We present the extended follow-up (≥ 20 years)
of a series of fully hydroxyapatite-coated femoral components used in
72 primary total hip replacements (THRs). Earlier results of this
cohort have been previously published. All procedures were performed
between 1986 and 1991. The series involved 45 women and 15 men with
12 bilateral procedures. Their mean age at the time of surgery was
60 years (46 to 80) and the mean duration of follow-up was 22.5
years (20 to 25). At final follow-up, the mean Merle d’Aubigné and
Postel hip scores were 5.5 (4.5 to 6), 3.8 (3.5 to 5) and 3.3 (3.0
to 5.0) for pain, mobility and function, respectively. Of the patients
92% were very satisfied at the time of final follow-up. There were seven revisions: six of the acetabular component for
aseptic loosening and one of both the stem and the acetabular component
for loosening due to deep infection. The survival of this prosthesis
at 22.5 years with revision for any reason as the endpoint was 91.7%
(95% confidence interval (CI) 84 to 99). Survival with aseptic loosening
of the stem as the endpoint was 100% (95% CI 90 to 100). This prosthesis provides pain relief in the
Aims. Limited implant survival due to aseptic cup loosening is most commonly responsible for revision total hip arthroplasty (THA). Advances in implant designs and materials have been crucial in addressing those challenges. Vitamin E-infused highly cross-linked polyethylene (VEPE) promises strong wear resistance, high oxidative stability, and superior mechanical strength. Although VEPE monoblock cups have shown good mid-term performance and excellent wear patterns, long-term results remain unclear. This study evaluated migration and wear patterns and clinical and radiological outcomes at a minimum of ten years’ follow-up. Methods. This prospective observational study investigated 101 cases of primary THA over a mean duration of 129 months (120 to 149). At last follow-up, 57 cases with complete clinical and radiological outcomes were evaluated. In all cases, the acetabular component comprised an uncemented titanium particle-coated VEPE monoblock cup. Patients were assessed clinically and radiologically using the Harris Hip Score, visual analogue scale (pain and satisfaction), and an anteroposterior radiograph. Cup migration and polyethylene wear were measured using Einzel-Bild-Röntgen-Analyze software. All complications and associated treatments were documented until final follow-up. Results. Clinical assessment showed persistent major improvement in all scores. On radiological assessment, only one case showed a lucent line (without symptoms). At last follow-up, wear and migration were below the critical thresholds. No cup-related revisions were needed, indicating an outstanding survival rate of 100%. Conclusion. Isoelastic VEPE cups offer high success rates and may prevent osteolysis, aseptic loosening, and the need for revision surgeries in the
Aims. Hyaline cartilage has a low capacity for regeneration. Untreated osteochondral lesions of the femoral head can lead to progressive and symptomatic osteoarthritis of the hip. The purpose of this study is to analyze the clinical and radiological long-term outcome of patients treated with osteochondral autograft transfer. To our knowledge, this study represents a series of osteochondral autograft transfer of the hip with the longest follow-up. Methods. We retrospectively evaluated 11 hips in 11 patients who underwent osteochondral autograft transfer in our institution between 1996 and 2012. The mean age at the time of surgery was 28.6 years (8 to 45). Outcome measurement included standardized scores and conventional radiographs. Kaplan-Meier survival curve was used to determine the failure of the procedures, with conversion to total hip arthroplasty (THA) defined as the endpoint. Results. The mean follow-up of patients treated with osteochondral autograft transfer was 18.5 years (9.3 to 24.7). Six patients developed osteoarthritis and had a THA at a mean of 10.3 years (1.1 to 17.3). The cumulative survivorship of the native hips was 91% (95% confidence interval (CI) 74 to 100) at five years, 62% (95% CI 33 to 92) at ten years, and 37% (95% CI 6 to 70) at 20 years. Conclusion. This is the first study analyzing the long-term results of osteochondral autograft transfer of the femoral head. Although most patients underwent conversion to THA in the
