Advertisement for orthosearch.org.uk
Results 1 - 50 of 285
Results per page:
Bone & Joint Open
Vol. 5, Issue 10 | Pages 825 - 831
3 Oct 2024
Afghanyar Y Afghanyar B Loweg L Drees P Gercek E Dargel J Rehbein P Kutzner KP

Aims

Limited implant survival due to aseptic cup loosening is most commonly responsible for revision total hip arthroplasty (THA). Advances in implant designs and materials have been crucial in addressing those challenges. Vitamin E-infused highly cross-linked polyethylene (VEPE) promises strong wear resistance, high oxidative stability, and superior mechanical strength. Although VEPE monoblock cups have shown good mid-term performance and excellent wear patterns, long-term results remain unclear. This study evaluated migration and wear patterns and clinical and radiological outcomes at a minimum of ten years’ follow-up.

Methods

This prospective observational study investigated 101 cases of primary THA over a mean duration of 129 months (120 to 149). At last follow-up, 57 cases with complete clinical and radiological outcomes were evaluated. In all cases, the acetabular component comprised an uncemented titanium particle-coated VEPE monoblock cup. Patients were assessed clinically and radiologically using the Harris Hip Score, visual analogue scale (pain and satisfaction), and an anteroposterior radiograph. Cup migration and polyethylene wear were measured using Einzel-Bild-Röntgen-Analyze software. All complications and associated treatments were documented until final follow-up.


The Bone & Joint Journal
Vol. 106-B, Issue 10 | Pages 1067 - 1073
1 Oct 2024
Lodge CJ Adlan A Nandra RS Kaur J Jeys L Stevenson JD

Aims. Periprosthetic joint infection (PJI) is a challenging complication of any arthroplasty procedure. We reviewed our use of static antibiotic-loaded cement spacers (ABLCSs) for staged management of PJI where segmental bone loss, ligamentous instability, or soft-tissue defects necessitate a static construct. We reviewed factors contributing to their failure and techniques to avoid these complications when using ABLCSs in this context. Methods. A retrospective analysis was conducted of 94 patients undergoing first-stage revision of an infected knee prosthesis between September 2007 and January 2020 at a single institution. Radiographs and clinical records were used to assess and classify the incidence and causes of static spacer failure. Of the 94 cases, there were 19 primary total knee arthroplasties (TKAs), ten revision TKAs (varus-valgus constraint), 20 hinged TKAs, one arthrodesis (nail), one failed spacer (performed elsewhere), 21 distal femoral endoprosthetic arthroplasties, and 22 proximal tibial arthroplasties. Results. A total of 35/94 patients (37.2%) had spacer-related complications, of which 26/35 complications (74.3%) were because of mechanical failure of the spacer construct, while 9/35 (25.7%) were due to recurrence of infection. Risk factors for internal failure were a construct where the total intramedullary spacer length was less than twice the length of the central osseous defect (p = 0.009), where proximal or distal intraosseous spacer contact was < 10%, and after tibial tubercle osteotomy (p = 0.005). The incidence of spacer complications significantly increased the time to second stage: mean 157 days (42 to 458) in those without complications versus 227 days (11 to 528) with complications (p = 0.014). Conclusion. The failure rate of static antibiotic-loaded cement spacers is much higher than anticipated. Complications of the spacer significantly increased the time to second-stage revision. The risk of mechanical failure is significantly increased if the spacer is less than double the size of the segmental defect, or if inadequate reinforcement is inserted into the residual bone. These findings serve as a guide for surgeons to avoid mechanical complications with static spacers. Cite this article: Bone Joint J 2024;106-B(10):1067–1073


The Bone & Joint Journal
Vol. 106-B, Issue 7 | Pages 646 - 655
1 Jul 2024
Longo UG Gulotta LV De Salvatore S Lalli A Bandini B Giannarelli D Denaro V

Aims

Proximal humeral fractures are the third most common fracture among the elderly. Complications associated with fixation include screw perforation, varus collapse, and avascular necrosis of the humeral head. To address these challenges, various augmentation techniques to increase medial column support have been developed. There are currently no recent studies that definitively establish the superiority of augmented fixation over non-augmented implants in the surgical treatment of proximal humeral fractures. The aim of this systematic review and meta-analysis was to compare the outcomes of patients who underwent locking-plate fixation with cement augmentation or bone-graft augmentation versus those who underwent locking-plate fixation without augmentation for proximal humeral fractures.

Methods

The search was carried out according to the Preferred Reporting Items for Systematic Reviews and Meta-analyses guidelines. Articles involving patients with complex proximal humeral fractures treated using open reduction with locking-plate fixation, with or without augmentation, were considered. A meta-analysis of comparative studies comparing locking-plate fixation with cement augmentation or with bone-graft augmentation versus locking-plate fixation without augmentation was performed.


Bone & Joint Open
Vol. 5, Issue 6 | Pages 489 - 498
12 Jun 2024
Kriechling P Bowley ALW Ross LA Moran M Scott CEH

Aims

The purpose of this study was to compare reoperation and revision rates of double plating (DP), single plating using a lateral locking plate (SP), or distal femoral arthroplasty (DFA) for the treatment of periprosthetic distal femur fractures (PDFFs).

Methods

All patients with PDFF primarily treated with DP, SP, or DFA between 2008 and 2022 at a university teaching hospital were included in this retrospective cohort study. The primary outcome was revision surgery for failure following DP, SP, or DFA. Secondary outcome measures included any reoperation, length of hospital stay, and mortality. All basic demographic and relevant implant and injury details were collected. Radiological analysis included fracture classification and evaluation of metaphyseal and medial comminution.


Aims

The aim of this study was to investigate the distribution of phenotypes in Asian patients with end-stage osteoarthritis (OA) and assess whether the phenotype affected the clinical outcome and survival of mechanically aligned total knee arthroplasty (TKA). We also compared the survival of the group in which the phenotype unintentionally remained unchanged with those in which it was corrected to neutral.

