Advertisement for orthosearch.org.uk
Results 1 - 20 of 29
Results per page:
The Bone & Joint Journal
Vol. 97-B, Issue 6 | Pages 786 - 792
1 Jun 2015
Hutt JRB Farhadnia P Massé V Lavigne M Vendittoli P

This randomised trial evaluated the outcome of a single design of unicompartmental arthroplasty of the knee (UKA) with either a cemented all-polyethylene or a metal-backed modular tibial component. A total of 63 knees in 45 patients (17 male, 28 female) were included, 27 in the all-polyethylene group and 36 in the metal-backed group. The mean age was 57.9 years (39.6 to 76.9). At a mean follow-up of 6.4 years (5 to 9.9), 11 all-polyethylene components (41%) were revised (at a mean of 5.8 years; 1.4 to 8.0) post-operatively and two metal-backed components were revised (at one and five years). One revision in both groups was for unexplained pain, one in the metal-backed group was for progression of osteoarthritis. The others in the all-polyethylene group were for aseptic loosening. The survivorship at seven years calculated by the Kaplan–Meier method for the all-polyethylene group was 56.5% (95% CI 31.9 to 75.2, number at risk 7) and for the metal-backed group was 93.8% (95% CI 77.3 to 98.4, number at risk 16) This difference was statistically significant (p <  0.001). At the most recent follow-up, significantly better mean Western Ontario and McMaster Universities Arthritis Index Scores were found in the all-polyethylene group (13.4 vs 23.0, p = 0.03) but there was no difference in the mean Knee injury and Osteoarthritis Outcome scores (68.8; 41.4 to 99.0 vs 62.6; 24.0 to 100.0), p = 0.36). There were no significant differences for range of movement (p = 0.36) or satisfaction (p = 0.23).

This randomised study demonstrates that all-polyethylene components in this design of fixed bearing UKA had unsatisfactory results with significantly higher rates of failure before ten years compared with the metal-back components.

Cite this article: Bone Joint J 2015;97-B:786–92.


Bone & Joint Research
Vol. 13, Issue 5 | Pages 201 - 213
1 May 2024
Hamoodi Z Gehringer CK Bull LM Hughes T Kearsley-Fleet L Sergeant JC Watts AC

Aims

The aims of this study were to identify and evaluate the current literature examining the prognostic factors which are associated with failure of total elbow arthroplasty (TEA).

Methods

Electronic literature searches were conducted using MEDLINE, Embase, PubMed, and Cochrane. All studies reporting prognostic estimates for factors associated with the revision of a primary TEA were included. The risk of bias was assessed using the Quality In Prognosis Studies (QUIPS) tool, and the quality of evidence was assessed using the modified Grading of Recommendations, Assessment, Development, and Evaluations (GRADE) framework. Due to low quality of the evidence and the heterogeneous nature of the studies, a narrative synthesis was used.


Bone & Joint Research
Vol. 9, Issue 11 | Pages 768 - 777
2 Nov 2020
Huang C Lu Y Hsu L Liau J Chang T Huang C

Aims. The material and design of knee components can have a considerable effect on the contact characteristics of the tibial post. This study aimed to analyze the stress distribution on the tibial post when using different grades of polyethylene for the tibial inserts. In addition, the contact properties of fixed-bearing and mobile-bearing inserts were evaluated. Methods. Three different grades of polyethylene were compared in this study; conventional ultra high molecular weight polyethylene (UHMWPE), highly cross-linked polyethylene (HXLPE), and vitamin E-stabilized polyethylene (VEPE). In addition, tibial baseplates with a fixed-bearing and a mobile-bearing insert were evaluated to understand differences in the contact properties. The inserts were implanted in neutral alignment and with a 10° internal malrotation. The contact stress, von Mises stress, and equivalent plastic strain (PEEQ) on the tibial posts were extracted for comparison. Results. The stress and strain on the tibial post for the three polyethylenes greatly increased when the insert was placed in malrotation, showing a 38% to 56% increase in von Mises stress and a 335% to 434% increase in PEEQ. The VEPE insert had the lowest PEEQ among the three materials. The mobile-bearing design exhibited a lower increase in stress and strain around the tibial posts than the fixed-bearing design. Conclusion. Using VEPE for the tibial component potentially eliminates the risk of material permanent deformation. The mobile-bearing insert can help to avoid a dramatic increase in plastic strain around the tibial post in cases of malrotation. The mobility allows the pressure to be distributed on the tibial post and demonstrated lower stresses with all three polyethylenes simulated. Cite this article: Bone Joint Res 2020;9(11):768–777