Aims. The aim of this study was to compare the mid-term patient-reported outcome, bone remodelling, and migration of a short stem (Collum Femoris Preserving; CFP) with a conventional uncemented stem (Corail). Methods. Of 81 patients who were initially enrolled, 71 were available at five years’ follow-up. The outcomes at two years have previously been reported. The primary outcome measure was the clinical result assessed using the Oxford Hip Score (OHS). Secondary outcomes were the migration of the stem, measured using radiostereometric analysis (RSA), change of bone mineral density (BMD) around the stem, the development of radiolucent lines, and additional patient-reported outcome measures (PROMs). Results. There were no statistically significant differences between the groups regarding PROMs (median OHS (CFP 45 (interquartile range (IQR) 35 to 48); Corail 45 (IQR 40 to 48); p = 0.568). RSA showed stable stems in both groups, with little or no further subsidence between two and five years. Resorption of the femoral neck was evident in nine patients in the CFP group and in none of the 15 Corail stems with a collar that could be studied. Dual X-ray absorbiometry showed a significantly higher loss of BMD in the proximal Gruen zones in the CFP group (mean changes in BMD: Gruen zone 1, CFP -9.5 (95% confidence interval (CI) -14.8 to -4.2), Corail 1.0 (95% CI 3.4 to 5.4); Gruen zone 7, CFP -23.0 (95% CI -29.4 to -16.6), Corail -7.2 (95% CI -15.9 to 1.4). Two CFP stems were revised before two years’ follow-up due to loosening, and one Corail stem was revised after two years due to chronic infection. Conclusion. The CFP stem has a similar clinical outcome and subsidence pattern when compared with the Corail stem. More pronounced proximal stress-shielding was seen with the CFP stem, suggesting diaphyseal fixation, and questioning its femoral neck-sparing properties in the
Aims. The aim of this study is to report the long-term outcomes of instrumented femoral revisions with impaction allograft bone grafting (IBG) using the X-change femoral revision system at 30 years after introduction of the technique. Methods. We updated the outcomes of our previous study, based on 208 consecutive revisions using IBG and the X-change femoral revision system in combination with a cemented polished stem, performed in our tertiary care institute between 1991 and 2007. Kaplan-Meier survival analyses were used to determine the survival rate of the revisions with endpoint revision for any reason and aseptic loosening. Secondary outcomes were radiological loosening and patient-reported outcome measures. Results. Mean age at revision total hip arthroplasty (THA) was 64.9 years (30 to 86). The most prevalent diagnosis for the femoral revision was aseptic loosening. At review in May 2021, 81 patients (85 hips) were still alive and 118 patients (120 hips; 58%) had died. Three patients (3 hips; 1%) were lost to follow-up at 11, 15, and 16 years after surgery, respectively. Data of all deceased and lost patients were included until final follow-up. The mean follow-up was 13.4 years (0 to 28). During the follow-up, 22 re-revisions were performed. The most common reason for re-revision was infection (n = 12; 54%). The survival with endpoint re-revision for any reason was 86% (95% confidence interval (CI) 79 to 91) at 20 years and 74% (95% CI 43 to 89) at 25 years after surgery. The survival for endpoint re-revision for aseptic loosening was 97% (95% CI 91 to 99) after both 20 and 25 years. Conclusion. We conclude that femoral IBG is a valuable technique that can reconstitute femoral bone loss in the
Aims. The aim of this prospective cohort study was to evaluate the early migration of the TriFit cementless proximally coated tapered femoral stem using radiostereometric analysis (RSA). Methods. A total of 21 patients (eight men and 13 women) undergoing primary total hip arthroplasty (THA) for osteoarthritis of the hip were recruited in this study and followed up for two years. Two patients were lost to follow-up. All patients received a TriFit stem and Trinity Cup with a vitamin E-infused highly cross-linked ultra-high molecular weight polyethylene liner. Radiographs for RSA were taken postoperatively and then at three, 12, and 24 months. Oxford Hip Score (OHS), EuroQol five-dimension questionnaire (EQ-5D), and adverse events were reported. Results. At two years, the mean subsidence of the head and tip for the TriFit stem was 0.38 mm (SD 0.32) and 0.52 mm (SD 0.36), respectively. The total migration of the head and tip was 0.55 mm (SD 0.32) and 0.71 mm (SD 0.38), respectively. There were no statistically significant differences between the three to 12 months' migration (p = 0.105) and 12 to 24 months' migration (p = 0.694). The OHS and EQ-5D showed significant improvements at two years. Conclusion. The results of this study suggest that the TriFit femoral stem achieves initial stability and is likely to be stable in the mid and
