Advertisement for orthosearch.org.uk
Results 1 - 14 of 14
Results per page:
Bone & Joint Research
Vol. 13, Issue 5 | Pages 226 - 236
9 May 2024
Jürgens-Lahnstein JH Petersen ET Rytter S Madsen F Søballe K Stilling M

Aims

Micromotion of the polyethylene (PE) inlay may contribute to backside PE wear in addition to articulate wear of total knee arthroplasty (TKA). Using radiostereometric analysis (RSA) with tantalum beads in the PE inlay, we evaluated PE micromotion and its relationship to PE wear.

Methods

A total of 23 patients with a mean age of 83 years (77 to 91), were available from a RSA study on cemented TKA with Maxim tibial components (Zimmer Biomet). PE inlay migration, PE wear, tibial component migration, and the anatomical knee axis were evaluated on weightbearing stereoradiographs. PE inlay wear was measured as the deepest penetration of the femoral component into the PE inlay.


Bone & Joint Research
Vol. 10, Issue 6 | Pages 363 - 369
1 Jun 2021
MacDonald DRW Neilly DW Elliott KE Johnstone AJ

Aims

Tourniquets have potential adverse effects including postoperative thigh pain, likely caused by their ischaemic and possible compressive effects. The aims of this preliminary study were to determine if it is possible to directly measure intramuscular pH in human subjects over time, and to measure the intramuscular pH changes resulting from tourniquet ischaemia in patients undergoing knee arthroscopy.

Methods

For patients undergoing short knee arthroscopic procedures, a sterile calibrated pH probe was inserted into the anterior fascial compartment of the leg after skin preparation, but before tourniquet inflation. The limb was elevated for three minutes prior to tourniquet inflation to 250 mmHg or 300 mmHg. Intramuscular pH was recorded at one-second intervals throughout the procedure and for 20 minutes following tourniquet deflation. Probe-related adverse events were recorded.


The Bone & Joint Journal
Vol. 103-B, Issue 1 | Pages 98 - 104
1 Jan 2021
van Ooij B Sierevelt IN van der Vis HM Hoornenborg D Haverkamp D

Aims

For many designs of total knee arthroplasty (TKA) it remains unclear whether cemented or uncemented fixation provides optimal long-term survival. The main limitation in most studies is a retrospective or non-comparative study design. The same is true for comparative trials looking only at the survival rate as extensive sample sizes are needed to detect true differences in fixation and durability. Studies using radiostereometric analysis (RSA) techniques have shown to be highly predictive in detecting late occurring aseptic loosening at an early stage. To investigate the difference in predicted long-term survival between cemented, uncemented, and hybrid fixation of TKA, we performed a randomized controlled trial using RSA.

Methods

A total of 105 patients were randomized into three groups (cemented, uncemented, and hybrid fixation of the ACS Mobile Bearing (ACS MB) knee system, implantcast). RSA examinations were performed on the first day after surgery and at scheduled follow-up visits at three months, six months, one year, and two years postoperatively. Patient-reported outcome measures (PROMs) were obtained preoperatively and after two years follow-up. Patients and follow-up investigators were blinded for the result of randomization.


The Bone & Joint Journal
Vol. 102-B, Issue 11 | Pages 1527 - 1534
3 Nov 2020
Orita K Minoda Y Sugama R Ohta Y Ueyama H Takemura S Nakamura H

Aims

Vitamin E-infused highly cross-linked polyethylene (E1) has recently been introduced in total knee arthroplasty (TKA). An in vitro wear simulator study showed that E1 reduced polyethylene wear. However there is no published information regarding in vivo wear. Previous reports suggest that newly introduced materials which reduce in vitro polyethylene wear do not necessarily reduce in vivo polyethylene wear. To assist in the evaluation of the newly introduced material before widespread use, we established an in vivo polyethylene wear particle analysis for TKA. The aim of this study was to compare in vivo polyethylene wear particle generation between E1 and conventional polyethylene (ArCom) in TKA.

