header advert
Results 1 - 11 of 11
Results per page:
Bone & Joint Research
Vol. 12, Issue 6 | Pages 362 - 371
1 Jun 2023
Xu D Ding C Cheng T Yang C Zhang X

Aims

The present study aimed to investigate whether patients with inflammatory bowel disease (IBD) undergoing joint arthroplasty have a higher incidence of adverse outcomes than those without IBD.

Methods

A comprehensive literature search was conducted to identify eligible studies reporting postoperative outcomes in IBD patients undergoing joint arthroplasty. The primary outcomes included postoperative complications, while the secondary outcomes included unplanned readmission, length of stay (LOS), joint reoperation/implant revision, and cost of care. Pooled odds ratios (ORs) and 95% confidence intervals (CIs) were calculated using a random-effects model when heterogeneity was substantial.


Bone & Joint Open
Vol. 3, Issue 9 | Pages 716 - 725
15 Sep 2022
Boulton C Harrison C Wilton T Armstrong R Young E Pegg D Wilkinson JM

Data of high quality are critical for the meaningful interpretation of registry information. The National Joint Registry (NJR) was established in 2002 as the result of an unexpectedly high failure rate of a cemented total hip arthroplasty. The NJR began data collection in 2003. In this study we report on the outcomes following the establishment of a formal data quality (DQ) audit process within the NJR, within which each patient episode entry is validated against the hospital unit’s Patient Administration System and vice-versa. This process enables bidirectional validation of every NJR entry and retrospective correction of any errors in the dataset. In 2014/15 baseline average compliance was 92.6% and this increased year-on-year with repeated audit cycles to 96.0% in 2018/19, with 76.4% of units achieving > 95% compliance. Following the closure of the audit cycle, an overall compliance rate of 97.9% was achieved for the 2018/19 period. An automated system was initiated in 2018 to reduce administrative burden and to integrate the DQ process into standard workflows. Our processes and quality improvement results demonstrate that DQ may be implemented successfully at national level, while minimizing the burden on hospitals.

Cite this article: Bone Jt Open 2022;3(9):716–725.


The Bone & Joint Journal
Vol. 102-B, Issue 12 | Pages 1682 - 1688
1 Dec 2020
Corona PS Vicente M Carrera L Rodríguez-Pardo D Corró S

Aims

The success rates of two-stage revision arthroplasty for infection have evolved since their early description. The implementation of internationally accepted outcome criteria led to the readjustment of such rates. However, patients who do not undergo reimplantation are usually set aside from these calculations. The aim of this study was to investigate the outcomes of two-stage revision arthroplasty when considering those who do not undergo reimplantation, and to investigate the characteristics of this subgroup.

Methods

A retrospective cohort study was conducted. Patients with chronic hip or knee periprosthetic joint infection (PJI) treated with two-stage revision between January 2010 and October 2018, with a minimum follow-up of one year, were included. Variables including demography, morbidity, microbiology, and outcome were collected. The primary endpoint was the eradication of infection. Patients who did not undergo reimplantation were analyzed in order to characterize this subgroup better.


The Bone & Joint Journal
Vol. 101-B, Issue 10 | Pages 1248 - 1255
1 Oct 2019
Pineda A Pabbruwe MB Kop AM Vlaskovsky P Hurworth M

Aims. The aim of this study was to conduct the largest low contact stress (LCS) retrieval study to elucidate the failure mechanisms of the Porocoat and Duofix femoral component. The latter design was voluntarily recalled by the manufacturer. Materials and Methods. Uncemented LCS explants were divided into three groups: Duofix, Porocoat, and mixed. Demographics, polyethylene wear, tissue ingrowth, and metallurgical analyses were performed. Results. In 104 implants, a decrease in the odds of loosening and an increase in metallosis and tissue staining in the Duofix group relative to Porocoat group was detected (p = 0.028). There was an increased presence of embedded metallic debris in the Duofix group (p < 0.001). Decreased tissue ingrowth was associated with the Duofix surface (p < 0.001). The attached beads had reduced microhardness, indicative of adverse thermal processing, which resulted in bead shedding, particulate debris, and metallosis. Conclusion. Hydroxyapatite coating of the LCS femoral component produced unexpected results and led to its recall. The root cause was likely a combination of retained alumina grit and a reduction in bead microhardness (mechanical strength) resulting in increased particle debris, metallosis, and early revision. The Duofix LCS femoral component was not equivalent to the Porocoat version despite its approval through the Food and Drug Administration (FDA) 510(k) equivalance approval process. Regulation of the introduction of modified existing devices needs to be improved and the Duofix LCS should have been considered to be a new device for which equivalence had not been demonstrated at the point of introduction. Cite this article: Bone Joint J 2019;101-B:1248–1255


The Bone & Joint Journal
Vol. 101-B, Issue 8 | Pages 960 - 969
1 Aug 2019
Odgaard A Laursen MB Gromov K Troelsen A Kristensen PW Schrøder H Madsen F Overgaard S

Aims

The aim of this study was to give estimates of the incidence of component incompatibility in hip and knee arthroplasty and to test the effect of an online, real-time compatibility check.

Materials and Methods

Intraoperative barcode registration of arthroplasty implants was introduced in Denmark in 2013. We developed a compatibility database and, from May 2017, real-time compatibility checking was implemented and became part of the registration. We defined four classes of component incompatibility: A-I, A-II, B-I, and B-II, depending on an assessment of the level of risk to the patient (A/B), and on whether incompatibility was knowingly accepted (I/II).


