We biomechanically investigated whether the standard dynamic hip screw (DHS) or the
Introduction. The objective of this study was to determine if a synthetic bone
substitute would provide results similar to bone from osteoporotic
femoral heads during in vitro testing with orthopaedic
implants. If the synthetic material could produce results similar
to those of the osteoporotic bone, it could reduce or eliminate
the need for testing of implants on bone. Methods. Pushout studies were performed with the dynamic hip screw (DHS)
and the
Although the importance of lateral femoral wall
integrity is increasingly being recognised in the treatment of intertrochanteric
fracture, little attention has been put on the development of a
secondary post-operative fracture of the lateral wall. Patients
with post-operative fractures of the lateral wall were reported
to have high rates of re-operation and complication. To date, no
predictors of post-operative lateral wall fracture have been reported.
In this study, we investigated the reliability of lateral wall thickness
as a predictor of lateral wall fracture after dynamic hip screw
(DHS) implantation. A total of 208 patients with AO/OTA 31-A1 and -A2 classified
intertrochanteric fractures who received internal fixation with
a
Fractures of the proximal femur are a common clinical problem, and a number of orthopaedic devices are available for the treatment of such fractures. The objective of this study was to assess the rotational stability, a common failure predictor, of three different rotational control design philosophies: a screw, a helical blade and a deployable crucifix. Devices were compared in terms of the mechanical work (W) required to rotate the implant by 6° in a bone substitute material. The substitute material used was Sawbones polyurethane foam of three different densities (0.08 g/cm3, 0.16 g/cm3 and 0.24 g/cm3). Each torsion test comprised a steady ramp of 1°/minute up to an angular displacement of 10°.Objectives
Methods
Aims. To evaluate if, for orthopaedic trainees, additional cadaveric simulation training or standard training alone yields superior radiological and clinical outcomes in patients undergoing dynamic hip screw (DHS) fixation or hemiarthroplasty for hip fracture. Methods. This was a preliminary, pragmatic, multicentre, parallel group randomized controlled trial in nine secondary and tertiary NHS hospitals in England. Researchers were blinded to group allocation. Overall, 40 trainees in the West Midlands were eligible: 33 agreed to take part and were randomized, five withdrew after randomization, 13 were allocated cadaveric training, and 15 were allocated standard training. The intervention was an additional two-day cadaveric simulation course. The control group received standard on-the-job training. Primary outcome was implant position on the postoperative radiograph: tip-apex distance (mm) (DHS) and leg length discrepancy (mm) (hemiarthroplasty). Secondary clinical outcomes were procedure time, length of hospital stay, acute postoperative complication rate, and 12-month mortality. Procedure-specific secondary outcomes were intraoperative radiation dose (for DHS) and postoperative blood transfusion requirement (hemiarthroplasty). Results. Eight female (29%) and 20 male trainees (71%), mean age 29.4 years, performed 317
Aims. To develop a core outcome set of measurements from postoperative radiographs that can be used to assess technical skill in performing dynamic hip screw (DHS) and hemiarthroplasty, and to validate these against Van der Vleuten’s criteria for effective assessment. Methods. A Delphi exercise was undertaken at a regional major trauma centre to identify candidate measurement items. The feasibility of taking these measurements was tested by two of the authors (HKJ, GTRP). Validity and reliability were examined using the radiographs of operations performed by orthopaedic resident participants (n = 28) of a multicentre randomized controlled educational trial (ISRCTN20431944). Trainees were divided into novice and intermediate groups, defined as having performed < ten or ≥ ten cases each for
We compared the outcome of patients treated for an intertrochanteric fracture of the femoral neck with a locked, long intramedullary nail with those treated with a dynamic hip screw (DHS) in a prospective randomised study. Each patient who presented with an extra-capsular hip fracture was randomised to operative stabilisation with either a long intramedullary Holland nail or a
The spiral blade modification of the Dynamic
Hip Screw (DHS) was designed for superior biomechanical fixation
in the osteoporotic femoral head. Our objective was to compare clinical
outcomes and in particular the incidence of loss of fixation. . In a series of 197 consecutive patients over the age of 50 years
treated with DHS-blades (blades) and 242 patients treated with conventional
DHS (screw) for AO/OTA 31.A1 or A2 intertrochanteric fractures were
identified from a prospectively compiled database in a level 1 trauma
centre. Using propensity score matching, two groups comprising 177
matched patients were compiled and radiological and clinical outcomes
compared. In each group there were 66 males and 111 females. Mean
age was 83.6 (54 to 100) for the conventional
Objectives. The use of two implants to manage concomitant ipsilateral femoral
shaft and proximal femoral fractures has been indicated, but no
studies address the relationship of dynamic hip screw (DHS) side