Methods

The study involved 945 TKAs, which were performed in 641 patients with primary OA, between January 2000 and January 2009. These were classified into 12 phenotypes based on the combined assessment of four categories of the arithmetic hip-knee-ankle angle and three categories of actual joint line obliquity. The rates of survival were analyzed using Kaplan-Meier methods and the log-rank test. The Hospital for Special Surgery score and survival of each phenotype were compared with those of the reference phenotype with neutral alignment and a parallel joint line. We also compared long-term survival between the unchanged phenotype group and the corrected to neutral alignment-parallel joint line group in patients with Type IV-b (mild to moderate varus alignment-parallel joint line) phenotype.


The Bone & Joint Journal
Vol. 106-B, Issue 5 Supple B | Pages 125 - 132
1 May 2024
Carender CN Perry KI Sierra RJ Trousdale RT Berry DJ Abdel MP

Aims

Uncemented implants are now commonly used at reimplantation of a two-stage revision total hip arthoplasty (THA) following periprosthetic joint infection (PJI). However, there is a paucity of data on the performance of the most commonly used uncemented femoral implants – modular fluted tapered (MFT) femoral components – in this setting. This study evaluated implant survival, radiological results, and clinical outcomes in a large cohort of reimplantation THAs using MFT components.

Methods

We identified 236 reimplantation THAs from a single tertiary care academic institution from September 2000 to September 2020. Two designs of MFT femoral components were used as part of an established two-stage exchange protocol for the treatment of PJI. Mean age at reimplantation was 65 years (SD 11), mean BMI was 32 kg/m2 (SD 7), and 46% (n = 109) were female. Mean follow-up was seven years (SD 4). A competing risk model accounting for death was used.


The Bone & Joint Journal
Vol. 106-B, Issue 5 Supple B | Pages 89 - 97
1 May 2024
Scholz J Perka C Hipfl C

Aims

There is little information in the literature about the use of dual-mobility (DM) bearings in preventing re-dislocation in revision total hip arthroplasty (THA). The aim of this study was to compare the use of DM bearings, standard bearings, and constrained liners in revision THA for recurrent dislocation, and to identify risk factors for re-dislocation.

Methods

We reviewed 86 consecutive revision THAs performed for dislocation between August 2012 and July 2019. A total of 38 revisions (44.2%) involved a DM bearing, while 39 (45.3%) and nine (10.5%) involved a standard bearing and a constrained liner, respectively. Rates of re-dislocation, re-revision for dislocation, and overall re-revision were compared. Radiographs were assessed for the positioning of the acetabular component, the restoration of the centre of rotation, leg length, and offset. Risk factors for re-dislocation were determined by Cox regression analysis. The modified Harris Hip Scores (mHHSs) were recorded. The mean age of the patients at the time of revision was 70 years (43 to 88); 54 were female (62.8%). The mean follow-up was 5.0 years (2.0 to 8.75).


The Bone & Joint Journal
Vol. 106-B, Issue 5 Supple B | Pages 47 - 53
1 May 2024
Jones SA Parker J Horner M

Aims

The aims of this study were to determine the success of a reconstruction algorithm used in major acetabular bone loss, and to further define the indications for custom-made implants in major acetabular bone loss.

Methods

We reviewed a consecutive series of Paprosky type III acetabular defects treated according to a reconstruction algorithm. IIIA defects were planned to use a superior augment and hemispherical acetabular component. IIIB defects were planned to receive either a hemispherical acetabular component plus augments, a cup-cage reconstruction, or a custom-made implant. We used national digital health records and registry reports to identify any reoperation or re-revision procedure and Oxford Hip Score (OHS) for patient-reported outcomes. Implant survival was determined via Kaplan-Meier analysis.


Bone & Joint Open
Vol. 5, Issue 4 | Pages 286 - 293
9 Apr 2024
Upadhyay PK Kumar V Mirza SB Shah N

Aims

This study reports the results of 38 total hip arthroplasties (THAs) in 33 patients aged less than 50 years, using the JRI Furlong hydroxyapatite ceramic (HAC)-coated femoral component.

Methods

We describe the survival, radiological, and functional outcomes of 33 patients (38 THAs) at a mean follow-up of 27 years (25 to 32) between 1988 and 2018.


The Bone & Joint Journal
Vol. 106-B, Issue 4 | Pages 312 - 318
1 Apr 2024
Sheth NP Jones SA Sanghavi SA Manktelow A

The advent of modular porous metal augments has ushered in a new form of treatment for acetabular bone loss. The function of an augment can be seen as reducing the size of a defect or reconstituting the anterosuperior/posteroinferior columns and/or allowing supplementary fixation. Depending on the function of the augment, the surgeon can decide on the sequence of introduction of the hemispherical shell, before or after the augment. Augments should always, however, be used with cement to form a unit with the acetabular component. Given their versatility, augments also allow the use of a hemispherical shell in a position that restores the centre of rotation and biomechanics of the hip. Progressive shedding or the appearance of metal debris is a particular finding with augments and, with other radiological signs of failure, should be recognized on serial radiographs. Mid- to long-term outcomes in studies reporting the use of augments with hemispherical shells in revision total hip arthroplasty have shown rates of survival of > 90%. However, a higher risk of failure has been reported when augments have been used for patients with chronic pelvic discontinuity.

Cite this article: Bone Joint J 2024;106-B(4):312–318.


Bone & Joint Open
Vol. 5, Issue 3 | Pages 218 - 226
15 Mar 2024
Voigt JD Potter BK Souza J Forsberg J Melton D Hsu JR Wilke B

Aims

Prior cost-effectiveness analyses on osseointegrated prosthesis for transfemoral unilateral amputees have analyzed outcomes in non-USA countries using generic quality of life instruments, which may not be appropriate when evaluating disease-specific quality of life. These prior analyses have also focused only on patients who had failed a socket-based prosthesis. The aim of the current study is to use a disease-specific quality of life instrument, which can more accurately reflect a patient’s quality of life with this condition in order to evaluate cost-effectiveness, examining both treatment-naïve and socket refractory patients.

Methods

Lifetime Markov models were developed evaluating active healthy middle-aged male amputees. Costs of the prostheses, associated complications, use/non-use, and annual costs of arthroplasty parts and service for both a socket and osseointegrated (OPRA) prosthesis were included. Effectiveness was evaluated using the questionnaire for persons with a transfemoral amputation (Q-TFA) until death. All costs and Q-TFA were discounted at 3% annually. Sensitivity analyses on those cost variables which affected a change in treatment (OPRA to socket, or socket to OPRA) were evaluated to determine threshold values. Incremental cost-effectiveness ratios (ICERs) were calculated.


