Aims. The aim of this study is to report the long-term outcomes of instrumented
Revision after failed femoral components may
be technically demanding due to loss of peri-prosthetic bone. This retrospective
study evaluated the long-term results of
Aims.
Aims.
Aims. Although good clinical outcomes have been reported for monolithic tapered, fluted, titanium stems (TFTS), early results showed high rates of subsidence. Advances in stem design may mitigate these concerns. This study reports on the use of a current monolithic TFTS for a variety of indications. Methods. A multi-institutional retrospective study of all consecutive total hip arthroplasty (THA) and revision total hip arthroplasty (rTHA) patients who received the monolithic TFTS was conducted. Surgery was performed by eight fellowship-trained arthroplasty surgeons at four institutions. A total of 157 hips in 153 patients at a mean follow-up of 11.6 months (SD7.8) were included. Mean patient age at the time of surgery was 67.4 years (SD 13.3) and mean body mass index (BMI) was 28.9 kg/m. 2. (SD 6.5). Outcomes included intraoperative complications, one-year all-cause re-revisions, and subsidence at postoperative time intervals (two weeks, six weeks, six months, nine months, and one year). Results. There were eight intraoperative complications (4.9%), six of which were intraoperative fractures; none occurred during stem insertion. Six hips (3.7%) underwent re-revision within one year; only one procedure involved removal of the prosthesis due to infection. Mean total subsidence at latest follow-up was 1.64 mm (SD 2.47). Overall, 17 of 144 stems (11.8%) on which measurements could be performed had >5 mm of subsidence, and 3/144 (2.1%) had >10 mm of subsidence within one year. A univariate regression analysis found that additional subsidence after three months was minimal. A multivariate regression analysis found that subsidence was not significantly associated with periprosthetic fracture as an indication for surgery, the presence of an extended trochanteric osteotomy (ETO), Paprosky classification of femoral bone loss, stem length, or type of procedure performed (i.e. full revision vs conversion/primary). Conclusion. Advances in implant design, improved trials, a range of stem lengths and diameters, and high offset options mitigate concerns of early subsidence and dislocation with monolithic TFTS, making them a valuable option for
Endoprosthetic reconstruction following resection of 31 tumours of the proximal femur in 30 patients was performed using a Wagner SL
Aims. Revision total hip arthroplasty in patients with Vancouver type B3 fractures with Paprosky type IIIA, IIIB, and IV femoral defects are difficult to treat. One option for Paprovsky type IIIB and IV defects involves modular cementless, tapered,
Cement-in-cement revision of the femoral component represents a widely practised technique for a variety of indications in revision total hip arthroplasty. In this study, we compare the clinical and radiological outcomes of two polished tapered femoral components. From our prospectively collated database, we identified all patients undergoing cement-in-cement revision from January 2005 to January 2013 who had a minimum of two years' follow-up. All cases were performed by the senior author using either an Exeter short revision stem or the C-Stem AMT high offset No. 1 prosthesis. Patients were followed-up annually with clinical and radiological assessment.Aims
Methods
Femoral impaction bone grafting was first developed in 1987 using
morselised cancellous bone graft impacted into the femoral canal
in combination with a cemented, tapered, polished stem. We describe
the evolution of this technique and instrumentation since that time. Between 1987 and 2005, 705 revision total hip arthroplasties
(56 bilateral) were performed with femoral impaction grafting using
a cemented femoral stem. All surviving patients were prospectively
followed for a mean of 14.7 years (9.8 to 28.3) with no loss to
follow-up. By the time of the final review, 404 patients had died.Aims
Patients and Methods
We present the clinical and radiological results at a minimum
follow-up of five years for patients who have undergone multiple
cement-in-cement revisions of their femoral component at revision
total hip arthroplasty (THA). We reviewed the outcome on a consecutive series of 24 patients
(10 men, 14 women) (51 procedures) who underwent more than one cement-in-cement
revision of the same femoral component. The mean age of the patients was
67.5 years (36 to 92) at final follow-up. Function was assessed using the original Harris hip score (HHS),
Oxford Hip Score (OHS) and the Merle D’Aubigné Postel score (MDP).Aims
Patients and Methods
We have tested the axial and torsional stability of femoral components after revision arthroplasty in a cadaver model, using impacted morsellised cancellous graft and cement. Each one of six matched pairs of fresh frozen human femora had either a primary or a revision prosthesis cemented in place. For the ‘revision’ experiments, all cancellous bone was removed from the proximal femur which was then over-reamed to create a smooth-walled cortical shell. An MTS servohydraulic test frame was used to apply axial and torsional loads to each specimen through the prosthetic femoral heads with the femur submerged in isotonic saline solution at 37°C. The mean subsidence was 0.27 ± 0.17 mm for the primary and 0.52 ± 0.30 mm for the revision groups. The difference was statistically significant (p <
0.025), but the mean subsidence was <
1 mm in both groups. The mean maximum torque before failure was 42.9 ± 26.9 N-m for the primary and 34.8± 20.7 N-m for the revision groups. This difference was not statistically significant (p >
0.015). Based on our results we suggest that revision of the femoral component using morsellised cancellous graft followed by cementing with a collarless prosthesis with a polished tapered stem restores the integrity of the proximal femur and provides immediate stability of the implant.