plate screws and the intramedullary nail where failure might occur
after union. This study compares different implant configurations
in order to investigate bridging the gap between the distal DHS
and tip of the intramedullary nail. Methods. A total of 29 left synthetic femora were tested in three groups:
1) gapped short nail (GSN); 2) unicortical short nail (USN), differing
from GSN by the use of two unicortical bridging screws; and 3) bicortical
long nail (BLN), with two angled bicortical and one unicortical bridging
screws. With these findings, five matched-pairs of cadaveric femora
were tested in two groups: 1) unicortical long nail (ULN), with
a longer nail than USN and three bridging unicortical screws; and
2) BLN. Specimens were axially loaded to 22.7 kg (50 lb), and internally
rotated 90°/sec until failure. Results. For synthetic femora, a difference was detected between GSN and
BLN in energy to failure (p = 0.04) and torque at failure (p = 0.02),
but not between USN and other groups for energy to failure (vs GSN,
p = 0.71; vs BLN, p = 0.19) and torque at failure
(vs GSN, p = 0.55; vs BLN, p =
0.15). For cadaveric femora, ULN and BLN performed similarly because
of the improvement provided by the bridging screws. Conclusions. Our study shows that bicortical angled screws in the
Total hip arthroplasty (THA) is considered the preferred treatment for displaced proximal femoral neck fractures. However, in many countries this option is economically unviable. To improve outcomes in financially disadvantaged populations, we studied the technique of concomitant valgus hip osteotomy and operative fixation (VOOF). This prospective serial study compares two treatment groups: VOOF versus operative fixation alone with cannulated compression screws (CCSs). In the first series, 98 hip fixation procedures were performed using CCS. After fluoroscopic reduction of the fracture, three CCSs were placed. In the second series, 105 VOOF procedures were performed using a closing wedge intertrochanteric osteotomy with a compression lag screw and lateral femoral plate. The alignment goal was to create a modified Pauwel’s fracture angle of 30°. After fluoroscopic reduction of fracture, lag screw was placed to achieve the calculated correction angle, followed by inter-trochanteric osteotomy and placement of barrel plate. Patients were followed for a minimum of two years.Aims
Methods
Prior to the availability of vaccines, mortality for hip fracture patients with concomitant COVID-19 infection was three times higher than pre-pandemic rates. The primary aim of this study was to determine the 30-day mortality rate of hip fracture patients in the post-vaccine era. A multicentre observational study was carried out at 19 NHS Trusts in England. The study period for the data collection was 1 February 2021 until 28 February 2022, with mortality tracing until 28 March 2022. Data collection included demographic details, data points to calculate the Nottingham Hip Fracture Score, COVID-19 status, 30-day mortality, and vaccination status.Aims
Methods
Despite limited clinical scientific backing, an additional trochanteric stabilizing plate (TSP) has been advocated when treating unstable trochanteric fractures with a sliding hip screw (SHS). We aimed to explore whether the TSP would result in less post operative fracture motion, compared to SHS alone. Overall, 31 patients with AO/OTA 31-A2 trochanteric fractures were randomized to either a SHS alone or a SHS with an additional TSP. To compare postoperative fracture motion, radiostereometric analysis (RSA) was performed before and after weightbearing, and then at four, eight, 12, 26, and 52 weeks. With the “after weightbearing” images as baseline, we calculated translations and rotations, including shortening and medialization of the femoral shaft.Aims
Methods
In this study, we aimed to explore surgical variations in the Femoral Neck System (FNS) used for stable fixation of Pauwels type III femoral neck fractures. Finite element models were established with surgical variations in the distance between the implant tip and subchondral bone, the gap between the plate and lateral femoral cortex, and inferior implant positioning. The models were subjected to physiological load.Aims
Methods
The use of fluoroscopy in orthopaedic surgery creates risk of radiation exposure to surgeons. Appropriate personal protective equipment (PPE) can help mitigate this. The primary aim of this study was to assess if current radiation protection in orthopaedic trauma is safe. The secondary aims were to describe normative data of radiation exposure during common orthopaedic procedures, evaluate ways to improve any deficits in protection, and validate the use of electronic personal dosimeters (EPDs) in assessing radiation dose in orthopaedic surgery. Radiation exposure to surgeons during common orthopaedic trauma operations was prospectively assessed using EPDs and thermoluminescent dosimeters (TLDs). Normative data for each operation type were calculated and compared to recommended guidelines.Aims
Methods
In UK there are around 76,000 hip fractures occur each year 10% to 15% of which are undisplaced intracapsular. There is considerable debate whether internal fixation is the most appropriate treatment for undisplaced fractures in older patients. This study describes cannulated hip screws survivorship analysis for patients aged ≥ 60 years with undisplaced intra-capsular fractures. This was a retrospective cohort study of consecutive patients aged ≥ 60 years who had cannulated screws fixation for Garden I and II fractures in a teaching hospital between March 2013 and March 2016. The primary outcome was further same-side hip surgery. Descriptive statistics were used and Kaplan-Meier estimates calculated for implant survival.Aims
