Periprosthetic fracture and implant loosening are two of the major reasons for revision surgery of cementless implants. Optimal implant fixation with minimal bone damage is challenging in this procedure. This pilot study investigates whether vibratory implant insertion is gentler compared to consecutive single blows for acetabular component implantation in a surrogate polyurethane (PU) model. Acetabular components (cups) were implanted into 1 mm nominal under-sized cavities in PU foams (15 and 30 per cubic foot (PCF)) using a vibratory implant insertion device and an automated impaction device for single blows. The impaction force, remaining polar gap, and lever-out moment were measured and compared between the impaction methods.Aims
Methods
The Birmingham Hip Resurfacing (BHR) was introduced in 1997 to address the needs of young active patients using a historically proven large-diameter metal-on-metal (MoM) bearing. A single designer surgeon’s consecutive series of 130 patients (144 hips) was previously reported at five and ten years, reporting three and ten failures, respectively. The aim of this study was to extend the follow-up of this original cohort at 25 years. The study extends the reporting on the first consecutive 144 resurfacing procedures in 130 patients for all indications. All operations were undertaken between August 1997 and May 1998. The mean age at operation was 52.1 years (SD 9.93; 17 to 76), and included 37 female patients (28.5%). Failure was defined as revision of either component for any reason. Kaplan-Meier survival analysis was performed. Routine follow-up with serum metal ion levels, radiographs, and Oxford Hip Scores (OHSs) was undertaken.Aims
Methods
The risk of mechanical failure of modular revision hip stems is frequently mentioned in the literature, but little is currently known about the actual clinical failure rates of this type of prosthesis. The current retrospective long-term analysis examines the distal and modular failure patterns of the Prevision hip stem from 18 years of clinical use. A design improvement of the modular taper was introduced in 2008, and the data could also be used to compare the original and the current design of the modular connection. We performed an analysis of the Prevision modular hip stem using the manufacturer’s vigilance database and investigated different mechanical failure patterns of the hip stem from January 2004 to December 2022.Aims
Methods
Metal-on-metal hip resurfacing (MoM-HR) has seen decreased usage due to safety and longevity concerns. Joint registries have highlighted the risks in females, smaller hips, and hip dysplasia. This study aimed to identify if reported risk factors are linked to revision in a long-term follow-up of MoM-HR performed by a non-designer surgeon. A retrospective review of consecutive MoM hip arthroplasties (MoM-HRAs) using Birmingham Hip Resurfacing was conducted. Data on procedure side, indication, implant sizes and orientation, highest blood cobalt and chromium ion concentrations, and all-cause revision were collected from local and UK National Joint Registry records.Aims
Methods
Although the Fitmore Hip Stem has been on the market for almost 15 years, it is still not well documented in randomized controlled trials. This study compares the Fitmore stem with the CementLeSs (CLS) in several different clinical and radiological aspects. The hypothesis is that there will be no difference in outcome between stems. In total, 44 patients with bilateral hip osteoarthritis were recruited from the outpatient clinic at a single tertiary orthopaedic centre. The patients were operated with bilateral one-stage total hip arthroplasty. The most painful hip was randomized to either Fitmore or CLS femoral component; the second hip was operated with the femoral component not used on the first side. Patients were evaluated at three and six months and at one, two, and five years postoperatively with patient-reported outcome measures, radiostereometric analysis, dual-energy X-ray absorptiometry, and conventional radiography. A total of 39 patients attended the follow-up visit at two years (primary outcome) and 35 patients at five years. The primary outcome was which hip the patient considered to have the best function at two years.Aims
Methods
To find out if there is an inverse association between estimated glomerular filtration rate (eGFR) and whole blood cobalt (Co) and chromium (Cr) levels in patients with metal-on-metal (MoM) hip arthroplasties and renal insufficiency, suggesting that renal insufficiency could cause accumulation of Co and Cr in blood. Out of 2,520 patients with 3,013 MoM hip arthroplasties, we identified 1,244 patients with whole blood Co, Cr, and creatinine measured within no more than a one-year interval. We analyzed the correlation of blood metal ion levels and eGFR to identify a potential trend of accumulating Co or Cr with decreasing eGFR.Aims
Methods
The optimum clearance between the bearing surfaces of hip arthroplasties is unknown. Theoretically, to minimize wear, it is understood that clearances must be low enough to maintain optimal contact pressure and fluid film lubrication, while being large enough to allow lubricant recovery and reduce contact patch size. This study aimed to identify the relationship between diametrical clearance and volumetric wear, through the analysis of retrieved components. A total of 81 metal-on-metal Pinnacle hips paired with 12/14 stems were included in this study. Geometrical analysis was performed on each component, using coordinate and roundness measuring machines. The relationship between their as-manufactured diametrical clearance and volumetric wear was investigated. The Mann-Whitney U test and unpaired Aims
Methods
The purpose of this study was to compare two different types
of metal-on-metal (MoM) bearing for total hip arthroplasty (THA):
one with a large femoral head (38 mm to 52 mm) and the other with
a conventional femoral head (28 mm or 32 mm). We compared clinical
outcome, blood metal ion levels, and the incidence of pseudotumour in
the two groups. Between December 2009 and December 2011, 62 patients underwent
MoM THA with a large femoral head (Magnum group) and 57 patients
an MoM THA with a conventional femoral head (conventional group).
