The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 4 | Pages 510 - 516
1 Apr 2012
Hickey BA Kempshall PJ Metcalfe AJ Forster MC

As part of the national initiative to reduce waiting times for joint replacement surgery in Wales, the Cardiff and Vale NHS Trust referred 224 patients to the NHS Treatment Centre in Weston-Super-Mare for total knee replacement (TKR). A total of 258 Kinemax TKRs were performed between November 2004 and August 2006. Of these, a total of 199 patients (232 TKRs, 90%) have been followed up for five years. This cohort was compared with 258 consecutive TKRs in 250 patients, performed at Cardiff and Vale Orthopaedic Centre (CAVOC) over a similar time period. The five year cumulative survival rate was 80.6% (95% confidence interval (CI) 74.0 to 86.0) in the Weston-Super-Mare cohort and 95.0% (95% CI 90.2 to 98.2) in the CAVOC cohort with revision for any reason as the endpoint. The relative risk for revision at Weston-Super-Mare compared with CAVOC was 3.88 (p < 0.001). For implants surviving five years, the mean Oxford knee scores (OKS) and mean EuroQol (EQ-5D) scores were similar (OKS: Weston-Super-Mare 29 (2 to 47) vs CAVOC 29.8 (3 to 48), p = 0.61; EQ-5D: Weston-Super-Mare 0.53 (-0.38 to 1.00) vs CAVOC 0.55 (-0.32 to 1.00), p = 0.79). Patients with revised TKRs had significantly lower Oxford knee and EQ-5D scores (p < 0.001). The results show a higher revision rate for patients operated at Weston-Super-Mare Treatment Centre, with a reduction in functional outcome and quality of life after revision. This further confirms that patients moved from one area to another for joint replacement surgery fare poorly


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 5 | Pages 604 - 611
1 May 2009
Reay E Wu J Holland J Deehan D

We describe a cohort of patients with a high rate of mid-term failure following Kinemax Plus total knee replacement inserted between 1998 and 2001. This implant has been recorded as having a survival rate of 96% at ten years. However, in our series the survival rate was 75% at nine years. This was also significantly lower than that of subsequent consecutive series of PFC Sigma knee replacements performed by the same surgeon. No differences were found in the clinical and radiological parameters between the two groups. At revision the most striking finding was polyethylene wear. An independent analysis of the polyethylene components was therefore undertaken. Scanning electron microscopy revealed type 2 fusion defects in the ultra-high molecular weight polyethylene (UHMWPE), which indicated incomplete boundary fusion. Other abnormalities consistent with weak UHMWPE particle interface strength were present in both the explanted inserts and in unused inserts from the same period. We consider that these type 2 fusion defects are the cause of the early failure of the Kinemax implants. This may represent a manufacturing defect resulting in a form of programmed polyethylene failure


The Journal of Bone & Joint Surgery British Volume
Vol. 83-B, Issue 3 | Pages 359 - 363
1 Apr 2001
Back DL Cannon SR Hilton A Bankes MJK Briggs TWR

We reviewed the outcome of 422 primary cemented Kinemax total knee arthroplasties implanted into 369 patients over a period of five years, from January 1989. The operations were carried out at two NHS district general hospitals and one teaching hospital by 31 surgeons. During the period of review, 49 patients died and ten knees were lost to follow-up (68 knees). The mean Knee Society score improved from 28 before to 89 after surgery, and the mean function score increased from 23 to 79. The range of flexion improved from 92° to 105°. These improvements were maintained throughout the period of study. At the latest review radiolucent lines of 1 mm were seen around 15% of tibial components, 1.4% of patellar components and 9.5% of femoral components. In no case were these changes progressive. Using revision as the endpoint, cumulative survival was 99% after five years and 96.95% after nine years. All revisions were undertaken for deep infection or secondary trauma. Our study has shown that the Kinemax total knee replacement, when carried out with retention of the posterior cruciate ligament by surgeons of varying experience, produces very satisfactory results in the medium term


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 2 | Pages 229 - 233
1 Feb 2009
Kempshall PJ Metcalfe A Forster MC

As part of the government’s initiative to reduce waiting times for major joint surgery in Wales, the Cardiff and Vale NHS Trust sent 224 patients (258 knees) to the NHS Treatment Centre in Weston-Super-Mare for total knee replacement. The Kinemax total knee replacement system was used in all cases. The cumulative survival rate at three years was 79.2% (95% confidence interval (CI) 69.2 to 86.8) using re-operation for any cause as an endpoint and 85.3% (95% CI 75.9 to 91.8) using aseptic revision as an endpoint. This is significantly worse than that recorded in the published literature. These poor results have resulted in a significant impact on our service


