Patients undergoing femoral lengthening by external fixation
tolerate treatment less well when compared to tibial lengthening.
Lengthening of the femur with an intramedullary device may have
advantages. We reviewed all cases of simple femoral lengthening performed
at our unit from 2009 to 2014. Cases of nonunions, concurrent deformities,
congenital limb deficiencies and lengthening with an unstable hip
were excluded, leaving 33 cases (in 22 patients; 11 patients had
bilateral procedures) for review. Healing index, implant tolerance
and complications were compared.Aims
Patients and Methods
Although gradual bone transport may permit the
restoration of large-diameter bones, complications are common owing
to the long duration of external fixation. In order to reduce such
complications, a new technique of bone transport involving the use
of an external fixator and a locking plate was devised for segmental
tibial bone defects. A total of ten patients (nine men, one woman) with a mean age
at operation of 40.4 years (16 to 64) underwent distraction osteogenesis
with a locking plate to treat previously infected post-traumatic
segmental tibial defects. The locking plate was fixed percutaneously
to bridge proximal and distal segments, and was followed by external fixation.
After docking, percutaneous screws were fixed at the transported
segment through plate holes. At the same time, bone grafting was
performed at the docking site with the external fixator removed. The mean defect size was 5.9 cm (3.8 to 9.3) and mean external
fixation index was
13.4 days/cm (11.8 to 19.5). In all cases, primary union of the
docking site and distraction callus was achieved, with an excellent
bony result. There was no recurrence of deep infection or osteomyelitis,
and with the exception of one patient with a pre-existing peroneal
nerve injury, all achieved an excellent or good functional result. With short external fixation times and low complication rates,
bone transport with a locking plate could be recommended for patients
with segmental tibial defects. Cite this article:
Internal lengthening devices in the femur lengthen
along the anatomical axis, potentially creating lateral shift of
the mechanical axis. We aimed to determine whether femoral lengthening
along the anatomical axis has an inadvertent effect on lower limb
alignment. Isolated femoral lengthening using the Intramedullary
Skeletal Kinetic Distractor was performed in 27 femora in 24 patients
(mean age 32 years (16 to 57)). Patients who underwent simultaneous realignment
procedures or concurrent tibial lengthening, or who developed mal-
or nonunion, were excluded. Pre-operative and six-month post-operative
radiographs were used to measure lower limb alignment. The mean lengthening
achieved was 4.4 cm (1.5 to 8.0). In 26 of 27 limbs, the mechanical
axis shifted laterally by a mean of 1.0 mm/cm of lengthening (0
to 3.5). In one femur that was initially in varus, a 3 mm medial
shift occurred during a lengthening of 2.2 cm. In a normally aligned
We report the results of intramedullary leg lengthening conducted between 2002 and 2009 using the Intramedullary Skeletal Kinetic Distractor in 69 unilateral lengthenings involving 58 femora and 11 tibiae. We identified difficulties that occurred during the treatment and assessed whether they were specifically due to the implant or independent of it. Paley’s classification for evaluating problems, obstacles and complications with external fixators was adopted, and implant-specific difficulties were continuously noted. There were seven failures requiring premature removal of the device, in four due to nail breakage and three for other reasons, and five unsuccessful outcomes after completion of the lengthening. In all, 116 difficulties were noted in 45 patients, with only 24 having problem-free courses. In addition to the difficulties arising from the use of external fixators, there were almost the same number again of implant-specific difficulties. Nevertheless, successful femoral lengthening was achieved in 52 of the 58 patients (90%). However, successful tibial lengthening was only achieved in five of 11 patients (45%).
We compared the complications and outcome of tibial lengthening using the Ilizarov method with and without the use of a supplementary intramedullary nail. In a retrospective case-matched series assembled from 176 patients with tibial lengthening, we matched 52 patients (26 pairs, group A with nail and group B without) according to the following criteria in order of importance: 1) difference in amount of lengthening (± 2 cm); 2) percentage difference in lengthening (± 5%); 3) difference in patient’s age (± seven years); 4) aetiology of the shortening, and 5) level of difficulty in obtaining the correction. The outcome was evaluated using the external fixator index, the healing index and an outcome score according to the criteria of Paley. It was found that some complications were specific to group A or B respectively, but others were common to both groups. The outcome was generally better in lengthenings with a nail, although there was a higher incidence of rectifiable equinus deformity in these patients.
Between October 2001 and September 2009 we lengthened 242 lower-limb segments in 180 patients using the Intramedullary Skeletal Kinetic Distractor (ISKD). Mechanical failure was defined either as breakage of the ISKD or failure of the internal mechanism to activate. Retrieved nails which failed mechanically were examined by the manufacturer for defects. In all, 15 ISKDs in 12 patients (13 limbs) failed mechanically representing an overall failure rate of 6.2%, with fracture of the device occurring in ten of the 15 failures. Two nails in one patient failed to lengthen and had to be replaced. The manufacturer detected an error in the assembly of the nail, which prompted a wide recall. One nail jammed after being forcefully inserted, and two nails failed to lengthen fully. Lengthening was achieved in all 12 patients, although three required a second operation to exchange a defective nail for a new, functioning device. The ISKD is a complex mechanical device which lengthens by the oscillation of two telescopic sections connected by a threaded rod. The junction between these sections is surrounded by a keyring collar. This keyring collar is the weakest part of the device.