Methods

A total of 34 knees undergoing TKA (17 each with ArCom or E1) were investigated. Except for the polyethylene insert material, the prostheses used for both groups were identical. Synovial fluid was obtained at a mean of 3.4 years (SD 1.3) postoperatively. The in vivo polyethylene wear particles were isolated from the synovial fluid using a previously validated method and examined by scanning electron microscopy.


The Bone & Joint Journal
Vol. 102-B, Issue 9 | Pages 1158 - 1166
14 Sep 2020
Kaptein BL den Hollander P Thomassen B Fiocco M Nelissen RGHH

Aims

The primary objective of this study was to compare migration of the cemented ATTUNE fixed bearing cruciate retaining tibial component with the cemented Press-Fit Condylar (PFC)-sigma fixed bearing cruciate retaining tibial component. The secondary objectives included comparing clinical and radiological outcomes and Patient Reported Outcome Measures (PROMs).

Methods

A single blinded randomized, non-inferiority study was conducted including 74 patients. Radiostereometry examinations were made after weight bearing, but before hospital discharge, and at three, six, 12, and 24 months postoperatively. PROMS were collected preoperatively and at three, six, 12, and 24 months postoperatively. Radiographs for measuring radiolucencies were collected at two weeks and two years postoperatively.


The Bone & Joint Journal
Vol. 101-B, Issue 5 | Pages 559 - 564
1 May 2019
Takemura S Minoda Y Sugama R Ohta Y Nakamura S Ueyama H Nakamura H

Aims

The use of vitamin E-infused highly crosslinked polyethylene (HXLPE) in total knee prostheses is controversial. In this paper we have compared the clinical and radiological results between conventional polyethylene and vitamin E-infused HXLPE inserts in total knee arthroplasty (TKA).

Patients and Methods

The study included 200 knees (175 patients) that underwent TKA using the same total knee prostheses. In all, 100 knees (77 patients) had a vitamin E-infused HXLPE insert (study group) and 100 knees (98 patients) had a conventional polyethylene insert (control group). There were no significant differences in age, sex, diagnosis, preoperative knee range of movement (ROM), and preoperative Knee Society Score (KSS) between the two groups. Clinical and radiological results were evaluated at two years postoperatively.


The Bone & Joint Journal
Vol. 99-B, Issue 12 | Pages 1596 - 1602
1 Dec 2017
Dunbar MJ Laende EK Collopy D Richardson CG

Aims

Hydroxyapatite coatings for uncemented fixation in total knee arthroplasty can theoretically provide a long-lasting biological interface with the host bone. The objective of this study was to test this hypothesis with propriety hydroxyapatite, peri-apatite, coated tibial components using component migration measured with radiostereometric analysis over two years as an indicator of long-term fixation.

Patients and Methods

A total of 29 patients at two centres received uncemented PA-coated tibial components and were followed for two years with radiostereometric analysis exams to quantify the migration of the component.


Bone & Joint Research
Vol. 5, Issue 8 | Pages 320 - 327
1 Aug 2016
van IJsseldijk EA Valstar ER Stoel BC Nelissen RGHH Baka N van’t Klooster R Kaptein BL

Objectives

An important measure for the diagnosis and monitoring of knee osteoarthritis is the minimum joint space width (mJSW). This requires accurate alignment of the x-ray beam with the tibial plateau, which may not be accomplished in practice. We investigate the feasibility of a new mJSW measurement method from stereo radiographs using 3D statistical shape models (SSM) and evaluate its sensitivity to changes in the mJSW and its robustness to variations in patient positioning and bone geometry.

