We compared lower limb coronal alignment measurements
obtained pre- and post-operatively with long-leg radiographs and
computer navigation in patients undergoing primary total knee replacement
(TKR). A series of 185 patients had their pre- and post-implant
radiological and computer-navigation system measurements of coronal alignment
compared using the Bland-Altman method. The study included 81 men
and 104 women with a mean age of 68.5 years (32 to 87) and a mean
body mass index of 31.7 kg/m. 2. (19 to 49). Pre-implant
Bland–Altman limits of agreement were -9.4° to 8.6° with a repeatability
coefficient of 9.0°. The Bland–Altman plot showed a tendency for the
radiological measurement to indicate a higher level of pre-operative
deformity than the corresponding navigation measurement. Post-implant
limits of agreement were -5.0° to 5.4° with a repeatability coefficient
of 5.2°. The tendency for valgus knees to have greater deformity
on the radiograph was still seen, but was weaker for
We present the results of the surgical correction of lower-limb deformities caused by metabolic bone disease. Our series consisted of 17 patients with a diagnosis of hypophosphataemic rickets and two with renal osteodystrophy; their mean age was 25.6 years (14 to 57). In all, 43 lower-limb segments (27 femora and 16 tibiae) were osteotomised and the deformity corrected using a monolateral external fixator. The segment was then stabilised with locked intramedullary nailing. In addition, six femora in three patients were subsequently lengthened by distraction osteogenesis. The mean follow-up was 60 months (18 to 120). The frontal alignment parameters (the mechanical axis deviation, the lateral distal femoral angle and the medial proximal tibial angle) and the sagittal alignment parameters (the posterior distal femoral angle and the posterior proximal tibial angle) improved post-operatively. The external fixator was removed either at the end of surgery or at the end of the lengthening period, allowing for early mobilisation and weight-bearing. We encountered five problems and four obstacles in the programme of treatment. The use of intramedullary nails prevented recurrence of deformity and refracture.