We aimed to investigate if the use of the largest possible cobalt-chromium head articulating with polyethylene acetabular inserts would increase the in vivo wear rate in total hip arthroplasty. In a single-blinded randomized controlled trial, 96 patients (43 females), at a median age of 63 years (interquartile range (IQR) 57 to 69), were allocated to receive either the largest possible modular femoral head (36 mm to 44 mm) in the thinnest possible insert or a standard 32 mm head. All patients received a vitamin E-doped cross-linked polyethylene insert and a cobalt-chromium head. The primary outcome was proximal head penetration measured with radiostereometric analysis (RSA) at two years. Secondary outcomes were volumetric wear, periacetabular radiolucencies, and patient-reported outcomes.Aims
Methods
Aims. Despite advances in the treatment of paediatric hip disease, adolescent and young adult patients can develop early onset end-stage osteoarthritis. The aims of this study were to address the indications and medium-term outcomes for total hip arthroplasty (THA) with
To determine the outcomes following revision surgery of metal-on-metal
hip arthroplasties (MoMHA) performed for adverse reactions to metal
debris (ARMD), and to identify factors predictive of re-revision. We performed a retrospective observational study using National
Joint Registry (NJR) data on 2535 MoMHAs undergoing revision surgery
for ARMD between 2008 and 2014. The outcomes studied following revision were
intra-operative complications, mortality and re-revision surgery.
Predictors of re-revision were identified using competing-risk regression
modelling.Aims
Patients and Methods
Few studies have assessed outcomes following non-metal-on-metal hip arthroplasty (non-MoMHA) revision surgery performed for adverse reactions to metal debris (ARMD). We assessed outcomes following non-MoMHA revision surgery performed for ARMD, and identified predictors of re-revision. We performed a retrospective observational study using data from the National Joint Registry for England and Wales. All non-MoMHAs undergoing revision surgery for ARMD between 2008 and 2014 were included (185 hips in 185 patients). Outcome measures following ARMD revision were intra-operative complications, mortality and re-revision surgery. Predictors of re-revision were identified using Cox regression.Objectives
Methods
To determine the effect of a change in design of a cementless
ceramic acetabular component in fixation and clinical outcome after
total hip arthroplasty We compared 342 hips (302 patients) operated between 1999 and
2005 with a relatively smooth hydroxyapatite coated acetabular component
(group 1), and 337 hips (310 patients) operated between 2006 and
2011 using a similar acetabular component with a macrotexture on
the entire outer surface of the component (group 2). The mean age of
the patients was 53.5 (14 to 70) in group 1 and 53.0 (15 to 70)
in group 2. The mean follow-up was 12.7 years (10 to 17) for group
1 and 7.2 years (4 to 10) for group 2.Aims
Patients and Methods
Femoral stems with exchangeable (modular) necks were introduced
to offer surgeons an increased choice when determining the version,
offset and length of the femoral neck during total hip arthroplasty
(THA). It was hoped that this would improve outcomes and reduce
complications, particularly dislocation. In 2010, the Australian Orthopaedic
Association National Joint Replacement Registry (AOANJRR) first
reported an increased rate of revision after primary THA using femoral
stems with an exchangeable neck. The aim of this study was to provide
a more comprehensive up-to-date analysis of primary THA using femoral
stems with exchangeable and fixed necks. The data included all primary THA procedures performed for osteoarthritis
(OA), reported to the AOANJRR between 01 September 1999 and 31 December
2014. There were 9289 femoral stems with an exchangeable neck and
253 165 femoral stems with a fixed neck. The characteristics of
the patients and prostheses including the bearing surface and stem/neck
metal combinations were examined using Cox proportional hazard ratios
(HRs) and Kaplan-Meier estimates of survivorship. Aims
Materials and Methods
Modular or custom-made femoral components have been preferred
for total hip arthroplasty (THA) in patients with a history of Perthes’
disease because of the distortion in the anatomy of the proximal
femur. However, it has not been established whether a monobloc cementless
stem will fit the distorted proximal femur or whether the results
of the procedure are satisfactory in this group of patients. We reviewed 68 consecutive patients who had undergone THA for
childhood Perthes’ disease between June 2003 and December 2008.
