The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 4 | Pages 439 - 442
1 Apr 2011
Sexton SA Yeung E Jackson MP Rajaratnam S Martell JM Walter WL Zicat BA Walter WK

We investigated factors that were thought to be associated with an increased incidence of squeaking of ceramic-on-ceramic total hip replacements. Between June 1997 and December 2008 the three senior authors implanted 2406 primary total hip replacements with a ceramic-on-ceramic bearing surface. The mean follow-up was 10.6 years. The diagnosis was primary osteoarthritis in each case, and no patient had undergone previous surgery to the hip. We identified 74 squeaking hips (73 patients) giving an incidence of 3.1% at a mean follow-up of 9.5 years (4.1 to 13.3).

Taller, heavier and younger patients were significantly more likely to have hips that squeaked. Squeaking hips had a significantly higher range of post-operative internal (p = 0.001) and external rotation (p = 0.003) compared with silent hips. Patients with squeaking hips had significantly higher activity levels (p = 0.009). A squeaking hip was not associated with a significant difference in patient satisfaction (p = 0.24) or Harris hip score (p = 0.34). Four implant position factors enabled good prediction of squeaking. These were high acetabular component inclination, high femoral offset, lateralisation of the hip centre and either high or low acetabular component anteversion.

This is the largest study to date to examine patient factors and implant position factors that predispose to squeaking of a ceramic-on-ceramic hip. The results suggest that factors which increase the mechanical forces across the hip joint and factors which increase the risk of neck-to-rim impingement, and therefore edge-loading, are those that predispose to squeaking.


The Bone & Joint Journal
Vol. 96-B, Issue 11_Supple_A | Pages 23 - 26
1 Nov 2014
Cooper HJ Della Valle CJ

Dislocation remains among the most common complications of, and reasons for, revision of both primary and revision total hip replacements (THR). Hence, there is great interest in maximising stability to prevent this complication. Head size has been recognised to have a strong influence on the risk of dislocation post-operatively. As femoral head size increases, stability is augmented, secondary to an increase in impingement-free range of movement. Larger head sizes also greatly increase the ‘jump distance’ required for the head to dislocate in an appropriately positioned cup. Level-one studies support the use of larger diameter heads as they decrease the risk of dislocation following primary and revision THR. Highly cross-linked polyethylene has allowed us to increase femoral head size, without a marked increase in wear. However, the thin polyethylene liners necessary to accommodate larger heads may increase the risk of liner fracture and larger heads have also been implicated in causing soft-tissue impingement resulting in groin pain. Larger diameter heads also impart larger forces on the femoral trunnion, which may contribute to corrosion, metal release, and adverse local tissue reactions. Alternative large bearings including large ceramic heads and dual mobility bearings may mitigate some of these risks, and several of these devices have been used with clinical success.

Cite this article: Bone Joint J 2014;96-B(11 Suppl A):23–6.


Bone & Joint Research
Vol. 4, Issue 5 | Pages 70 - 77
1 May 2015
Gupta A Liberati TA Verhulst SJ Main BJ Roberts MH Potty AGR Pylawka TK El-Amin III SF

Objectives

The purpose of this study was to evaluate in vivo biocompatibility of novel single-walled carbon nanotubes (SWCNT)/poly(lactic-co-glycolic acid) (PLAGA) composites for applications in bone and tissue regeneration.

Methods

A total of 60 Sprague-Dawley rats (125 g to 149 g) were implanted subcutaneously with SWCNT/PLAGA composites (10 mg SWCNT and 1gm PLAGA 12 mm diameter two-dimensional disks), and at two, four, eight and 12 weeks post-implantation were compared with control (Sham) and PLAGA (five rats per group/point in time). Rats were observed for signs of morbidity, overt toxicity, weight gain and food consumption, while haematology, urinalysis and histopathology were completed when the animals were killed.


