Advertisement for orthosearch.org.uk
Results 1 - 3 of 3
Results per page:
The Bone & Joint Journal
Vol. 98-B, Issue 7 | Pages 952 - 960
1 Jul 2016
Muderis MA Tetsworth K Khemka A Wilmot S Bosley B Lord SJ Glatt V

Aims

This study describes the Osseointegration Group of Australia’s Accelerated Protocol two-stage strategy (OGAAP-1) for the osseointegrated reconstruction of amputated limbs.

Patients and Methods

We report clinical outcomes in 50 unilateral trans-femoral amputees with a mean age of 49.4 years (24 to 73), with a minimum one-year follow-up. Outcome measures included the Questionnaire for persons with a Trans-Femoral Amputation, the health assessment questionnaire Short-Form-36 Health Survey, the Amputation Mobility Predictor scores presented as K-levels, 6 Minute Walk Test and timed up and go tests. Adverse events included soft-tissue problems, infection, fractures and failure of the implant.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 5 | Pages 644 - 649
1 May 2011
Yonekura Y Miyamoto H Shimazaki T Ando Y Noda I Mawatari M Hotokebuchi T

A silver-containing hydroxyapatite (Ag-HA) coating has been developed using thermal spraying technology. We evaluated the osteoconductivity of this coating on titanium (Ti) implants in rat tibiae in relation to bacterial infection in joint replacement.

At 12 weeks, the mean affinity indices of bone formation of a Ti, an HA, a 3%Ag-HA and a 50%Ag-HA coating were 97.3%, 84.9%, 81.0% and 40.5%, respectively. The mean affinity indices of bone contact of these four coatings were 18.8%, 83.7%, 77.2% and 40.5%, respectively. The indices of bone formation and bone contact around the implant of the 3%Ag-HA coating were similar to those of the HA coating, and no significant differences were found between them (bone formation, p = 0.99; bone contact, p = 0.957). However, inhibition of bone formation was observed with the 50%Ag-HA coating.

These results indicate that the 3%Ag-HA coating has low toxicity and good osteoconductivity, and that the effect of silver toxicity on osteoconductivity depends on the dose.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 5 | Pages 639 - 643
1 May 2011
Burghardt RD Herzenberg JE Specht SC Paley D

Between October 2001 and September 2009 we lengthened 242 lower-limb segments in 180 patients using the Intramedullary Skeletal Kinetic Distractor (ISKD). Mechanical failure was defined either as breakage of the ISKD or failure of the internal mechanism to activate. Retrieved nails which failed mechanically were examined by the manufacturer for defects. In all, 15 ISKDs in 12 patients (13 limbs) failed mechanically representing an overall failure rate of 6.2%, with fracture of the device occurring in ten of the 15 failures. Two nails in one patient failed to lengthen and had to be replaced. The manufacturer detected an error in the assembly of the nail, which prompted a wide recall. One nail jammed after being forcefully inserted, and two nails failed to lengthen fully. Lengthening was achieved in all 12 patients, although three required a second operation to exchange a defective nail for a new, functioning device.

The ISKD is a complex mechanical device which lengthens by the oscillation of two telescopic sections connected by a threaded rod. The junction between these sections is surrounded by a keyring collar. This keyring collar is the weakest part of the device.