Clinical outcome was assessed using the Harris Hip score, University
of California, Los Angeles (UCLA) activity score and EuroQol-5D
(EQ-5D). Blood metal ion levels were measured and MRI scans were
analyzed at a minimum of five years postoperatively.Aims
Patients and Methods
The pelvis rotates in the sagittal plane during daily activities.
These rotations have a direct effect on the functional orientation
of the acetabulum. The aim of this study was to quantify changes
in pelvic tilt between different functional positions. Pre-operatively, pelvic tilt was measured in 1517 patients undergoing
total hip arthroplasty (THA) in three functional positions – supine,
standing and flexed seated (the moment when patients initiate rising
from a seated position). Supine pelvic tilt was measured from CT
scans, standing and flexed seated pelvic tilts were measured from standardised
lateral radiographs. Anterior pelvic tilt was assigned a positive
value.Aims
Patients and Methods
Objectives. High failure rates of metal-on-metal hip arthroplasty implants have highlighted the need for more careful introduction and monitoring of new implants and for the evaluation of the safety of
We conducted a retrospective study to assess
the prevalence of adverse reactions to metal debris (ARMD) in patients
operated on at our institution with metal-on-metal (MoM) total hip
replacements with 36 mm heads using a Pinnacle acetabular shell.
A total of 326 patients (150 males, 175 hips; 176 females, 203 hips)
with a mean age of 62.7 years (28 to 85) and mean follow-up of 7.5
years (0.1 to 10.8) participating in our in-depth modern MoM follow-up
programme were included in the study, which involved recording whole
blood cobalt and chromium ion measurements, Oxford hip scores (OHS)
and plain radiographs of the hip and targeted cross-sectional imaging. Elevated
blood metal ion levels (>
5 parts per billion) were seen in 32 (16.1%)
of the 199 patients who underwent unilateral replacement. At 23
months after the start of our modern MoM follow-up programme, 29
new cases of ARMD had been revealed. Hence, the nine-year survival
of this cohort declined from 96% (95% CI 95 to 98) with the old
surveillance routine to 86% (95% CI 82 to 90) following the new
protocol. Although ARMD may not be as common in 36 mm MoM THRs as
in those with larger heads, these results support the Medicines
and Healthcare Products Regulatory Agency guidelines on regular
reviews and further investigations, and emphasise the need for specific
a follow-up programme for patients with MoM THRs. Cite this article:
In this paper, we will consider the current role
of metal-on-metal bearings by looking at three subtypes of MoM hip
arthroplasty separately: Hip resurfacing, large head (>
36 mm) MoM
THA and MoM THA with traditional femoral head sizes.
Plasma levels of cobalt and chromium ions and
Metal Artefact Reduction Sequence (MARS)-MRI scans were performed
on patients with 209 consecutive, unilateral, symptomatic metal-on-metal
(MoM) hip arthroplasties. There was wide variation in plasma cobalt
and chromium levels, and MARS-MRI scans were positive for adverse reaction
to metal debris (ARMD) in 84 hips (40%). There was a significant
difference in the median plasma cobalt and chromium levels between
those with positive and negative MARS-MRI scans (p <
0.001).
Compared with MARS-MRI as the potential reference standard for the
diagnosis of ARMD, the sensitivity of metal ion analysis for cobalt
or chromium with a cut-off of >
7 µg/l was 57%. The specificity was
65%, positive predictive value was 52% and the negative predictive
value was 69% in symptomatic patients. A lowered threshold of >
3.5 µg/l for cobalt and chromium ion levels improved the sensitivity
and negative predictive value to 86% and 74% but at the expense
of specificity (27%) and positive predictive value (44%). Metal ion analysis is not recommended as a sole indirect screening
test in the surveillance of symptomatic patients with a MoM arthroplasty.
The investigating clinicians should have a low threshold for obtaining
cross-sectional imaging in these patients, even in the presence
of low plasma metal ion levels.