The Journal of Bone & Joint Surgery British Volume
Vol. 74-B, Issue 2 | Pages 243 - 246
1 Mar 1992
Nafei A Nielsen S Kristensen O Hvid I

We report the results of 75 consecutive primary press-fit Kinemax arthroplasties with an average follow-up of 14 months (three to 28). We reviewed 26 cemented and 49 non-cemented tibial components implanted into 72 patients (30 men and 42 women, median age 70 years). At the latest follow-up the overall evaluation (Hospital for Special Surgery knee rating scale) for cemented cases was excellent in 54%, good in 42% and poor in 4%. No cemented prosthesis loosened. Of the non-cemented cases 55% were excellent, 23% good, and 2% fair. Ten tibial implants (20%) loosened and required revision. Residual pain marred the result in 24% of the non-cemented prostheses and in 4% of the cemented group. We do not recommend the press-fit, smooth-surfaced Kinemax prosthesis for non-cemented use


The Journal of Bone & Joint Surgery British Volume
Vol. 74-B, Issue 3 | Pages 472 - 473
1 May 1992
Grigoris P Treacy R McMinn D


Bone & Joint Open
Vol. 5, Issue 10 | Pages 911 - 919
21 Oct 2024
Clement N MacDonald DJ Hamilton DF Gaston P

Aims. The aims were to assess whether joint-specific outcome after total knee arthroplasty (TKA) was influenced by implant design over a 12-year follow-up period, and whether patient-related factors were associated with loss to follow-up and mortality risk. Methods. Long-term follow-up of a randomized controlled trial was undertaken. A total of 212 patients were allocated a Triathlon or a Kinemax TKA. Patients were assessed preoperatively, and one, three, eight, and 12 years postoperatively using the Oxford Knee Score (OKS). Reasons for patient lost to follow-up, mortality, and revision were recorded. Results. A total of 94 patients completed 12-year functional follow-up (62 females, mean age 66 years (43 to 82) at index surgery). There was a clinically significantly greater improvement in the OKS at one year (mean difference (MD) 3.0 (95% CI 0.4 to 5.7); p = 0.027) and three years (MD 4.7 (95% CI 1.9 to 7.5); p = 0.001) for the Triathlon group, but no differences were observed at eight (p = 0.331) or 12 years’ (p = 0.181) follow-up. When assessing the OKS in the patients surviving to 12 years, the Triathlon group had a clinically significantly greater improvement in the OKS (marginal mean 3.8 (95% CI 0.2 to 7.4); p = 0.040). Loss to functional follow-up (53%, n = 109/204) was independently associated with older age (p = 0.001). Patient mortality was the major reason (56.4%, n = 62/110) for loss to follow-up. Older age (p < 0.001) and worse preoperative OKS (p = 0.043) were independently associated with increased mortality risk. An age at time of surgery of ≥ 72 years was 75% sensitive and 74% specific for predicting mortality with an area under the curve of 78.1% (95% CI 70.9 to 85.3; p < 0.001). Conclusion. The Triathlon TKA was associated with clinically meaningful greater improvement in knee-specific outcome when compared to the Kinemax TKA. Loss to follow-up at 12 years was a limitation, and studies planning longer-term functional assessment could limit their cohort to patients aged under 72 years. Cite this article: Bone Jt Open 2024;5(10):911–919


The Bone & Joint Journal
Vol. 102-B, Issue 4 | Pages 434 - 441
1 Apr 2020
Hamilton DF Burnett R Patton JT MacPherson GJ Simpson AHRW Howie CR Gaston P