Methods
This study assessed the impact of COVID-19 on hip and distal femur fracture patient outcomes across three successive UK lockdown periods over one year. A single-centre retrospective cohort study was performed at an acute NHS Trust. Hip and distal femur fracture patients admitted within the first month from each of the three starting dates of each national lockdown were included and compared to a control group in March 2019. Data were collected as per the best practice tariff outcomes including additional outcomes as required. Data collection included COVID-19 status, time to theatre, 30-day mortality, presence of acute kidney injury (AKI) and pneumonia, and do not attempt cardiopulmonary resuscitation (DNACPR) status. Data were analyzed using an independent-samples Aims
Methods
We have prospectively compared the fixation of 100 intertrochanteric fractures of the proximal femur in elderly patients with random use of either a Dynamic Hip Screw (DHS) or a new intramedullary device, the Gamma nail. We found no difference in operating time, blood loss, wound complications, stay in hospital, place of eventual discharge, or the patients' mobility at final review. There was no difference in failure of proximal fixation: cut-out occurred in three cases with the
We aim to objectively assess the impact of COVID-19 on mean total operative cases for all indicative procedures (as outlined by the Joint Committee on Surgical Training (JCST)) experienced by orthopaedic trainees in the deanery of the Republic of Ireland. Subjective experiences were reported for each trainee using questionnaires. During the first four weeks of the nationwide lockdown due to COVID-19, the objective impact of the pandemic on each trainee’s surgical caseload exposure was assessed using data from individual trainee logbook profiles in the deanery of the Republic of Ireland. Independent predictor variables included the trainee grade (ST 3 to 8), the individual trainee, the unit that the logbook was reported from, and the year in which the logbook was recorded. We used the analysis of variance (ANOVA) test to assess for any statistically significant predictor variables. The subjective experience of each trainee was captured using an electronic questionnaire.Aims
Methods
Our rural orthopaedic service has undergone service restructure during the COVID-19 pandemic in order to sustain hip fracture care. All adult trauma care has been centralised to the Royal Shrewsbury Hospital for assessment and medical input, before transferring those requiring operative intervention to the Robert Jones and Agnes Hunt Orthopaedic Hospital. We aim to review the impact of COVID-19 on hip fracture workload and service changes upon management of hip fractures. We reviewed our prospectively maintained trust database and National Hip Fracture Database records for the months of March and April between the years 2016 and 2020. Our assessment included fracture pattern (intrascapular vs extracapsular hip fracture), treatment intervention, length of stay and mortality.Aims
Methods
The new COVID-19 variant was reported by the authorities of the UK to the World Health Organization (WHO) on 14 December 2020. We aim to describe the clinical characteristics and nosocomial infection rates in major trauma and orthopaedic patients comparing the first and second wave of COVID-19 infection. A retrospective analysis of a prospectively collected trauma database was reviewed at a level 1 major trauma centre from 1 December 2020 to 18 February 2021 looking at demographics, clinical characteristics, and nosocomial infections and compared to our previously published first wave data (26 January 2020 to 14 April 2020).Aims
Methods
There are reports of a marked increase in perioperative mortality in patients admitted to hospital with a fractured hip during the COVID-19 pandemic in the UK, USA, Spain, and Italy. Our study aims to describe the risk of mortality among patients with a fractured neck of femur in England during the early stages of the COVID-19 pandemic. We completed a multicentre cohort study across ten hospitals in England. Data were collected from 1 March 2020 to 6 April 2020, during which period the World Health Organization (WHO) declared COVID-19 to be a pandemic. Patients ≥ 60 years of age admitted with hip fracture and a minimum follow-up of 30 days were included for analysis. Primary outcome of interest was mortality at 30 days post-surgery or postadmission in nonoperative patients. Secondary outcomes included length of hospital stay and discharge destination.Aims
Methods
Despite the COVID-19 pandemic, incidence of hip fracture has not changed. Evidence has shown increased mortality rates associated with COVID-19 infection. However, little is known about the outcomes of COVID-19 negative patients in a pandemic environment. In addition, the impact of vitamin D levels on mortality in COVID-19 hip fracture patients has yet to be determined. This multicentre observational study included 1,633 patients who sustained a hip fracture across nine hospital trusts in North West England. Data were collected for three months from March 2020 and for the same period in 2019. Patients were matched by Nottingham Hip Fracture Score (NHFS), hospital, and fracture type. We looked at the mortality outcomes of COVID-19 positive and COVID-19 negative patients sustaining a hip fracture. We also looked to see if vitamin D levels had an impact on mortality.Aims
