The Journal of Bone & Joint Surgery British Volume
Vol. 39-B, Issue 3 | Pages 477 - 486
1 Aug 1957
McKee GK

A method of intra-articular arthrodesis of the hip is described in which fixation is obtained with a lag-screw. Two main advantages are claimed: first, that it is a compression arthrodesis and gives early bony fusion; and second, that it gives adequate internal fixation in most cases so that the disadvantages of a plaster hip spica are avoided, and the patient may walk with crutches two or three weeks after the operation. The results were satisfactory in forty-seven of the first fifty cases in which this method of arthrodesis was employed. There were two post-operative deaths and one failure to secure fusion in a case of massive necrosis of the femoral head


The Bone & Joint Journal
Vol. 95-B, Issue 12 | Pages 1662 - 1666
1 Dec 2013
Parker L Garlick N McCarthy I Grechenig S Grechenig W Smitham P

The AO Foundation advocates the use of partially threaded lag screws in the fixation of fractures of the medial malleolus. However, their threads often bypass the radiodense physeal scar of the distal tibia, possibly failing to obtain more secure purchase and better compression of the fracture.

We therefore hypothesised that the partially threaded screws commonly used to fix a medial malleolar fracture often provide suboptimal compression as a result of bypassing the physeal scar, and proposed that better compression of the fracture may be achieved with shorter partially threaded screws or fully threaded screws whose threads engage the physeal scar.

We analysed compression at the fracture site in human cadaver medial malleoli treated with either 30 mm or 45 mm long partially threaded screws or 45 mm fully threaded screws. The median compression at the fracture site achieved with 30 mm partially threaded screws (0.95 kg/cm2 (interquartile range (IQR) 0.8 to 1.2) and 45 mm fully threaded screws (1.0 kg/cm2 (IQR 0.7 to 2.8)) was significantly higher than that achieved with 45 mm partially threaded screws (0.6 kg/cm2 (IQR 0.2 to 0.9)) (p = 0.04 and p < 0.001, respectively). The fully threaded screws and the 30mm partially threaded screws were seen to engage the physeal scar under an image intensifier in each case.

The results support the use of 30 mm partially threaded or 45 mm fully threaded screws that engage the physeal scar rather than longer partially threaded screws that do not. A 45 mm fully threaded screw may in practice offer additional benefit over 30 mm partially threaded screws in increasing the thread count in the denser paraphyseal region.

Cite this article: Bone Joint J 2013;95-B:1662–6.


Bone & Joint Research
Vol. 6, Issue 8 | Pages 481 - 488
1 Aug 2017
Caruso G Bonomo M Valpiani G Salvatori G Gildone A Lorusso V Massari L

Objectives. Intramedullary fixation is considered the most stable treatment for pertrochanteric fractures of the proximal femur and cut-out is one of the most frequent mechanical complications. In order to determine the role of clinical variables and radiological parameters in predicting the risk of this complication, we analysed the data pertaining to a group of patients recruited over the course of six years. Methods. A total of 571 patients were included in this study, which analysed the incidence of cut-out in relation to several clinical variables: age; gender; the AO Foundation and Orthopaedic Trauma Association classification system (AO/OTA); type of nail; cervical-diaphyseal angle; surgical wait times; anti-osteoporotic medication; complete post-operative weight bearing; and radiological parameters (namely the lag-screw position with respect to the femoral head, the Cleveland system, the tip-apex distance (TAD), and the calcar-referenced tip-apex distance (CalTAD)). Results. The incidence of cut-out across the sample was 5.6%, with a higher incidence in female patients. A significantly higher risk of this complication was correlated with lag-screw tip positioning in the upper part of the femoral head in the anteroposterior radiological view, posterior in the latero-lateral radiological view, and in the Cleveland peripheral zones. The tip-apex distance and the calcar-referenced tip-apex distance were found to be highly significant predictors of the risk of cut-out at cut-offs of 30.7 mm and 37.3 mm, respectively, but the former appeared more reliable than the latter in predicting the occurrence of this complication. Conclusion. The tip-apex distance remains the most accurate predictor of cut-out, which is significantly greater above a cut-off of 30.7 mm. Cite this article: G. Caruso, M. Bonomo, G. Valpiani, G. Salvatori, A. Gildone, V. Lorusso, L. Massari. A six-year retrospective analysis of cut-out risk predictors in cephalomedullary nailing for pertrochanteric fractures: Can the tip-apex distance (TAD) still be considered the best parameter?. Bone Joint Res 2017;6:481–488. DOI: 10.1302/2046-3758.68.BJR-2016-0299.R1


