Kasten P Pape G Raiss P Bruckner T Rickert M Zeifang F Loew M

We have investigated the mid-term outcome of total shoulder replacement using a keeled cemented glenoid component and a modern cementing technique with regard to the causes of failure and loosening of the components.

Between 1997 and 2003 we performed 96 total shoulder replacements on 88 patients, 24 men and 64 women with a mean age of 69.7 years (31 to 82). The minimum follow-up was five years and at the time of review 87 shoulders (77 patients) were examined at a mean follow-up of 89.1 months (60 to 127). Cumulative survival curves were generated with re-operations (accomplished and planned), survivorship of the proshesis, loosening of the glenoid (defined as tilt > 5° or subsidence > 5 mm), the presence of radiolucent lines and a Constant score of < 30 as the endpoints.

There were two re-operations not involving revision of the implants and the survival rate of the prosthesis was 100.0% for the follow-up period, with an absolute Constant score of > 30 as the endpoint the survival rate was 98%. Radiological glenoid loosening was 9% after five years, and 33% after nine years. There was an incidence of 8% of radiolucent lines in more than three of six zones in the immediate post-operative period, of 37.0% after the first year which increased to 87.0% after nine years. There was no correlation between the score of Boileau and the total Constant score at the latest follow-up, but there was correlation between glenoid loosening and pain (p = 0.001).

We found that total shoulder replacement had an excellent mid-term survivorship and clinical outcome. The surgical and cementing techniques were related to the decrease in radiolucent lines around the glenoid compared with earlier studies. One concern, however, was the fact that radiolucent lines increased over time and there was a rate of glenoid loosening of 9% after five years and 33% after nine years. This suggests that the design of the glenoid component, and the implantation and cementing techniques may need further improvement.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 7 | Pages 920 - 924
1 Jul 2006
Mori T Kudo H Iwano K Juji T

We studied 11 patients (14 elbows) with gross rheumatoid deformity of the elbow, treated by total arthroplasty using the Kudo type-5 unlinked prosthesis, and who were evaluated between five and 11 years after operation. Massive bone defects were augmented by autogenous bone grafts. There were no major complications such as infection, subluxation or loosening. In most elbows relief from pain and stability were achieved. The results, according to the Mayo Elbow Performance Score, were excellent in eight, good in five and fair in one. In most elbows there was minimal or no resorption of the grafted bone. There were no radiolucent lines around the stems of the cementless components. This study shows that even highly unstable rheumatoid elbows can be replaced successfully using an unlinked prosthesis, with augmentation by grafting for major defects of bone.