Aims. We have previously demonstrated raised cobalt and chromium levels in patients with larger diameter femoral heads, following metal-on-polyethylene uncemented total hip arthroplasty. Further data have been collected, to see whether these associations have altered with time and to determine the long-term implications for these patients and our practice. Methods. Patients from our previous study who underwent Trident-Accolade primary total hip arthroplasties using a metal-on-polyethylene bearing in 2009 were reviewed. Patients were invited to have their cobalt and chromium levels retested, and were provided an Oxford Hip Score. Serum ion levels were then compared between groups (28 mm, 36 mm, and 40 mm heads) and over time. Results. Metal ion levels were repeated in 33 patients. When comparing the results of serum metal ion levels over time, regardless of head size, there was a significant increase in both cobalt and chromium levels (p < 0.001). Two patients with larger head sizes had undergone revision arthroplasty with evidence of trunnion damage at surgery. Two patients within the 40 mm subgroup had metal ion levels above the MHRA (Medicines and Healthcare Products Regulatory Agency) threshold for detailed investigation. The increase in cobalt and chromium, when comparing the 36 mm and 40 mm groups with those of the 28 mm group, was not significant (36 mm vs 28 mm; p = 0.092/p = 0.191; 40 mm vs 28 mm; p = 0.200/p = 0.091, respectively). There was no difference, between femoral head sizes, when comparing outcome as measured by the Oxford Hip Score. Conclusion. This study shows an increase in cobalt and chromium levels over time for all modular femoral head sizes in patients with metal-on-polyethylene bearings, with two patients demonstrating ion levels above the MHRA threshold for failure, and a further two patients requiring revision surgery. These results may have clinical implications regarding
Aims. Total hip arthroplasty (THA) patients undergoing or having a prior lumbar spine fusion (LSF) have an increased risk of mechanical complications. The aim of this registry-based, retrospective comparative cohort study is to assess the
The Exeter femoral stem is a double-tapered highly
polished collarless cemented implant with good long-term clinical
results. In order to determine why the stem functions well we have
undertaken a long-term radiostereometric analysis (RSA) study. A total of 20 patients undergoing primary Exeter total hip replacement
for osteoarthritis using the Hardinge approach were recruited and
followed with RSA for ten years. The stems progressively subsided
and internally rotated with posterior head migration. The mean subsidence
was 0.7 mm (95% confidence interval (CI) 0.5 to 0.9) at two years
and 1.3 mm (95% CI 1.0 to 1.6) at ten years. The mean posterior
migration of the head was 0.7 mm (95% CI 0.5 to 0.9) at two years
and 1.2 mm (95% CI 1.0 to 1.4) at ten years. There was no significant
cement restrictor migration. The Exeter stem continues to subside slowly into the cement mantle
in the
Aims. We investigated the long-term performance of the Tripolar Trident acetabular component used for recurrent dislocation in revision total hip arthroplasty. We assessed: 1) rate of re-dislocation; 2) incidence of complications requiring re-operation; and 3) Western Ontario and McMaster Universities osteoarthritis index (WOMAC) pain and functional scores. Patients and Methods. We retrospectively identified 111 patients who had 113 revision tripolar constrained liners between 1994 and 2008. All patients had undergone revision hip arthroplasty before the constrained liner was used: 13 after the first revision, 17 after the second, 38 after the third, and 45 after more than three revisions. A total of 75 hips (73 patients) were treated with Tripolar liners due to recurrent instability with abductor deficiency, In addition, six patients had associated cerebral palsy, four had poliomyelitis, two had multiple sclerosis, two had spina bifida, two had spondyloepiphyseal dysplasia, one had previous reversal of an arthrodesis, and 21 had proximal femoral replacements. The mean age of patients at time of Tripolar insertions was 72 years (53 to 89); there were 69 female patients (two bilateral) and 42 male patients. All patients were followed up for a mean of 15 years (10 to 24). Overall, 55 patients (57 hips) died between April 2011 and February 2018, at a mean of 167 months (122 to 217) following their tripolar liner implantation. We extracted demographics, implant data, rate of dislocations, and incidence of other complications. Results. At ten years, the Kaplan–Meier survivorship for dislocation was 95.6% (95% confidence interval (CI) 90 to 98), with 101 patients at risk. At 20 years, the survivorship for dislocation was 90.6% (95% CI 81.0 to 95.5), with one patient at risk. Eight patients (7.2%) had a dislocation of their constrained liners. At ten years, the survival to any event was 89.4% (95% CI 82 to 93.8), with 96 patients at risk. At 20 years, the survival to any event was 82.5% (95% CI 71.9 to 89.3), with one patient at risk. Five hips (4.4%) had deep infection. Two patients (1.8%) developed dissociated constraining rings with pain but without dislocation, which required re-operation. Two patients (1.8%) had periprosthetic femoral fractures, without dislocation, that were treated by revision stems along with exchange of the well-functioning constrained liners. Conclusion. Constrained tripolar liners used at revision hip arthroplasty provided favourable results in the