Bone & Joint 360
Vol. 4, Issue 1 | Pages 2 - 5
1 Feb 2015
Wright GM Porteous MJ


The Bone & Joint Journal
Vol. 96-B, Issue 11_Supple_A | Pages 73 - 77
1 Nov 2014
Brown NM Hellman M Haughom BH Shah RP Sporer SM Paprosky WG

A pelvic discontinuity occurs when the superior and inferior parts of the hemi-pelvis are no longer connected, which is difficult to manage when associated with a failed total hip replacement. Chronic pelvic discontinuity is found in 0.9% to 2.1% of hip revision cases with risk factors including severe pelvic bone loss, female gender, prior pelvic radiation and rheumatoid arthritis. Common treatment options include: pelvic plating with allograft, cage reconstruction, custom triflange implants, and porous tantalum implants with modular augments. The optimal technique is dependent upon the degree of the discontinuity, the amount of available bone stock and the likelihood of achieving stable healing between the two segments. A method of treating pelvic discontinuity using porous tantalum components with a distraction technique that achieves both initial stability and subsequent long-term biological fixation is described.

Cite this article: Bone Joint J 2014;96-B(11 Suppl A):73–7.


The Bone & Joint Journal
Vol. 96-B, Issue 11_Supple_A | Pages 43 - 47
1 Nov 2014
Su EP Su SL

Surgical interventions consisting of internal fixation (IF) or total hip replacement (THR) are required to restore patient mobility after hip fractures. Conventionally, this decision was based solely upon the degree of fracture displacement. However, in the last ten years, there has been a move to incorporate patient characteristics into the decision making process. Research demonstrating that joint replacement renders superior functional results when compared with IF, in the treatment of displaced femoral neck fractures, has swayed the pendulum in favour of THR. However, a high risk of dislocation has always been the concern. Fortunately, there are newer technologies and alternative surgical approaches that can help reduce the risk of dislocation. The authors propose an algorithm for the treatment of femoral neck fractures: if minimally displaced, in the absence of hip joint arthritis, IF should be performed; if arthritis is present, or the fracture is displaced, then THR is preferred.

Cite this article: Bone Joint J 2014;96-B(11 Suppl A):43–7.


The Bone & Joint Journal
Vol. 96-B, Issue 11_Supple_A | Pages 87 - 92
1 Nov 2014
Kwong LM Nielsen ESN Ruiz DR Hsu AH Dines MD Mellano CM

A retrospective review was performed of patients undergoing primary cementless total knee replacement (TKR) using porous tantalum performed by a group of surgical trainees. Clinical and radiological follow-up involved 79 females and 26 males encompassing 115 knees. The mean age was 66.9 years (36 to 85). Mean follow-up was 7 years (2 to 11). Tibial and patellar components were porous tantalum monoblock implants, and femoral components were posterior stabilised (PS) in design with cobalt–chromium fibre mesh. Radiological assessments were made for implant positioning, alignment, radiolucencies, lysis, and loosening. There was 95.7% survival of implants. There was no radiological evidence of loosening and no osteolysis found. No revisions were performed for aseptic loosening. Average tibial component alignment was 1.4° of varus (4°of valgus to 9° varus), and 6.2° (3° anterior to 15° posterior) of posterior slope. Mean femoral component alignment was 6.6° (1° to 11°) of valgus. Mean tibiofemoral alignment was 5.6° of valgus (7° varus to 16° valgus). Patellar tilt was a mean of 2.4° lateral (5° medial to 28° lateral). Patient satisfaction with improvement in pain was 91%. Cementless TKR incorporating porous tantalum yielded good clinical and radiological outcomes at a mean of follow-up of seven-years.

Cite this article: Bone Joint J 2014;96-B(11 Suppl A):87–92.


The Bone & Joint Journal
Vol. 97-B, Issue 2 | Pages 177 - 184
1 Feb 2015
Felden A Vaz G Kreps S Anract P Hamadouche M Biau DJ

Conventional cemented acetabular components are reported to have a high rate of failure when implanted into previously irradiated bone. We recommend the use of a cemented reconstruction with the addition of an acetabular reinforcement cross to improve fixation.

We reviewed a cohort of 45 patients (49 hips) who had undergone irradiation of the pelvis and a cemented total hip arthroplasty (THA) with an acetabular reinforcement cross. All hips had received a minimum dose of 30 Gray (Gy) to treat a primary nearby tumour or metastasis. The median dose of radiation was 50 Gy (Q1 to Q3: 45 to 60; mean: 49.57, 32 to 72).