The Bone & Joint Journal
Vol. 99-B, Issue 7 | Pages 880 - 886
1 Jul 2017
Mohaddes M Shareghi B Kärrholm J

Aims. The aim of this study was to compare the incidence of aseptic loosening after the use of a cemented acetabular component and a Trabecular Metal (TM) acetabular component (Zimmer Inc., Warsaw, Indiana) at acetabular revision with bone impaction grafting. Patients and Methods. A total of 42 patients were included in the study. Patients were randomised to receive an all- polyethylene cemented acetabular component (n = 19) or a TM component (n = 23). Radiostereometric analysis and conventional radiographic examinations were performed regularly up to two years post-operatively or until further revision. Results. The proximal migration was significantly higher in the cemented group. At two years, the median proximal migration was 1.45 mm and 0.25 mm in the cemented and TM groups, respectively (p = 0.02). One cemented component was revised due to dislocation. There were no revisions in the TM group. Conclusion. Lower proximal migration in the TM group suggests that this design might be associated with a lower risk of aseptic loosening in the long term compared with an all polyethylene cemented component. Longer follow-up is required to confirm the clinical advantages of using this component at acetabular revision. Cite this article: Bone Joint J 2017;99-B:880–6


The Bone & Joint Journal
Vol. 103-B, Issue 10 | Pages 1555 - 1560
4 Oct 2021
Phillips JRA Tucker K

Aims

Knee arthroplasty surgery is a highly effective treatment for arthritis and disorders of the knee. There are a wide variety of implant brands and types of knee arthroplasty available to surgeons. As a result of a number of highly publicized failures, arthroplasty surgery is highly regulated in the UK and many other countries through national registries, introduced to monitor implant performance, surgeons, and hospitals. With time, the options available within many brand portfolios have grown, with alternative tibial or femoral components, tibial insert materials, or shapes and patella resurfacings. In this study we have investigated the effect of the expansion of implant brand portfolios and where there may be a lack of transparency around a brand name. We also aimed to establish the potential numbers of compatible implant construct combinations.

Methods

Hypothetical implant brand portfolios were proposed, and the number of compatible implant construct combinations was calculated.


Bone & Joint 360
Vol. 10, Issue 3 | Pages 20 - 23
1 Jun 2021


Bone & Joint Open
Vol. 2, Issue 5 | Pages 278 - 292
3 May 2021
Miyamoto S Iida S Suzuki C Nakatani T Kawarai Y Nakamura J Orita S Ohtori S

Aims

The main aims were to identify risk factors predictive of a radiolucent line (RLL) around the acetabular component with an interface bioactive bone cement (IBBC) technique in the first year after THA, and evaluate whether these risk factors influence the development of RLLs at five and ten years after THA.

Methods

A retrospective review was undertaken of 980 primary cemented THAs in 876 patients using cemented acetabular components with the IBBC technique. The outcome variable was any RLLs that could be observed around the acetabular component at the first year after THA. Univariate analyses with univariate logistic regression and multivariate analyses with exact logistic regression were performed to identify risk factors for any RLLs based on radiological classification of hip osteoarthritis.


The Bone & Joint Journal
Vol. 102-B, Issue 7 | Pages 852 - 860
1 Jul 2020
Zamora T Garbuz DS Greidanus NV Masri BA

Aims

Our objective is to describe our early and mid-term results with the use of a new simple primary knee prosthesis as an articulating spacer in planned two-stage management for infected knee arthroplasty. As a second objective, we compared outcomes between the group with a retained first stage and those with a complete two-stage revision.

Methods

We included 47 patients (48 knees) with positive criteria for infection, with a minimum two-year follow-up, in which a two-stage approach with an articulating spacer with new implants was used. Patients with infection control, and a stable and functional knee were allowed to retain the initial first-stage components. Outcomes recorded included: infection control rate, reoperations, final range of motion (ROM), and quality of life assessment (QoL) including Western Ontario and McMaster Universities osteoarthritis index, Knee Injury and Osteoarthritis Outcome Score, Oxford Knee Score, 12-Item Short-Form Health Survey questionnaire, and University of California Los Angeles (UCLA) activity score and satisfaction score. These outcomes were evaluated and compared to additional cohorts of patients with retained first-stage interventions and those with a complete two-stage revision. Mean follow-up was 3.7 years (2.0 to 6.5).


