The aim of this study was to identify the optimal lip position for total hip arthroplasties (THAs) using a lipped liner. There is a lack of consensus on the optimal position, with substantial variability in surgeon practice. A model of a THA was developed using a 20° lipped liner. Kinematic analyses included a physiological range of motion (ROM) analysis and a provocative dislocation manoeuvre analysis. ROM prior to impingement was calculated and, in impingement scenarios, the travel distance prior to dislocation was assessed. The combinations analyzed included nine cup positions (inclination 30-40-50°, anteversion 5-15-25°), three stem positions (anteversion 0-15-30°), and five lip orientations (right hip 7 to 11 o’clock).Aims
Methods
Instability is a common cause of failure after total hip arthroplasty. A novel reverse total hip has been developed, with a femoral cup and acetabular ball, creating enhanced mechanical stability. The purpose of this study was to assess the implant fixation using radiostereometric analysis (RSA), and the clinical safety and efficacy of this novel design. Patients with end-stage osteoarthritis were enrolled in a prospective cohort at a single centre. The cohort consisted of 11 females and 11 males with mean age of 70.6 years (SD 3.5) and BMI of 31.0 kg/m2 (SD 5.7). Implant fixation was evaluated using RSA as well as Western Ontario and McMaster Universities Osteoarthritis Index, Harris Hip Score, Oxford Hip Score, Hip disability and Osteoarthritis Outcome Score, 38-item Short Form survey, and EuroQol five-dimension health questionnaire scores at two-year follow-up. At least one acetabular screw was used in all cases. RSA markers were inserted into the innominate bone and proximal femur with imaging at six weeks (baseline) and six, 12, and 24 months. Independent-samples Aims
Methods
Whether patient-reported pain differs among surgical approaches in total hip arthroplasty (THA) remains unclear. This study’s purposes were to determine differences in pain based on surgical approach (direct anterior (DA) This was a retrospective investigation from two centres and seven surgeons (three DA, three PL, one both) of primary THAs. PL patients were categorized for incision length (6 cm to 8 cm, 8 cm to 12 cm, 12 cm to 15 cm). All patients had cementless femoral and acetabular fixation, at least one year’s follow-up, and well-fixed components. Patients completed a pain-drawing questionnaire identifying the location and intensity of pain on an anatomical diagram. Power analysis indicated 800 patients in each cohort for adequate power to detect a 4% difference in pain (alpha = 0.05, beta = 0.80).Aims
Patients and Methods
The primary objective of this study was to compare accuracy in restoring the native centre of hip rotation in patients undergoing conventional manual total hip arthroplasty (THA) This prospective cohort study included 50 patients undergoing conventional manual THA and 25 patients receiving robotic-arm assisted THA. Patients undergoing conventional manual THA and robotic-arm assisted THA were well matched for age (mean age, 69.4 years (Objectives
Materials and Methods
The objective of this five-year prospective, blinded, randomised
controlled trial (RCT) was to compare femoral head penetration into
a vitamin E diffused highly cross-linked polyethylene (HXLPE) liner
with penetration into a medium cross-linked polyethylene control
liner using radiostereometric analysis. Patients scheduled for total hip arthroplasty (THA) were randomised
to receive either the study E1 (32 patients) or the control ArComXL
polyethylene (35 patients). The median age (range) of the overall
cohort was 66 years (40 to 76).Aims
Patients and Methods
One goal of total hip arthroplasty is to restore normal hip anatomy.
