Advertisement for orthosearch.org.uk
Results 1 - 9 of 9
Results per page:
Bone & Joint Research
Vol. 7, Issue 6 | Pages 422 - 429
1 Jun 2018
Acklin YP Zderic I Inzana JA Grechenig S Schwyn R Richards RG Gueorguiev B

Aims

Plating displaced proximal humeral fractures is associated with a high rate of screw perforation. Dynamization of the proximal screws might prevent these complications. The aim of this study was to develop and evaluate a new gliding screw concept for plating proximal humeral fractures biomechanically.

Methods

Eight pairs of three-part humeral fractures were randomly assigned for pairwise instrumentation using either a prototype gliding plate or a standard PHILOS plate, and four pairs were fixed using the gliding plate with bone cement augmentation of its proximal screws. The specimens were cyclically tested under progressively increasing loading until perforation of a screw. Telescoping of a screw, varus tilting and screw migration were recorded using optical motion tracking.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 10 | Pages 1382 - 1389
1 Oct 2012
Sewell MD Kang SN Al-Hadithy N Higgs DS Bayley I Falworth M Lambert SM

There is little information about the management of peri-prosthetic fracture of the humerus after total shoulder replacement (TSR). This is a retrospective review of 22 patients who underwent a revision of their original shoulder replacement for peri-prosthetic fracture of the humerus with bone loss and/or loose components. There were 20 women and two men with a mean age of 75 years (61 to 90) and a mean follow-up 42 months (12 to 91): 16 of these had undergone a previous revision TSR. Of the 22 patients, 12 were treated with a long-stemmed humeral component that bypassed the fracture. All their fractures united after a mean of 27 weeks (13 to 94). Eight patients underwent resection of the proximal humerus with endoprosthetic replacement to the level of the fracture. Two patients were managed with a clam-shell prosthesis that retained the original components. The mean Oxford shoulder score (OSS) of the original TSRs before peri-prosthetic fracture was 33 (14 to 48). The mean OSS after revision for fracture was 25 (9 to 31). Kaplan-Meier survival using re-intervention for any reason as the endpoint was 91% (95% confidence interval (CI) 68 to 98) and 60% (95% CI 30 to 80) at one and five years, respectively.

There were two revisions for dislocation of the humeral head, one open reduction for modular humeral component dissociation, one internal fixation for nonunion, one trimming of a prominent screw and one re-cementation for aseptic loosening complicated by infection, ultimately requiring excision arthroplasty. Two patients sustained nerve palsies.

Revision TSR after a peri-prosthetic humeral fracture associated with bone loss and/or loose components is a salvage procedure that can provide a stable platform for elbow and hand function. Good rates of union can be achieved using a stem that bypasses the fracture. There is a high rate of complications and function is not as good as with the original replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11 | Pages 1540 - 1545
1 Nov 2012
Reigstad O Lütken T Grimsgaard C Bolstad B Thorkildsen R Røkkum M

The Motec cementless modular metal-on-metal ball-and-socket wrist arthroplasty was implanted in 16 wrists with scaphoid nonunion advanced collapse (SNAC; grades 3 or 4) and 14 wrists with scapholunate advanced collapse (SLAC) in 30 patients (20 men) with severe (grades 3 or 4) post-traumatic osteoarthritis of the wrist. The mean age of the patients was 52 years (31 to 71). All prostheses integrated well radiologically. At a mean follow-up of 3.2 years (1.1 to 6.1) no luxation or implant breakage occurred. Two wrists were converted to an arthrodesis for persistent pain. Loosening occurred in one further wrist at five years post-operatively. The remainder demonstrated close bone–implant contact. The clinical results were good, with markedly decreased Disabilities of the Arm Shoulder and Hand (DASH) and pain scores, and increased movement and grip strength. No patient used analgesics and most had returned to work.