The August 2023 Hip & Pelvis Roundup. 360. looks at: Using machine learning to predict venous thromboembolism and major bleeding events following total joint arthroplasty; Antibiotic length in revision total hip arthroplasty; Preoperative colonization and worse outcomes; Short stem cemented total hip arthroplasty; What are the outcomes of one- versus two-stage revisions in the UK?; To cement or not to cement? The best approach in hemiarthroplasty; Similar re-revisions in cemented and cementless
The April 2024 Hip & Pelvis Roundup. 360. looks at: Impaction bone grafting for
Aims. The aim of this study was to report the long-term follow-up of cemented short Exeter femoral components when used in primary total hip arthroplasty (THA). Methods. We included all primary 394 THAs with a cemented short Exeter femoral component (≤ 125 mm) used in our tertiary referral centre between October 1993 and December 2021. A total of 83 patients (21%) were male. The median age of the patients at the time of surgery was 42 years (interquartile range (IQR) 30 to 55). The main indication for THA was a childhood hip disease (202; 51%). The median follow-up was 6.7 years (IQR 3.1 to 11.0). Kaplan-Meier survival analyses were performed to determine the rates of survival with
Aims. There are limited long-term studies reporting on outcomes of the Zimmer Modular Revision (ZMR) stem, and concerns remain regarding failure. Our primary aim was to determine long-term survival free from all-cause revision and stem-related failure for this modular revision stem in revision total hip arthroplasty (THA). Secondary aims included evaluating radiological and functional outcomes. Methods. We retrospectively identified all patients in our institutional database who underwent revision THA using the ZMR system from January 2000 to December 2007. We included 106 patients (108 hips) with a mean follow-up of 14.5 years (2.3 to 22.3). Mean patient age was 69.2 years (37.0 to 89.4), and 51.9% were female (n = 55). Indications for index revision included aseptic loosening (73.1%), infection (16.7%), fracture (9.3%), and stem fracture (0.9%). Kaplan-Meier analysis was used to determine the all-cause and stem-related failure revision-free survival. At most recent follow-up, Oxford Hip Scores (OHS) were collected, and radiological stem stability was determined using the Engh classification. Results. A total of 17 hips (15.7%) underwent re-revision of any component. Indications for re-revision were stem failure (35.3%; n = 6), infection (29.4%; n = 5), instability (29.4%; n = 5), and acetabular aseptic loosening (5.9%; n = 1). The five- and 15-year all-cause survival was 89.7% (95% confidence interval (CI) 86.7 to 92.7) and 83.3% (95% CI 79.6 to 87.0), respectively. There were six re-revisions (5.6%) for stem failure; five for stem fracture and one for aseptic loosening. The five- and 15-year survival free from stem-related failure was 97.2% (95% CI 95.6 to 98.8) and 94.0% (95% CI 91.6 to 96.4), respectively. At final follow-up, the mean OHS was 36.9 (8.0 to 48.0) and 95.7% (n = 66) of surviving modular revision stems were well-fixed in available radiographs. Conclusion.
Aims. This study reports the results of 38 total hip arthroplasties (THAs) in 33 patients aged less than 50 years, using the JRI Furlong hydroxyapatite ceramic (HAC)-coated femoral component. Methods. We describe the survival, radiological, and functional outcomes of 33 patients (38 THAs) at a mean follow-up of 27 years (25 to 32) between 1988 and 2018. Results. Of the surviving 30 patients (34 THAs), there were four periprosthetic fractures: one underwent
Aims. We have evaluated the survivorship, outcomes, and failures of an interlocking, reconstruction-mode stem-sideplate implant used to preserve the native hip joint and achieve proximal fixation when there is little residual femur during large endoprosthetic reconstruction of the distal femur. Methods. A total of 14 patients underwent primary or revision reconstruction of a large femoral defect with a short remaining proximal femur using an interlocking, reconstruction-mode stem-sideplate for fixation after oncological distal femoral and diaphyseal resections. The implant was attached to a standard endoprosthetic reconstruction system. The implant was attached to a standard endoprosthetic reconstruction system. None of the
We report the clinical and radiographic outcomes
of 208 consecutive femoral revision arthroplasties performed in 202
patients (119 women, 83 men) between March 1991 and December 2007
using the X-change
The purpose of this study is to report our updated results at a minimum follow-up of 30 years using a first generation uncemented tapered femoral component in primary total hip arthroplasty (THA). The original cohort consisted of 145 consecutive THAs performed by a single surgeon in 138 patients. A total of 37 patients (40 hips) survived a minimum of 30 years, and are the focus of this review. The femoral component used in all cases was a first-generation Taperloc with a non-modular 28 mm femoral head. Clinical follow-up at a minimum of 30 years was obtained on every living patient. Radiological follow-up at 30 years was obtained on all but four.Aims
Methods