The mean follow-up after THA was 51 months (17 to 137). The cumulative probability of revision of the acetabular component for a mechanical reason was 0% (0 to 0%) at 24 months, 2.9% (0.2 to 13.3%) at 60 months and 2.9% (0.2% to 13.3%) at 120 months, respectively. One hip was revised for mechanical failure and three for infection.

Cemented acetabular components with a reinforcement cross provide good medium-term fixation after pelvic irradiation. These patients are at a higher risk of developing infection of their THA.

Cite this article: Bone Joint J 2015;97-B:177–84.


The Bone & Joint Journal
Vol. 97-B, Issue 1 | Pages 76 - 82
1 Jan 2015
Siebachmeyer M Boddu K Bilal A Hester TW Hardwick T Fox TP Edmonds M Kavarthapu V

We report the outcomes of 20 patients (12 men, 8 women, 21 feet) with Charcot neuro-arthropathy who underwent correction of deformities of the ankle and hindfoot using retrograde intramedullary nail arthrodesis. The mean age of the patients was 62.6 years (46 to 83); their mean BMI was 32.7 (15 to 47) and their median American Society of Anaesthetists score was 3 (2 to 4). All presented with severe deformities and 15 had chronic ulceration. All were treated with reconstructive surgery and seven underwent simultaneous midfoot fusion using a bolt, locking plate or a combination of both. At a mean follow-up of 26 months (8 to 54), limb salvage was achieved in all patients and 12 patients (80%) with ulceration achieved healing and all but one patient regained independent mobilisation. There was failure of fixation with a broken nail requiring revision surgery in one patient. Migration of distal locking screws occurred only when standard screws had been used but not with hydroxyapatite-coated screws. The mean American Academy of Orthopaedic Surgeons Foot and Ankle (AAOS-FAO) score improved from 50.7 (17 to 88) to 65.2 (22 to 88), (p = 0.015). The mean Short Form (SF)-36 Health Survey Physical Component Score improved from 25.2 (16.4 to 42.8) to 29.8 (17.7 to 44.2), (p = 0.003) and the mean Euroqol EQ‑5D‑5L score improved from 0.63 (0.51 to 0.78) to 0.67 (0.57 to 0.84), (p = 0.012).

Single-stage correction of deformity using an intramedullary hindfoot arthrodesis nail is a good form of treatment for patients with severe Charcot hindfoot deformity, ulceration and instability provided a multidisciplinary care plan is delivered.

Cite this article: Bone Joint J 2015;97-B:76–82.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 5 | Pages 603 - 608
1 May 2012
Vasukutty NL Middleton RG Matthews EC Young PS Uzoigwe CE Minhas THA

We present our experience with a double-mobility acetabular component in 155 consecutive revision total hip replacements in 149 patients undertaken between 2005 and 2009, with particular emphasis on the incidence of further dislocation. The mean age of the patients was 77 years (42 to 89) with 59 males and 90 females. In all, five patients died and seven were lost to follow-up. Indications for revision were aseptic loosening in 113 hips, recurrent instability in 29, peri-prosthetic fracture in 11 and sepsis in two. The mean follow-up was 42 months (18 to 68). Three hips (2%) in three patients dislocated within six weeks of surgery; one of these dislocated again after one year. All three were managed successfully with closed reduction. Two of the three dislocations occurred in patients who had undergone revision for recurrent dislocation. All three were found at revision to have abductor deficiency. There were no dislocations in those revised for either aseptic loosening or sepsis.

These results demonstrate a good mid-term outcome for this component. In the 29 patients revised for instability, only two had a further dislocation, both of which were managed by closed reduction.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 5 | Pages 571 - 571
1 May 2011
Faux JC

This year is the centenary of the birth of John Charnley. Some of his legacy to Orthopaedic Surgery is described in this editorial.