Good short-term function was achieved using this wrist arthroplasty in a high-demand group of patients with post-traumatic osteoarthritis.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 5 | Pages 672 - 676
1 May 2005
Thomas S Price AJ Sankey RA Thomas M

Replacement of the shoulder in juvenile idiopathic arthritis is not often performed and there have been no published series to date.

We present nine glenohumeral hemiarthroplasties in eight patients with systemic or polyarticular juvenile idiopathic arthritis. The mean follow-up was six years (59 to 89 months). The mean age at the time of surgery was 32 years. Surgery took place at a mean of 27 years after diagnosis. The results indicated excellent relief from pain. There was restoration of useful function which deteriorated with time, in part because of progression of the systemic disease in this severely affected group. No patient has required revision to date and there has been no radiological evidence of loosening or osteolysis around the implants.

We discuss the pathoanatomical challenges unique to this group. There was very little space for a prosthetic joint and, in some cases, bony deformity and the small size necessitated the use of custom-made implants.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 1 | Pages 57 - 61
1 Jan 2011
Naveed MA Kitson J Bunker TD

The combination of an irreparable tear of the rotator cuff and destructive arthritis of the shoulder joint may cause severe pain, disability and loss of independence in the aged. Standard anatomical shoulder replacements depend on a functioning rotator cuff, and hence may fail in the presence of tears in the cuff. Many designs of non-anatomical constrained or semi-constrained prostheses have been developed for cuff tear arthropathy, but have proved unsatisfactory and were abandoned. The DePuy Delta III reverse prosthesis, designed by Grammont, medialises and stabilises the centre of rotation of the shoulder joint and has shown early promise. This study evaluated the mid-term clinical and radiological results of this arthroplasty in a consecutive series of 50 shoulders in 43 patients with a painful pseudoparalysis due to an irreparable cuff tear and destructive arthritis, performed over a period of seven years by a single surgeon. A follow-up of 98% was achieved, with a mean duration of 39 months (8 to 81). The mean age of the patients at the time of surgery was 81 years (59 to 95). The female to male ratio was 5:1. During the seven years, six patients died of natural causes. The clinical outcome was assessed using the American Shoulder and Elbow score, the Oxford Shoulder Score and the Short-form 36 score. A radiological review was performed using the Sirveaux score for scapular notching.

The mean American Shoulder and Elbow score was 19 (95% confidence interval (CI) 14 to 23) pre-operatively, and 65 (95% CI 48 to 82) (paired t-test, p < 0.001) at final follow-up. The mean Oxford score was 44 (95% CI 40 to 51) pre-operatively and 23 (95% CI 18 to 28) (paired t-test, p < 0.001) at final follow-up.

The mean maximum elevation improved from 55° pre-operatively to 105° at final follow-up. There were seven complications during the whole series, although only four patients required further surgery.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 10 | Pages 1410 - 1415
1 Oct 2010
Castagna A Randelli M Garofalo R Maradei L Giardella A Borroni M

Total shoulder replacement is a successful procedure for degenerative or some inflammatory diseases of the shoulder. However, fixation of the glenoid seems to be the main weakness with a high rate of loosening. The results using all-polyethylene components have been better than those using metal-backed components. We describe our experience with 35 consecutive total shoulder replacements using a new metal-backed glenoid component with a mean follow-up of 75.4 months (48 to 154).

Our implant differs from others because of its mechanism of fixation. It has a convex metal-backed bone interface and the main stabilising factor is a large hollow central peg. The patients were evaulated with standard radiographs and with the Constant Score, the Simple Shoulder Test and a visual analogue scale. All the scores improved and there was no loosening, no polyethylene-glenoid disassembly and no other implant-related complications.

We conclude that a metal-backed glenoid component is a good option in total shoulder replacement with no worse results than of those using a cemented all-polyethylene prosthesis.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 10 | Pages 1403 - 1409
1 Oct 2010
Pape G Zeifang F Bruckner T Raiss P Rickert M Loew M

Fractures of the proximal humerus can lead to malalignment of the humeral head, necrosis and post-traumatic osteoarthritis. In such cases surface replacement might be a promising option.