There were 35 men and 33 women with a mean age of 48 years (16 to
73) at the time of index arthroplasty. Their mean body mass index
was 24.4 (18.3 to 32.9). Of the 68 hips, 32 were classified as Stulberg
class III and 36 as class IV. The mean pre-operative shortening
of the affected leg was 17.2 mm (5 to 34). The minimum follow-up
was five years (mean 8.5 years; 5.2 to 10).Aims
Patients and Methods
We evaluated the short-term functional outcome and prevalence
of bearing-specific generation of audible noise in 301 patients
(336 hips) operated on with fourth generation (Delta) medium diameter
head, ceramic-on-ceramic (CoC) total hip arthroplasties (THAs). There were 191 female (63%) and 110 male patients (37%) with
a mean age of 61 years (29 to 78) and mean follow-up of 2.1 years
(1.3 to 3.4). Patients completed three questionnaires: Oxford Hip
Score (OHS), Research and Development 36-item health survey (RAND-36)
and a noise-specific symptom questionnaire. Plain radiographs were also
analysed. A total of three hips (0.9%) were revised. Aims
Patients and Methods
Studies reporting specifically on squeaking in total hip arthroplasty have focused on cementless, and not on hybrid, fixation. We hypothesised that the cement mantle of the femur might have a damping effect on the sound transmitted through the metal stem. The objective of this study was to test the effect of cement on sound propagation along different stem designs and under different fixation conditions. An Objectives
Methods
Modular junctions are ubiquitous in contemporary hip arthroplasty. The head-trunnion junction is implicated in the failure of large diameter metal-on-metal (MoM) hips which are the currently the topic of one the largest legal actions in the history of orthopaedics (estimated costs are stated to exceed $4 billion). Several factors are known to influence the strength of these press-fit modular connections. However, the influence of different head sizes has not previously been investigated. The aim of the study was to establish whether the choice of head size influences the initial strength of the trunnion-head connection. Ti-6Al-4V trunnions (n = 60) and two different sizes of cobalt-chromium (Co-Cr) heads (28 mm and 36 mm; 30 of each size) were used in the study. Three different levels of assembly force were considered: 4 kN; 5 kN; and 6 kN (n = 10 each). The strength of the press-fit connection was subsequently evaluated by measuring the pull-off force required to break the connection. The statistical differences in pull-off force were examined using a Kruskal–Wallis test and two-sample Mann–Whitney U test. Finite element and analytical models were developed to understand the reasons for the experimentally observed differences.Objectives
Materials and Methods
As adverse events related to metal on metal hip
arthroplasty have been better understood, there has been increased
interest in toxicity related to the high circulating levels of cobalt ions.
However, distinguishing true toxicity from benign elevations in
cobalt levels can be challenging. The purpose of this review is
to examine the use of cobalt alloys in total hip arthroplasty, to
review the methods of measuring circulating cobalt levels, to define
a level of cobalt which is considered pathological and to review
the pathophysiology, risk factors and treatment of cobalt toxicity.
To the best of our knowledge, there are 18 published cases where
cobalt metal ion toxicity has been attributed to the use of cobalt-chromium
alloys in hip arthroplasty. Of these cases, the great majority reported
systemic toxic reactions at serum cobalt levels more than 100 μg/L.
This review highlights some of the clinical features of cobalt toxicity,
with the goal that early awareness may decrease the risk factors
for the development of cobalt toxicity and/or reduce its severity. Take home message: Severe adverse events can arise from the release
of cobalt from metal-on-metal arthroplasties, and as such, orthopaedic
surgeons should not only be aware of the presenting problems, but
also have the knowledge to treat appropriately. Cite this article:
Oxidised zirconium (OxZi) has been developed
as an alternative bearing surface for femoral heads in total hip arthroplasty
(THA). This study has investigated polyethylene wear, functional
outcomes and complications, comparing OxZi and cobalt–chrome (CoCr)
as part of a three-arm, multicentre randomised controlled trial.
