Advertisement for orthosearch.org.uk
Results 1 - 3 of 3
Results per page:
The Bone & Joint Journal
Vol. 103-B, Issue 2 | Pages 398 - 404
1 Feb 2021
Christ AB Fujiwara T Yakoub MA Healey JH

Aims

We have evaluated the survivorship, outcomes, and failures of an interlocking, reconstruction-mode stem-sideplate implant used to preserve the native hip joint and achieve proximal fixation when there is little residual femur during large endoprosthetic reconstruction of the distal femur.

Methods

A total of 14 patients underwent primary or revision reconstruction of a large femoral defect with a short remaining proximal femur using an interlocking, reconstruction-mode stem-sideplate for fixation after oncological distal femoral and diaphyseal resections. The implant was attached to a standard endoprosthetic reconstruction system. The implant was attached to a standard endoprosthetic reconstruction system. None of the femoral revisions were amenable to standard cemented or uncemented stem fixation. Patient and disease characteristics, surgical history, final ambulatory status, and Musculoskeletal Tumor Society (MSTS) score were recorded. The percentage of proximal femur remaining was calculated from follow-up radiographs.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 10 | Pages 1367 - 1372
1 Oct 2006
Gupta A Pollock R Cannon SR Briggs TWR Skinner J Blunn G

We used a knee-sparing distal femoral endoprosthesis in young patients with malignant bone tumours of the distal femur in whom it was possible to resect the tumour and to preserve the distal femoral condyles. The proximal shaft of the endoprosthesis had a coated hydroxyapatite collar, while the distal end had hydroxyapatite-coated extracortical plates to secure it to the small residual femoral condylar fragment. We reviewed the preliminary results of this endoprosthesis in eight patients with primary bone tumours of the distal femur. Their mean age at surgery was 17.years (14 to 21). The mean follow-up was 24 months (20 to 31). At final follow-up the mean flexion at the knee was 102° (20° to 120°) and the mean Musculoskeletal Tumour Society score was 80% (57% to 96.7%).

There was excellent osteointegration at the prosthesis-proximal bone interface with formation of new bone around the hydroxyapatite collar. The prosthesis allowed preservation of the knee and achieved a good functional result. Formation of new bone and remodelling at the interface make the implant more secure. Further follow-up is required to determine the long-term structural integrity of the prosthesis.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 4 | Pages 534 - 539
1 Apr 2005
Cobb JP Ashwood N Robbins G Witt JD Unwin PS Blunn G

Massive endoprostheses using a cemented intramedullary stem are widely used to allow early resumption of activity after surgery for tumours. The survival of the prosthesis varies with the anatomical site, the type of prosthesis and the mode of fixation. Revision surgery is required in many cases because of aseptic loosening. Insertion of a second cemented endoprosthesis may be difficult because of the poor quality of the remaining bone, and loosening recurs quickly.

We describe a series of 14 patients with triplate fixation in difficult revision or joint-sparing tumour surgery with a minimum follow-up of four years. The triplate design incorporated well within a remodelled cortex to achieve osseomechanical integration with all patients regaining their original level of function within five months.

Our preliminary results suggest that this technique may provide an easy, biomechanically friendly alternative to insertion of a further device with an intramedullary stem, which has a shorter lifespan in revision or joint-sparing tumour surgery. A short segment of bone remaining after resection of a tumour will not accept an intramedullary stem, but may be soundly fixed using this method.