Bone & Joint 360
Vol. 3, Issue 2 | Pages 8 - 9
1 Apr 2014

The April 2014 Hip & Pelvis Roundup360 looks at: Recent arthroplasty and flight; whether that squeak could be a fracture; diagnosing early infected hip replacement; impaction grafting at a decade; whether squeaking is more common than previously thought; femoral offset associated with post THR outcomes; and periprosthetic fracture stabilisation.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 9 | Pages 1193 - 1201
1 Sep 2012
Hamilton HW Jamieson J

It is probable that both genetic and environmental factors play some part in the aetiology of most cases of degenerative hip disease. Geneticists have identified some single gene disorders of the hip, but have had difficulty in identifying the genetics of many of the common causes of degenerative hip disease. The heterogeneity of the phenotypes studied is part of the problem. A detailed classification of phenotypes is proposed. This study is based on careful documentation of 2003 consecutive total hip replacements performed by a single surgeon between 1972 and 2000. The concept that developmental problems may initiate degenerative hip disease is supported. The influences of gender, age and body mass index are outlined. Biomechanical explanations for some of the radiological appearances encountered are suggested. The body weight lever, which is larger than the abductor lever, causes the abductor power to be more important than body weight. The possibility that a deficiency in joint lubrication is a cause of degenerative hip disease is discussed. Identifying the phenotypes may help geneticists to identify genes responsible for degenerative hip disease, and eventually lead to a definitive classification.


The Bone & Joint Journal
Vol. 96-B, Issue 1 | Pages 3 - 4
1 Jan 2014
Barrack RL


The Bone & Joint Journal
Vol. 96-B, Issue 4 | Pages 508 - 512
1 Apr 2014
van Amerongen EA Creemers LB Kaoui N Bekkers JEJ Kon M Schuurman AH

Damage to the cartilage of the distal radioulnar joint frequently leads to pain and limitation of movement, therefore repair of this joint cartilage would be highly desirable. The purpose of this study was to investigate the fixation of scaffold in cartilage defects of this joint as part of matrix-assisted regenerative autologous cartilage techniques. Two techniques of fixation of collagen scaffolds, one involving fibrin glue alone and one with fibrin glue and sutures, were compared in artificially created cartilage defects of the distal radioulnar joint in a human cadaver. After being subjected to continuous passive rotation, the methods of fixation were evaluated for cover of the defect and pull out force.

No statistically significant differences were found between the two techniques for either cover of the defect or integrity of the scaffold. However, a significantly increased mean pull out force was found for the combined procedure, 0.665 N (0.150 to 1.160) versus 0.242 N (0.060 to 0.730) for glue fixation (p = 0.001).

This suggests that although successful fixation of a collagen type I/III scaffold in a distal radioulnar joint cartilage defect is feasible with both forms of fixation, fixation with glue and sutures is preferable.

Cite this article: Bone Joint J 2014;96-B:508–12.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 9 | Pages 1259 - 1264
1 Sep 2011
Wähnert D Windolf M Brianza S Rothstock S Radtke R Brighenti V Schwieger K

We investigated the static and cyclical strength of parallel and angulated locking plate screws using rigid polyurethane foam (0.32 g/cm3) and bovine cancellous bone blocks. Custom-made stainless steel plates with two conically threaded screw holes with different angulations (parallel, 10° and 20° divergent) and 5 mm self-tapping locking screws underwent pull-out and cyclical pull and bending tests. The bovine cancellous blocks were only subjected to static pull-out testing. We also performed finite element analysis for the static pull-out test of the parallel and 20° configurations. In both the foam model and the bovine cancellous bone we found the significantly highest pull-out force for the parallel constructs. In the finite element analysis there was a 47% more damage in the 20° divergent constructs than in the parallel configuration. Under cyclical loading, the mean number of cycles to failure was significantly higher for the parallel group, followed by the 10° and 20° divergent configurations.

In our laboratory setting we clearly showed the biomechanical disadvantage of a diverging locking screw angle under static and cyclical loading.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 4 | Pages 479 - 483
1 Apr 2011
Chang C Lai K Yang C Lan S

Between April 2004 and July 2007, we performed 241 primary total knee replacements in 204 patients using the e.motion posterior cruciate-retaining, multidirectional mobile-bearing prosthesis. Of these, 100 were carried out using an image-free navigation system, and the remaining 141 with the conventional technique. We conducted a retrospective study from the prospectively collected data of these patients to assess the early results of this new mobile-bearing design.