Aims. There are comparatively few randomized studies evaluating knee arthroplasty prostheses, and fewer still that report longer-term functional outcomes. The aim of this study was to evaluate mid-term outcomes of an existing implant trial cohort to document changing patient function over time following total knee arthroplasty using longitudinal analytical techniques and to determine whether implant design chosen at time of surgery influenced these outcomes. Methods. A mid-term follow-up of the remaining 125 patients from a randomized cohort of total knee arthroplasty patients (initially comprising 212 recruited patients), comparing modern (Triathlon) and traditional (Kinemax) prostheses was undertaken. Functional outcomes were assessed with the Oxford Knee Score (OKS), knee range of movement, pain numerical rating scales, lower limb power output, timed functional assessment battery, and satisfaction survey. Data were linked to earlier assessment timepoints, and analyzed by repeated measures analysis of variance (ANOVA) mixed models, incorporating longitudinal change over all assessment timepoints. Results. The mean follow-up of the 125 patients was 8.12 years (7.3 to 9.4). There was a reduction in all assessment parameters relative to earlier assessments. Longitudinal models highlight changes over time in all parameters and demonstrate large effect sizes. Significant between-group differences were seen in measures of knee flexion (medium-effect size), lower limb power output (large-effect size), and report of worst daily pain experienced (large-effect size) favouring the Triathlon group. No longitudinal between-group differences were observed in mean OKS, average daily pain report, or timed performance test. Satisfaction with outcome in surviving patients at eight years was 90.5% (57/63) in the Triathlon group and 82.8% (48/58) in the Kinemax group, with no statistical difference between groups (p = 0.321). Conclusion. At a mean 8.12 years, this mid-term follow-up of a randomized controlled trial cohort highlights a general reduction in measures of patient function with patient age and follow-up duration, and a comparative preservation of function based on implant received at time of surgery. Cite this article: Bone Joint J 2020;102-B(4):434–441


The Bone & Joint Journal
Vol. 97-B, Issue 1 | Pages 64 - 70
1 Jan 2015
Hamilton DF Burnett R Patton JT Howie CR Moran M Simpson AHRW Gaston P

Total knee arthroplasty (TKA) is an established and successful procedure. However, the design of prostheses continues to be modified in an attempt to optimise the functional outcome of the patient. . The aim of this study was to determine if patient outcome after TKA was influenced by the design of the prosthesis used. A total of 212 patients (mean age 69; 43 to 92; 131 female (62%), 81 male (32%)) were enrolled in a single centre double-blind trial and randomised to receive either a Kinemax (group 1) or a Triathlon (group 2) TKA. . Patients were assessed pre-operatively, at six weeks, six months, one year and three years after surgery. The outcome assessments used were the Oxford Knee Score; range of movement; pain numerical rating scales; lower limb power output; timed functional assessment battery and a satisfaction survey. Data were assessed incorporating change over all assessment time points, using repeated measures analysis of variance longitudinal mixed models. Implant group 2 showed a significantly greater range of movement (p = 0.009), greater lower limb power output (p = 0.026) and reduced report of ‘worst daily pain’ (p = 0.003) over the three years of follow-up. Differences in Oxford Knee Score (p = 0.09), report of ‘average daily pain’ (p = 0.57) and timed functional performance tasks (p = 0.23) did not reach statistical significance. Satisfaction with outcome was significantly better in group 2 (p = 0.001). These results suggest that patient outcome after TKA can be influenced by the prosthesis used. Cite this article: Bone Joint J 2015;97-B:64–70


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 9 | Pages 1228 - 1233
1 Sep 2012
Baliga S McNair CJ Barnett KJ MacLeod J Humphry RW Finlayson D

The incidence of anterior knee pain following total knee replacement (TKR) is reported to be as high as 49%. The source of the pain is poorly understood but the soft tissues around the patella have been implicated. In theory circumferential electrocautery denervates the patella thereby reducing efferent pain signals. However, there is mixed evidence that this practice translates into improved outcomes. We aimed to investigate the clinical effect of intra-operative circumpatellar electrocautery in patients undergoing TKR using the LCS mobile bearing or Kinemax fixed bearing TKR. A total of 200 patients were randomised to receive either circumpatellar electrocautery (diathermy) or not (control). Patients were assessed by visual analogue scale (VAS) for anterior knee pain and Oxford knee score (OKS) pre-operatively and three months, six months and one year post-operatively. Patients and assessors were blinded. There were 91 patients in the diathermy group and 94 in the control. The mean VAS improvement at one year was 3.9 in both groups (control; -10 to 6, diathermy;. -9 to 8, p < 0.001 in both cases, paired, two-tailed t-test). There was no significant difference in VAS between the groups at any other time. The mean OKS improvement was 17.7 points (0 to 34) in the intervention group and 16.6 (0 to 42) points in the control (p = 0.36). There was no significant difference between the two groups in OKS at any other time. We found no relevant effect of patellar electrocautery on either VAS anterior knee pain or OKS for patients undergoing LCS and Kinemax TKR


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 9 | Pages 1172 - 1179
1 Sep 2008
Wylde V Learmonth I Potter A Bettinson K Lingard E