The aim of this study was to compare displacement of the centre
of rotation (COR) using a standard reaming technique with a technique
in which the acetabulum was reamed immediately peripherally and
referenced off the rim. In the first cohort the acetabulum was reamed to the floor followed
by sequentially larger reamers. In the second cohort the acetabulum
was only reamed peripherally, starting with a reamer the same size
as the native femoral head. Anteroposterior pelvic radiographs were
analysed for acetabular floor depth and vertical and horizontal position
of the COR.Aims
Patients and Methods
Given the increasing number of total hip arthroplasty
procedures being performed annually, it is imperative that orthopaedic
surgeons understand factors responsible for instability. In order
to treat this potentially complex problem, we recommend correctly
classifying the type of instability present based on component position, abductor
function, impingement, and polyethylene wear. Correct classification
allows the treating surgeon to choose the appropriate revision option
that ultimately will allow for the best potential outcome. Cite this article:
A high radiographic inclination angle (RI) contributes
to accelerated wear and has been associated with dislocation after
total hip arthroplasty (THA). With freehand positioning of the acetabular
component there is a lack of accuracy, with a trend towards a high
radiographic inclination angle. The aim of this study was to investigate
whether the use of a digital protractor to measure the operative
inclination angle (OI) could improve the positioning of the acetabular
component in relation to a ‘safe zone’. We measured the radiographic inclination angles of 200 consecutive
uncemented primary THAs. In the first 100 the component was introduced
freehand and in the second 100 a digital protractor was used to
measure the operative inclination angle. The mean difference between the operative and the radiographic
inclination angles
(∆RI–OI) in the second cohort was 12.3° (3.8° to 19.8°). There was
a strong correlation between the circumference of the hip and ∆RI–OI.
The number of RI outliers was significantly reduced in the protractor
group (p = 0.002). Adjusting the OI, using a digital protractor and taking into
account the circumference of the patient’s hip, improves the RI
significantly (p <
0.001) and does not require additional operating
time. Cite this article:
Dislocation remains among the most common complications
of, and reasons for, revision of both primary and revision total
hip replacements (THR). Hence, there is great interest in maximising
stability to prevent this complication. Head size has been recognised
to have a strong influence on the risk of dislocation post-operatively.
As femoral head size increases, stability is augmented, secondary
to an increase in impingement-free range of movement. Larger head
sizes also greatly increase the ‘jump distance’ required for the
head to dislocate in an appropriately positioned cup. Level-one
studies support the use of larger diameter heads as they decrease
the risk of dislocation following primary and revision THR. Highly cross-linked
polyethylene has allowed us to increase femoral head size, without
a marked increase in wear. However, the thin polyethylene liners
necessary to accommodate larger heads may increase the risk of liner
fracture and larger heads have also been implicated in causing soft-tissue
impingement resulting in groin pain. Larger diameter heads also
impart larger forces on the femoral trunnion, which may contribute
to corrosion, metal release, and adverse local tissue reactions.
Alternative large bearings including large ceramic heads and dual
mobility bearings may mitigate some of these risks, and several
of these devices have been used with clinical success. Cite this article:
We have reviewed 42 patients who had revision of metal-on-metal resurfacing procedures, mostly because of problems with the acetabular component. The revisions were carried out a mean of 26.2 months (1 to 76) after the initial operation and most of the patients (30) were female. Malpositioning of the acetabular component resulted in 27 revisions, mostly because of excessive abduction (mean 69.9°; 56° to 98°) or insufficient or excessive anteversion. Seven patients had more than one reason for revision. The mean increase in the diameter of the component was 1.8 mm (0 to 4) when exchange was needed. Malpositioning of the components was associated with metallosis and a high level of serum ions. The results of revision of the femoral component to a component with a modular head were excellent, but four patients had dislocation after revision and four required a further revision.
We studied 33 third generation, alumina ceramic-on-ceramic bearings retrieved from cementless total hip replacements after more than six months in situ. Wear volume was measured with a Roundtest machine, and acetabular orientation from the anteroposterior pelvic radiograph. The overall median early wear rate was 0.1 mm3/yr for the femoral heads, and 0.04 mm3/yr for the acetabular liners. We then excluded hips where the components had migrated. In this stable subgroup of 22 bearings, those with an acetabular anteversion of <
15° (seven femoral heads) had a median femoral head wear rate of 1.2 mm3/yr, compared with 0 mm3/yr for those with an anteversion of ≥15° (15 femoral heads, p <
0.001). Even under edge loading, wear volumes with ceramic-on-ceramic bearings are small in comparison to other bearing materials. Low acetabular anteversion is associated with greater wear.