End caps are intended to prevent nail migration
(push-out) in elastic stable intramedullary nailing. The aim of
this study was to investigate the force at failure with and without
end caps, and whether different insertion angles of nails and end caps
would alter that force at failure. Simulated oblique fractures of the diaphysis were created in
15 artificial paediatric femurs. Titanium Elastic Nails with end
caps were inserted at angles of 45°, 55° and 65° in five specimens
for each angle to create three study groups. Biomechanical testing
was performed with axial compression until failure. An identical
fracture was created in four small adult cadaveric femurs harvested
from two donors (both female, aged 81 and 85 years, height 149 cm and
156 cm, respectively). All femurs were tested without and subsequently
with end caps inserted at 45°. In the artificial femurs, maximum force was not significantly
different between the three groups (p = 0.613). Push-out force was
significantly higher in the cadaveric specimens with the use of
end caps by an up to sixfold load increase (830 N, standard deviation
(SD) 280 These results indicate that the nail and end cap insertion angle
can be varied within 20° without altering construct stability and
that the risk of elastic stable intramedullary nailing push–out
can be effectively reduced by the use of end caps. Cite this article:
We treated 39 patients with fractures of the humeral shaft by closed retrograde locked intramedullary nailing, using Russell-Taylor humeral nails. The mean healing time of all fractures was 13.7 weeks. After consolidation, shoulder function was excellent in 92.3% and elbow function excellent in 87.2%. Functional end-results were excellent in 84.6% of patients, moderate in 10.3% and bad in 5.1%. One patient had a postoperative radial nerve palsy, which recovered within three months. There was additional comminution at the fracture site in three patients (7.7%) which did not affect healing, and slight
Experimental studies indicate that non-steroidal anti-inflammatory drugs (NSAIDs) may have negative effects on fracture healing. This study aimed to assess the effect of immediate and delayed short-term administration of clinically relevant parecoxib doses and timing on fracture healing using an established animal fracture model. A standardized closed tibia shaft fracture was induced and stabilized by reamed intramedullary nailing in 66 Wistar rats. A ‘parecoxib immediate’ (Pi) group received parecoxib (3.2 mg/kg bodyweight twice per day) on days 0, 1, and 2. A ‘parecoxib delayed’ (Pd) group received the same dose of parecoxib on days 3, 4, and 5. A control group received saline only. Fracture healing was evaluated by biomechanical tests, histomorphometry, and dual-energy x-ray absorptiometry (DXA) at four weeks.Objectives
Methods
The Precice intramedullary limb-lengthening system has demonstrated significant benefits over external fixation lengthening methods, leading to a paradigm shift in limb lengthening. This study compares outcomes following antegrade and retrograde femoral lengthening in both adolescent and adult patients. A retrospective review of prospectively collected data was undertaken of a consecutive series of 107 femoral lengthening operations in 92 patients. In total, 73 antegrade nails and 34 retrograde nails were inserted. Outcome was assessed by the regenerate healing index (HI), hip and knee range of movement (ROM), and the presence of any complications.Aims
Patients and Methods
This review is aimed at clinicians appraising
preclinical trauma studies and researchers investigating compromised bone
healing or novel treatments for fractures. It categorises the clinical
scenarios of poor healing of fractures and attempts to match them
with the appropriate animal models in the literature. We performed an extensive literature search of animal models
of long bone fracture repair/nonunion and grouped the resulting
studies according to the clinical scenario they were attempting
to reflect; we then scrutinised them for their reliability and accuracy
in reproducing that clinical scenario. Models for normal fracture repair (primary and secondary), delayed
union, nonunion (atrophic and hypertrophic), segmental defects and
fractures at risk of impaired healing were identified. Their accuracy
in reflecting the clinical scenario ranged greatly and the reliability
of reproducing the scenario ranged from 100% to 40%. It is vital to know the limitations and success of each model
when considering its application.