We compared patient-reported outcomes of the Kinemax fixed- and mobile-bearing total knee replacement in a multi-centre randomised controlled trial. Patients were randomised to the fixed- or the mobile-bearing prosthesis via a sealed envelope method after the bone cuts had been made in the operating theatre. Randomisation was stratified by centre and diagnosis. Patients were assessed pre-operatively and at eight to 12 weeks, one year and two years post-operatively. Validated questionnaires were used which included the Western Ontario MacMasters University, Short-Form 12, Mental Health Index-5, Knee Injury and Osteoarthritis Outcome Score for Knee-Related Quality of Life and Function in Sport and Recreation scales and a validated scale of satisfaction post-operatively. In total, 242 patients (250 knees) with a mean age of 68 years (40 to 80) were recruited from four NHS orthopaedic centres. Of these, 132 patients (54.5%) were women. No statistically significant differences could be identified in any of the patient-reported outcome scores between patients who received the fixed-bearing or the mobile-bearing knee up to two-years post-operatively


The Bone & Joint Journal
Vol. 103-B, Issue 12 | Pages 1791 - 1801
1 Dec 2021
Bhalekar RM Nargol ME Shyam N Nargol AVF Wells SR Collier R Pabbruwe M Joyce TJ Langton DJ

Aims

The aim of this study was to investigate whether wear and backside deformation of polyethylene (PE) tibial inserts may influence the cement cover of tibial trays of explanted total knee arthroplasties (TKAs).

Methods

At our retrieval centre, we measured changes in the wear and deformation of PE inserts using coordinate measuring machines and light microscopy. The amount of cement cover on the backside of tibial trays was quantified as a percentage of the total surface. The study involved data from the explanted fixed-bearing components of four widely used contemporary designs of TKA (Attune, NexGen, Press Fit Condylar (PFC), and Triathlon), revised for any indication, and we compared them with components that used previous generations of PE. Regression modelling was used to identify variables related to the amount of cement cover on the retrieved trays.


The Bone & Joint Journal
Vol. 102-B, Issue 10 | Pages 1331 - 1340
3 Oct 2020
Attard V Li CY Self A Mann DA Borthwick LA O’Connor P Deehan DJ Kalson NS

Aims

Stiffness is a common complication after total knee arthroplasty (TKA). Pathogenesis is not understood, treatment options are limited, and diagnosis is challenging. The aim of this study was to investigate if MRI can be used to visualize intra-articular scarring in patients with stiff, painful knee arthroplasties.

Methods

Well-functioning primary TKAs (n = 11), failed non-fibrotic TKAs (n = 5), and patients with a clinical diagnosis of fibrosis1 (n = 8) underwent an MRI scan with advanced metal suppression (Slice Encoding for Metal Artefact Correction, SEMAC) with gadolinium contrast. Fibrotic tissue (low intensity on T1 and T2, low-moderate post-contrast enhancement) was quantified (presence and tissue thickness) in six compartments: supra/infrapatella, medial/lateral gutters, and posterior medial/lateral.


Bone & Joint Research
Vol. 6, Issue 3 | Pages 172 - 178
1 Mar 2017
Clement ND MacDonald DJ Hamilton DF Burnett R

Objectives

Preservation of posterior condylar offset (PCO) has been shown to correlate with improved functional results after primary total knee arthroplasty (TKA). Whether this is also the case for revision TKA, remains unknown. The aim of this study was to assess the independent effect of PCO on early functional outcome after revision TKA.

Methods

A total of 107 consecutive aseptic revision TKAs were performed by a single surgeon during an eight-year period. The mean age was 69.4 years (39 to 85) and there were 59 female patients and 48 male patients. The Oxford Knee Score (OKS) and Short-form (SF)-12 score were assessed pre-operatively and one year post-operatively. Patient satisfaction was also assessed at one year. Joint line and PCO were assessed radiographically at one year.


Bone & Joint 360
Vol. 2, Issue 3 | Pages 6 - 14
1 Jun 2013
Wallace WA

In the UK we have many surgeon inventors – surgeons who innovate and create new ways of doing things, who invent operations, who design new instruments to facilitate surgery or design new implants for using in patients. However truly successful surgeon inventors are a rare breed and they need to develop additional knowledge and skills during their career in order to push forward their devices and innovations. This article reviews my own experiences as a surgeon inventor and the highs and lows over the whole of my surgical career.