Materials and Methods

A validation study was performed using five cadaver specimens. The actual mJSW was varied and images were acquired with variation in the cadaver positioning. For comparison purposes, the mJSW was also assessed from plain radiographs. To study the influence of SSM model accuracy, the 3D mJSW measurement was repeated with models from the actual bones, obtained from CT scans.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 12 | Pages 1575 - 1578
1 Dec 2009
Jaiswal PK Macmull S Bentley G Carrington RWJ Skinner JA Briggs TWR

Smoking is known to have an adverse effect on wound healing and musculoskeletal conditions. This case-controlled study looked at whether smoking has a deleterious effect in the outcome of autologous chondrocyte implantation for the treatment of full thickness chondral defects of the knee.

The mean Modified Cincinatti Knee score was statistically significantly lower in smokers (n = 48) than in non-smokers (n = 66) both before and after surgery (p < 0.05). Smokers experienced significantly less improvement in the knee score two years after surgery (p < 0.05). Graft failures were only seen in smokers (p = 0.016). There was a strong negative correlation between the number of cigarettes smoked and the outcome following surgery (Pearson’s correlation coefficient −0.65, p = 0.004).

These results suggest that patients who smoke have worse pre-operative function and obtain less benefit from this procedure than non-smokers. The counselling of patients undergoing autologous chondrocyte implantation should include smoking, not only as a general cardiopulmonary risk but also because poorer results can be expected in smokers following this procedure.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 8 | Pages 1027 - 1031
1 Aug 2006
Karim A Pandit H Murray J Wandless F Thomas NP

We sought to determine whether smoking affected the outcome of reconstruction of the anterior cruciate ligament. We analysed the results of 66 smokers (group 1 with a mean follow-up of 5.67 years (1.1 to 12.7)) and 238 non-smokers (group 2 with a mean follow-up of 6.61 years (1.2 to 11.5)), who were statistically similar in age, gender, graft type, fixation and associated meniscal and chondral pathology. The assessment was performed using the International Knee Documentation Committee form and serial cruciometer readings.

Poor outcomes were reported in group 1 for the mean subjective International Knee Documentation Committee score (p < 0.001), the frequency (p = 0.005) and intensity (p = 0.005) of pain, a side-to-side difference in knee laxity (p = 0.001) and the use of a four-strand hamstring graft (p = 0.015). Patients in group 1 were also less likely to return to their original level of pre-injury sport (p = 0.003) and had an overall worse final 7 International Knee Documentation Committee grade score (p = 0.007).

Despite the well-known negative effects of smoking on tissue healing, the association with an inferior outcome after reconstruction of the anterior cruciate ligament has not previously been described and should be included in the pre-operative counselling of patients undergoing the procedure.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 7 | Pages 893 - 899
1 Jul 2006
Petrou G Chardouvelis C Kouzoupis A Dermon A Petrou H Tilkeridis C Gavras M

We carried out a prospective study of 71 patients who had undergone reconstruction of the anterior cruciate ligament with the ABC scaffold. Their mean age was 28 years (18 to 50). All had either sub-acute or chronic traumatic deficiency of the ligament. The mean period of follow-up was five years (four to seven). Assessment included the use of the International Knee Documentation Committee score, the modified Lysholm score, the Tegner Activity score, the Knee Injury and Osteoarthritis Outcome score and measurement with the KT-1000 arthrometer. Two patients had mild recurrent synovitis. There were no infections and no failures of the ligament. During the period of study, two patients sustained a traumatic fracture of a femoral condyle. The implants retained their integrity in both cases. All patients returned to their previous or enhanced levels of daily activity by three months after operation and 56 (79%) achieved their pre-injury level of sporting activity by six months. The patients who were competing in National level sports returned to play at one level less after operation than before. The Lysholm score showed that 58% of the patients (41) were excellent, 34% (24) good, and 8% (6) fair, with a mean post-operative score of 93. According to the International Knee Documentation Committee score, 35% of knees (25) were ‘normal’, 52% (37) ‘nearly normal’ and 13% (9) ‘abnormal’. Complete satisfaction was noted in 90% of patients (64). The development of osteoarthritis and the management of anterior cruciate deficiency associated with laxity of the medial collateral ligament remains uncertain. Our results indicate that in the medium-term, the ABC ligament scaffold is suitable and effective when early and safe return to unrestricted activities is demanded. We acknowledge the current general hostility towards reconstruction of the anterior cruciate ligament with artificial materials following reports of early failure and chronic synovitis associatiated with the production of particulate debris. We did not encounter these problems.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 1 | Pages 54 - 60
1 Jan 2006
Pandit H Jenkins C Barker K Dodd CAF Murray DW

