We aimed to obtain anthropometric data on Korean
knees and to compare these with data on commonly available total
knee arthroplasties (TKAs). The dimensions of the femora and tibiae
of 1168 knees were measured intra-operatively. The femoral components
were found to show a tendency toward mediolateral (ML) under-coverage
in small femurs and ML overhang in the large femurs. The ML under-coverage
was most prominent for the small prostheses. The ML/anteroposterior
(ML/AP) ratio of Korean tibiae was greater than that of tibial components. This study shows that, for different reasons, current TKAs do
not provide a reasonable fit for small or large Korean knees, and
that the ‘gender-specific’ and ‘stature-specific’ components help
for large Korean femurs but offer less satisfactory fits for small
femurs. Specific modifications of prostheses are needed for Asian
knees.