The long-term survival of the cementless Spotorno
CLS femoral component in patients aged >
50 years at the time of
arthroplasty was investigated. Survivorship analysis of a consecutive
series of 85 patients (100 hips; under 50 years of age at a mean
follow-up of 18.4 years (16.3 to 20.8)) was performed. The clinical
and radiographic outcomes were satisfactory. The overall rate of
survival of the femoral component was 93.5% (95% confidence interval
(CI), 90.9 to 96.1) after 19 years. Survival with revision for aseptic
loosening as the end point was 95.7% (95% CI 93.6 to 97.8%) at 19
years. This study demonstrates an excellent long-term survival of the
Spotorno CLS femoral component after 16 to 20 years in young patients
undergoing total hip arthroplasty. Cite this article:
In 2012 we reviewed a consecutive series of 92
uncemented THRs performed between 1986 and 1991 at our institution
using the CLS
We followed the first 354 consecutive implantations of a cementless, double-tapered straight femoral stem in 326 patients. Follow-up was at a mean of 12 years (10 to 15). The mean age of the patients was 57 years (13 to 81). At follow-up, 56 patients (59 hips) had died, and eight (eight hips) had been lost to follow-up. Twenty-five hips underwent femoral revision, eight for infection, three for periprosthetic fracture and 14 for aseptic loosening. The overall survival was 92% at 12 years (95% CI 88 to 95). Survival with femoral revision for aseptic loosening as an endpoint was 95% (95% CI 92 to 98). The median Harris hip score at follow-up was 84 points (23 to 100). Radiolucent lines (<
2 mm) in Gruen zones 1 and 7 were present in 38 (16%) and 34 hips (14%), respectively. Radiolucencies in zones 2 to 6 were found in five hips (2%). The results for mid- to long-term survival with this femoral component are encouraging and compare with those achieved in primary cemented total hip arthroplasty. The high rate of loosening of the cup and the high rate of pain are, however, a source of concern.
Peri-prosthetic femoral fracture after total hip replacement (THR) is associated with a poor outcome and high mortality. However, little is known about its long-term incidence after uncemented THR. We retrospectively reviewed a consecutive series of 326 patients (354 hips) who had received a CLS
We have updated our previous randomised controlled trial comparing release of chromium (Cr) and cobalt (Co) ions and included levels of titanium (Ti) ions. We have compared the findings from 28 mm metal-on-metal total hip replacement, performed using titanium CLS/
Although the Fitmore Hip Stem has been on the market for almost 15 years, it is still not well documented in randomized controlled trials. This study compares the Fitmore stem with the CementLeSs (CLS) in several different clinical and radiological aspects. The hypothesis is that there will be no difference in outcome between stems. In total, 44 patients with bilateral hip osteoarthritis were recruited from the outpatient clinic at a single tertiary orthopaedic centre. The patients were operated with bilateral one-stage total hip arthroplasty. The most painful hip was randomized to either Fitmore or CLS femoral component; the second hip was operated with the femoral component not used on the first side. Patients were evaluated at three and six months and at one, two, and five years postoperatively with patient-reported outcome measures, radiostereometric analysis, dual-energy X-ray absorptiometry, and conventional radiography. A total of 39 patients attended the follow-up visit at two years (primary outcome) and 35 patients at five years. The primary outcome was which hip the patient considered to have the best function at two years.Aims
Methods
The aim of this study was to evaluate the survival of a collarless, straight, hydroxyapatite-coated femoral stem in total hip arthroplasty (THA) at a minimum follow-up of 20 years. We reviewed the results of 165 THAs using the Omnifit HA system in 138 patients, performed between August 1993 and December 1999. The mean age of the patients at the time of surgery was 46 years (20 to 77). Avascular necrosis was the most common indication for THA, followed by ankylosing spondylitis and primary osteoarthritis. The mean follow-up was 22 years (20 to 31). At 20 and 25 years, 113 THAs in 91 patients and 63 THAs in 55 patients were available for review, respectively, while others died or were lost to follow-up. Kaplan-Meier analysis was performed to evaluate the survival of the stem. Radiographs were reviewed regularly, and the stability of the stem was evaluated using the Engh classification.Aims
Methods