The Bone & Joint Journal
Vol. 106-B, Issue 2 | Pages 151 - 157
1 Feb 2024
Dreyer L Bader C Flörkemeier T Wagner M

Aims. The risk of mechanical failure of modular revision hip stems is frequently mentioned in the literature, but little is currently known about the actual clinical failure rates of this type of prosthesis. The current retrospective long-term analysis examines the distal and modular failure patterns of the Prevision hip stem from 18 years of clinical use. A design improvement of the modular taper was introduced in 2008, and the data could also be used to compare the original and the current design of the modular connection. Methods. We performed an analysis of the Prevision modular hip stem using the manufacturer’s vigilance database and investigated different mechanical failure patterns of the hip stem from January 2004 to December 2022. Results. Two mechanical failure patterns were identified: fractures in the area of the distal fluted profile (distal stem fracture) and failure of the modular taper (modular fracture). A failure rate of 0.07% was observed for distal stem fracture, and modular fracture rates of 1.74% for the original and 0.013% for the current taper design. Conclusion. A low risk of mechanical failure for both fracture types was observed compared to other known complications in revision hip arthroplasty. In addition, the data show that a design change did significantly reduce the risk of a modular fracture. Cite this article: Bone Joint J 2024;106-B(2):151–157


Bone & Joint Open
Vol. 5, Issue 1 | Pages 37 - 45
19 Jan 2024
Alm CE Karlsten A Madsen JE Nordsletten L Brattgjerd JE Pripp AH Frihagen F Röhrl SM

Aims

Despite limited clinical scientific backing, an additional trochanteric stabilizing plate (TSP) has been advocated when treating unstable trochanteric fractures with a sliding hip screw (SHS). We aimed to explore whether the TSP would result in less post operative fracture motion, compared to SHS alone.

Methods

Overall, 31 patients with AO/OTA 31-A2 trochanteric fractures were randomized to either a SHS alone or a SHS with an additional TSP. To compare postoperative fracture motion, radiostereometric analysis (RSA) was performed before and after weightbearing, and then at four, eight, 12, 26, and 52 weeks. With the “after weightbearing” images as baseline, we calculated translations and rotations, including shortening and medialization of the femoral shaft.


The Bone & Joint Journal
Vol. 106-B, Issue 1 | Pages 93 - 98
1 Jan 2024
Broida SE Tsoi KM Rose PS Ferguson PC Griffin AM Wunder JS Houdek MT

Aims

The sacroiliac joint (SIJ) is the only mechanical connection between the axial skeleton and lower limbs. Following iliosacral resection, there is debate on whether reconstruction of the joint is necessary. There is a paucity of data comparing the outcomes of patients undergoing reconstruction and those who are not formally reconstructed.

Methods

A total of 60 patients (25 females, 35 males; mean age 39 years (SD 18)) undergoing iliosacral resection were reviewed. Most resections were performed for primary malignant tumours (n = 54; 90%). The mean follow-up for surviving patients was nine years (2 to 19).


The Bone & Joint Journal
Vol. 106-B, Issue 1 | Pages 6 - 8
1 Jan 2024
Stevenson J Cool P Ashford R

Cite this article: Bone Joint J 2024;106-B(1):6–8.


Bone & Joint Open
Vol. 4, Issue 12 | Pages 948 - 956
15 Dec 2023
Vella-Baldacchino M Webb J Selvarajah B Chatha S Davies A Cobb JP Liddle AD

Aims

With up to 40% of patients having patellofemoral joint osteoarthritis (PFJ OA), the two arthroplasty options are to replace solely the patellofemoral joint via patellofemoral arthroplasty (PFA), or the entire knee via total knee arthroplasty (TKA). The aim of this study was to assess postoperative success of second-generation PFAs compared to TKAs for patients treated for PFJ OA using patient-reported outcome measures (PROMs) and domains deemed important by patients following a patient and public involvement meeting.

Methods

MEDLINE, EMBASE via OVID, CINAHL, and EBSCO were searched from inception to January 2022. Any study addressing surgical treatment of primary patellofemoral joint OA using second generation PFA and TKA in patients aged above 18 years with follow-up data of 30 days were included. Studies relating to OA secondary to trauma were excluded. ROB-2 and ROBINS-I bias tools were used.


Bone & Joint Open
Vol. 4, Issue 10 | Pages 801 - 807
23 Oct 2023
Walter N Szymski D Kurtz SM Lowenberg DW Alt V Lau EC Rupp M

Aims

This work aimed at answering the following research questions: 1) What is the rate of mechanical complications, nonunion and infection for head/neck femoral fractures, intertrochanteric fractures, and subtrochanteric fractures in the elderly USA population? and 2) Which factors influence adverse outcomes?

Methods

Proximal femoral fractures occurred between 1 January 2009 and 31 December 2019 were identified from the Medicare Physician Service Records Data Base. The Kaplan-Meier method with Fine and Gray sub-distribution adaptation was used to determine rates for nonunion, infection, and mechanical complications. Semiparametric Cox regression model was applied incorporating 23 measures as covariates to identify risk factors.


Bone & Joint Open
Vol. 4, Issue 10 | Pages 776 - 781
16 Oct 2023
Matar HE Bloch BV James PJ

Aims

The aim of this study was to evaluate medium- to long-term outcomes and complications of the Stanmore Modular Individualised Lower Extremity System (SMILES) rotating hinge implant in revision total knee arthroplasty (rTKA) at a tertiary unit. It is hypothesized that this fully cemented construct leads to satisfactory clinical outcomes.

Methods

A retrospective consecutive study of all patients who underwent a rTKA using the fully cemented SMILES rotating hinge prosthesis between 2005 to 2018. Outcome measures included aseptic loosening, reoperations, revision for any cause, complications, and survivorship. Patients and implant survivorship data were identified through both prospectively collected local hospital electronic databases and linked data from the National Joint Registry/NHS Personal Demographic Service. Kaplan-Meier survival analysis was used at ten years.