Patients undergoing THA from four institutions were prospectively
randomised into three groups. Group A received a CoCr femoral head
and highly cross-linked polyethylene (XLPE) liner; Group B received
an OxZi femoral head and XLPE liner; Group C received an OxZi femoral
head and ultra-high molecular weight polyethylene (UHMWPE) liner.
At five years, 368 patients had no statistically significant differences
in short-form-36 (p = 0.176 mental, p = 0.756 physical), Western
Ontario and McMaster Universities Osteoarthritis Index (p = 0.847),
pain scores
(p = 0.458) or complications. The mean rate of linear wear was 0.028
mm/year (standard deviation ( Cite this article:
We report the clinical and radiological outcomes
of a series of contemporary cementless ceramic-on-ceramic total hip
replacements (THRs) at ten years in patients aged ≤ 55 years of
age. Pre- and post-operative activity levels are described. A total
of 120 consecutive ceramic cementless THRs were performed at a single
centre in 110 patients from 1997 to 1999. The mean age of the patients
at operation was 45 years (20 to 55). At ten years, four patients
had died and six were lost to follow-up, comprising ten hips. The
mean post-operative Harris hip score was 94.7 (55 to 100). Radiological
analysis was undertaken in 90 available THRs of the surviving 106
hips at final review: all had evidence of stable bony ingrowth,
with no cases of osteolysis. Wear was undetectable. There were four
revisions. The survival for both components with revision for any
cause as an endpoint was 96.5% (95% confidence interval 94.5 to
98.7). The mean modified University of California, Los Angeles activity
level rose from a mean of 6.4 (4 to 10) pre-operatively to 9.0 (6
to 10) at the ten-year post-operative period. Alumina ceramic-on-ceramic bearings in cementless primary THR
in this series have resulted in good clinical and radiological outcomes
with undetectable rates of wear and excellent function in the demanding
younger patient group at ten years. Cite this article:
The practice of removing a well-fixed cementless
femoral component is associated with high morbidity. Ceramic bearing
couples are low wearing and their use minimises the risk of subsequent
further revision due to the production of wear debris. A total of
165 revision hip replacements were performed, in which a polyethylene-lined acetabular
component was revised to a new acetabular component with a ceramic
liner, while retaining the well-fixed femoral component. A titanium
sleeve was placed over the used femoral trunnion, to which a ceramic
head was added. There were 100 alumina and 65 Delta bearing couples
inserted. The mean Harris hip score improved significantly from 71.3 (9.0
to 100.0) pre-operatively to 91.0 (41.0 to 100.0) at a mean follow
up of 4.8 years (2.1 to 12.5) (p <
0.001). No patients reported
squeaking of the hip. There were two fractures of the ceramic head, both in alumina
bearings. No liners were seen to fracture. No fractures were observed
in components made of Delta ceramic. At 8.3 years post-operatively
the survival with any cause of failure as the endpoint was 96.6%
(95% confidence interval (CI) 85.7 to 99.3) for the acetabular component and
94.0% (95% CI 82.1 to 98.4) for the femoral component. The technique of revising the acetabular component in the presence
of a well-fixed femoral component with a ceramic head placed on
a titanium sleeve over the used trunnion is a useful adjunct in
revision hip practice. The use of Delta ceramic is recommended. Cite this article:
Noise generation has been reported with ceramic-on-ceramic
articulations in total hip replacement (THR). This study evaluated
208 consecutive Delta Motion THRs at a mean follow-up of 21 months
(12 to 35). There were 141 women and 67 men with a mean age of 59
years (22 to 84). Patients were reviewed clinically and radiologically,
and the incidence of noise was determined using a newly described
assessment method. Noise production was examined against range of
movement, ligamentous laxity, patient-reported outcome scores, activity
level and orientation of the acetabular component. There were 143
silent hips (69%), 22 (11%) with noises other than squeaking, 17
(8%) with unreproducible squeaking and 26 (13%) with reproducible
squeaking. Hips with reproducible squeaking had a greater mean range
of movement (p <
0.001) and mean ligament laxity (p = 0.004), smaller
median head size (p = 0.01) and decreased mean acetabular component
inclination (p = 0.02) and anteversion angle (p = 0.02) compared