We carried out 71 primary total hip arthroplasties using porous-coated, hemispherical press-fit Duraloc ‘100 Series’ cups in 68 consecutive patients; 61 were combined with the cementless
The aim of this study was to determine whether total hip arthroplasty (THA) for chronic hip pain due to unilateral primary osteoarthritis (OA) has a beneficial effect on cognitive performance. A prospective cohort study was conducted with 101 patients with end-stage hip OA scheduled for THA (mean age 67.4 years (SD 9.5), 51.5% female (n = 52)). Patients were assessed at baseline as well as after three and months. Primary outcome was cognitive performance measured by d2 Test of Attention at six months, Trail Making Test (TMT), FAS-test, Rivermead Behavioural Memory Test (RBMT; story recall subtest), and Rey-Osterrieth Complex Figure Test (ROCF). The improvement of cognitive performance was analyzed using repeated measures analysis of variance.Aims
Methods
Preoperative nasal All primary total hip arthroplasties (THA) and total knee arthroplasties (TKA) performed from January 2006 to April 2018 were retrospectively reviewed for the incidence of early PJI. Demographic parameters, risk factors for PJI (American Society of Anaesthesiologists classification, body mass index, smoking status, and diabetes mellitus) and implant types were collected. A preoperative screening and eradication protocol for nasal colonization of Aims
Methods
There are limited published data detailing the volumetric material loss from tapers of conventional metal-on-polyethylene (MoP) total hip arthroplasties (THAs). Our aim was to address this by comparing the taper wear rates measured in an explanted cohort of the widely used Exeter THA with those measured in a group of metal-on-metal (MoM) THAs. We examined an existing retrieval database to identify all Exeter V40 and Universal MoP THAs. Volumetric wear analysis of the taper surfaces was conducted using previously validated methodology. These values were compared with those obtained from a series of MoM THAs using non-parametric statistical methodology. A number of patient and device variables were accounted for using multiple regression modelling.Aims
Patients and Methods
Periprosthetic femoral fractures (PFF) following total hip arthroplasty
(THA) are devastating complications that are associated with functional
limitations and increased overall mortality. Although cementless
implants have been associated with an increased risk of PFF, the
precise contribution of implant geometry and design on the risk
of both intra-operative and post-operative PFF remains poorly investigated.
A systematic review was performed to aggregate all of the PFF literature
with specific attention to the femoral implant used. A systematic search strategy of several journal databases and
recent proceedings from the American Academy of Orthopaedic Surgeons
was performed. Clinical articles were included for analysis if sufficient
implant description was provided. All articles were reviewed by
two reviewers. A review of fundamental investigations of implant
load-to-failure was performed, with the intent of identifying similar
conclusions from the clinical and fundamental literature.Aims
Patients and Methods
This study reports the mid-term results of total hip arthroplasty (THA) performed using a monoblock acetabular component with a large-diameter head (LDH) ceramic-on-ceramic (CoC) bearing. Of the 276 hips (246 patients) included in this study, 264 (96%) were reviewed at a mean of 67 months (48 to 79) postoperatively. Procedures were performed with a mini posterior approach. Clinical and radiological outcomes were recorded at regular intervals. A noise assessment questionnaire was completed at last follow-up.Aims
Patients and Methods
We present a minimum 20-year follow-up study of 382 cemented
Exeter Universal total hip arthroplasties (350 patients) operated
on at a mean age of 66.3 years (17 to 94). All patients received the same design of femoral component, regardless
of the original diagnosis. Previous surgery had been undertaken
for 33 hips (8.6%). During the study period 218 patients with 236
hips (62%) died, 42 hips (11%) were revised and 110 hips (29%) in
96 patients were available for review. The acetabular components
were varied and some designs are now obsolete, however they were
all cemented.Aims
Patients and Methods
The aim of this study was to assess the effect
of frictional torque and bending moment on fretting corrosion at
the taper interface of a modular femoral component and to investigate
whether different combinations of material also had an effect. The
combinations we examined were 1) cobalt–chromium (CoCr) heads on
CoCr stems 2) CoCr heads on titanium alloy (Ti) stems and 3) ceramic
heads on CoCr stems. In test 1 increasing torque was imposed by offsetting the stem