The Bone & Joint Journal
Vol. 100-B, Issue 12 | Pages 1592 - 1599
1 Dec 2018
Galea VP Connelly JW Shareghi B Kärrholm J Sköldenberg O Salemyr M Laursen MB Muratoglu O Bragdon C Malchau H

Aims

The primary aim of this study was to compare the wear properties of vitamin E-diffused, highly crosslinked polyethylene (VEPE) and one formulation of moderately crosslinked and mechanically annealed ultra-high molecular weight polyethylene (ModXLPE) in patients five years after primary total hip arthroplasty (THA). The secondary aim was to assess the clinical results of patients treated with VEPE by evaluating patient-reported outcome measures (PROMs), radiological evidence of fixation, and the incidence of mechanical failure.

Patients and Methods

A total of 208 patients (221 THAs) from four international centres were recruited into a prospective study involving radiostereometric analysis (RSA) and the assessment of clinical outcomes. A total of 193 hips (87%) were reviewed at the five-year follow-up. Of these, 136 (70%) received VEPE (vs ModXLPE) liners and 68 (35%) received ceramic (vs metal) femoral heads. PROMs and radiographs were collected preoperatively and at one, two, and five years postoperatively. In addition, RSA images were collected to measure PE wear postoperatively and at one, two, and five years after surgery.


The Bone & Joint Journal
Vol. 98-B, Issue 8 | Pages 1062 - 1068
1 Aug 2016
Singh G Deutloff N Maertens N Meyer H Awiszus F Feuerstein B Roessner A Lohmann CH

Aims

Tissue responses to debris formed by abrasion of polymethylmethacrylate (PMMA) spacers at two-stage revision arthroplasty for prosthetic joint infection are not well described. We hypothesised that PMMA debris induces immunomodulation in periprosthetic tissues.

Patients and Methods

Samples of tissue were taken during 35 two-stage revision arthroplasties (nine total hip and 26 total knee arthroplasties) in patients whose mean age was 67 years (44 to 85). Fourier transform infrared microscopy was used to confirm the presence of PMMA particles. Histomorphometry was performed using Sudan Red and Haematoxylin-Eosin staining. CD-68, CD-20, CD-11(c), CD-3 and IL-17 antibodies were used to immunophenotype the inflammatory cells. All slides were scored semi-quantitatively using the modified Willert scoring system.


Bone & Joint Research
Vol. 5, Issue 4 | Pages 122 - 129
1 Apr 2016
Small SR Rogge RD Malinzak RA Reyes EM Cook PL Farley KA Ritter MA

Objectives

Initial stability of tibial trays is crucial for long-term success of total knee arthroplasty (TKA) in both primary and revision settings. Rotating platform (RP) designs reduce torque transfer at the tibiofemoral interface. We asked if this reduced torque transfer in RP designs resulted in subsequently reduced micromotion at the cemented fixation interface between the prosthesis component and the adjacent bone.

Methods

Composite tibias were implanted with fixed and RP primary and revision tibial trays and biomechanically tested under up to 2.5 kN of axial compression and 10° of external femoral component rotation. Relative micromotion between the implanted tibial tray and the neighbouring bone was quantified using high-precision digital image correlation techniques.


Bone & Joint Research
Vol. 3, Issue 3 | Pages 60 - 68
1 Mar 2014
Langton DJ Sidaginamale RP Holland JP Deehan D Joyce TJ Nargol AVF Meek RD Lord JK

Objectives

Wear debris released from bearing surfaces has been shown to provoke negative immune responses in the recipient. Excessive wear has been linked to early failure of prostheses. Analysis using coordinate measuring machines (CMMs) can provide estimates of total volumetric material loss of explanted prostheses and can help to understand device failure. The accuracy of volumetric testing has been debated, with some investigators stating that only protocols involving hundreds of thousands of measurement points are sufficient. We looked to examine this assumption and to apply the findings to the clinical arena.

Methods

We examined the effects on the calculated material loss from a ceramic femoral head when different CMM scanning parameters were used. Calculated wear volumes were compared with gold standard gravimetric tests in a blinded study.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 9 | Pages 1225 - 1228
1 Sep 2007
van de Pol GJ Sturm PDJ van Loon CJ Verhagen C Schreurs BW

Allografts of bone from the femoral head are often used in orthopaedic procedures. Although the donated heads are thoroughly tested microscopically before release by the bone bank, some surgeons take additional cultures in the operating theatre before implantation. There is no consensus about the need to take these cultures. We retrospectively assessed the clinical significance of the implantation of positive-cultured bone allografts. The contamination rate at retrieval of the allografts was 6.4% in our bone bank. Intra-operative cultures were taken from 426 femoral head allografts before implantation; 48 (11.3%) had a positive culture. The most frequently encountered micro-organism was coagulase-negative staphylococcus. Deep infection occurred in two of the 48 patients (4.2%). In only one was it likely that the same micro-organism caused the contamination and the subsequent infection. In our study, the rate of infection in patients receiving positive-cultured allografts at implantation was not higher than the overall rate of infection in allograft surgery suggesting that the positive cultures at implantation probably represent contamination and that the taking of additional cultures is not useful


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 11 | Pages 1524 - 1526
1 Nov 2005
Kwong FNK Ibrahim T Power RA

Implantation of allograft bone is an integral part of revision surgery of the hip. One major concern with its use is the risk of transmission of infective agents. There are a number of methods of processing allograft bone in order to reduce this risk. One method requires washing the tissue using pulsed irrigation immediately before implantation. We report the incidence of deep bacterial infection in 138 patients (144 revision hip arthroplasties) who had undergone implantation of allograft bone. The bone used was fresh-frozen, non-irradiated and pulse-washed with normal saline before implantation. The deep infection rate at a minimum follow-up of one year was 0.7%. This method of processing appears to be associated with a very low risk of allograft-related bacterial infection.