Bone & Joint 360
Vol. 12, Issue 4 | Pages 38 - 41
1 Aug 2023

The August 2023 Children’s orthopaedics Roundup360 looks at: DDH: What can patients expect after open reduction?; Femoral head deformity associated with hip displacement in non-ambulatory cerebral palsy; Bony hip reconstruction for displaced hips in patients with cerebral palsy: is postoperative immobilization indicated?; Opioid re-prescriptions after ACL reconstruction in adolescents are associated with subsequent opioid use disorder; Normative femoral and tibial lengths in a modern population of USA children; Retrospective analysis of associated anomalies in 636 patients with operatively treated congenital scoliosis; Radiological hip shape and patient-reported outcome measures in healed Perthes’ disease; Significantly displaced adolescent posterior sternoclavicular joint injuries.


The Bone & Joint Journal
Vol. 105-B, Issue 8 | Pages 888 - 894
1 Aug 2023
Murray J Jeyapalan R Davies M Sheehan C Petrie M Harrison T

Aims

Total femoral arthroplasty (TFA) is a rare procedure used in cases of significant femoral bone loss, commonly from cancer, infection, and trauma. Low patient numbers have resulted in limited published work on long-term outcomes, and even less regarding TFA undertaken for non-oncological indications. The aim of this study was to evaluate the long-term clinical outcomes of all TFAs in our unit.

Methods

Data were collected retrospectively from a large tertiary referral revision arthroplasty unit’s database. Inclusion criteria included all patients who underwent TFA in our unit. Preoperative demographics, operative factors, and short- and long-term outcomes were collected for analysis. Outcome was defined using the Musculoskeletal Infection Society (MSIS) outcome reporting tool.


The Bone & Joint Journal
Vol. 105-B, Issue 8 | Pages 905 - 911
1 Aug 2023
Giannicola G Amura A Sessa P Prigent S Cinotti G

Aims

The aim of this study was to analyze how proximal radial neck resorption (PRNR) starts and progresses radiologically in two types of press-fit radial head arthroplasties (RHAs), and to investigate its clinical relevance.

Methods

A total of 97 patients with RHA were analyzed: 56 received a bipolar RHA (Group 1) while 41 received an anatomical implant (Group 2). Radiographs were performed postoperatively and after three, six, nine, and 12 weeks, six, nine, 12, 18, and 24 months, and annually thereafter. PRNR was measured in all radiographs in the four radial neck quadrants. The Mayo Elbow Performance Score (MEPS), the abbreviated version of the Disabilities of the Arm, Shoulder, and Hand questionnaire (QuickDASH), and the patient-assessed American Shoulder and Elbow Surgeons score - Elbow (pASES-E) were used for the clinical assessment. Radiological signs of implant loosening were investigated.


Bone & Joint Open
Vol. 4, Issue 5 | Pages 329 - 337
8 May 2023
Khan AQ Chowdhry M Sherwani MKA McPherson EJ

Aims

Total hip arthroplasty (THA) is considered the preferred treatment for displaced proximal femoral neck fractures. However, in many countries this option is economically unviable. To improve outcomes in financially disadvantaged populations, we studied the technique of concomitant valgus hip osteotomy and operative fixation (VOOF). This prospective serial study compares two treatment groups: VOOF versus operative fixation alone with cannulated compression screws (CCSs).

Methods

In the first series, 98 hip fixation procedures were performed using CCS. After fluoroscopic reduction of the fracture, three CCSs were placed. In the second series, 105 VOOF procedures were performed using a closing wedge intertrochanteric osteotomy with a compression lag screw and lateral femoral plate. The alignment goal was to create a modified Pauwel’s fracture angle of 30°. After fluoroscopic reduction of fracture, lag screw was placed to achieve the calculated correction angle, followed by inter-trochanteric osteotomy and placement of barrel plate. Patients were followed for a minimum of two years.


Aims

The primary objective of this study was to compare the five-year tibial component migration and wear between highly crosslinked polyethylene (HXLPE) inserts and conventional polyethylene (PE) inserts of the uncemented Triathlon fixed insert cruciate-retaining total knee arthroplasty (TKA). Secondary objectives included clinical outcomes and patient-reported outcome measures (PROMs).

Methods

A double-blinded, randomized study was conducted including 96 TKAs. Tibial component migration and insert wear were measured with radiostereometric analysis (RSA) at three, six, 12, 24, and 60 months postoperatively. PROMS were collected preoperatively and at all follow-up timepoints.


The Bone & Joint Journal
Vol. 105-B, Issue 5 | Pages 526 - 533
1 May 2023
Harmer JR Wyles CC Duong SQ Morgan III RJ Maradit-Kremers H Abdel MP

Aims

The aim of this study was to determine the prevalence of depressive and anxiety disorders prior to total hip (THA) and total knee arthroplasty (TKA) and to assess their impact on the rates of any infection, revision, or reoperation.

Methods

Between January 2000 and March 2019, 21,469 primary and revision arthroplasties (10,011 THAs; 11,458 TKAs), which were undertaken in 15,504 patients at a single academic medical centre, were identified from a 27-county linked electronic medical record (EMR) system. Depressive and anxiety disorders were identified by diagnoses in the EMR or by using a natural language processing program with subsequent validation from review of the medical records. Patients with mental health diagnoses other than anxiety or depression were excluded.


The Bone & Joint Journal
Vol. 105-B, Issue 5 | Pages 467 - 470
1 May 2023
McBryde CW Prakash R Haddad FS


The Bone & Joint Journal
Vol. 105-B, Issue 3 | Pages 277 - 283
1 Mar 2023
Gausden EB Puri S Chiu Y Figgie MP Sculco TP Westrich G Sculco PK Chalmers BP