Methods
Now that we are in the deceleration phase of the COVID-19 pandemic, the focus has shifted to how to safely reinstate elective operating. Regional and speciality specific data is important to guide this decision-making process. This study aimed to review 30-day mortality for all patients undergoing orthopaedic surgery during the peak of the pandemic within our region. This multicentre study reviewed data on all patients undergoing trauma and orthopaedic surgery in a region from 18 March 2020 to 27 April 2020. Information was collated from regional databases. Patients were COVID-19-positive if they had positive laboratory testing and/or imaging consistent with the infection. 30-day mortality was assessed for all patients. Secondly, 30-day mortality in fracture neck of femur patients was compared to the same time period in 2019.Aims
Methods
Elective surgery has been severely curtailed as a result of the COVID-19 pandemic. There is little evidence to guide surgeons in assessing what processes should be put in place to restart elective surgery safely in a time of endemic COVID-19 in the community. We used data from a stand-alone hospital admitting and operating on 91 trauma patients. All patients were screened on admission and 100% of patients have been followed-up after discharge to assess outcome.Aims
Methods
Hip fracture patients are at higher risk of severe COVID-19 illness, and admission into hospital puts them at further risk. We implemented a two-site orthopaedic trauma service, with ‘COVID’ and ‘COVID-free’ hubs, to deliver urgent and infection-controlled trauma care for hip fracture patients, while increasing bed capacity for medical patients during the COVID-19 pandemic. A vacated private elective surgical centre was repurposed to facilitate a two-site, ‘COVID’ and ‘COVID-free’, hip fracture service. Patients were screened for COVID-19 infection and either kept at our ‘COVID’ site or transferred to our ‘COVID-free’ site. We collected data for 30 days on patient demographics, Clinical Frailty Scale (CFS), Nottingham Hip Fracture Scores (NHFS), time to surgery, COVID-19 status, mortality, and length of stay (LOS).Aims
Methods
The first death in the UK caused by COVID-19 occurred on 5 March 2020. We aim to describe the clinical characteristics and outcomes of major trauma and orthopaedic patients admitted in the early COVID-19 era. A prospective trauma registry was reviewed at a Level 1 Major Trauma Centre. We divided patients into Group A, 40 days prior to 5 March 2020, and into Group B, 40 days after.Aims
Methods
The aim of this study is to determine the effects of the UK lockdown during the COVID-19 pandemic on the orthopaedic admissions, operations, training opportunities, and theatre efficiency in a large district general hospital. The number of patients referred to the orthopaedic team between 1 April 2020 and 30 April 2020 were collected. Other data collected included patient demographics, number of admissions, number and type of operations performed, and seniority of primary surgeon. Theatre time was collected consisting of anaesthetic time, surgical time, time to leave theatre, and turnaround time. Data were compared to the same period in 2019.Aims
Methods
The aim of this study was to report the medium-term outcomes of impaction bone allograft and fibular grafting for osteonecrosis of the femoral head (ONFH) and to define the optimal indications. A total of 67 patients (77 hips) with ONFH were enrolled in a single centre retrospective review. Success of the procedure was assessed using the Harris Hip Score (HHS) and rate of revision to total hip arthroplasty (THA). Risk factors were studied, including age, aetiology, duration of hip pain, as well as two classification systems (Association Research Circulation Osseous (ARCO) and Japanese Investigation Committee (JIC) systems).Aims
Methods
The aim of this study was to analyze drain fluid, blood, and urine simultaneously to follow the long-term release of vancomycin from a biphasic ceramic carrier in major hip surgery. Our hypothesis was that there would be high local vancomycin concentrations during the first week with safe low systemic trough levels and a complete antibiotic release during the first month. Nine patients (six female, three male; mean age 75.3 years (sd 12.3; 44 to 84)) with trochanteric hip fractures had internal fixations. An injectable ceramic bone substitute, with hydroxyapatite in a calcium sulphate matrix, containing 66 mg of vancomycin per millilitre, was inserted to augment the fixation. The vancomycin elution was followed by simultaneously collecting drain fluid, blood, and urine.Objectives
Methods
Cephalomedullary nails (CMNs) are commonly used for the treatment of intertrochanteric hip fractures. Total hip arthroplasty (THA) may be used as a salvage procedure when fixation fails in these patients. The aim of this study was to analyze the complications of THA following failed intertrochanteric hip fracture fixation using a CMN. Patients who underwent THA were identified from the 5% subset of Medicare Parts A/B between 2002 and 2015. A subgroup involving those with an intertrochanteric fracture that was treated using a CMN during the previous five years was identified and compared with the remaining patients who underwent THA. The length of stay (LOS) was compared using both univariate and multivariate analysis. The incidence of infection, dislocation, revision, and re-admission was compared between the two groups, using multivariate analysis adjusted for demographic, hospital, and clinical factors.Aims