with the other groups. There was no relationship between squeaking
and age (p = 0.13), height (p = 0.263), weight (p = 0.333), body
mass index (p = 0.643), gender (p = 0.07) or patient outcome score
(p = 0.422). There were no revisions during follow-up. Despite the
surprisingly high incidence of squeaking, all patients remain satisfied
with their hip replacement. Cite this article:
We analysed 54 alumina ceramic-on-ceramic bearings
from total hip replacements retrieved at one centre after a mean
duration of 3.5 years (0.2 to 10.6) in situ. These
implants were obtained from 54 patients (16 men and 38 women) with
a mean age of 67 years (33 to 88) who underwent revision for a variety
of reasons. Posterior edge loading was found in the majority of
these retrievals (32 out of 54). Anterosuperior edge loading occurred
less often but produced a higher rate of wear. Stripe wear on the
femoral heads had a median volumetric wear rate of 0.2 mm. 3. /year
(0 to 7.2). The wear volume on the femoral heads corresponded to
the width of edge wear on the matching liner. Anteversion of the
acetabular component was found to be a more important determinant
than inclination for wear in
The ideal bearing surface for young patients
undergoing total hip replacement (THR) remains controversial. We report
the five-year results of a randomised controlled trial comparing
the clinical and radiological outcomes of 102 THRs in 91 patients
who were <
65 years of age. These patients were randomised to
receive a cobalt–chrome on ultra-high-molecular-weight polyethylene,
cobalt–chrome on highly cross-linked polyethylene, or a ceramic-on-ceramic
bearing. In all, 97 hip replacements in 87 patients were available
for review at five years. Two hips had been revised, one for infection
and one for peri-prosthetic fracture. At the final follow-up there were no significant differences
between the groups for the mean Western Ontario and McMaster Universities
osteoarthritis index (pain, p = 0.543; function, p = 0.10; stiffness,
p = 0.99), Short Form-12 (physical component, p = 0.878; mental
component, p = 0.818) or Harris hip scores (p = 0.22). Radiological
outcomes revealed no significant wear in the ceramic group. Comparison of standard and highly cross-linked polyethylene,
however, revealed an almost threefold difference in the mean annual
linear wear rates (0.151 mm/year
Fracture of a ceramic component in total hip
replacement is a rare but potentially catastrophic complication.
The incidence is likely to increase as the use of ceramics becomes
more widespread. We describe such a case, which illustrates how
inadequate initial management will lead to further morbidity and
require additional surgery. We present the case as a warning that
fracture of a ceramic component should be revised to another ceramic-on-ceramic
articulation in order to minimise the risk of further catastrophic
wear.
We report the use of a 15° face-changing cementless
acetabular component in patients undergoing total hip replacement
for osteoarthritis secondary to developmental dysplasia of the hip.
The rationale behind its design and the surgical technique used
for its implantation are described. It is distinctly different from
a standard cementless hemispherical component as it is designed
to position the bearing surface at the optimal angle of inclination,
that is, <
45°, while maximising the cover of the component by
host bone.
Ceramic-on-ceramic bearings in hip replacement have low rates of wear and are increasingly being used in young adults. Our aim was to determine the incidence of audible phenomena or other bearing-related complications. We retrospectively analysed 250 ceramic-on-ceramic hip replacements in 224 patients which had been implanted between April 2000 and December 2007. The mean age of the patients at operation was 44 years (14 to 83) and all the operations were performed using the same surgical technique at a single centre. At a mean follow-up of 59 months (24 to 94), the mean Oxford hip score was 40.89 (11 to 48). There were six revisions, three of which were for impingement-related complications. No patient reported squeaking, but six described grinding or clicking, which was usually associated with deep flexion. No radiological evidence of osteolysis or migration of the components was observed in any hip. The early to mid-term results of contemporary ceramic-on-ceramic hip replacement show promising results with few concerns in terms of noise and squeaking. Positioning of the acetabular component remains critical in regard to the reduction of other impingement-related complications.