The Journal of Bone & Joint Surgery British Volume
Vol. 71-B, Issue 5 | Pages 798 - 803
1 Nov 1989
Scuderi G Insall J Windsor R Moran M

The survivorship method of analysis has been used to compare the failure rate and overall success of 1,430 cemented primary total knee arthroplasties performed at The Hospital for Special Surgery over a 15-year period. There were 224 total condylar prostheses with a polyethylene tibia, 289 of the posterior stabilised type with an all polyethylene tibia, and 917 posterior stabilised with a metal-backed tibial component. There were 12 failures in the total condylar series, giving an average annual failure rate of 0.65% and a 15-year success rate of 90.56%. The posterior stabilised prosthesis with a polyethylene tibia showed an average annual failure rate of 0.27% and a 10-year success rate of 97.34%, and this prosthesis with a metal-backed tibial component gave an annual failure rate of 0.19% and a seven-year success rate of 98.75%. The overall survival rate was not influenced by sex or age, diagnosis or the percentage of ideal body weight. No metal-backed tibial components have yet needed revision for loosening. It seems that infection will be the major cause of failure


Bone & Joint Research
Vol. 9, Issue 1 | Pages 15 - 22
1 Jan 2020
Clement ND Bell A Simpson P Macpherson G Patton JT Hamilton DF

Aims

The primary aim of the study was to compare the knee-specific functional outcome of robotic unicompartmental knee arthroplasty (rUKA) with manual total knee arthroplasty (mTKA) for the management of isolated medial compartment osteoarthritis. Secondary aims were to compare length of hospital stay, general health improvement, and satisfaction between rUKA and mTKA.

Methods

A powered (1:3 ratio) cohort study was performed. A total of 30 patients undergoing rUKA were propensity score matched to 90 patients undergoing mTKA for isolated medial compartment arthritis. Patients were matched for age, sex, body mass index (BMI), and preoperative function. The Oxford Knee Score (OKS) and EuroQol five-dimension questionnaire (EQ-5D) were collected preoperatively and six months postoperatively. The Forgotten Joint Score (FJS) and patient satisfaction were collected six months postoperatively. Length of hospital stay was also recorded.


The Bone & Joint Journal
Vol. 102-B, Issue 1 | Pages 108 - 116
1 Jan 2020
Burger JA Kleeblad LJ Laas N Pearle AD

Aims

Limited evidence is available on mid-term outcomes of robotic-arm assisted (RA) partial knee arthroplasty (PKA). Therefore, the purpose of this study was to evaluate mid-term survivorship, modes of failure, and patient-reported outcomes of RA PKA.

Methods

A retrospective review of patients who underwent RA PKA between June 2007 and August 2016 was performed. Patients received a fixed-bearing medial or lateral unicompartmental knee arthroplasty (UKA), patellofemoral arthroplasty (PFA), or bicompartmental knee arthroplasty (BiKA; PFA plus medial UKA). All patients completed a questionnaire regarding revision surgery, reoperations, and level of satisfaction. Knee Injury and Osteoarthritis Outcome Scores (KOOS) were assessed using the KOOS for Joint Replacement Junior survey.


The Bone & Joint Journal
Vol. 102-B, Issue 4 | Pages 414 - 422
1 Apr 2020
Gwynne-Jones DP Gray AR

Aims

To compare long-term survival of all-cemented and hybrid total hip arthroplasty (THA) using the Exeter Universal stem.

Methods

Details of 1,086 THAs performed between 1999 and 2005 using the Exeter stem and either a cemented (632) or uncemented acetabular component (454) were collected from local records and the New Zealand Joint Registry. A competing risks regression survival analysis was performed with death as the competing risk with adjustments made for age, sex, approach, and bearing.


Bone & Joint 360
Vol. 7, Issue 5 | Pages 2 - 7
1 Oct 2018
Palan J Bloch BV Shannak O James P


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 8 | Pages 1106 - 1113
1 Aug 2008
Richards L Brown C Stone MH Fisher J Ingham E Tipper JL

Nanometre-sized particles of ultra-high molecular weight polyethylene have been identified in the lubricants retrieved from hip simulators. Tissue samples were taken from seven failed Charnley total hip replacements, digested using strong alkali and analysed using high-resolution field emission gun-scanning electron microscopy to determine whether nanometre-sized particles of polyethylene debris were generated in vivo. A randomised method of analysis was used to quantify and characterise all the polyethylene particles isolated. We isolated nanometre-sized particles from the retrieved tissue samples. The smallest identified was 30 nm and the majority were in the 0.1 μm to 0.99 μm size range. Particles in the 1.0 μm to 9.99 μm size range represented the highest proportion of the wear volume of the tissue samples, with 35% to 98% of the total wear volume comprised of particles of this size. The number of nanometre-sized particles isolated from the tissues accounted for only a small proportion of the total wear volume. Further work is required to assess the biological response to nanometre-sized polyethylene particles


Bone & Joint Research
Vol. 6, Issue 1 | Pages 22 - 30
1 Jan 2017
Scott CEH Eaton MJ Nutton RW Wade FA Evans SL Pankaj P

Objectives

Up to 40% of unicompartmental knee arthroplasty (UKA) revisions are performed for unexplained pain which may be caused by elevated proximal tibial bone strain. This study investigates the effect of tibial component metal backing and polyethylene thickness on bone strain in a cemented fixed-bearing medial UKA using a finite element model (FEM) validated experimentally by digital image correlation (DIC) and acoustic emission (AE).