At a mean follow-up of 49 months (32 to 71), 18 knees (7.5%) had mechanical complications of which 13 required revision. Three of these had a peri-prosthetic fracture, and were removed from the study. The indication for revision in the remaining ten was loosening of the femoral component in two, tibiofemoral dislocation in three, disassociation of the polyethylene liner in four, and a broken polyethyene liner in one. There were eight further mechanically unstable knees which presented with recurrent disassociation of the polyethylene liner. There was no significant difference in the incidence of mechanical instability between the navigation-assisted procedures (8 of 99, 8.1%) and the conventionally implanted knees (10 of 139, 7.2%).

In our view, the relatively high rate of mechanical complications and revision within 30 months precludes the further use of new design of knee replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 7 | Pages 949 - 954
1 Jul 2011
Bisseling P Zeilstra DJ Hol AM van Susante JLC

The purpose of this study was to evaluate whether concerns about the release of metal ions in metal-on-metal total hip replacements (THR) should be extended to patients with metal-bearing total disc replacements (TDR).

Cobalt and chromium levels in whole blood and serum were measured in ten patients with a single-level TDR after a mean follow-up of 34.5 months (13 to 61) using inductively-coupled plasma mass spectrometry. These metal ion levels were compared with pre-operative control levels in 81 patients and with metal ion levels 12 months after metal-on-metal THR (n = 21) and resurfacing hip replacement (n = 36). Flexion-extension radiographs were used to verify movement of the TDR.

Cobalt levels in whole blood and serum were significantly lower in the TDR group than in either the THR (p = 0.007) or the resurfacing group (p < 0.001). Both chromium levels were also significantly lower after TDR versus hip resurfacing (p < 0.001), whereas compared with THR this difference was only significant for serum levels (p = 0.008). All metal ion levels in the THR and resurfacing groups were significantly higher than in the control group (p < 0.001). In the TDR group only cobalt in whole blood appeared to be significantly higher (p < 0.001). The median range of movement of the TDR was 15.5° (10° to 22°).

These results suggest that there is minimal cause for concern about high metal ion concentrations after TDR, as the levels appear to be only moderately elevated. However, spinal surgeons using a metal-on-metal TDR should still be aware of concerns expressed in the hip replacement literature about toxicity from elevated metal ion levels, and inform their patients appropriately.


The Bone & Joint Journal
Vol. 95-B, Issue 11_Supple_A | Pages 103 - 108
1 Nov 2013
Abolghasemian M Tangsataporn S Sternheim A Backstein DJ Safir OA Gross AE

The conventional method for reconstructing acetabular bone loss at revision surgery includes using structural bone allograft. The disadvantages of this technique promoted the advent of metallic but biocompatible porous implants to fill bone defects enhancing initial and long-term stability of the acetabular component. This paper presents the indications, surgical technique and the outcome of using porous metal acetabular augments for reconstructing acetabular defects.

Cite this article: Bone Joint J 2013;95-B, Supple A:103–8.


The Bone & Joint Journal
Vol. 95-B, Issue 11_Supple_A | Pages 3 - 6
1 Nov 2013
Wassef AJ Schmalzried TP

A modular femoral head–neck junction has practical advantages in total hip replacement. Taper fretting and corrosion have so far been an infrequent cause of revision. The role of design and manufacturing variables continues to be debated. Over the past decade several changes in technology and clinical practice might result in an increase in clinically significant taper fretting and corrosion. Those factors include an increased usage of large diameter (36 mm) heads, reduced femoral neck and taper dimensions, greater variability in taper assembly with smaller incision surgery, and higher taper stresses due to increased patient weight and/or physical activity. Additional studies are needed to determine the role of taper assembly compared with design, manufacturing and other implant variables.

Cite this article: Bone Joint J 2013;95-B, Supple A:3–6.