Patients and Methods
The aim of this study was to quantify the stability
of fracture-implant complex in fractures after fixation. A total
of 15 patients with an undisplaced fracture of the femoral neck,
treated with either a dynamic hip screw or three cannulated hip
screws, and 16 patients with an AO31-A2 trochanteric fracture treated
with a dynamic hip screw or a Gamma Nail, were included. Radiostereometric
analysis was used at six weeks, four months and 12 months post-operatively
to evaluate shortening and rotation. Migration could be assessed in ten patients with a fracture of
the femoral neck and seven with a trochanteric fracture. By four
months post-operatively, a mean shortening of 5.4 mm (-0.04 to 16.1)
had occurred in the fracture of the femoral neck group and 5.0 mm
(-0.13 to 12.9) in the trochanteric fracture group. A wide range
of rotation occurred in both types of fracture. Right-sided trochanteric
fractures seem more rotationally stable than left-sided fractures. This prospective study shows that migration at the fracture site
occurs continuously during the first four post-operative months,
after which stabilisation occurs. This information may allow the
early recognition of patients at risk of failure of fixation. Cite this article:
Fractures of the distal femur can be challenging to manage and
are on the increase in the elderly osteoporotic population. Management
with casting or bracing can unacceptably limit a patient’s ability
to bear weight, but historically, operative fixation has been associated
with a high rate of re-operation. In this study, we describe the outcomes
of fixation using modern implants within a strategy of early return
to function. All patients treated at our centre with lateral distal femoral
locking plates (LDFLP) between 2009 and 2014 were identified. Fracture
classification and operative information including weight-bearing
status, rates of union, re-operation, failure of implants and mortality
rate, were recorded.Aims
Patients and Methods
The objective of this study was to investigate bone healing after
internal fixation of displaced femoral neck fractures (FNFs) with
the Dynamic Locking Blade Plate (DLBP) in a young patient population
treated by various orthopaedic (trauma) surgeons. We present a multicentre prospective case series with a follow-up
of one year. All patients aged ≤ 60 years with a displaced FNF treated
with the DLBP between 1st August 2010 and December 2014 were included.
Patients with pathological fractures, concomitant fractures of the
lower limb, symptomatic arthritis, local infection or inflammation,
inadequate local tissue coverage, or any mental or neuromuscular
disorder were excluded. Primary outcome measure was failure in fracture
healing due to nonunion, avascular necrosis, or implant failure
requiring revision surgery.Aims
Patients and Methods
The aim of this study was to assess the current evidence relating
to the benefits of virtual reality (VR) simulation in orthopaedic
surgical training, and to identify areas of future research. A literature search using the MEDLINE, Embase, and Google Scholar
databases was performed. The results’ titles, abstracts, and references
were examined for relevance.Aims
Materials and Methods
Because of the contradictory body of evidence related to the
potential benefits of helical blades in trochanteric fracture fixation,
we studied the effect of bone compaction resulting from the insertion
of a proximal femoral nail anti-rotation (PFNA). We developed a subject-specific computational model of a trochanteric
fracture (31-A2 in the AO classification) with lack of medial support
and varied the bone density to account for variability in bone properties
among hip fracture patients.Objectives
Methods
The February 2013 Trauma Roundup360 looks at: the risk of ankle fractures; absorbable implants; minimally invasive heel fracture fixation; pertrochanteric fractures; arthroplasty and intracapsular hip fractures; and extensor mechanism disruption.
We treated 108 patients with a pertrochanteric femoral fracture using either the dynamic hip screw or the proximal femoral nail in this prospective, randomised series. We compared walking ability before fracture, intra-operative variables and return to their residence. Patients treated with the proximal femoral nail (n = 42) had regained their pre-operative walking ability significantly (p = 0.04) more often by the four-month review than those treated with the dynamic hip screw (n = 41). Peri-operative or immediate post-operative measures of outcome did not differ between the groups, with the exception of operation time. The dynamic hip screw allowed a significantly greater compression of the fracture during the four-month follow-up, but consolidation of the fracture was comparable between the two groups. Two major losses of reduction were observed in each group, resulting in a total of four revision operations. Our results suggest that the use of the proximal femoral nail may allow a faster postoperative restoration of walking ability, when compared with the dynamic hip screw.