The Journal of Bone & Joint Surgery British Volume
Vol. 77-B, Issue 2 | Pages 277 - 283
1 Mar 1995
Choueka J Koval K Kummer F Crawford G Zuckerman J

We studied the biomechanical behaviour of three sliding fixation devices for trochanteric femoral fractures. These were a titanium alloy sideplate and lag screw, a titanium alloy sideplate and dome plunger with cement augmentation, and a stainless-steel sideplate and lag screw. We used 18 mildly osteoporotic cadaver femora, randomly assigned to one of the three fixation groups. Four displacement and two strain gauges were fixed to each specimen, and each femur was first tested intact (control), then as a two-part fracture and then as a four-part intertrochanteric fracture. A range of physiological loads was applied to determine load-bearing, load-sharing and head displacement. The four-part-fracture specimens were subsequently tested to failure to determine maximum fixation strengths and modes of failure. The dome-plunger group failed at a load 50% higher than that of the stainless-steel lag-screw group (p < 0.05) and at a load 20% higher than that of the titanium-alloy lag-screw group (NS). All 12 lag-screw specimens failed by cut-out through the femoral head or neck, but none of the dome-plunger group showed movement within the femoral head when tested to failure. Strain-gauge analysis showed that the dome plunger produced considerably less strain in the inferior neck and calcar region than either of the lag screws. Inferior displacement of the femoral head was greatest for the dome-plunger group, and was due to sliding of the plunger. The dome plunger with cement augmentation was able to support higher loads and did not fail by cut-out through the femoral head.(ABSTRACT TRUNCATED AT 250 WORDS)


The Journal of Bone & Joint Surgery British Volume
Vol. 78-B, Issue 3 | Pages 387 - 390
1 May 1996
Lunsjö K Ceder L Stigsson L Hauggaard A

The Medoff sliding plate (MSP) is a new device used to treat intertrochanteric and subtrochanteric fractures. There are three options for sliding; either along the shaft or the neck of the femur, or a combination of both. In a prospective series of 108 consecutive displaced intertrochanteric fractures we used combined dynamic compression. The patients were followed clinically and radiologically for one year. All fractures healed during the follow-up period. The only postoperative technical failure was one lag-screw penetration. Combined compression of the MSP gives increased dynamic capacity which reduces the risk of complications. The low rate of technical failure in our series compares favourably with that of the sliding hip screw or the Gamma nail but randomised trials comparing the MSP with other hip screw systems are necessary to find the true role of the MSP with its various sliding modes


The Journal of Bone & Joint Surgery British Volume
Vol. 54-B, Issue 3 | Pages 432 - 441
1 Aug 1972
Maudsley RH Chen SC

1. A modification of the McLaughlin technique of lag screw fixation of the fractured scaphoid is described. 2. Fifty-six patients operated upon between 1956 and 1966 have been reviewed and their fractures classified under the headings recent, delayed union and non-union. 3. Plaster casts were avoided and early return to work encouraged; 95 per cent did so within two months. 4. Of twenty-two recent fractures, including three perilunar trans-scaphoid dislocations, nineteen united; of fifteen showing delayed union, eight united: and of nineteen cases of non-union, only two united. 5. The function of the wrist in the seventeen cases of persistent non-union stabilised by lag-screw was surprisingly good; only one patient has required arthrodesis to date. 6. The indications and contra-indications for the operation are discussed


Bone & Joint Open
Vol. 5, Issue 2 | Pages 123 - 131
12 Feb 2024
Chen B Duckworth AD Farrow L Xu YJ Clement ND

Aims

This study aimed to determine whether lateral femoral wall thickness (LWT) < 20.5 mm was associated with increased revision risk of intertrochanteric fracture (ITF) of the hip following sliding hip screw (SHS) fixation when the medial calcar was intact. Additionally, the study assessed the association between LWT and patient mortality.

Methods

This retrospective study included ITF patients aged 50 years and over treated with SHS fixation between 2019 and 2021 at a major trauma centre. Demographic information, fracture type, delirium status, American Society of Anesthesiologists grade, and length of stay were collected. LWT and tip apex distance were measured. Revision surgery and mortality were recorded at a mean follow-up of 19.5 months (1.6 to 48). Cox regression was performed to evaluate independent risk factors associated with revision surgery and mortality.


Bone & Joint Open
Vol. 5, Issue 8 | Pages 652 - 661
8 Aug 2024
Taha R Davis T Montgomery A Karantana A

Aims

The aims of this study were to describe the epidemiology of metacarpal shaft fractures (MSFs), assess variation in treatment and complications following standard care, document hospital resource use, and explore factors associated with treatment modality.