Aims. The purpose of this study was to assess mid-term survivorship following primary total knee arthroplasty (TKA) with Optetrak Logic components and identify the most common revision indications at a single institution. Methods. We identified a retrospective cohort of 7,941 Optetrak primary TKAs performed from January 2010 to December 2018. We reviewed the intraoperative findings of 369 TKAs that required revision TKA from January 2010 to December 2021 and the details of the revision implants used. Kaplan-Meier analysis was used to determine survivorship. Cox regression analysis was used to examine the impact of patient variables and year of implantation on survival time. Results. The estimated survivorship free of all-cause revision was 98% (95% confidence interval (CI) 97% to 98%), 95% (95% CI 95% to 96%), and 86% (95% CI 83% to 88%) at two, five, and ten years, respectively. In 209/369 revisions there was a consistent constellation of findings with varying severity that included polyethylene wear and associated synovitis, osteolysis, and component loosening. This failure mode, which we refer to as aseptic mechanical failure, was the most common revision indication. The mean time from primary TKA to revision for aseptic mechanical failure was five years (5 months to 11 years). Conclusion. In this series of nearly 8,000 primary TKAs performed with a specific implant, we identified a lower-than-expected mid-term survivorship and a high number of revisions with a unique presentation. This study, along with the recent recall of the implant, confirms the need for frequent monitoring of patients with Optetrak TKAs given the incidence of polyethylene failure, osteolysis, and component loosening. Cite this article: Bone Joint J 2023;105-B(3):277–283


Bone & Joint Open
Vol. 4, Issue 2 | Pages 53 - 61
1 Feb 2023
Faraj S de Windt TS van Hooff ML van Hellemondt GG Spruit M

Aims

The aim of this study was to assess the clinical and radiological results of patients who were revised using a custom-made triflange acetabular component (CTAC) for component loosening and pelvic discontinuity (PD) after previous total hip arthroplasty (THA).

Methods

Data were extracted from a single centre prospective database of patients with PD who were treated with a CTAC. Patients were included if they had a follow-up of two years. The Hip Disability and Osteoarthritis Outcome Score (HOOS), modified Oxford Hip Score (mOHS), EurQol EuroQoL five-dimension three-level (EQ-5D-3L) utility, and Numeric Rating Scale (NRS), including visual analogue score (VAS) for pain, were gathered at baseline, and at one- and two-year follow-up. Reasons for revision, and radiological and clinical complications were registered. Trends over time are described and tested for significance and clinical relevance.


Bone & Joint Open
Vol. 4, Issue 1 | Pages 19 - 26
13 Jan 2023
Nishida K Nasu Y Hashizume K Okita S Nakahara R Saito T Ozaki T Inoue H

Aims

There are concerns regarding complications and longevity of total elbow arthroplasty (TEA) in young patients, and the few previous publications are mainly limited to reports on linked elbow devices. We investigated the clinical outcome of unlinked TEA for patients aged less than 50 years with rheumatoid arthritis (RA).

Methods

We retrospectively reviewed the records of 26 elbows of 21 patients with RA who were aged less than 50 years who underwent primary TEA with an unlinked elbow prosthesis. The mean patient age was 46 years (35 to 49), and the mean follow-up period was 13.6 years (6 to 27). Outcome measures included pain, range of motion, Mayo Elbow Performance Score (MEPS), radiological evaluation for radiolucent line and loosening, complications, and revision surgery with or without implant removal.


The Bone & Joint Journal
Vol. 105-B, Issue 1 | Pages 29 - 34
1 Jan 2023
Fransen BL Bengoa FJ Neufeld ME Sheridan GA Garbuz DS Howard LC

Aims

Several short- and mid-term studies have shown minimal liner wear of highly cross-linked polyethylene (HXLPE) in total hip arthroplasty (THA), but the safety of using thinner HXLPE liners to maximize femoral head size remains uncertain. The objective of this study was to analyze clinical survival and radiological wear rates of patients with HXLPE liners, a 36 mm femoral head, and a small acetabular component with a minimum of ten years’ follow-up.

Methods

We retrospectively identified 55 patients who underwent primary THA performed at a single centre, using HXLPE liners with 36 mm cobalt-chrome heads in acetabular components with an outer diameter of 52 mm or smaller. Patient demographic details, implant details, death, and all-cause revisions were recorded. Cox regression and Kaplan-Meier survival was used to determine all-cause and liner-specific revision. Of these 55 patients, 22 had a minimum radiological follow-up of seven years and were assessed radiologically for linear and volumetric wear.


The Bone & Joint Journal
Vol. 104-B, Issue 11 | Pages 1249 - 1255
1 Nov 2022
Williamson TK Passfall L Ihejirika-Lomedico R Espinosa A Owusu-Sarpong S Lanre-Amos T Schoenfeld AJ Passias PG

Aims

Postoperative complication rates remain relatively high after adult spinal deformity (ASD) surgery. The extent to which modifiable patient-related factors influence complication rates in patients with ASD has not been effectively evaluated. The aim of this retrospective cohort study was to evaluate the association between modifiable patient-related factors and complications after corrective surgery for ASD.

Methods

ASD patients with two-year data were included. Complications were categorized as follows: any complication, major, medical, surgical, major mechanical, major radiological, and reoperation. Modifiable risk factors included smoking, obesity, osteoporosis, alcohol use, depression, psychiatric diagnosis, and hypertension. Patients were stratified by the degree of baseline deformity (low degree of deformity (LowDef)/high degree of deformity (HighDef): below or above 20°) and age (Older/Younger: above or below 65 years). Complication rates were compared for modifiable risk factors in each age/deformity group, using multivariable logistic regression analysis to adjust for confounders.


The Bone & Joint Journal
Vol. 104-B, Issue 7 | Pages 875 - 883
1 Jul 2022
Mills K Wymenga AB van Hellemondt GG Heesterbeek PJC

Aims

Both the femoral and tibial component are usually cemented at revision total knee arthroplasty (rTKA), while stems can be added with either cemented or press-fit (hybrid) fixation. The aim of this study was to compare the long-term stability of rTKA with cemented and press-fitted stems, using radiostereometric analysis (RSA).

Methods

This is a follow-up of a randomized controlled trial, initially involving 32 patients, of whom 19 (nine cemented, ten hybrid) were available for follow-up ten years postoperatively, when further RSA measurements were made. Micromotion of the femoral and tibial components was assessed using model-based RSA software (RSAcore). The clinical outcome was evaluated using the Knee Society Score (KSS), the Knee injury and Osteoarthritis Outcome Score (KOOS), and visual analogue scale (pain and satisfaction).


