Advertisement for orthosearch.org.uk
Results 1 - 20 of 47
Results per page:
The Bone & Joint Journal
Vol. 103-B, Issue 8 | Pages 1405 - 1413
1 Aug 2021
Ogura K Fujiwara T Morris CD Boland PJ Healey JH

Aims. Rotating-hinge knee prostheses are commonly used to reconstruct the distal femur after resection of a tumour, despite the projected long-term burden of reoperation due to complications. Few studies have examined the factors that influence their failure and none, to our knowledge, have used competing risk models to do so. The purpose of this study was to determine the risk factors for failure of a rotating-hinge knee distal femoral arthroplasty using the Fine-Gray competing risk model. Methods. We retrospectively reviewed 209 consecutive patients who, between 1991 and 2016, had undergone resection of the distal femur for tumour and reconstruction using a rotating-hinge knee prosthesis. The study endpoint was failure of the prosthesis, defined as removal of the femoral component, the tibial component, or the bone-implant fixation; major revision (exchange of the femoral component, tibial component, or the bone-implant fixation); or amputation. Results. Multivariate Fine-Gray regression analyses revealed different hazards for each Henderson failure mode: percentage of femoral resection (p = 0.001) and extent of quadriceps muscle resection (p = 0.005) for overall prosthetic failure; extent of quadriceps muscle resection (p = 0.002) and fixation of femoral component (p = 0.011) for type 2 failure (aseptic loosening); age (p = 0.009) and percentage of femoral resection (p = 0.019) for type 3 failure (mechanical failure); and type of joint resection (p = 0.037) for type 4 (infection) were independent predictors. A bone stem ratio of > 2.5 reliably predicted aseptic loosening. Conclusion. We identified independent risk factors for overall and cause-specific prosthetic failure after rotating-hinge knee distal femoral arthroplasty using a competing risk Fine-Gray model. A bone stem ratio > 2.5 reliably predicts aseptic loosening. An accurate knowledge of the risks of distal femoral arthroplasty after resection for tumour assists surgical planning and managing patient expectations. Cite this article: Bone Joint J 2021;103-B(8):1405–1413


The Bone & Joint Journal
Vol. 106-B, Issue 6 | Pages 565 - 572
1 Jun 2024
Resl M Becker L Steinbrück A Wu Y Perka C

Aims

This study compares the re-revision rate and mortality following septic and aseptic revision hip arthroplasty (rTHA) in registry data, and compares the outcomes to previously reported data.

Methods

This is an observational cohort study using data from the German Arthroplasty Registry (EPRD). A total of 17,842 rTHAs were included, and the rates and cumulative incidence of hip re-revision and mortality following septic and aseptic rTHA were analyzed with seven-year follow-up. The Kaplan-Meier estimates were used to determine the re-revision rate and cumulative probability of mortality following rTHA.


Bone & Joint Open
Vol. 4, Issue 2 | Pages 72 - 78
9 Feb 2023
Kingsbury SR Smith LKK Pinedo-Villanueva R Judge A West R Wright JM Stone MH Conaghan PG

Aims

To review the evidence and reach consensus on recommendations for follow-up after total hip and knee arthroplasty.

Methods

A programme of work was conducted, including: a systematic review of the clinical and cost-effectiveness literature; analysis of routine national datasets to identify pre-, peri-, and postoperative predictors of mid-to-late term revision; prospective data analyses from 560 patients to understand how patients present for revision surgery; qualitative interviews with NHS managers and orthopaedic surgeons; and health economic modelling. Finally, a consensus meeting considered all the work and agreed the final recommendations and research areas.


Bone & Joint Open
Vol. 3, Issue 9 | Pages 733 - 740
21 Sep 2022
Sacchetti F Aston W Pollock R Gikas P Cuomo P Gerrand C

Aims

The proximal tibia (PT) is the anatomical site most frequently affected by primary bone tumours after the distal femur. Reconstruction of the PT remains challenging because of the poor soft-tissue cover and the need to reconstruct the extensor mechanism. Reconstructive techniques include implantation of massive endoprosthesis (megaprosthesis), osteoarticular allografts (OAs), or allograft-prosthesis composites (APCs).

Methods

This was a retrospective analysis of clinical data relating to patients who underwent proximal tibial arthroplasty in our regional bone tumour centre from 2010 to 2018.


The Bone & Joint Journal
Vol. 104-B, Issue 6 | Pages 747 - 757
1 Jun 2022
Liang H Yang Y Guo W Yan L Tang X Li D Qu H Zang J Du Z

Aims

The aim of this study was to investigate the feasibility of application of a 3D-printed megaprosthesis with hemiarthroplasty design for defects of the distal humerus or proximal ulna following tumour resection.