Methods

A multicentre, cross-sectional retrospective study of MSFs at six centres in the UK. We collected and analyzed healthcare records, operative notes, and radiographs of adults presenting within ten days of a MSF affecting the second to fifth metacarpal between 1 August 2016 and 31 July 2017. Total emergency department (ED) attendances were used to estimate prevalence.


The Bone & Joint Journal
Vol. 100-B, Issue 1 | Pages 95 - 100
1 Jan 2018
Evers J Fischer M Zderic I Wähnert D Richards RG Gueorguiev B Raschke MJ Ochman S

Aims

The aim of this study was to investigate the effect of a posterior malleolar fragment (PMF), with < 25% ankle joint surface, on pressure distribution and joint-stability. There is still little scientific evidence available to advise on the size of PMF, which is essential to provide treatment. To date, studies show inconsistent results and recommendations for surgical treatment date from 1940.

Materials and Methods

A total of 12 cadaveric ankles were assigned to two study groups. A trimalleolar fracture was created, followed by open reduction and internal fixation. PMF was fixed in Group I, but not in Group II. Intra-articular pressure was measured and cyclic loading was performed.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 12 | Pages 1548 - 1557
1 Dec 2008
Brinkman J Lobenhoffer P Agneskirchner JD Staubli AE Wymenga AB van Heerwaarden RJ

New developments in osteotomy techniques and methods of fixation have caused a revival of interest of osteotomies around the knee. The current consensus on the indications, patient selection and the factors influencing the outcome after high tibial osteotomy is presented. This paper highlights recent research aimed at joint pressure redistribution, fixation stability and bone healing that has led to improved surgical techniques and a decrease of post-operative time to full weight-bearing.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 2 | Pages 145 - 154
1 Feb 2007
Giannoudis PV Tzioupis CC Pape H Roberts CS

With the development of systems of trauma care the management of pelvic disruption has evolved and has become increasingly refined. The goal is to achieve an anatomical reduction and stable fixation of the fracture. This requires adequate visualisation for reduction of the fracture and the placement of fixation. Despite the advances in surgical approach and technique, the functional outcomes do not always produce the desired result. New methods of percutaneous treatment in conjunction with innovative computer-based imaging have evolved in an attempt to overcome the existing difficulties. This paper presents an overview of the technical aspects of percutaneous surgery of the pelvis and acetabulum.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 1 | Pages 68 - 75
1 Jan 2005
Miedel R Ponzer S Törnkvist H Söderqvist A Tidermark J

We studied 217 patients with an unstable trochanteric or subtrochanteric fracture who had been randomly allocated to treatment by either internal fixation with a standard Gamma nail (SGN) or a Medoff sliding plate (MSP, biaxial dynamisation mode). Their mean age was 84 years (65 to 99) and they were reviewed at four and 12 months after surgery. Assessments of outcome included general complications, technical failures, revision surgery, activities of daily living (ADL), hip function (Charnley score) and the health-related quality of life (HRQOL, EQ-5D).

The rate of technical failure in patients with unstable trochanteric fractures was 6.5% (6/93) (including intra-operative femoral fractures) in the SGN group and 5.2% (5/96) in the MSP group. In patients with subtrochanteric fractures, there were no failures in the SGN group (n = 16) and two in the MSP group (n = 12). In the SGN group, there were intra-operative femoral fractures in 2.8% (3/109) and no post-operative fractures. There was a reduced need for revision surgery in the SGN group compared with the MSP group (8.3%; 9/108; p = 0.072). The SGN group also showed a lower incidence of severe general complications (p < 0.05) and a trend towards a lower incidence of wound infections (p = 0.05). There were no differences between the groups regarding the outcome of ADL, hip function or the HRQOL. The reduction in the HRQOL (EQ-5Dindexscore) was significant in both groups compared with that before the fracture (p < 0.005).

Our findings indicate that the SGN showed good results in both trochanteric and subtrochanteric fractures. The limited number of intra-operative femoral fractures did not influence the outcome or the need for revision surgery. Moreover, the SGN group had a reduced number of serious general complications and wound infections compared with the MSP group. The MSP in the biaxial dynamisation mode had a low rate of failure in trochanteric fractures but an unacceptably high rate when used in the biaxial dynamisation mode in subtrochanteric fractures.

The negative influence of an unstable trochanteric or subtrochanteric fracture on the quality of life was significant regardless of the surgical method.