in the anteroposterior plane in increments of 0 mm, 4 mm, 6 mm and
8 mm when the torque generated was equivalent to 0 Nm, 9 Nm, 14
Nm and 18 Nm. In test 2 we investigated the effect of increasing the bending
moment by offsetting the application of axial load from the midline
in the mediolateral plane. Increments of offset equivalent to head
+ 0 mm, head + 7 mm and head + 14 mm were used. Significantly higher currents and amplitudes were seen with increasing
torque for all combinations of material. However, Ti stems showed
the highest corrosion currents. Increased bending moments associated
with using larger offset heads produced more corrosion: Ti stems
generally performed worse than CoCr stems. Using ceramic heads did
not prevent corrosion, but reduced it significantly in all loading
configurations. Cite this article:
Many different lengths of stem are available
for use in primary total hip replacement, and the morphology of
the proximal femur varies greatly. The more recently developed shortened
stems provide a distribution of stress which closely mimics that
of the native femur. Shortening the femoral component potentially
comes at the cost of decreased initial stability. Clinical studies
on the performance of shortened cemented and cementless stems are promising,
although long-term follow-up studies are lacking. We provide an
overview of the current literature on the anatomical features of
the proximal femur and the biomechanical aspects and clinical outcomes
associated with the length of the femoral component in primary hip
replacement, and suggest a classification system for the length
of femoral stems. Cite this article:
Large-head metal-on-metal (MoM) total hip replacements
(THR) have given rise to concern. Comparative studies of small-head
MoM THRs over a longer follow-up period are lacking. Our objective
was to compare the incidence of complications such as infection,
dislocation, revision, adverse local tissue reactions, mortality
and radiological and clinical outcomes in small-head (28 mm) MoM
and ceramic-on-polyethylene (CoP) THRs up to 12 years post-operatively. A prospective cohort study included 3341 THRs in 2714 patients.
The mean age was 69.1 years (range 24 to 98) and 1848 (55.3%) were
performed in women, with a mean follow-up of 115 months (18 to 201).
There were 883 MoM and 2458 CoP bearings. Crude incidence rates
(cases/1000 person-years) were: infection 1.3 In conclusion, we found similar results for small-head MoM and
CoP bearings up to ten years post-operatively, but after ten years
MoM THRs had a higher risk of all-cause revision. Furthermore, the
presence of an adverse response to metal debris seen in the small-head
MOM group at revision is a cause for concern. Cite this article:
The need to demonstrate probity and fair market competition has increased scrutiny of the relationships between orthopaedic surgeons and the industry that supplies them with their tools and devices. Investigations and judgements from the US Department of Justice and the introduction of the AdvaMed and Eucomed codes have defined new boundaries for interactions between these groups. This article summarises the current interplay between orthopaedic surgeons and industry, and provides recommendations for the future.
The Morscher-Spotorno (MS-30) femoral stem is a stainless-steel, straight, three-dimensionally tapered, collarless implant for cemented fixation in total hip replacement. We report the results at ten years of a consecutive series of 124 total hip replacements in 121 patients with the matt-surfaced MS-30 stem and an alumina ceramic head of 28-mm diameter. All the stems were fixed with Palacos bone cement with gentamicin using a modern cementing technique. They were combined with an uncemented, press-fit cup. The mean period of observation was 10.2 years (8.3 to 12.1) and no patient was lost to follow-up. Twenty-seven patients (22%) died with the implant None of the stems or cups had been revised. The Harris hip score was excellent or good in 97% (85 hips) and moderate in 3% (three hips). Radiologically, six hips (6.8%) had osteolysis adjacent to the stem, mostly in Gruen zone 7. Twenty (22.7%) showed one or more radiolucent lines. Twenty-two stems (25%) had subsided by 2 mm to 5 mm. In these cases two showed osteolysis (9.1%) with subsidence and four without (6.1%). Radiolucent lines were seen in seven with migration (31.8%) and in 13 without (19.7%). No infections and no acetabular osteolysis were observed. The clinical results were excellent with survivorship after ten years of 100% and only a slightly statistically non-significant higher rate of osteolysis and radiolucency in cases of subsidence.