Materials and Methods

A total of ten composite tibias implanted with all-polyethylene (AP) and metal-backed (MB) tibial components were loaded to 2500 N. Cortical strain was measured using DIC and cancellous microdamage using AE. FEMs were created and validated and polyethylene thickness varied from 6 mm to 10 mm. The volume of cancellous bone exposed to < -3000 µε (pathological loading) and < -7000 µε (yield point) minimum principal (compressive) microstrain and > 3000 µε and > 7000 µε maximum principal (tensile) microstrain was computed.


The Bone & Joint Journal
Vol. 99-B, Issue 1_Supple_A | Pages 25 - 30
1 Jan 2017
Waddell BS Della Valle AG

This review summarises the technique of impaction grafting with mesh augmentation for the treatment of uncontained acetabular defects in revision hip arthroplasty.

The ideal acetabular revision should restore bone stock, use a small socket in the near-anatomic position, and provide durable fixation. Impaction bone grafting, which has been in use for over 40 years, offers the ability to achieve these goals in uncontained defects. The precepts of modern, revision impaction grafting are that the segmental or cavitary defects must be supported with a mesh; the contained cavity is filled with vigorously impacted morselised fresh-frozen allograft; and finally, acrylic cement is used to stabilise the graft and provide rigid, long-lasting fixation of the revised acetabular component.

Favourable results have been published with this technique. While having its limitations, it is a viable option to address large acetabular defects in revision arthroplasty.

Cite this article: Bone Joint J 2017;99-B(1 Supple A):25–30.


Bone & Joint 360
Vol. 5, Issue 4 | Pages 1 - 1
1 Aug 2016
Ollivere B


The Bone & Joint Journal
Vol. 98-B, Issue 4 | Pages 437 - 441
1 Apr 2016
Middleton S Toms A

We explored the literature surrounding whether allergy and hypersensitivity has a clinical basis for implant selection in total knee arthroplasty (TKA). In error, the terms hypersensitivity and allergy are often used synonymously. Although a relationship is present, we could not find any evidence of implant failure due to allergy. There is however increasing basic science that suggests a link between loosening and metal ion production. This is not an allergic response but is a potential problem. With a lack of evidence logically there can be no justification to use ‘hypoallergenic’ implants in patients who have pre-existing skin sensitivity to the metals used in TKA.

Cite this article: Bone Joint J 2016;98-B:437–41.


The Bone & Joint Journal
Vol. 98-B, Issue 3 | Pages 307 - 312
1 Mar 2016
Maggs JL Smeatham A Whitehouse SL Charity J Timperley AJ Gie GA

Aims

We report on the outcome of the Exeter Contemporary flanged cemented all-polyethylene acetabular component with a mean follow-up of 12 years (10 to 13.9). This study reviewed 203 hips in 194 patients. 129 hips in 122 patients are still in situ; 66 hips in 64 patients were in patients who died before ten years, and eight hips (eight patients) were revised. Clinical outcome scores were available for 108 hips (104 patients) and radiographs for 103 hips (100 patients).

Patients and Methods

A retrospective review was undertaken of a consecutive series of 203 routine primary cemented total hip arthroplasties (THA) in 194 patients.


The Bone & Joint Journal
Vol. 98-B, Issue 1 | Pages 75 - 80
1 Jan 2016
Merolla G Chin P Sasyniuk TM Paladini P Porcellini G

Aims

We evaluated clinical and radiographic outcomes of total shoulder arthroplasty (TSA) using the second-generation Trabecular Metal (TM) Glenoid component. The first generation component was withdrawn in 2005 after a series of failures were reported. Between 2009 and 2012, 40 consecutive patients with unilateral TSA using the second-generation component were enrolled in this clinical study. The mean age of the patients was 63.8 years (40 to 75) and the mean follow-up was 38 months (24 to 42).

Methods

Patients were evaluated using the Constant score (CS), the American Shoulder and Elbow Surgeons (ASES) score and routine radiographs.