Methods

From June 2018 to January 2020, 13 patients with aggressive or malignant tumours involving the distal humerus (n = 8) or proximal ulna (n = 5) were treated by en bloc resection and reconstruction with a 3D-printed megaprosthesis with hemiarthroplasty, designed in our centre. In this paper, we summarize the baseline and operative data, oncological outcome, complication profiles, and functional status of these patients.


The Journal of Bone & Joint Surgery British Volume
Vol. 75-B, Issue 4 | Pages 572 - 576
1 Jul 1993
Knight D Rymaszewski L Amis A Miller J

Compression testing of cadaver specimens showed that excision of the radial head allowed proximal radial displacement. The insertion of a metallic radial head restored normal mechanics, while a silicone rubber implant did not. We reviewed 31 of 36 comminuted fractures of the radial head, 21 associated with dislocation or ulnar fracture, which had been treated by primary replacement with a Vitallium prosthesis. At a mean follow-up of 4.5 years, there was reliable restoration of stability and prevention of proximal radial migration. There had been no dislocations or prosthetic failures, but two implants had been removed for loosening. The prosthesis is recommended for use as a spacer to stabilise the elbow after severe injuries while the soft tissues heal


The Journal of Bone & Joint Surgery British Volume
Vol. 78-B, Issue 3 | Pages 461 - 467
1 May 1996
Gaine WJ McCreath SW

We assessed 46 Syme’s amputees attending our prosthetic clinics in terms of the clinical and radiological condition of their stumps, their level of function and problems with the prosthesis. Twenty-five were compared with a matching group of 25 transtibial amputees in regard to activity, function and prosthetic behaviour. Function was similar in the two groups, but Syme’s amputees had a higher incidence of prosthetic failure. Overall, Syme’s amputees were pleased with their prostheses and their function. Childhood amputations were associated with fewer long-term problems in terms of function and stump problems. Syme’s amputation is indicated for congenital foot deformities, fibular hemimelia and severe injury to the foot as long as the heel pad remains viable


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 9 | Pages 1143 - 1151
1 Sep 2008
Langton DJ Jameson SS Joyce TJ Webb J Nargol AVF

Increased concentrations of metal ions after metal-on-metal resurfacing arthroplasty of the hip remain a concern. Although there has been no proven link to long-term health problems or early prosthetic failure, variables associated with high metal ion concentrations should be identified and, if possible, corrected. Our study provides data on metal ion levels from a series of 76 consecutive patients (76 hips) after resurfacing arthroplasty with the Articular Surface Replacement. Chromium and cobalt ion concentrations in the whole blood of patients with smaller (≤ 51 mm) femoral components were significantly higher than in those with the larger (≥ 53 mm) components (p < 0.01). Ion concentrations in the former group were significantly related to the inclination (p = 0.01) and anteversion (p = 0.01) of the acetabular component. The same relationships were not significant in the patients with larger femoral components (p = 0.61 and p = 0.49, respectively). Accurate positioning of the acetabular component intra-operatively is essential in order to reduce the concentration of metal ions in the blood after hip resurfacing arthroplasty with the Articular Surface Replacement implant


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 12 | Pages 1653 - 1657
1 Dec 2005
Wedin R Bauer HCF

We report positive and negative factors associated with the most commonly-used methods of reconstruction after pathological fracture of the proximal femur. The study was based on 142 patients treated surgically for 146 metastatic lesions between 1996 and 2003. The local rate of failure was 10.3% (15 of 146). Of 37 operations involving osteosynthetic devices, six failed (16.2%) compared with nine (8.3%) in 109 operations involving endoprostheses. Of nine cases of prosthetic failure, four were due to periprosthetic fractures and three to recurrent dislocation. In the osteosynthesis group, three (13.6%) of 22 reconstruction nails failed. The two-year risk of re-operation after any type of osteosynthesis was 0.35 compared with 0.18 after any type of endoprosthetic reconstruction (p = 0.07). Endoprosthetic reconstructions are preferable to the use of reconstruction nails and other osteosynthetic devices for the treatment of metastatic lesions in the proximal third of the femur


The Journal of Bone & Joint Surgery British Volume
Vol. 72-B, Issue 5 | Pages 784 - 787
1 Sep 1990
Franzen H Nilsson L Stromqvist B Johnsson R Herrlin K

We studied the rate of revision in 84 consecutive total hip replacements performed for failed osteosynthesis of femoral neck fractures and compared it with that for primary arthroplasty for osteoarthritis. The age and sex adjusted risk of prosthetic failure was 2.5 times higher after failure of fixation, but all the excess risk was in patients over 70 years of age. There were radiographic signs of loosening of the femoral component at five to 12 years after secondary arthroplasty in six of 33 survivors. In general, the results of secondary replacement were no worse than those obtained after primary arthroplasty for femoral neck fracture. We consider that internal fixation should be the primary procedure: total hip replacement is a safe secondary procedure when osteosynthesis fails


The Journal of Bone & Joint Surgery British Volume
Vol. 79-B, Issue 6 | Pages 927 - 937
1 Nov 1997
Schindler OS Cannon SR Briggs TWR Blunn GW

The use of extendible distal femoral replacements is a relatively new treatment alternative for malignant bone tumours in growing individuals. Although their appearance was widely appreciated, questions about functional practicality and longevity remain unclear. With longer follow-up, advantages of immediate functional restoration and beneficial psychological aspects seem to be overshadowed by an increase in complications such as aseptic loosening, infection or prosthetic failure. We have reviewed 18 children with such tumours who were treated between 1983 and 1990 by custom-made Stanmore extendible distal femoral replacements. Four died from metastatic disease within 2.5 years of operation and two required amputation for local recurrence or chronic infection. The remaining 12 patients were followed for a mean of 8.7 years (6 to 13.2). A mean total lengthening of 5.2 cm was achieved, requiring, on average, 4.3 operations. Using the Musculoskeletal Tumor Society rating score the functional result at review was, on average, 77% of the expected normal function, with seven patients achieving ≥ 80%. Revision of the prosthesis was required in ten patients, in six for aseptic loosening, at a mean of 6.2 years after the initial procedure


Bone & Joint Open
Vol. 2, Issue 9 | Pages 721 - 727
1 Sep 2021
Zargaran A Zargaran D Trompeter AJ

Aims

Orthopaedic infection is a potentially serious complication of elective and emergency trauma and orthopaedic procedures, with a high associated burden of morbidity and cost. Optimization of vitamin D levels has been postulated to be beneficial in the prevention of orthopaedic infection. This study explores the role of vitamin D in orthopaedic infection through a systematic review of available evidence.

Methods

A comprehensive search was conducted on databases including Medline and Embase, as well as grey literature such as Google Scholar and The World Health Organization Database. Pooled analysis with weighted means was undertaken.


Bone & Joint 360
Vol. 9, Issue 6 | Pages 5 - 11
1 Dec 2020
Sharma V Turmezei T Wain J McNamara I


The Bone & Joint Journal
Vol. 102-B, Issue 11 | Pages 1549 - 1554
1 Nov 2020
Schwartz AM Farley KX Boden SH Wilson JM Daly CA Gottschalk MB Wagner ER

Aims

The impact of tobacco use on readmission and medical and surgical complications has been documented in hip and knee arthroplasty. However, there remains little information about the effect of smoking on the outcome after total shoulder arthroplasty (TSA). We hypothesized that active smokers are at an increased risk of poor medical and surgial outcomes after TSA.

Methods

Data for patients who underwent arthroplasty of the shoulder in the USA between January 2011 and December 2015 were obtained from the National Readmission Database, and 90-day readmissions and complications were documented using validated coding methods. Multivariate regression analysis was performed to quantify the risk of smoking on the outcome after TSA, while controlling for patient demographics, comorbidities, and hospital-level confounding factors.


Bone & Joint Research
Vol. 9, Issue 9 | Pages 587 - 592
5 Sep 2020
Qin L Li X Wang J Gong X Hu N Huang W

Aims

This study aimed to explore whether serum combined with synovial interleukin-6 (IL-6) measurement can improve the accuracy of prosthetic joint infection (PJI) diagnosis, and to establish the cut-off values of IL-6 in serum and synovial fluid in detecting chronic PJI.

Methods

Patients scheduled to have a revision surgery for indications of chronic infection of knee and hip arthroplasties or aseptic loosening of an implant were prospectively screened before being enrolled into this study. The Musculoskeletal Infection Society (MSIS) definition of PJI was used for the classification of cases as aseptic or infected. Serum CRP, ESR, IL-6, and percentage of polymorphonuclear neutrophils (PMN%) and IL-6 in synovial fluid were analyzed. Statistical tests were performed to compare these biomarkers in the two groups, and receiver operating characteristic (ROC) curves and area under the curve (AUC) were analyzed for each biomarker.


Bone & Joint Open
Vol. 1, Issue 12 | Pages 737 - 742
1 Dec 2020
Mihalič R Zdovc J Brumat P Trebše R

Aims

Synovial fluid white blood cell (WBC) count and percentage of polymorphonuclear cells (%PMN) are elevated at periprosthetic joint infection (PJI). Leucocytes produce different interleukins (IL), including IL-6, so we hypothesized that synovial fluid IL-6 could be a more accurate predictor of PJI than synovial fluid WBC count and %PMN. The main aim of our study was to compare the predictive performance of all three diagnostic tests in the detection of PJI.

Methods

Patients undergoing total hip or knee revision surgery were included. In the perioperative assessment phase, synovial fluid WBC count, %PMN, and IL-6 concentration were measured. Patients were labeled as positive or negative according to the predefined cut-off values for IL-6 and WBC count with %PMN. Intraoperative samples for microbiological and histopathological analysis were obtained. PJI was defined as the presence of sinus tract, inflammation in histopathological samples, and growth of the same microorganism in a minimum of two or more samples out of at least four taken.


The Bone & Joint Journal
Vol. 101-B, Issue 9 | Pages 1144 - 1150
1 Sep 2019
Tsuda Y Fujiwara T Sree D Stevenson JD Evans S Abudu A

Aims

The aim of this study was to report the results of custom-made endoprostheses with extracortical plates plus or minus a short, intramedullary stem aimed at preserving the physis after resection of bone sarcomas in children.

Patients and Methods

Between 2007 and 2017, 18 children aged less than 16 years old who underwent resection of bone sarcomas, leaving ≤ 5 cm of bone from the physis, and reconstruction with a custom-made endoprosthesis were reviewed. Median follow-up was 67 months (interquartile range 45 to 91). The tumours were located in the femur in 11 patients, proximal humerus in six, and proximal tibia in one.


The Bone & Joint Journal
Vol. 101-B, Issue 6 | Pages 724 - 731
1 Jun 2019
Bernthal NM Upfill-Brown A Burke ZDC Ishmael CR Hsiue P Hori K Hornicek F Eckardt JJ

Aims

Aseptic loosening is a major cause of failure in cemented endoprosthetic reconstructions. This paper presents the long-term outcomes of a custom-designed cross-pin fixation construct designed to minimize rotational stress and subsequent aseptic loosening in selected patients. The paper will also examine the long-term survivorship and modes of failure when using this technique.

Patients and Methods

A review of 658 consecutive, prospectively collected cemented endoprosthetic reconstructions for oncological diagnoses at a single centre between 1980 and 2017 was performed. A total of 51 patients were identified with 56 endoprosthetic implants with cross-pin fixation, 21 of which were implanted following primary resection of tumour. Locations included distal femoral (n = 36), proximal femoral (n = 7), intercalary (n = 6), proximal humeral (n = 3), proximal tibial (n = 3), and distal humeral (n = 1).


The Bone & Joint Journal
Vol. 101-B, Issue 5 | Pages 522 - 528
1 May 2019
Medellin MR Fujiwara T Clark R Stevenson JD Parry M Jeys L

Aims

The aim of this study was to evaluate the prosthesis characteristics and associated conditions that may modify the survival of total femoral endoprosthetic replacements (TFEPR).

Patients and Methods

In all, 81 patients treated with TFEPR from 1976 to 2017 were retrospectively evaluated and failures were categorized according to the Henderson classification. There were 38 female patients (47%) and 43 male patients (53%) with a mean age at diagnosis of 43 years (12 to 86). The mean follow-up time was 10.3 years (0 to 31.7). A survival analysis was performed followed by univariate and multivariate Cox regression to identify independent implant survival factors.


The Bone & Joint Journal
Vol. 101-B, Issue 4 | Pages 372 - 377
1 Apr 2019
Zagra L Villa F Cappelletti L Gallazzi E Materazzi G De Vecchi E

Aims

Leucocyte esterase (LE) has been shown to be an accurate marker of prosthetic joint infection (PJI), and has been proposed as an alternative to frozen section (FS) histology for intraoperative diagnosis. In this study, the intraoperative assessment of LE was compared with FS histology for the diagnosis of prosthetic hip infection.

Patients and Methods

A total of 119 patients undergoing revision total hip arthroplasty (THA) between June 2015 and December 2017 were included in the study. There were 56 men and 63 women with a mean age of 66.2 years (27 to 88). Synovial fluid was collected before arthrotomy for the assessment of LE using enzymatic colourimetric strips. Between five and six samples were stained with haematoxylin and eosin for FS histology, and considered suggestive of infection when at least five polymorphonuclear leucocytes were found in five high-power fields.