Little biomechanical information is available about kinematically aligned (KA) total knee arthroplasty (TKA). The purpose of this study was to simulate the kinematics and kinetics after KA TKA and mechanically aligned (MA) TKA with four different limb alignments. Bone models were constructed from one volunteer (normal) and three patients with three different knee deformities (slight, moderate and severe varus). A dynamic musculoskeletal modelling system was used to analyse the kinematics and the tibiofemoral contact force. The contact stress on the tibial insert, and the stress to the resection surface and medial tibial cortex were examined by using finite element analysis.Objectives
Materials and Methods
We report the results of revision total knee
replacement (TKR) in 26 patients with major metaphyseal osteolytic defects
using 29 trabecular metal cones in conjunction with a rotating hinged
total knee prosthesis. The osteolytic defects were types II and
III (A or B) according to the Anderson Orthopaedic Research Institute
(AORI) classification. The mean age of the patients was 72 years
(62 to 84) and there were 15 men and 11 women. In this series patients had
undergone a mean of 2.34 previous total knee arthroplasties. The
main objective was to restore anatomy along with stability and function
of the knee joint to allow immediate full weight-bearing and active
knee movement. Outcomes were measured using Knee Society scores,
Oxford knee scores, range of movement of the knee and serial radiographs.
Patients were followed for a mean of 36 months (24 to 49). The mean
Oxford knee clinical scores improved from 12.83 (10 to 15) to 35.20
(32 to 38) (p <
0.001) and mean American Knee Society scores
improved from 33.24 (13 to 36) to 81.12 (78 to 86) (p <
0.001).
No radiolucent lines suggestive of loosening were seen around the trabecular
metal cones, and by one year all the radiographs showed good osteo-integration.
There was no evidence of any collapse or implant migration. Our
early results confirm the findings of others that trabecular metal
cones offer a useful way of managing severe bone loss in revision
TKR. Cite this article:
We carried out a prospective study to assess the clinical outcome, complications and survival of the NexGen Legacy posterior-stabilised-Flex total knee replacement (TKR) in a consecutive series of 278 knees between May 2003 and February 2005. Mean follow-up for 259 TKRs (98.2%) was 3.8 years (3.0 to 4.8). Annual follow-up showed improvement in the Knee Society scores (paired This relatively large study indicates that the legacy posterior stabilised-Flex design provides excellent short-term outcome but warrants ongoing evaluation to confirm the long-term durability and functioning of the implant.