We report positive and negative factors associated with the most commonly-used methods of reconstruction after pathological fracture of the proximal femur. The study was based on 142 patients treated surgically for 146 metastatic lesions between 1996 and 2003. The local rate of failure was 10.3% (15 of 146). Of 37 operations involving osteosynthetic devices, six failed (16.2%) compared with nine (8.3%) in 109 operations involving endoprostheses. Of nine cases of prosthetic failure, four were due to periprosthetic fractures and three to recurrent dislocation. In the osteosynthesis group, three (13.6%) of 22 reconstruction nails failed. The two-year risk of re-operation after any type of osteosynthesis was 0.35 compared with 0.18 after any type of endoprosthetic reconstruction (p = 0.07). Endoprosthetic reconstructions are preferable to the use of reconstruction nails and other osteosynthetic devices for the treatment of metastatic lesions in the proximal third of the femur.
We report the outcome of 39 patients who underwent
a modified Pauwels’ intertrochanteric osteotomy for nonunion of
a femoral neck fracture following failed osteosynthesis. There were
31 men and eight women with a mean age of 47.2 years (34 to 59).
By Pauwels’ classification, there were 11 Type II fractures and
28 Type III fractures. The mean follow-up was 7.9 years (2 to 19).
In the 11 patients whose initial treatment had been osteotomy, union was
achieved in nine (81.8%). In 28 patients whose initial treatment
had been with a lag screw or a dynamic hip screw, union was achieved
in 27 (96.4%). Limb lengths were equalised in 14 of 16 patients
(87.5%) with pre-operative shortening. The mean neck-shaft angle
improved significantly from 100.5° (80° to 120°) to 131.6° (120°
to 155°) (p = 0.004). The mean modified Harris hip score was 85.6
points (70 to 97) and the mean modified Merle d’Aubigné score was
14.3 (11 to 18). Good to excellent functional outcomes were achieved
in 32 patients (88.8%). A modified Pauwels’ intertrochanteric osteotomy
is a reliable method of treating ununited fractures of the femoral neck
following failed osteosynthesis: coxa vara and shortening can also
simultaneously be addressed. Cite this article
We have reviewed our experience of the removal of deep extremity orthopaedic implants in children to establish the nature, rate and risk of complications associated with this procedure. A retrospective review was performed of 801 children who had 1223 implants inserted and subsequently removed over a period of 17 years. Bivariate analysis of possible predictors including clinical factors, complications associated with implant insertion and indications for removal and the complications encountered at removal was performed. A logistical regression model was then constructed using those predictors which were significantly associated with surgical complications from the bivariate analyses. Odds ratios estimated in the logistical regression models were converted to risk ratios. The overall rate of complications after removal of the implant was 12.5% (100 complications in 801 patients), with 48 (6.0%) major and 52 (6.5%) minor. Children with a complication after insertion of the initial implant or with a non-elective indication for removal, a neuromuscular disease associated with a seizure disorder or a neuromuscular disease in those unable to walk, had a significantly greater chance of having a major complication after removal of the implant. Children with all four of these predictors were 14.6 times more likely to have a major complication.
We investigated the clinical outcome of internal
fixation for pathological fracture of the femur after primary excision of
a soft-tissue sarcoma that had been treated with adjuvant radiotherapy. A review of our database identified 22 radiation-induced fractures
of the femur in 22 patients (seven men, 15 women). We noted the
mechanism of injury, fracture pattern and any complications after
internal fixation, including nonunion, hardware failure, secondary
fracture or deep infection. The mean age of the patients at primary excision of the tumour
was 58.3 years (39 to 86). The mean time from primary excision to
fracture was 73.2 months (2 to 195). The mean follow-up after fracture
fixation was 65.9 months (12 to 205). Complications occurred in
19 patients (86%). Nonunion developed in 18 patients (82%), of whom
11 had a radiological nonunion at 12 months, five a nonunion and
hardware failure and two an infected nonunion. One patient developed
a second radiation-associated fracture of the femur after internal
fixation and union of the initial fracture. A total of 13 patients
(59%) underwent 24 revision operations. Internal fixation of a pathological fracture of the femur after
radiotherapy for a soft-tissue sarcoma has an extremely high rate
of complication and requires specialist attention. Cite this article:
An MR scan was performed on all patients who presented to our hospital with a clinical diagnosis of a fracture of the proximal femur, but who had no abnormality on plain radiographs. This was a prospective study of 102 consecutive patients over a ten-year period. There were 98 patients who fulfilled our inclusion criteria, of whom 75 were scanned within 48 hours of admission, with an overall mean time between admission and scanning of 2.4 days (0 to 10). A total of 81 patients (83%) had abnormalities detected on MRI; 23 (23%) required operative management. The use of MRI led to the early diagnosis and treatment of occult hip pathology. We recommend that incomplete intertrochanteric fractures are managed non-operatively with protected weight-bearing. The study illustrates the high incidence of fractures which are not apparent on plain radiographs, and shows that MRI is useful when diagnosing other pathology such as malignancy, which may not be apparent on plain films.
