The surgical helmet system (SHS) was developed to reduce the risk of periprosthetic joint infection (PJI), but the evidence is contradictory, with some studies suggesting an increased risk of PJI due to potential leakage through the glove-gown interface (GGI) caused by its positive pressure. We assumed that SHS and glove exchange had an impact on the leakage via GGI. There were 404 arthroplasty simulations with fluorescent gel, in which SHS was used (H+) or not (H-), and GGI was sealed (S+) or not (S-), divided into four groups: H+S+, H+S-, H-S+, and H-S-, varying by exposure duration (15 to 60 minutes) and frequency of glove exchanges (0 to 6 times). The intensity of fluorescent leakage through GGI was quantified automatically with an image analysis software. The effect of the above factors on fluorescent leakage via GGI were compared and analyzed.Aims
Methods
Aims. The aim of this study was to conduct the largest low contact stress (LCS) retrieval study to elucidate the failure mechanisms of the Porocoat and Duofix femoral component. The latter design was voluntarily recalled by the manufacturer. Materials and Methods. Uncemented LCS explants were divided into three groups: Duofix, Porocoat, and mixed. Demographics, polyethylene wear, tissue ingrowth, and metallurgical analyses were performed. Results. In 104 implants, a decrease in the odds of loosening and an increase in metallosis and tissue staining in the Duofix group relative to Porocoat group was detected (p = 0.028). There was an increased presence of embedded metallic debris in the Duofix group (p < 0.001). Decreased tissue ingrowth was associated with the Duofix surface (p < 0.001). The attached beads had reduced microhardness, indicative of adverse thermal processing, which resulted in bead shedding, particulate debris, and metallosis. Conclusion. Hydroxyapatite coating of the LCS femoral component produced unexpected results and led to its recall. The root cause was likely a combination of retained alumina grit and a reduction in bead microhardness (mechanical strength) resulting in increased
The aim of this study was to evaluate the diagnostic accuracy of the synovial alpha-defensin enzyme-linked immunosorbent assay (ELISA) for the diagnosis of prosthetic joint infection (PJI) in the work-up prior to revision of total hip (THA) and knee arthroplasty (TKA). Inclusion criteria for this prospective cohort study were acute or chronic symptoms of the index joint without specific exclusion criteria. Synovial fluid aspirates of 202 patients were analyzed and semiquantitative laboratory alpha-defensin ELISA was performed. Final diagnosis of PJI was established by examination of samples obtained during revision surgery.Aims
Patients and Methods
Increasing innovation in rapid prototyping (RP)
and additive manufacturing (AM), also known as 3D printing, is bringing
about major changes in translational surgical research. This review describes the current position in the use of additive
manufacturing in orthopaedic surgery. Cite this article:
The interaction between surgical lighting and laminar airflow
is poorly understood. We undertook an experiment to identify any
effect contemporary surgical lights have on laminar flow and recommend
practical strategies to limit any negative effects. Neutrally buoyant bubbles were introduced into the surgical field
of a simulated setup for a routine total knee arthroplasty in a
laminar flow theatre. Patterns of airflow were observed and the
number of bubbles remaining above the surgical field over time identified.