The Bone & Joint Journal
Vol. 104-B, Issue 7 | Pages 801 - 810
1 Jul 2022
Krull P Steinbrück A Grimberg AW Melsheimer O Morlock M Perka C

Aims. Registry studies on modified acetabular polyethylene (PE) liner designs are limited. We investigated the influence of standard and modified PE acetabular liner designs on the revision rate for mechanical complications in primary cementless total hip arthroplasty (THA). Methods. We analyzed 151,096 primary cementless THAs from the German Arthroplasty Registry (EPRD) between November 2012 and November 2020. Cumulative incidence of revision for mechanical complications for standard and four modified PE liners (lipped, offset, angulated/offset, and angulated) was determined using competing risk analysis at one and seven years. Confounders were investigated with a Cox proportional-hazards model. Results. Median follow-up was 868 days (interquartile range 418 to 1,364). The offset liner design reduced the risk of revision (hazard ratio (HR) 0.68 (95% confidence interval (CI) 0.50 to 0.92)), while the angulated/offset liner increased the risk of revision for mechanical failure (HR 1.81 (95% CI 1.38 to 2.36)). The cumulative incidence of revision was lowest for the offset liner at one and seven years (1.0% (95% CI 0.7 to 1.3) and 1.8% (95% CI 1.0 to 3.0)). No difference was found between standard, lipped, and angulated liner designs. Higher age at index primary THA and an Elixhauser Comorbidity Index greater than 0 increased the revision risk in the first year after surgery. Implantation of a higher proportion of a single design of liner in a hospital reduced revision risk slightly but significantly (p = 0.001). Conclusion. The use of standard acetabular component liners remains a good choice in primary uncemented THA, as most modified liner designs were not associated with a reduced risk of revision for mechanical failure. Offset liner designs were found to be beneficial and angulated/offset liner designs were associated with higher risks of revision. Cite this article: Bone Joint J 2022;104-B(7):801–810


Bone & Joint Open
Vol. 3, Issue 6 | Pages 485 - 494
13 Jun 2022
Jaubert M Le Baron M Jacquet C Couvreur A Fabre-Aubrespy M Flecher X Ollivier M Argenson J

Aims. Two-stage exchange revision total hip arthroplasty (THA) performed in case of periprosthetic joint infection (PJI) has been considered for many years as being the gold standard for the treatment of chronic infection. However, over the past decade, there have been concerns about its safety and its effectiveness. The purposes of our study were to investigate our practice, collecting the overall spacer complications, and then to analyze their risk factors. Methods. We retrospectively included 125 patients with chronic hip PJI who underwent a staged THA revision performed between January 2013 and December 2019. All spacer complications were systematically collected, and risk factors were analyzed. Statistical evaluations were performed using the Student's t-test, Mann-Whitney U test, and Fisher's exact test. Results. Our staged exchange practice shows poor results, which means a 42% mechanical spacer failure rate, and a 20% recurrent infection rate over the two years average follow-up period. Moreover, we found a high rate of spacer dislocation (23%) and a low rate of spacer fracture (8%) compared to the previous literature. Our findings stress that the majority of spacer complications and failures is reflecting a population with high comorbid burden, highlighted by the American Society of Anesthesiology grade, Charlson Comorbidity Index, and Lee score associations, as well as the cardiac, pulmonary, kidney, or hepatic chronic conditions. Conclusion. Our experience of a two-stage hip exchange revision noted important complication rates associated with high failure rates of polymethylmethacrylate spacers. These findings must be interpreted in the light of the patient’s comorbidity profiles, as the elective population for staged exchange has an increasing comorbid burden leading to poor results. In order to provide better results for this specific population, our conclusion suggests that comparative strategy studies are required to improve our therapeutic indication. Cite this article: Bone Jt Open 2022;3(6):485–494


Bone & Joint Open
Vol. 3, Issue 5 | Pages 359 - 366
1 May 2022
Sadekar V Watts AT Moulder E Souroullas P Hadland Y Barron E Muir R Sharma HK

Aims. The timing of when to remove a circular frame is crucial; early removal results in refracture or deformity, while late removal increases the patient morbidity and delay in return to work. This study was designed to assess the effectiveness of a staged reloading protocol. We report the incidence of mechanical failure following both single-stage and two stage reloading protocols and analyze the associated risk factors. Methods. We identified consecutive patients from our departmental database. Both trauma and elective cases were included, of all ages, frame types, and pathologies who underwent circular frame treatment. Our protocol is either a single-stage or two-stage process implemented by defunctioning the frame, in order to progressively increase the weightbearing load through the bone, and promote full loading prior to frame removal. Before progression, through the process we monitor patients for any increase in pain and assess radiographs for deformity or refracture. Results. There were 244 frames (230 patients) included in the analyses, of which 90 were Ilizarov type frames and 154 were hexapods. There were 149 frames which underwent single-stage reloading and 95 frames which underwent a two-stage reloading protocol. Mechanical failure occurred after frame removal in 13 frames (5%), which suffered refracture. There were no cases of change in alignment. There was no difference between refracture patients who underwent single-stage or two-stage reloading protocols (p = 0.772). In all, 14 patients had failure prevented through identification with the reloading protocol. Conclusion. Our reloading protocol is a simple and effective way to confirm the timing of frame removal and minimize the rate of mechanical failure. Similar failure rates occurred between patients undergoing single-stage and two-stage reloading protocols. If the surgeon is confident with clinical and radiological assessment, it may be possible to progress directly to stage two and decrease frame time and patient morbidity. Cite this article: Bone Jt Open 2022;3(5):359–366


Bone & Joint 360
Vol. 11, Issue 2 | Pages 18 - 20
1 Apr 2022


The Bone & Joint Journal
Vol. 104-B, Issue 4 | Pages 464 - 471
1 Apr 2022
Veerman K Raessens J Telgt D Smulders K Goosen JHM

Aims

Debridement, antibiotics, and implant retention (DAIR) is a widely accepted form of surgical treatment for patients with an early periprosthetic joint infection (PJI) after primary arthroplasty. The outcome of DAIR after revision arthroplasty, however, has not been reported. The aim of this study was to report the success rate of DAIR after revision arthroplasty with a follow-up of two years.

Methods

This retrospective study, conducted at the Sint Maartenskliniek, Nijmegen, the Netherlands, included 88 patients who underwent DAIR within 90 days of revision total hip or total knee arthroplasty between 2012 and 2019. Details of the surgical procedures and PJI were collected. Univariate analysis and a subgroup analysis of the culture-positive group were performed. Kaplan-Meier survivorship curves were constructed.