We reviewed the long-term radiological outcome,
complications and revision operations in 19 children with quadriplegic
cerebral palsy and hip dysplasia who underwent combined peri-iliac
osteotomy and femoral varus derotation osteotomy. They had a mean
age of 7.5 years (1.6 to 10.9) and comprised 22 hip dislocations
and subluxations. We also studied the outcome for the contralateral
hip. At a mean follow-up of 11.7 years (10 to 15.1) the Melbourne
cerebral palsy (CP) hip classification was grade 2 in 16 hips, grade
3 in five, and grade 5 in one. There were five complications seen
in four hips (21%, four patients), including one dislocation, one
subluxation, one coxa vara with adduction deformity, one subtrochanteric
fracture and one infection. A recurrent soft-tissue contracture occurred
in five hips and ten required revision surgery. In pre-adolescent children with quadriplegic cerebral palsy good
long-term outcomes can be achieved after reconstruction of the hip;
regular follow-up is required.
Cite this article:
In this study of patients who underwent internal fixation without
fusion for a burst thoracolumbar or lumbar fracture, we compared
the serial changes in the injured disc height (DH), and the fractured
vertebral body height (VBH) and kyphotic angle between patients
in whom the implants were removed and those in whom they were not. Radiological
parameters such as injured DH, fractured VBH and kyphotic angle
were measured. Functional outcomes were evaluated using the Greenough
low back outcome scale and a VAS scale for pain. Between June 1996 and May 2012, 69 patients were analysed retrospectively;
47 were included in the implant removal group and 22 in the implant
retention group. After a mean follow-up of 66 months (48 to 107),
eight patients (36.3%) in the implant retention group had screw
breakage. There was no screw breakage in the implant removal group.
All radiological and functional outcomes were similar between these
two groups. Although solid union of the fractured vertebrae was
achieved, the kyphotic angle and the anterior third of the injured
DH changed significantly with time (p <
0.05). Methods
Results
The April 2014 Research Roundup360 looks at: scientific writing needed in orthopaedic papers; antiseptics and osteoblasts; thromboembolic management in orthopaedic patients; nicotine and obesity in post-operative complications; defining the “Patient Acceptable Symptom State”; and cheap and nasty implants of poor quality.
There have been recent reports linking alendronate and a specific pattern of subtrochanteric insufficiency fracture. We performed a retrospective review of all subtrochanteric fractures admitted to our institution between 2001 and 2007. There were 20 patients who met the inclusion criteria, 12 of whom were on long-term alendronate. Alendronate-associated fractures tend to be bilateral (Fisher’s exact test, p = 0.018), have unique radiological features (p <
0.0005), be associated radiologically with a pre-existing ellipsoid thickening of the lateral femoral cortex and are likely to be preceded by prodromal pain. Biomechanical investigations did not suggest overt metabolic bone disease. Only one patient on alendronate had osteoporosis prior to the start of therapy. We used these findings to develop a management protocol to optimise fracture healing. We also advocate careful surveillance in individuals at-risk, and present our experience with screening and prophylactic fixation in selected patients.
We report our experience of treating 17 patients with benign lesions of the proximal femur with non-vascularised, autologous fibular strut grafts, without osteosynthesis. The mean age of the patients at presentation was 16.5 years (5 to 33) and they were followed up for a mean of 2.9 years (0.4 to 19.5). Histological diagnoses included simple bone cyst, fibrous dysplasia, aneurysmal bone cysts and giant cell tumour. Local recurrence occurred in two patients (11.7%) and superficial wound infection, chronic hip pain and deep venous thrombosis occurred in three. Pathological fracture did not occur in any patient following the procedure. We conclude that non-vascularised fibular strut grafts are a safe and satisfactory method of treating benign lesions of the proximal femur.
The December 2013 Hip &
Pelvis Roundup360 looks at: Enhanced recovery works; Acetabular placement; Exercise better than rest in osteoarthritis patients; if Birmingham hip resurfacing is immune from pseudotumour; HIV and arthroplasty; Labral tears revisited; Prophylactic surgery for FAI; and Ceramics and impaction grafting
The October 2013 Hip &
Pelvis Roundup360 looks at: Young and impinging; Clothes, weather and femoral heads?; Go long, go cemented; Surgical repair of the abductors?; Aspirin for DVT prophylaxis?; Ceramic-on-polyethylene: a low-wear solution?; ALVAL and ASR™: the story continues….; Salvaging Legg-Calve-Perthes’ disease
Iontophoresis is a novel technique which may be used to facilitate the movement of antibiotics into the substance of bone using an electrical potential applied externally. We have examined the rate of early infection in allografts following application of this technique in clinical practice. A total of 31 patients undergoing revision arthroplasty or surgery for limb salvage received 34 iontophoresed sequential allografts, of which 26 survived for a minimum of two years. The mean serum antibiotic levels after operation were low (gentamicin 0.37 mg/l (0.2 to 0.5); flucloxacillin 1 mg/l (0 to 1) and the levels in the drains were high (gentamicin 40 mg/l (2.5 to 131); flucloxacillin 17 mg/l (1 to 43). There were no early deep infections. Two late infections were presumed to be haemotogenous; 28 of the 34 allografts were retained. In 12 patients with pre-existing proven infection further infection has not occurred at a mean follow-up of 51 months (24 to 82).
