Endoprosthetic reconstruction following distal femur tumour resection has been widely advocated. In this paper, we present the design of an uncemented endoprosthesis system featuring a short, curved stem, with the goal of enhancing long-term survivorship and functional outcomes. This study involved patients who underwent implantation of an uncemented distal femoral endoprosthesis with a short and curved stem between 2014 and 2019. Functional outcomes were assessed using the 1993 version of the Musculoskeletal Tumour Society (MSTS-93) score. Additionally, we quantified five types of complications and assessed osseointegration radiologically. The survivorship of the endoprosthesis was evaluated according to two endpoints. A total of 134 patients with a median age of 26 years (IQR 16 to 41) were included in our study. The median follow-up time was 61 months (IQR 56 to 76), and the median functional MSTS-93 was 83% (IQR 73 to 91) postoperatively.Aims
Methods
Massive endoprostheses rely on extra-cortical bone bridging (ECBB)
to enhance fixation. The aim of this study was to investigate the
role of selective laser sintered (SLS) porous collars in augmenting
the osseointegration of these prostheses. The two novel designs of porous SLS collars, one with small pores
(Ø700 μm, SP) and one with large pores (Ø1500 μm, LP), were compared
in an ovine tibial diaphyseal model. Osseointegration of these collars
was compared with that of a clinically used solid, grooved design
(G). At six months post-operatively, the ovine tibias were retrieved and
underwent radiological and histological analysis.Aims
Materials and Methods
Endoprosthetic reconstruction following resection of 31 tumours of the proximal femur in 30 patients was performed using a Wagner SL femoral revision stem. The mean follow-up was 25.6 months (0.6 to 130.0). Of the 28 patients with a metastasis, 27 died within a mean follow-up period of 18.1 months (0.6 to 56.3) after the operation, and the remaining patient was excluded from the study 44.4 months post-operatively when the stem was removed. The two patients with primary bone tumours were still alive at the latest follow-up of 81.0 and 130.0 months, respectively. One stem only was removed for suspected low-grade infection 44.4 months post-operatively. The worst-case survival rate with removal of the stem for any cause and/or loss to follow-up was 80.0% (95% confidence interval 44.9 to 100) at 130.0 months. The mean Karnofsky index increased from 44.2% (20% to 70%) pre-operatively to 59.7% (0% to 100%) post-operatively, and the mean Merle d’Aubigné score improved from 4.5 (0 to 15) to 12.0 (0 to 18). The mean post-operative Musculoskeletal Tumour Society score was 62.4% (3.3% to 100%). The Wagner SL femoral revision stem offers an alternative to special tumour prostheses for the treatment of primary and secondary tumours of the proximal femur. The mid-term results are very promising, but long-term experience is necessary.
We reviewed 25 patients in whom a MUTARS megaprosthesis with a conical fluted stem had been implanted. There were three types of stem: a standard stem was used in 17 cases (three in the proximal femur, nine in the distal femur and five proximal tibia), a custom-made proximal femoral stem in four cases and a custom-made distal femoral stem in four cases. The mean age of the patients was 40.1 years (17 to 70) and the mean follow-up was for 2.5 years (0.9 to 7.4). At follow-up two patients had died from their disease: one was alive with disease and 22 were disease-free. One of 23 prostheses had been removed for infection and another revised to a cemented stem. The mean Musculoskeletal Tumor Society score was 24.9 (12 to 30) and the mean Karnofsky index was 82% (60% to 100%). There was no radiological evidence of loosening or subsidence. Stem
We present a retrospective study of patients suffering from a variety of benign tumours in whom external fixators were used to treat deformity and limb-length discrepancy, and for the reconstruction of bone defects. A total of 43 limbs in 31 patients (12 male and 19 female) with a mean age of 14 years (2 to 54) were treated. The diagnosis was Ollier’s disease in 12 limbs, fibrous dysplasia in 11, osteochondroma in eight, giant cell tumour in five, osteofibrous dysplasia in five and non-ossifying fibroma in two. The lesions were treated in the tibia in 19 limbs, in the femur in 16, and in the forearm in eight. The Ilizarov frame was used in 25 limbs, the Taylor Spatial Frame in seven, the Orthofix fixator in six, the Monotube in four and the Heidelberg fixator in one. The mean follow-up was 72 months (22 to 221). The mean external fixation period was 168 days (71 to 352). The mean external fixation index was 42 days/cm (22.2 to 102.0) in the 22 patients who required limb lengthening. The mean correction angle for those with angular deformity was 23° (7° to 45°). At final follow-up all patients had returned to normal activities. Four patients required a second operation for recurrent deformity of further limb lengthening. Local recurrence occurred in one patient, requiring further surgery.
We describe 25 patients who were treated for a tumour of the proximal femur by resection and replacement with an uncemented, bipolar, modular prosthesis. When followed up after more than ten years four prostheses (16%) had required revision. Two joints showed wear and another necrosis of the acetabulum. One patient with loosening of the stem had been treated by radiotherapy to the femur. Articular cartilage seemed to be a reliable barrier to acetabular wear. Very few signs of the formation of particulate debris were observed. The most obvious feature in the bone-stem relationship was