Bone & Joint 360
Vol. 4, Issue 2 | Pages 12 - 14
1 Apr 2015

The April 2015 Knee Roundup360 looks at: Genetic determinants of ACL strength; TKA outcomes influenced by prosthesis; Single- or two-stage revision for infected TKA?; Arthroscopic meniscectomy: a problem that just won’t go away!; Failure in arthroscopic ACL reconstruction; ACL reconstruction in the over 50s?; Knee arthroplasty for early osteoarthritis; All inside meniscal repair; Steroids, thrombogenic markers and TKA


The Bone & Joint Journal
Vol. 98-B, Issue 12 | Pages 1625 - 1634
1 Dec 2016
Scott CEH Oliver WM MacDonald D Wade FA Moran M Breusch SJ

Aims

Risk of revision following total knee arthroplasty (TKA) is higher in patients under 55 years, but little data are reported regarding non-revision outcomes. This study aims to identify predictors of dissatisfaction in these patients.

Patients and Methods

We prospectively assessed 177 TKAs (157 consecutive patients, 99 women, mean age 50 years; 17 to 54) from 2008 to 2013. Age, gender, implant, indication, body mass index (BMI), social deprivation, range of movement, Kellgren-Lawrence (KL) grade of osteoarthritis (OA) and prior knee surgery were recorded. Pre- and post-operative Oxford Knee Score (OKS) as well as Short Form-12 physical (PCS) and mental component scores were obtained. Post-operative range of movement, complications and satisfaction were measured at one year.


The Bone & Joint Journal
Vol. 95-B, Issue 9 | Pages 1209 - 1216
1 Sep 2013
van der Voort P Pijls BG Nouta KA Valstar ER Jacobs WCH Nelissen RGHH

Mobile-bearing (MB) total knee replacement (TKR) was introduced to reduce the risk of aseptic loosening and wear of polyethylene inserts. However, no consistent clinical advantages of mobile- over fixed-bearing (FB) TKR have been found. In this study we evaluated whether mobile bearings have an advantage over fixed bearings with regard to revision rates and clinical outcome scores. Furthermore, we determined which modifying variables affected the outcome.

A systematic search of the literature was conducted to collect clinical trials comparing MB and FB in primary TKR. The primary outcomes were revision rates for any reason, aseptic loosening and wear. Secondary outcomes included range of movement, Knee Society score (KSS), Oxford knee score (OKS), Short-Form 12 (SF-12) score and radiological parameters. Meta-regression techniques were used to explore factors modifying the observed effect.

Our search yielded 1827 publications, of which 41 studies met our inclusion criteria, comprising over 6000 TKRs. Meta-analyses showed no clinically relevant differences in terms of revision rates, clinical outcome scores or patient-reported outcome measures between MB and FB TKRs. It appears that theoretical assumptions of superiority of MB over FB TKR are not borne out in clinical practice.

Cite this article: Bone Joint J 2013;95-B:1209–16.


The Bone & Joint Journal
Vol. 95-B, Issue 8 | Pages 1057 - 1063
1 Aug 2013
Zeng Y Shen B Yang J Zhou ZK Kang PD Pei FX

The purpose of this study was to undertake a meta-analysis to determine whether there is lower polyethylene wear and longer survival when using mobile-bearing implants in total knee replacement when compared with fixed-bearing implants. Of 975 papers identified, 34 trials were eligible for data extraction and meta-analysis comprising 4754 patients (6861 knees). We found no statistically significant differences between the two designs in terms of the incidence of radiolucent lines, osteolysis, aseptic loosening or survival. There is thus currently no evidence to suggest that the use of mobile-bearing designs reduce polyethylene wear and prolong survival after total knee replacement.

Cite this article: Bone Joint J 2013;95-B:1057–63.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 8 | Pages 1009 - 1015
1 Aug 2012
Scott CEH Biant LC

Stems improve the mechanical stability of tibial components in total knee replacement (TKR), but come at a cost of stress shielding along their length. Their advantages include resistance to shear, reduced tibial lift-off and increased stability by reducing micromotion. Longer stems may have disadvantages including stress shielding along the length of the stem with associated reduction in bone density and a theoretical risk of subsidence and loosening, peri-prosthetic fracture and end-of-stem pain. These features make long stems unattractive in the primary TKR setting, but often desirable in revision surgery with bone loss and instability. In the revision scenario, stems are beneficial in order to convey structural stability to the construct and protect the reconstruction of bony defects. Cemented and uncemented long stemmed implants have different roles depending on the nature of the bone loss involved.

This review discusses the biomechanics of the design of tibial components and stems to inform the selection of the component and the technique of implantation.