The Articular Surface Replacement (ASR) hip resurfacing arthroplasty has a failure rate of 12.0% at five years, compared with 4.3% for the Birmingham Hip Resurfacing (BHR). We analysed 66 ASR and 64 BHR explanted metal-on-metal hip replacements with the aim of understanding their mechanisms of failure. We measured the linear wear rates of the acetabular and femoral components and analysed the clinical cause of failure, pre-revision blood metal ion levels and orientation of the acetabular component. There was no significant difference in metal ion levels (chromium, p = 0.82; cobalt, p = 0.40) or head wear rate (p = 0.14) between the two groups. The ASR had a significantly increased rate of wear of the acetabular component (p = 0.03) and a significantly increased occurrence of edge loading (p <
0.005), which can be attributed to differences in design between the ASR and BHR. The effects of differences in design on the
This study compared component wear rates and pre-revision blood metal ions levels in two groups of failed metal-on-metal hip arthroplasties: hip resurfacing and modular total hip replacement (THR). There was no significant difference in the median rate of linear wear between the groups for both acetabular (p = 0.4633) and femoral (p = 0.0872) components. There was also no significant difference in the median linear wear rates when failed hip resurfacing and modular THR hips of the same type (ASR and Birmingham hip resurfacing (BHR)) were compared. Unlike other studies of well-functioning hips, there was no significant difference in pre-revision blood metal ion levels between hip resurfacing and modular THR. Edge loading was common in both groups, but more common in the resurfacing group (67%) than in the modular group (57%). However, this was not significant (p = 0.3479). We attribute this difference to retention of the neck in resurfacing of the hip, leading to impingement-type edge loading. This was supported by visual evidence of impingement on the femur. These findings show that failed metal-on-metal hip resurfacing and modular THRs have similar component wear rates and are both associated with raised pre-revision blood levels of metal ions.
We report serum metal ion level data in patients with unilateral and bilateral hip resurfacing over a ten-year period. In these patients there is an increase in both cobalt and chromium levels above the accepted reference ranges during the first 18 months after operation. Metal ion levels remain elevated, but decline slowly for up to five years. However, the levels then appear to start rising again in some patients up to the ten-year mark. There was no significant difference in cobalt or chromium levels between men and women. These findings appear to differ from much of the current literature. The clinical significance of a raised metal ion level remains under investigation.
Metal-on-metal total hip replacement has been targeted at younger patients with anticipated long-term survival, but the effect of the production of metal ions is a concern because of their possible toxicity to cells. We have reviewed the results of the use of the Ultima hybrid metal-on-metal total hip replacement, with a cemented polished tapered femoral component with a 28 mm diameter and a cobalt-chrome (CoCr) modular head, articulating with a 28 mm CoCr acetabular bearing surface secured in a titanium alloy uncemented shell. Between 1997 and 2004, 545 patients with 652 affected hips underwent replacement using this system. Up to 31 January 2008, 90 (13.8%) hips in 82 patients had been revised. Pain was the sole reason for revision in 44 hips (48.9%) of which 35 had normal plain radiographs. Peri-prosthetic fractures occurred in 17 hips (18.9%) with early dislocation in three (3.3%) and late dislocation in 16 (17.8%). Infection was found in nine hips (10.0%). At operation, a range of changes was noted including cavities containing cloudy fluid under pressure, necrotic soft tissues with avulsed tendons and denuded osteonecrotic upper femora. Corrosion was frequently observed on the retrieved cemented part of the femoral component. Typically, the peri-operative findings confirmed those found on pre-operative metal artefact reduction sequence MRI and histological examination showed severe necrosis. Metal artefact reduction sequence MRI proved to be useful when investigating these patients with pain in the absence of adverse plain radiological features.
Deficiencies of acetabular bone stock at revision hip replacement were reconstructed with two different types of allograft using impaction bone grafting and a Burch-Schneider reinforcement ring. We compared a standard frozen non-irradiated bone bank allograft (group A) with a freeze-dried irradiated bone allograft, vitalised with autologous marrow (group B). We studied 78 patients (79 hips), of whom 87% (69 hips) had type III acetabular defects according to the American Academy of Orthopaedic Surgeons classification at a mean of 31.4 months (14 to 51) after surgery. At the latest follow-up, the mean Harris hip score was 69.9 points (13.5 to 97.1) in group A and 71.0 points (11.5 to 96.5) in group B. Each hip showed evidence of trabeculation and incorporation of the allograft with no acetabular loosening. These results suggest that the use of an acetabular reinforcement ring and a living composite of sterile allograft and autologous marrow appears to be a method of reconstructing acetabular deficiencies which gives comparable results to current forms of treatment.
We describe the results at five years of a prospective study of a new tri-tapered polished, cannulated, cemented femoral stem implanted in 51 patients (54 hips) with osteoarthritis. The mean age and body mass index of the patients was 74 years and 27.9, respectively. Using the anterolateral approach, half of the stems were implanted by a consultant orthopaedic surgeon and half by six different registrars. There were three withdrawals from the study because of psychiatric illness, a deep infection and a recurrent dislocation. Five deaths occurred prior to five-year follow-up and one patient withdrew from clinical review. In the remaining 51 hips the mean pre-operative Oxford hip score was 47 points which decreased to 19 points at five years (45 hips). Of the stems 49 (98%) were implanted within 1° of neutral in the femoral canal. The mean migration of the stem at five years was 1.9 mm and the survivorship for aseptic loosening was 100%. There was no significant difference in outcome between the consultant and registrar groups. At five years, the results were comparable with those of other polished, tapered, cemented stems. Long-term surveillance continues.