Bone & Joint Open
Vol. 3, Issue 3 | Pages 173 - 181
1 Mar 2022
Sobol KR Fram BR Strony JT Brown SA

Aims. Endoprosthetic reconstruction with a distal femoral arthroplasty (DFA) can be used to treat distal femoral bone loss from oncological and non-oncological causes. This study reports the short-term implant survivorship, complications, and risk factors for patients who underwent DFA for non-neoplastic indications. Methods. We performed a retrospective review of 75 patients from a single institution who underwent DFA for non-neoplastic indications, including aseptic loosening or mechanical failure of a previous prosthesis (n = 25), periprosthetic joint infection (PJI) (n = 23), and native or periprosthetic distal femur fracture or nonunion (n = 27). Patients with less than 24 months’ follow-up were excluded. We collected patient demographic data, complications, and reoperations. Reoperation for implant failure was used to calculate implant survivorship. Results. Overall one- and five-year implant survivorship was 87% and 76%, respectively. By indication for DFA, mechanical failure had one- and five-year implant survivorship of 92% and 68%, PJI of 91% and 72%, and distal femur fracture/nonunion of 78% and 70% (p = 0.618). A total of 37 patients (49%) experienced complications and 27 patients (36%) required one or more reoperation. PJI (n = 16, 21%), aseptic loosening (n = 9, 12%), and wound complications (n = 8, 11%) were the most common complications. Component revision (n = 10, 13.3%) and single-stage exchange for PJI (n = 9, 12.0 %) were the most common reoperations. Only younger age was significantly associated with increased complications (mean 67 years (SD 9.1)) with complication vs 71 years (SD 9.9) without complication; p = 0.048). Conclusion. DFA is a viable option for distal femoral bone loss from a range of non-oncological causes, demonstrating acceptable short-term survivorship but with high overall complication rates. Cite this article: Bone Jt Open 2022;3(3):173–181


Bone & Joint Research
Vol. 11, Issue 2 | Pages 82 - 90
7 Feb 2022
Eckert JA Bitsch RG Sonntag R Reiner T Schwarze M Jaeger S

Aims

The cemented Oxford unicompartmental knee arthroplasty (OUKA) features two variants: single and twin peg OUKA. The purpose of this study was to assess the stability of both variants in a worst-case scenario of bone defects and suboptimal cementation.

Methods

Single and twin pegs were implanted randomly allocated in 12 pairs of human fresh-frozen femora. We generated 5° bone defects at the posterior condyle. Relative movement was simulated using a servohydraulic pulser, and analyzed at 70°/115° knee flexion. Relative movement was surveyed at seven points of measurement on implant and bone, using an optic system.


The Bone & Joint Journal
Vol. 104-B, Issue 2 | Pages 257 - 264
1 Feb 2022
Tahir M Mehta D Sandhu C Jones M Gardner A Mehta JS

Aims

The aim of this study was to compare the clinical and radiological outcomes of patients with early-onset scoliosis (EOS), who had undergone spinal fusion after distraction-based spinal growth modulation using either traditional growing rods (TGRs) or magnetically controlled growing rods (MCGRs).

Methods

We undertook a retrospective review of skeletally mature patients who had undergone fusion for an EOS, which had been previously treated using either TGRs or MCGRs. Measured outcomes included sequential coronal T1 to S1 height and major curve (Cobb) angle on plain radiographs and any complications requiring unplanned surgery before final fusion.


The Bone & Joint Journal
Vol. 104-B, Issue 2 | Pages 297 - 301
1 Feb 2022
Jamshidi K Bagherifard A Mohaghegh MR Mirzaei A

Aims

Giant cell tumours (GCTs) of the proximal femur are rare, and there is no consensus about the best method of filling the defect left by curettage. In this study, we compared the outcome of using a fibular strut allograft and bone cement to reconstruct the bone defect after extended curettage of a GCT of the proximal femur.

Methods

In a retrospective study, we reviewed 26 patients with a GCT of the proximal femur in whom the bone defect had been filled with either a fibular strut allograft (n = 12) or bone cement (n = 14). Their demographic details and oncological and nononcological complications were retrieved from their medical records. Limb function was assessed using the Musculoskeletal Tumor Society (MSTS) score.


The Bone & Joint Journal
Vol. 104-B, Issue 1 | Pages 19 - 26
1 Jan 2022
Sevaldsen K Schnell Husby O Lian ØB Farran KM Schnell Husby V

Aims

Highly polished stems with force-closed design have shown satisfactory clinical results despite being related to relatively high early migration. It has been suggested that the minimal thickness of cement mantles surrounding the femoral stem should be 2 mm to 4 mm to avoid aseptic loosening. The line-to-line cementing technique of the femoral stem, designed to achieve stem press-fit, challenges this opinion. We compared the migration of a highly polished stem with force-closed design by standard and line-to-line cementing to investigate whether differences in early migration of the stems occur in a clinical study.

Methods

In this single-blind, randomized controlled, clinical radiostereometric analysis (RSA) study, the migration pattern of the cemented Corail hip stem was compared between line-to-line and standard cementing in 48 arthroplasties. The primary outcome measure was femoral stem migration in terms of rotation and translation around and along with the X-, Y-, and Z- axes measured using model-based RSA at three, 12, and 24 months. A linear mixed-effects model was used for statistical analysis.


The Bone & Joint Journal
Vol. 103-B, Issue 11 | Pages 1695 - 1701
1 Nov 2021
Currier JH Currier BH Abdel MP Berry DJ Titus AJ Van Citters DW

Aims

Wear of the polyethylene (PE) tibial insert of total knee arthroplasty (TKA) increases the risk of revision surgery with a significant cost burden on the healthcare system. This study quantifies wear performance of tibial inserts in a large and diverse series of retrieved TKAs to evaluate the effect of factors related to the patient, knee design, and bearing material on tibial insert wear performance.

Methods

An institutional review board-approved retrieval archive was surveyed for modular PE tibial inserts over a range of in vivo duration (mean 58 months (0 to 290)). Five knee designs, totalling 1,585 devices, were studied. Insert wear was estimated from measured thickness change using a previously published method. Linear regression statistical analyses were used to test association of 12 patient and implant design variables with calculated wear rate.


