Advertisement for orthosearch.org.uk
Results 1 - 3 of 3
Results per page:
Bone & Joint 360
Vol. 7, Issue 2 | Pages 40 - 42
1 Apr 2018
Foy MA


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 5 | Pages 692 - 695
1 May 2006
Karataglis D Kapetanos G Lontos A Christodoulou A Christoforides J Pournaras J

The aim of this biomechanical study was to investigate the role of the dorsal vertebral cortex in transpedicular screw fixation. Moss transpedicular screws were introduced into both pedicles of each vertebra in 25 human cadaver vertebrae. The dorsal vertebral cortex and subcortical bone corresponding to the entrance site of the screw were removed on one side and preserved on the other. Biomechanical testing showed that the mean peak pull-out strength for the inserted screws, following removal of the dorsal cortex, was 956.16 N. If the dorsal cortex was preserved, the mean peak pullout strength was 1295.64 N. The mean increase was 339.48 N (26.13%; p = 0.033). The bone mineral density correlated positively with peak pull-out strength.

Preservation of the dorsal vertebral cortex at the site of insertion of the screw offers a significant increase in peak pull-out strength. This may result from engagement by the final screw threads in the denser bone of the dorsal cortex and the underlying subcortical area. Every effort should be made to preserve the dorsal vertebral cortex during insertion of transpedicular screws.


The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 3 | Pages 457 - 461
1 Apr 2004
Sandén B Olerud C Petrén-Mallmin M Johansson C Larsson S

We examined the radiographs from a prospective clinical study of fixation by pedicle screws and those from an experimental study in a sheep model. In the clinical study, instruments were removed from 21 patients after implantation for 11 to 16 months and the extraction torques of the screws were recorded. A structured protocol was used for the radiological examinations. In the experimental study, loaded pedicle screw instrumentations were implanted in the sheep for six or 12 weeks. After radiological examination the pull-out resistance and the histological characteristics were studied. In the clinical study, all screws with radiolucent zones had a significantly reduced mean extraction torque compared with screws without radiolucent zones (16 ± 10 Ncm v 403 ± 220 Ncm; p < 0.0001). In the experimental study the mean maximum pull-out resistance for the screws with radiolucent zones was significantly lower than for those with no radiolucency (243 ± 156 N v 2214 ± 578 N; p = 0.0006) and the mean bone-to-screw contact was reduced for screws with zones compared with those without zones (8 ± 9% v 55 ± 29%; p = 0.0002). Our findings showed that all screws with radiolucent zones had low extraction torques or low pull-out resistance. A radiolucent zone is a good indicator of loosening of a pedicle screw