Aims. This study reports the mid-term results of total hip arthroplasty (THA) performed using a monoblock acetabular component with a large-diameter head (LDH) ceramic-on-ceramic (CoC) bearing. Patients and Methods. Of the 276 hips (246 patients) included in this study, 264 (96%) were reviewed at a mean of 67 months (48 to 79) postoperatively. Procedures were performed with a mini posterior approach. Clinical and radiological outcomes were recorded at regular intervals. A noise assessment questionnaire was completed at last follow-up. Results. There were four re-operations (1%) including one early revision for insufficient primary fixation (0.4%). No hip dislocation was reported. The mean University of California, Los Angeles (UCLA) activity score, 12-Item Short-Form Health Survey (SF-12) Mental Component Summary (MCS) score, SF-12 Physical Component Summary (PCS) score, Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC) score, and Forgotten Joint Score (FJS) were 6.6 (2 to 10), 52.8 (25.5 to 65.7), 53.0 (27.2 to 66.5), 7.7 (0 to 63), and 88.5 (23 to 100), respectively. No signs of loosening or osteolysis were observed on radiological review. The incidence of squeaking was 23% (n = 51/225). Squeaking was significantly associated with larger head diameter (p < 0.001), younger age (p < 0.001), higher SF-12 PCS (p < 0.001), and UCLA scores (p < 0.001). Squeaking did not affect patient satisfaction, with 100% of the squeaking hips satisfied with the surgery. Conclusion. LDH CoC THAs have demonstrated excellent functional outcomes at medium-term follow-up, with very low revision rate and no dislocations. The high incidence of squeaking did not affect patient satisfaction or function. LDH CoC with a monoblock acetabular component has the potential to provide
The localization of necrotic areas has been reported to impact the prognosis and treatment strategy for osteonecrosis of the femoral head (ONFH). Anteroposterior localization of the necrotic area after a femoral neck fracture (FNF) has not been properly investigated. We hypothesize that the change of the weight loading direction on the femoral head due to residual posterior tilt caused by malunited FNF may affect the location of ONFH. We investigate the relationship between the posterior tilt angle (PTA) and anteroposterior localization of osteonecrosis using lateral hip radiographs. Patients aged younger than 55 years diagnosed with ONFH after FNF were retrospectively reviewed. Overall, 65 hips (38 males and 27 females; mean age 32.6 years (SD 12.2)) met the inclusion criteria. Patients with stage 1 or 4 ONFH, as per the Association Research Circulation Osseous classification, were excluded. The ratios of anterior and posterior viable areas and necrotic areas of the femoral head to the articular surface were calculated by setting the femoral head centre as the reference point. The PTA was measured using Palm’s method. The association between the PTA and viable or necrotic areas of the femoral head was assessed using Spearman’s rank correlation analysis (median PTA 6.0° (interquartile range 3 to 11.5)).Aims
Methods
The purpose of this study is to report our updated results at a minimum follow-up of 30 years using a first generation uncemented tapered femoral component in primary total hip arthroplasty (THA). The original cohort consisted of 145 consecutive THAs performed by a single surgeon in 138 patients. A total of 37 patients (40 hips) survived a minimum of 30 years, and are the focus of this review. The femoral component used in all cases was a first-generation Taperloc with a non-modular 28 mm femoral head. Clinical follow-up at a minimum of 30 years was obtained on every living patient. Radiological follow-up at 30 years was obtained on all but four.Aims
Methods
The primary outcome was investigating differences in wear, as measured by femoral head penetration, between cross-linked vitamin E-diffused polyethylene (vE-PE) and cross-linked polyethylene (XLPE) acetabular component liners and between 32 and 36 mm head sizes at the ten-year follow-up. Secondary outcomes included acetabular component migration and patient-reported outcome measures (PROMs) such as the EuroQol five-dimension questionnaire, 36-Item Short-Form Health Survey, Harris Hip Score, and University of California, Los Angeles Activity Scale (UCLA). A single-blinded, multi-arm, 2 × 2 factorial randomized controlled trial was undertaken. Patients were recruited between May 2009 and April 2011. Radiostereometric analyses (RSAs) were performed from baseline to ten years. Of the 220 eligible patients, 116 underwent randomization, and 82 remained at the ten-year follow-up. Eligible patients were randomized into one of four interventions: vE-PE acetabular liner with either 32 or 36 mm femoral head, and XLPE acetabular liner with either 32 or 36 mm femoral head. Parameters were otherwise identical except for acetabular liner material and femoral head size.Aims
Methods
