In computer simulations, the shape of the range of motion (ROM) of a stem with a cylindrical neck design will be a perfect cone. However, many modern stems have rectangular/oval-shaped necks. We hypothesized that the rectangular/oval stem neck will affect the shape of the ROM and the prosthetic impingement. Total hip arthroplasty (THA) motion while standing and sitting was simulated using a MATLAB model (one stem with a cylindrical neck and one stem with a rectangular neck). The primary predictor was the geometry of the neck (cylindrical vs rectangular) and the main outcome was the shape of ROM based on the prosthetic impingement between the neck and the liner. The secondary outcome was the difference in the ROM provided by each neck geometry and the effect of the pelvic tilt on this ROM. Multiple regression was used to analyze the data.Aims
Methods
Large femoral heads have been used with increasing
frequency over the last decade. The prime reason is likely the effect
of large heads on stability. The larger head neck ratio, combined
with the increased jump distance of larger heads result in a greater
arc of impingement free motion, and greater resistance to dislocation
in a provocative position. Multiple studies have demonstrated clear
clinical efficacy in diminishing dislocation rates with the use
of large femoral heads. With crosslinked polyethylene, wear has
been shown to be equivalent between larger and smaller heads. However,
the stability advantages of increasing diameter beyond 38 mm have
not been clearly demonstrated. More importantly, recent data implicates
large heads in the increasing prevalence of groin pain and psoas impingement.
There are clear benefits with larger femoral head diameters, but
the advantages of diameters beyond 38 mm have not yet been demonstrated
clinically.
In this prospective study we studied the effect
of the inclination angle of the acetabular component on polyethylene wear
and component migration in cemented acetabular sockets using radiostereometric
analysis. A total of 120 patients received either a cemented Reflection
All-Poly ultra-high-molecular-weight polyethylene or a cemented
Reflection All-Poly highly cross-linked polyethylene acetabular
component, combined with either cobalt–chrome or Oxinium femoral
heads. Femoral head penetration and migration of the acetabular
component were assessed with repeated radiostereometric analysis
for two years. The inclination angle was measured on a standard
post-operative anteroposterior pelvic radiograph. Linear regression
analysis was used to determine the relationship between the inclination
angle and femoral head penetration and migration of the acetabular component. We found no relationship between the inclination angle and penetration
of the femoral head at two years’ follow-up (p = 0.9). Similarly,
our data failed to reveal any statistically significant correlation
between inclination angle and migration of these cemented acetabular
components (p = 0.07 to p = 0.9).
The purpose of this study was to define immediate post-operative ‘quality’ in total hip replacements and to study prospectively the occurrence of failure based on these definitions of quality. The evaluation and assessment of failure were based on ten radiological and clinical criteria. The cumulative summation (CUSUM) test was used to study 200 procedures over a one-year period. Technical criteria defined failure in 17 cases (8.5%), those related to the femoral component in nine (4.5%), the acetabular component in 32 (16%) and those relating to discharge from hospital in five (2.5%). Overall, the procedure was considered to have failed in 57 of the 200 total hip replacements (28.5%). The use of a new design of acetabular component was associated with more failures. For the CUSUM test, the level of adequate performance was set at a rate of failure of 20% and the level of inadequate performance set at a failure rate of 40%; no alarm was raised by the test, indicating that there was no evidence of inadequate performance. The use of a continuous monitoring statistical method is useful to ensure that the quality of total hip replacement is maintained, especially as newer implants are introduced.