Five different lighting configurations were assessed. Data were analysed
using simple linear regression after logarithmic transformation.Aims
Materials and Methods
Infection following total hip or knee arthroplasty is a serious
complication. We noted an increase in post-operative infection in
cases carried out in temporary operating theatres. We therefore
compared those cases performed in standard and temporary operating
theatres and examined the deep periprosthetic infection rates. A total of 1223 primary hip and knee arthroplasties were performed
between August 2012 and June 2013. A total of 539 (44%) were performed
in temporary theatres. The two groups were matched for age, gender,
body mass index and American Society of Anesthesiologists grade.Aims
Patients and methods
Aims. Tissue responses to debris formed by abrasion of polymethylmethacrylate
(PMMA) spacers at two-stage revision arthroplasty for prosthetic
joint infection are not well described. We hypothesised that PMMA
debris induces immunomodulation in periprosthetic tissues. Patients and Methods. Samples of tissue were taken during 35 two-stage revision arthroplasties
(nine total hip and 26 total knee arthroplasties) in patients whose
mean age was 67 years (44 to 85). Fourier transform infrared microscopy
was used to confirm the presence of PMMA
Computed tomography (CT) plays an important role in evaluating wear and periacetabular osteolysis (PAO) in total hip replacements. One concern with CT is the high radiation exposure since standard pelvic CT provides approximately 3.5 millisieverts (mSv) of radiation exposure, whereas a planar radiographic examination with three projections totals approximately 0.5 mSv. The objective of this study was to evaluate the lowest acceptable radiation dose for dual-energy CT (DECT) images when measuring wear and periacetabular osteolysis in uncemented metal components. A porcine pelvis with bilateral uncemented hip prostheses and with known linear wear and acetabular bone defects was examined in a third-generation multidetector DECT scanner. The examinations were performed with four different radiation levels both with and without iterative reconstruction techniques. From the high and low peak kilo voltage acquisitions, polychrmoatic images were created together with virtual monochromatic images of energies 100 kiloelectron volts (keV) and 150 keV.Objectives
Materials and Methods
The aim of this study was to evaluate whether
coating titanium discs with selenium in the form of sodium selenite decreased
bacterial adhesion of In order to evaluate bacterial adhesion, sterile titanium discs
were coated with increasing concentrations of selenium and incubated
with bacterial solutions of The tested Selenium coating is a promising method to reduce bacterial attachment
on prosthetic material. Cite this article:
We have recently shown that waste heat from forced-air
warming blankets can increase the temperature and concentration
of airborne
Patient warming significantly decreases the risk
of surgical site infection. Recently there have been concerns that forced
air warming may interfere with unidirectional airflow, potentially
posing an increased risk of infection. Our null hypothesis was that
forced air and radiant warming devices do not increase the temperature
and the number of
We have investigated whether the use of laminar-flow theatres and space suits reduced the rate of revision for early deep infection after total hip (THR) and knee (TKR) replacement by reviewing the results of the New Zealand Joint Registry at ten years. Of the 51 485 primary THRs and 36 826 primary TKRs analysed, laminar-flow theatres were used in 35.5% and space suits in 23.5%. For THR there was a significant increase in early infection in those procedures performed with the use of a space suit compared with those without (p <
0.0001), in those carried out in a laminar-flow theatre compared with a conventional theatre (p <
0.003) and in those undertaken in a laminar-flow theatre with a space suit (p <
0.001) when compared with conventional theatres without such a suit. The results were similar for TKR with the use of a space suit (p <
0.001), in laminar-flow theatres (p <
0.019) and when space suits were used in those theatres (p <
0.001). These findings were independent of age, disease and operating time and were unchanged when the surgeons and hospital were analysed individually. The rate of revision for early deep infection has not been reduced by using laminar flow and space suits. Our results question the rationale for their increasing use in routine joint replacement, where the added cost to the health system seems to be unjustified.
This prospective study evaluates the role of new laboratory markers in the diagnosis of deep implant infection in 78 patients (41 men and 37 women) with a revision total knee or hip replacement. The mean age at the time of operation was 64.0 years (19 to 90). Intra-operative cultures showed that 21 patients had a septic and 57 an aseptic total joint replacement. The white blood cell count, the erythrocyte sedimentation rate and levels of C-reactive protein, interleukin-6, procalcitonin and tumour necrosis factor (TNF)-α were measured in blood samples before operation. The diagnostic cut-off values were determined by Received Operating Characteristic curve analysis. C-reactive protein (>
3.2 md/dl) and interleukin-6 (>
12 pg/ml) have the highest sensitivity (0.95). Interleukin-6 is less specific than C-reactive protein (0.87 The combination of C-reactive protein and interleukin-6 measurement provide excellent screening tests for infection of a deep implant. A highly specific marker such as procalcitonin and pre-operative aspiration of the joint might be useful in identifying patients with true positive C-reactive protein and/or interleukin-6 levels.