This conversation represents an attempt by several
arthroplasty surgeons to critique several abstracts presented over
the last year as well as to use them as a jumping off point for trying
to figure out where they fit in into our current understanding of
multiple issues in modern hip and knee arthroplasty.
Incomplete intertrochanteric fractures do not extend across to the medial femoral cortex and are stable, without rotational deformity or shortening of the lower limb. The aim of our study was to establish whether they can be successfully managed conservatively. A total of 68 patients over a five-year period presented with a suspected fracture of the femoral neck and underwent an MRI scan for further assessment. From these, we retrospectively reviewed eight patients with normal plain radiographs but with an incomplete, intertrochanteric fracture on MRI scan. Five were managed conservatively and three operatively. The mean length of hospital stay was 16 days for the conservatively-treated group and 15 days for those who underwent surgery; this was not statistically significant (p >
0.5) and all patients were mobilised on discharge. Although five patients were readmitted at a mean of 3.2 years after discharge, none had progressed to a complete fracture. We believe that patients with incomplete intertrochanteric fractures should be considered for conservative treatment.
We undertook a trial on 60 patients with AO 31A2 fractures of the hip who were randomised after stabilisation of the fracture into two equal groups, one of which received post-operative treatment using a non-invasive interactive neurostimulation device and the other with a sham device. All other aspects of their rehabilitation were the same. The treatment was continued for ten days after operation. Outcome measurements included the use of a visual analogue scale for pain, the brief pain inventory and Ketorolac for post-operative control of pain, and an overall assessment of outcome by the surgeon. There were significantly better results for the patients receiving treatment by active electrical stimulation (repeated measures analysis of variance, p <
0.001). The findings of this pilot trial justify a larger study to determine if these results are more generally applicable.
Failure of fixation is a common problem in the treatment of osteoporotic fractures around the hip. The reinforcement of bone stock or of fixation of the implant may be a solution. Our study assesses the existing evidence for the use of bone substitutes in the management of these fractures in osteoporotic patients. Relevant publications were retrieved through Medline research and further scrutinised. Of 411 studies identified, 22 met the inclusion criteria, comprising 12 experimental and ten clinical reports. The clinical studies were evaluated with regard to their level of evidence. Only four were prospective and randomised. Polymethylmethacrylate and calcium-phosphate cements increased the primary stability of the implant-bone construct in all experimental and clinical studies, although there was considerable variation in the design of the studies. In randomised, controlled studies, augmentation of intracapsular fractures of the neck of the femur with calcium-phosphate cement was associated with poor long-term results. There was a lack of data on the long-term outcome for trochanteric fractures. Because there were only a few, randomised, controlled studies, there is currently poor evidence for the use of bone cement in the treatment of fractures of the hip.
Prospective data on hip fracture from 3686 patients at a United Kingdom teaching hospital were analysed to investigate the risk factors, financial costs and outcomes associated with deep or superficial wound infections after hip fracture surgery. In 1.2% (41) of patients a deep wound infection developed, and 1.1% (39) had a superficial wound infection. A total of 57 of 80 infections (71.3%) were due to No statistically significant pre-operative risk factors were detected. Length of stay, cost of treatment and pre-discharge mortality all significantly increased with deep wound infection. The one-year mortality was 30%, and this increased to 50% in those who developed an infection (p <
0.001). A deep infection resulted in doubled operative costs, tripled investigation costs and quadrupled ward costs. MRSA infection increased costs, length of stay, and pre-discharge mortality compared with non-MRSA infection.
Our aim was to determine the total blood loss associated with surgery for fracture of the hip and to identify risk factors for increased blood loss. We prospectively studied 546 patients with hip fracture. The total blood loss was calculated on the basis of the haemoglobin difference, the number of transfusions and the estimated blood volume. The hidden blood loss, in excess of that observed during surgery, varied from 547 ml (screws/ pins) to 1473 ml (intramedullary hip nail and screw) and was significantly associated with medical complications and increased hospital stay. The type of surgery, treatment with aspirin, intra-operative hypotension and gastro-intestinal bleeding or ulceration were all independent predictors of blood loss. We conclude that total blood loss after surgery for hip fracture is much greater than that observed intra-operatively. Frequent post-operative measurements of haemoglobin are necessary to avoid anaemia.