This prospective study describes the complications and survival of the first 688 Phase 3 Oxford medial unicompartmental knee replacements implanted using a minimally-invasive technique by two surgeons and followed up independently. None was lost to follow-up. We had carried out 132 of the procedures more than five years ago. The clinical assessment of 101 of these which were available for review at five years is also presented.

Nine of the 688 knees were revised: four for infection, three for dislocation of the bearing and two for unexplained pain. A further seven knees (1%) required other procedures: four had a manipulation under anaesthesia, two an arthroscopy and one a debridement for superficial infection. The survival rate at seven years was 97.3% (95% confidence interval 5.3). At five years, 96% of the patients had a good or excellent American Knee Society score, the mean Oxford knee score was 39 and the mean flexion was 133°. This study demonstrates that the minimally-invasive Oxford unicompartmental knee replacement is a reliable and effective procedure.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 5 | Pages 640 - 645
1 May 2005
Bartlett W Skinner JA Gooding CR Carrington RWJ Flanagan AM Briggs TWR Bentley G

Autologous chondrocyte implantation (ACI) is used widely as a treatment for symptomatic chondral and osteochondral defects of the knee. Variations of the original periosteum-cover technique include the use of porcine-derived type I/type III collagen as a cover (ACI-C) and matrix-induced autologous chondrocyte implantation (MACI) using a collagen bilayer seeded with chondrocytes. We have performed a prospective, randomised comparison of ACI-C and MACI for the treatment of symptomatic chondral defects of the knee in 91 patients, of whom 44 received ACI-C and 47 MACI grafts.

Both treatments resulted in improvement of the clinical score after one year. The mean modified Cincinnati knee score increased by 17.6 in the ACI-C group and 19.6 in the MACI group (p = 0.32). Arthroscopic assessments performed after one year showed a good to excellent International Cartilage Repair Society score in 79.2% of ACI-C and 66.6% of MACI grafts. Hyaline-like cartilage or hyaline-like cartilage with fibrocartilage was found in the biopsies of 43.9% of the ACI-C and 36.4% of the MACI grafts after one year. The rate of hypertrophy of the graft was 9% (4 of 44) in the ACI-C group and 6% (3 of 47) in the MACI group. The frequency of re-operation was 9% in each group.

We conclude that the clinical, arthroscopic and histological outcomes are comparable for both ACI-C and MACI. While MACI is technically attractive, further long-term studies are required before the technique is widely adopted.


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 6 | Pages 965 - 970
1 Nov 1998
Meister K Cobb A Bentley G

Between 1989 and 1993 we implanted a matrix support prosthesis made of carbon fibre for the treatment of chronic painful articular defects of the patella in 27 patients. The mean period of follow-up was 33 months (11 to 54) with clinical and arthroscopic evaluation of the implant. Overall, there were four excellent, three good, seven fair and 13 poor results. Nine patients subsequently had a patellectomy for persistent pain at a mean of 27 months after surgery (14 to 47). The mean visual analogue pain scores in those who retained their patella were 7.6 before operation and 5.5 at the time of the latest follow-up. Patient satisfaction for the entire group was 41%. There appeared to be good incorporation of the prosthesis and a satisfactory ingrowth of fibrocartilage, but the poor results in 48% and low patient satisfaction discourage us from recommending the procedure for lesions of the articular cartilage of the patella. The consistent seeding of the joint with carbon-fibre debris and a histiocytic giant-cell reaction in the synovium are also a cause for concern