A significant reduction in wear at five and ten years was previously reported when comparing Durasul highly cross-linked polyethylene with nitrogen-sterilized Sulene polyethylene in total hip arthroplasty (THA). We investigated whether the improvement observed at the earlier follow-up continued, resulting in decreased osteolysis and revision surgery rates over the second decade. Between January 1999 and December 2001, 90 patients underwent surgery using the same acetabular and femoral components with a 28 mm metallic femoral head and either a Durasul or Sulene liner. A total of 66 hips of this prospective randomized study were available for a minimum follow-up of 20 years. The linear femoral head penetration rate was measured at six weeks, one year, and annually thereafter, using the Dorr method on digitized radiographs with a software package.Aims
Methods
This study reports the results of 38 total hip arthroplasties (THAs) in 33 patients aged less than 50 years, using the JRI Furlong hydroxyapatite ceramic (HAC)-coated femoral component. We describe the survival, radiological, and functional outcomes of 33 patients (38 THAs) at a mean follow-up of 27 years (25 to 32) between 1988 and 2018.Aims
Methods
Radiostereometric analysis (RSA) is the most accurate radiological method to measure in vivo wear of highly cross-linked polyethylene (XLPE) acetabular components. We have previously reported very low wear rates for a sequentially irradiated and annealed X3 XLPE liner (Stryker Orthopaedics, USA) when used in conjunction with a 32 mm femoral heads at ten-year follow-up. Only two studies have reported the long-term wear rate of X3 liners used in conjunction with larger heads using plain radiographs which have poor sensitivity. The aim of this study was to measure the ten-year wear of thin X3 XLPE liners against larger 36 or 40 mm articulations with RSA. We prospectively reviewed 19 patients who underwent primary cementless THA with the XLPE acetabular liner (X3) and a 36 or 40 mm femoral head with a resultant liner thickness of at least 5.8 mm. RSA radiographs at one week, six months, and one, two, five, and ten years postoperatively and femoral head penetration within the acetabular component were measured with UmRSA software. Of the initial 19 patients, 12 were available at the ten-year time point.Aims
Methods
To clarify the mid-term results of transposition osteotomy of the acetabulum (TOA), a type of spherical periacetabular osteotomy, combined with structural allograft bone grafting for severe hip dysplasia. We reviewed patients with severe hip dysplasia, defined as Severin IVb or V (lateral centre-edge angle (LCEA) < 0°), who underwent TOA with a structural bone allograft between 1998 and 2019. A medical chart review was conducted to extract demographic data, complications related to the osteotomy, and modified Harris Hip Score (mHHS). Radiological parameters of hip dysplasia were measured on pre- and postoperative radiographs. The cumulative probability of TOA failure (progression to Tönnis grade 3 or conversion to total hip arthroplasty) was estimated using the Kaplan–Meier product-limited method, and a multivariate Cox proportional hazard model was used to identify predictors for failure.Aims
Methods
Periprosthetic proximal femoral fractures (PFFs) are a major complication after total hip arthroplasty (THA). Health status after PFF is not specifically investigated. The aim of this study is to evaluate the health status pattern over two years after sustaining a PFF. A cohort of patients with PFF after THA was derived from the Brabant Injury Outcomes Surveillance (BIOS) study. The BIOS study, a prospective, observational, multicentre follow-up cohort study, was conducted to obtain data by questionnaires pre-injury and at one week, and one, three, six, 12, and 24 months after trauma. Primary outcome measures were the EuroQol five-dimension three-level questionnaire (EQ-5D-3L), the Health Utility Index 2 (HUI2), and the Health Utility Index 3 (HUI3). Secondary outcome measures were general measurements such as duration of hospital stay and mortality.Aims
Methods
The Exeter short stem was designed for patients with Dorr type A femora and short-term results are promising. The aim of this study was to evaluate the minimum five-year stem migration pattern of Exeter short stems in comparison with Exeter standard stems. In this case-control study, 25 patients (22 female) at mean age of 78 years (70 to 89) received cemented Exeter short stem (case group). Cases were selected based on Dorr type A femora and matched first by Dorr type A and then age to a control cohort of 21 patients (11 female) at mean age of 74 years (70 to 89) who received with cemented Exeter standard stems (control group). Preoperatively, all patients had primary hip osteoarthritis and no osteoporosis as confirmed by dual X-ray absorptiometry scanning. Patients were followed with radiostereometry for evaluation of stem migration (primary endpoint), evaluation of cement quality, and Oxford Hip Score. Measurements were taken preoperatively, and at three, 12, and 24 months and a minimum five-year follow-up.Aims
Methods