Bone & Joint 360
Vol. 10, Issue 5 | Pages 35 - 37
1 Oct 2021


Bone & Joint Research
Vol. 10, Issue 9 | Pages 619 - 628
27 Sep 2021
Maestro-Paramio L García-Rey E Bensiamar F Saldaña L

Aims

To investigate whether idiopathic osteonecrosis of the femoral head (ONFH) is related to impaired osteoblast activities.

Methods

We cultured osteoblasts isolated from trabecular bone explants taken from the femoral head and the intertrochanteric region of patients with idiopathic ONFH, or from the intertrochanteric region of patients with osteoarthritis (OA), and compared their viability, mineralization capacity, and secretion of paracrine factors.


Bone & Joint Research
Vol. 10, Issue 8 | Pages 488 - 497
10 Aug 2021
Cleemann R Sorensen M West A Soballe K Bechtold JE Baas J

Aims

We wanted to evaluate the effects of a bone anabolic agent (bone morphogenetic protein 2 (BMP-2)) on an anti-catabolic background (systemic or local zoledronate) on fixation of allografted revision implants.

Methods

An established allografted revision protocol was implemented bilaterally into the stifle joints of 24 canines. At revision surgery, each animal received one BMP-2 (5 µg) functionalized implant, and one raw implant. One group (12 animals) received bone graft impregnated with zoledronate (0.005 mg/ml) before impaction. The other group (12 animals) received untreated bone graft and systemic zoledronate (0.1 mg/kg) ten and 20 days after revision surgery. Animals were observed for an additional four weeks before euthanasia.


The Bone & Joint Journal
Vol. 103-B, Issue 8 | Pages 1333 - 1338
2 Aug 2021
Kankanalu P Borton ZM Morgan ML Cresswell T Espag MP Tambe AA Clark DI

Aims. Reverse total shoulder arthroplasty (RTSA) using trabecular metal (TM)-backed glenoid implants has been introduced with the aim to increase implant survival. Only short-term reports on the outcomes of TM-RTSA have been published to date. We aim to present the seven-year survival of TM-backed glenoid implants along with minimum five-year clinical and radiological outcomes. Methods. All consecutive elective RTSAs performed at a single centre between November 2008 and October 2014 were reviewed. Patients who had primary TM-RTSA for rotator cuff arthropathy and osteoarthritis with deficient cuff were included. A total of 190 shoulders in 168 patients (41 male, 127 female) were identified for inclusion at a mean of 7.27 years (SD 1.4) from surgery. The primary outcome was survival of the implant with all-cause revision and aseptic glenoid loosening as endpoints. Secondary outcomes were clinical, radiological, and patient-related outcomes with a five-year minimum follow-up. Results. The implant was revised in ten shoulders (5.2%) with a median time to revision of 21.2 months (interquartile range (IQR) 9.9 to 41.8). The Kaplan-Meier survivorship estimate at seven years was 95.9% (95% confidence interval (CI) 91.7 to 98; 35 RTSAs at risk) for aseptic mechanical failure of the glenoid and 94.8% (95% CI 77.5 to 96.3; 35 RTSAs at risk) for all-cause revision. Minimum five-year clinical and radiological outcomes were available for 103 and 98 RTSAs respectively with a median follow-up time of six years (IQR 5.2 to 7.0). Median postoperative Oxford Shoulder Score was 38 (IQR 31 to 45); median Constant and Murley score was 60 (IQR 47.5 to 70); median forward flexion 115° (IQR 100° to 125°); median abduction 95° (IQR 80° to 120°); and external rotation 25° (IQR 15° to 40°) Scapular notching was seen in 62 RTSAs (63.2%). Conclusion. We present the largest and longest-term series of TM-backed glenoid implants demonstrating 94.8% all-cause survivorship at seven years. Specifically pertaining to glenoid loosening, survival of the implant increased to 95.9%. In addition, we report satisfactory minimum five-year clinical and radiological outcomes. Cite this article: Bone Joint J 2021;103-B(8):1333–1338


The Bone & Joint Journal
Vol. 103-B, Issue 8 | Pages 1331 - 1332
1 Aug 2021
Kankanalu P Borton ZM Morgan ML Cresswell T Espag MP Tambe AA Clark DI


The Bone & Joint Journal
Vol. 103-B, Issue 8 | Pages 1405 - 1413
1 Aug 2021
Ogura K Fujiwara T Morris CD Boland PJ Healey JH

Aims

Rotating-hinge knee prostheses are commonly used to reconstruct the distal femur after resection of a tumour, despite the projected long-term burden of reoperation due to complications. Few studies have examined the factors that influence their failure and none, to our knowledge, have used competing risk models to do so. The purpose of this study was to determine the risk factors for failure of a rotating-hinge knee distal femoral arthroplasty using the Fine-Gray competing risk model.

Methods

We retrospectively reviewed 209 consecutive patients who, between 1991 and 2016, had undergone resection of the distal femur for tumour and reconstruction using a rotating-hinge knee prosthesis. The study endpoint was failure of the prosthesis, defined as removal of the femoral component, the tibial component, or the bone-implant fixation; major revision (exchange of the femoral component, tibial component, or the bone-implant fixation); or amputation.


Bone & Joint 360
Vol. 10, Issue 4 | Pages 40 - 42
1 Aug 2021


The Bone & Joint Journal
Vol. 103-B, Issue 8 | Pages 1421 - 1427
1 Aug 2021
Li J Lu Y Chen G Li M Xiao X Ji C Wang Z Guo Z

Aims

We have previously reported cryoablation-assisted joint-sparing surgery for osteosarcoma with epiphyseal involvement. However, it is not clear whether this is a comparable alternative to conventional joint arthroplasty in terms of oncological and functional outcomes.

Methods

A total of 22 patients who had localized osteosarcoma with epiphyseal involvement around the knee and underwent limb salvage surgery were allocated to joint preservation (JP) group and joint arthroplasty (JA) group. Subjects were followed with radiographs, Musculoskeletal Tumor Society (MSTS) score, and clinical evaluations at one, three, and five years postoperatively.