A total of 28 shoulders with glenohumeral arthritis subsequent to a fracture underwent surface replacement arthroplasty of the humeral head in patients with a mean age of 60 years (35 to 83). On the basis of the inclination of the impacted head, post-traumatic arthritis was divided into three types: type 1, an impacted fracture of the head in an anatomical position (seven cases); type 2, a valgus impacted fracture (13 cases); type 3, a varus impacted fracture (eight cases). The outcome was measured by means of the Constant score.

According to the Boileau classification of the sequelae of fractures of the proximal humerus, all 28 patients had a final result of intra-capsular category 1. The mean Constant score for the 28 shoulders increased from 23.2 points (2 to 45) pre-operatively to 55.1 points (20 to 89) at a mean of 31 months (24 to 66) post-operatively. Valgus impacted fractures had significantly better results (p < 0.039).

Surface replacement arthroplasty can provide good results for patients with post-traumatic osteoarthritis of the shoulder. Their use avoids post-operative complications of the humeral shaft, such as peri-prosthetic fractures. Further surgery can be undertaken more easily as the bone stock is preserved.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 10 | Pages 1335 - 1340
1 Oct 2009
Kontakis GM Tosounidis TI Christoforakis Z Hadjipavlou AG

We reviewed the outcome of 28 patients who had been treated using the Aequalis fracture prosthesis for an acute fracture of the proximal humerus at a mean follow-up of 39.3 months (24 to 63). The mean age of the patients at the time of the fracture was 66.3 years (38 to 80). The mean Constant score was 68.2 (37 to 84) for the operated shoulder, which represented 89.5% of the mean score for the uninjured side (p < 0.001). The quality of the reconstruction as shown on the immediate post-operative radiographs was categorised into three types, anatomical, acceptable, and unacceptable, depending on the position of the tuberosities relative to the prosthetic head and the humeral shaft. Anatomical reconstruction was associated with a higher mean Constant score as well as higher mean values of anterior forward elevation, abduction and external rotation than the other types, but the differences were not statistically significant (p > 0.231).

A total of 18 patients had active anterior elevation ≥150°. Their mean active abduction and external rotation were 163.6° and 31.3°, respectively. In seven of the 28 patients, the mean active anterior elevation, abduction and external rotation were 130.7°, 129.2° and 22.8°, respectively. In all, 12 patients were very satisfied with the results, 12 were satisfied, two were dissatisfied and two were disappointed; 26 reported no or only mild pain while only two had moderate pain. In five patients proximal migration of the humeral head was shown on the anteroposterior radiographs of the shoulder. No evidence of loosening was found in any prosthesis.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 11 | Pages 1466 - 1469
1 Nov 2007
Mullett H Levy O Raj D Even T Abraham R Copeland SA

We describe the results of Copeland surface replacement shoulder arthroplasty using the mark III prosthesis in patients over 80 years of age. End-stage arthritis of the shoulder is a source of significant pain and debilitating functional loss in the elderly. An arthroplasty offers good relief of pain and may allow the patient to maintain independence. The risk-benefit ratio of shoulder replacement may be felt to be too high in an elderly age group, but there is no published evidence to support this theory. We have assessed whether the procedure was as reliable and safe as previously seen in a younger cohort of patients. Between 1993 and 2003, 213 Copeland surface replacement arthroplasty procedures were performed in our unit, of which 29 (13.6%) were undertaken in patients over the age of 80. This group of patients was followed up for a mean of 4.5 years (2.1 to 9.3). Their mean age was 84.3 years (81 to 93), the mean operating time was 40 minutes (30 to 45) and the mean in-patient stay was five days (2 to 21). There were no peri-operative deaths or significant complications. The mean Constant score adjusted for age and gender, improved from 15.1% to 77%. Copeland surface replacement shoulder arthroplasty may be performed with minimal morbidity and rapid rehabilitation in the elderly.