 42° (0° to 140°) to 89° (0° to 165°), mean external rotation from 30° (0° to 90°) to 43° (0° to 90°), and mean internal rotation from the sacrum to L5. There were 14 complications. Our study suggests that two-stage re-implantation for an infected shoulder replacement is associated with a high rate of unsatisfactory results, marginal success at eradicating infection and a high complication rate.
In this retrospective study, we investigated
the results of revision total hip replacement (THR) using a cemented long-stemmed
Exeter femoral component, with a minimum length of 205 mm in patients
with extensive femoral bone defects. The study included 37 consecutive
patients with a mean age of 76 years (39 to 93) and a mean follow-up
of nine years (5 to 16). A total of 26 patients (70%) had a pre-operative
Endo-Klinik score of 3 or 4. Impaction bone grafting was used in
24 patients (65%). At the time of evaluation, 22 patients (59%)
were still alive and were evaluated clinically and radiologically.
A total of 14 patients died during follow-up and their data were
included until the time of their death. One reconstruction failed
after five years and five months owing to recurrent dislocation:
the hip was converted to an excision arthroplasty. Intra-operative
fractures or fissures were encountered in nine patients (24%), but
none occurred during impaction of the bone graft. Post-operative
peri-prosthetic fractures occurred in two patients (5%); both were
treated with plate fixation. At nine years, survival with the endpoint
of all-cause re-revision was 96.3% (95% CI 76.4 to 99.5); including
re-operations for any reason, it was 80.7% (95% CI 56.3 to 92.3%).
There were no re-revisions for aseptic loosening. The survival of long stem cemented femoral components following
revision THR is satisfactory in a fragile population with extensive
femoral defects. Cite this article: