Aims. There are concerns regarding nail/medullary canal mismatch and initial stability after
Objectives. The paradoxical migration of the femoral neck element (FNE) superomedially against gravity, with respect to the intramedullary component of the cephalomedullary device, is a poorly understood phenomenon increasingly seen in the management of pertrochanteric hip fractures with the intramedullary nail. The aim of this study was to investigate the role of bidirectional loading on the medial migration phenomenon, based on unique wear patterns seen on scanning electron microscopy of retrieved implants suggestive of FNE toggling. Methods. A total of 18 synthetic femurs (Sawbones, Vashon Island, Washington) with comminuted pertrochanteric fractures were divided into three groups (n = 6 per group). Fracture fixation was performed using the Proximal Femoral Nail Antirotation (PFNA) implant (Synthes, Oberdorf, Switzerland; n = 6). Group 1 was subjected to unidirectional compression loading (600 N), with an elastomer (70A durometer) replacing loose fracture fragments to simulate surrounding soft-tissue tensioning. Group 2 was subjected to bidirectional loading (600 N compression loading, 120 N tensile loading), also with the elastomer replacing loose fracture fragments. Group 3 was subjected to bidirectional loading (600 N compression loading, 120 N tensile loading) without the elastomer. All constructs were tested at 2 Hz for 5000 cycles or until cut-out occurred. The medial migration distance (MMD) was recorded at the end of the testing cycles. Results. The MMDs for Groups 1, 2, and 3 were 1.02 mm, 6.27 mm, and 5.44 mm respectively, with reliable reproduction of medial migration seen in all groups. Bidirectional loading groups showed significantly higher MMDs compared with the unidirectional loading group (p < 0.01). Conclusion. Our results demonstrate significant contributions of bidirectional cyclic loading to the medial migration phenomenon in
Aims.
Objectives. Intramedullary fixation is considered the most stable treatment for pertrochanteric fractures of the proximal femur and cut-out is one of the most frequent mechanical complications. In order to determine the role of clinical variables and radiological parameters in predicting the risk of this complication, we analysed the data pertaining to a group of patients recruited over the course of six years. Methods. A total of 571 patients were included in this study, which analysed the incidence of cut-out in relation to several clinical variables: age; gender; the AO Foundation and Orthopaedic Trauma Association classification system (AO/OTA); type of nail; cervical-diaphyseal angle; surgical wait times; anti-osteoporotic medication; complete post-operative weight bearing; and radiological parameters (namely the lag-screw position with respect to the femoral head, the Cleveland system, the tip-apex distance (TAD), and the calcar-referenced tip-apex distance (CalTAD)). Results. The incidence of cut-out across the sample was 5.6%, with a higher incidence in female patients. A significantly higher risk of this complication was correlated with lag-screw tip positioning in the upper part of the femoral head in the anteroposterior radiological view, posterior in the latero-lateral radiological view, and in the Cleveland peripheral zones. The tip-apex distance and the calcar-referenced tip-apex distance were found to be highly significant predictors of the risk of cut-out at cut-offs of 30.7 mm and 37.3 mm, respectively, but the former appeared more reliable than the latter in predicting the occurrence of this complication. Conclusion. The tip-apex distance remains the most accurate predictor of cut-out, which is significantly greater above a cut-off of 30.7 mm. Cite this article: G. Caruso, M. Bonomo, G. Valpiani, G. Salvatori, A. Gildone, V. Lorusso, L. Massari. A six-year retrospective analysis of cut-out risk predictors in
Aims. To compare the outcomes for trochanteric fractures treated with
a sliding hip screw (SHS) or a
Aims. Restoration of proximal medial femoral support is the keystone in the treatment of intertrochanteric fractures. None of the available implants are effective in constructing the medial femoral support. Medial sustainable nail (MSN-II) is a novel
The purpose of this study was to identify factors
that predict implant cut-out after
The February 2014 Trauma Roundup. 360 . looks at: predicting nonunion; compartment Syndrome; octogenarian RTCs; does HIV status affect decision making in open tibial fractures?; flap timing and related complications; proximal humeral fractures under the spotlight; restoration of hip architecture with bipolar hemiarthroplasty in the elderly; and short versus long
Aims. We reviewed all patients who sustained a fracture of the hip
and were treated in Northern Ireland over a period of 15 years to
identify trends in incidence, the demographics of the patients,
the rates of mortality, the configuration of the fracture and the
choice of implant. Patients and Methods. Since 01 January 2001 data about every fracture of the hip sustained
in an adult have been collected centrally in Northern Ireland. All
adults with such a fracture between 2000 and 2015 were included
in the study. Temporal changes in their demographics, the mode of
treatment, and outcomes including mortality were analysed. Results. The incidence of fractures of the hip, in Northern Ireland, rose
from 54 in 100 000 in 2000 to 86 in 100 000 in 2015. If these trends
continue, we predict this rising to 128 in 100 000 in 2030. We found
that these patients are becoming older and increasingly frail, as
assessed by the American Association of Anesthesiology grade. Complex extracapsular
fractures have become more common since 2009, which may explain
the increased use of
Fractures of the proximal femur are one of the
greatest challenges facing the medical community, constituting a
heavy socioeconomic burden worldwide. Controversy exists regarding
the optimal treatment for patients with unstable trochanteric proximal
femoral fractures. The recognised treatment alternatives are extramedullary
fixation usually with a sliding hip screw and intramedullary fixation
with a
The use of fluoroscopy in orthopaedic surgery creates risk of radiation exposure to surgeons. Appropriate personal protective equipment (PPE) can help mitigate this. The primary aim of this study was to assess if current radiation protection in orthopaedic trauma is safe. The secondary aims were to describe normative data of radiation exposure during common orthopaedic procedures, evaluate ways to improve any deficits in protection, and validate the use of electronic personal dosimeters (EPDs) in assessing radiation dose in orthopaedic surgery. Radiation exposure to surgeons during common orthopaedic trauma operations was prospectively assessed using EPDs and thermoluminescent dosimeters (TLDs). Normative data for each operation type were calculated and compared to recommended guidelines.Aims
Methods
The December 2024 Trauma Roundup360 looks at: Percutaneous lumbopelvic fixation is effective in the management of unstable transverse sacral fractures; A systematic review on autologous matrix-induced chondrogenesis (AMIC) for chondral knee defects; Stable clinical and radiological outcomes at medium and over five-year follow-up of calcaneus fracture open reduction internal fixation using a sinus tarsi approach; Right or left? It might make a difference; Suprapatellar versus infrapatellar tibial nailing – is there a difference in anterior knee pain and function?; Can patients safely weightbear following ankle fracture fixation?; Anterior-to-posterior or a plate fixation for posterior malleous fractures?; Audio distraction for traction pin insertion: a prospective randomized controlled study; Is intramedullary nailing of femoral diaphyseal fractures in the lateral decubitus position as safe and effective as on a traction table?
The aim of this study was to describe the current pathways of care for patients with a fracture of the hip in five low- and middle-income countries (LMIC) in South Asia (Nepal and Sri Lanka) and Southeast Asia (Malaysia, Thailand, and the Philippines). The World Health Organization Service Availability and Readiness Assessment tool was used to collect data on the care of hip fractures in Malaysia, Thailand, the Philippines, Sri Lanka, and Nepal. Respondents were asked to provide details about the current pathway of care for patients with hip fracture, including pre-hospital transport, time to admission, time to surgery, and time to weightbearing, along with healthcare professionals involved at different stages of care, information on discharge, and patient follow-up.Aims
Methods
The aim of this study was to evaluate the outcomes of a salvage procedure using a 95° angled blade plate for failed osteosynthesis of atypical subtrochanteric femoral fractures associated with the long-term use of bisphosphonates. These were compared with those for failed osteosynthesis of subtrochanteric fractures not associated with bisphosphonate treatment. Between October 2008 and July 2016, 14 patients with failed osteosynthesis of an atypical subtrochanteric femoral fracture were treated with a blade plate (atypical group). Their mean age was 67.8 years (60 to 74); all were female. During the same period, 21 patients with failed osteosynthesis of a typical subtrochanteric fracture underwent restabilization using a blade plate (typical group). Outcome variables included the time of union, postoperative complications, Harris Hip Score, and Sanders functional rating scale.Aims
Patients and Methods
The June 2015 Oncology Roundup360 looks at: Infection in megaprosthesis; Impressive results for mid femoral reconstruction; Revered teaching or old myth? Femoral neck protection in metastatic disease; Megaprosthesis about the knee; Malignant transformation in multiple hereditary exostoses; Fracture of intercalary bone allograft; Comorbidity and outcomes in sarcoma; A worrying turn? Use of denosumab for giant cell tumour of bone
Elective surgery has been severely curtailed as a result of the COVID-19 pandemic. There is little evidence to guide surgeons in assessing what processes should be put in place to restart elective surgery safely in a time of endemic COVID-19 in the community. We used data from a stand-alone hospital admitting and operating on 91 trauma patients. All patients were screened on admission and 100% of patients have been followed-up after discharge to assess outcome.Aims
Methods
We wished to assess the feasibility of a future randomised controlled
trial of parathyroid hormone (PTH) supplements to aid healing of
trochanteric fractures of the hip, by an open label prospective
feasibility and pilot study with a nested qualitative sub study.
This aimed to inform the design of a future powered study comparing
the functional recovery after trochanteric hip fracture in patients
undergoing standard care, We undertook a pilot study comparing the functional recovery
after trochanteric hip fracture in patients 60 years or older, admitted
with a trochanteric hip fracture, and potentially eligible to be
randomised to either standard care or the administration of subcutaneous
PTH for six weeks. Our desired outcomes were functional testing
and measures to assess the feasibility and acceptability of the
study.Aims
Patients and Methods
The aims of this study were to evaluate the incidence of postoperatively restricted weight-bearing and its association with outcome in patients who undergo surgery for a fracture of the hip. Patient aged > 60 years undergoing surgery for a hip fracture were identified in the 2016 National Surgical Quality Improvement Program (NSQIP) Hip Fracture Targeted Procedure Dataset. Analysis of the effect of restricted weight-bearing on adverse events, delirium, infection, transfusion, length of stay, return to the operating theatre, readmission and mortality within 30 days postoperatively were assessed. Multivariate regression analysis was used to adjust for confounding demographic, comorbid and procedural characteristics.Aims
Patients and Methods
Fractures of the distal femur can be challenging to manage and
are on the increase in the elderly osteoporotic population. Management
with casting or bracing can unacceptably limit a patient’s ability
to bear weight, but historically, operative fixation has been associated
with a high rate of re-operation. In this study, we describe the outcomes
of fixation using modern implants within a strategy of early return
to function. All patients treated at our centre with lateral distal femoral
locking plates (LDFLP) between 2009 and 2014 were identified. Fracture
classification and operative information including weight-bearing
status, rates of union, re-operation, failure of implants and mortality
rate, were recorded.Aims
Patients and Methods
Fractures of the proximal femur are a common clinical problem, and a number of orthopaedic devices are available for the treatment of such fractures. The objective of this study was to assess the rotational stability, a common failure predictor, of three different rotational control design philosophies: a screw, a helical blade and a deployable crucifix. Devices were compared in terms of the mechanical work (W) required to rotate the implant by 6° in a bone substitute material. The substitute material used was Sawbones polyurethane foam of three different densities (0.08 g/cm3, 0.16 g/cm3 and 0.24 g/cm3). Each torsion test comprised a steady ramp of 1°/minute up to an angular displacement of 10°.Objectives
Methods
We performed a retrospective review of all patients
admitted to two large University Hospitals in the United Kingdom
over a 24-month period from January 2008 to January 2010 to identify
the incidence of atypical subtrochanteric and femoral shaft fractures
and their relationship to bisphosphonate treatment. Of the 3515 patients
with a fracture of the proximal femur, 156 fractures were in the
subtrochanteric region. There were 251 femoral shaft fractures.
The atypical fracture pattern was seen in 27 patients (7%) with
29 femoral shaft or subtrochanteric fractures. A total of 22 patients
with 24 atypical fractures were receiving bisphosphonate treatment at
the time of fracture. Prodromal pain was present in nine patients
(11 fractures); 11 (50%) of the patients on bisphosphonates suffered
12 spontaneous fractures, and healing of these fractures was delayed
in a number of patients. This large dual-centre review has established
the incidence of atypical femoral fractures at 7% of the study population,
81% of whom had been on bisphosphonate treatment for a mean of 4.6
years (0.04 to 12.1). This study does not advocate any change in the use of bisphosphonates
to prevent fragility fractures but attempts to raise awareness of
this possible problem so symptomatic patients will be appropriately
investigated. However, more work is required to identify the true
extent of this new and possibly increasing problem.
The December 2014 Trauma Roundup360 looks at: infection and temporising external fixation; Vitamin C in distal radial fractures; DRAFFT: Cheap and cheerful Kirschner wires win out; femoral neck fractures not as stable as they might be; displaced sacral fractures give high morbidity and mortality; sanders and calcaneal fractures: a 20-year experience; bleeding and pelvic fractures; optimising timing for acetabular fractures; and tibial plateau fractures.
Most hip fractures treated with modern internal
fixation techniques will heal. However, failures occasionally occur and
require revision procedures. Salvage strategies employed during
revision are based on whether the fixation failure occurs in the
femoral neck, or in the intertrochanteric region. Patient age and
remaining bone stock also influence decision making. For fractures
in young patients, efforts are generally focused on preserving the
native femoral head via osteotomies and repeat internal fixation.
For failures in older patients, some kind of hip replacement is
usually selected. Disuse osteopenia, deformity, bone loss, and stress-risers
from previous internal fixation devices all pose technical challenges
to successful reconstruction. Attention to detail is important in
order to minimise complications. In the majority of cases, good
outcomes have been reported for the various salvage strategies. Cite this article:
Many tumors metastasise to bone, therefore, pathologic
fracture and impending pathologic fractures are common reasons for
orthopedic consultation. Having effective treatment strategies is
important to avoid complications, and relieve pain and preserve
function. Thorough pre-operative evaluation is recommended for medical
optimization and to ensure that the lesion is in fact a metastasis
and not a primary bone malignancy. For impending fractures, various scoring
systems have been proposed to determine the risk of fracture, and
therefore the need for prophylactic stabilisation. Lower score lesions
can often be treated with radiation, while more problematic lesions
may require internal fixation. Intramedullary fixation is generally
preferred due to favorable biomechanics. Arthroplasty may be required
for lesions with massive bony destruction where internal fixation
attempts are likely to fail. Radiation may also be useful postoperatively
to minimise construct failure due to tumor progression.
The August 2013 Trauma Roundup360 looks at: reverse oblique fractures do better with a cephalomedullary device; locking screws confer no advantage in tibial plateau fractures; it’s all about the radius of curvature; radius of curvature revisited; radial head replacement in complex elbow reconstruction; stem cells in early fracture haematoma; heterotrophic ossification in forearms; and Boston in perspective.
The use of two implants to manage concomitant ipsilateral femoral
shaft and proximal femoral fractures has been indicated, but no
studies address the relationship of dynamic hip screw (DHS) side
plate screws and the intramedullary nail where failure might occur
after union. This study compares different implant configurations
in order to investigate bridging the gap between the distal DHS
and tip of the intramedullary nail. A total of 29 left synthetic femora were tested in three groups:
1) gapped short nail (GSN); 2) unicortical short nail (USN), differing
from GSN by the use of two unicortical bridging screws; and 3) bicortical
long nail (BLN), with two angled bicortical and one unicortical bridging
screws. With these findings, five matched-pairs of cadaveric femora
were tested in two groups: 1) unicortical long nail (ULN), with
a longer nail than USN and three bridging unicortical screws; and
2) BLN. Specimens were axially loaded to 22.7 kg (50 lb), and internally
rotated 90°/sec until failure.Objectives
Methods
We report a 30-year-old patient who was involved in a high-velocity road traffic accident and developed a left-sided hemiparesis, which was noted in the post-operative period following bilateral femoral intramedullary nailing. CT scanning of the brain revealed infarcts in the right frontal and parietal lobes in the distribution of the right middle cerebral artery. CT angiography showed occlusion of the right internal carotid artery consistent with internal carotid artery dissection. He was anticoagulated and nine months later was able to walk independently. An awareness of this injury is needed to diagnose blunt trauma to the internal carotid artery. Even in the absence of obvious neck trauma, carotid artery dissection should be suspected in patients with a neurological deficit in the peri-operative period.
The results of the treatment of 31 open femoral fractures (29 patients) with significant bone loss in a single trauma unit were reviewed. A protocol of early soft-tissue and bony debridement was followed by skeletal stabilisation using a locked intramedullary nail or a dynamic condylar plate for diaphyseal and metaphyseal fractures respectively. Soft-tissue closure was obtained within 48 hours then followed, if required, by elective bone grafting with or without exchange nailing. The mean time to union was 51 weeks (20 to 156). The time to union and functional outcome were largely dependent upon the location and extent of the bone loss. It was achieved more rapidly in fractures with wedge defects than in those with segmental bone loss. Fractures with metaphyseal defects healed more rapidly than those of comparable size in the diaphysis. Complications were more common in fractures with greater bone loss, and included stiffness of the knee, malunion and limb-length discrepancy. Based on our findings, we have produced an algorithm for the treatment of these injuries. We conclude that satisfactory results can be achieved in most femoral fractures with bone loss using initial debridement and skeletal stabilisation to maintain length, with further procedures as required.