In 2015, we published the results of our ceramic-on-metal (CoM) total hip arthroplasties (THAs) performed between October 2007 and July 2009 with a mean follow-up of 34 months (23 to 45) and a revision rate of 3.1%. The aim of this paper is to present the longer-term outcomes. A total of 264 patients were reviewed at a mean of 5.8 years (4.6 to 7.2) and 10.1 years (9.2 to 10.6) to determine revision rate, pain, outcome scores, radiological analysis, and blood ion levels. Those who were unwilling or unable to travel were contacted by telephone.Aims
Methods
The aim of this study was to evaluate the suitability of the tapered cone stem in total hip arthroplasty (THA) in patients with excessive femoral anteversion and after femoral osteotomy. We included patients who underwent THA using Wagner Cone due to proximal femur anatomical abnormalities between August 2014 and January 2019 at a single institution. We investigated implant survival time using the endpoint of dislocation and revision, and compared the prevalence of prosthetic impingements between the Wagner Cone, a tapered cone stem, and the Taperloc, a tapered wedge stem, through simulation. We also collected Oxford Hip Score (OHS), visual analogue scale (VAS) satisfaction, and VAS pain by postal survey in August 2023 and explored variables associated with those scores.Aims
Methods
This study evaluates the association between consultant and hospital volume and the risk of re-revision and 90-day mortality following first-time revision of primary hip arthroplasty for aseptic loosening. We conducted a cohort study of first-time, single-stage revision hip arthroplasties (RHAs) performed for aseptic loosening and recorded in the National Joint Registry (NJR) data for England, Wales, Northern Ireland, and the Isle of Man between 2003 and 2019. Patient identifiers were used to link records to national mortality data, and to NJR data to identify subsequent re-revision procedures. Multivariable Cox proportional hazard models with restricted cubic splines were used to define associations between volume and outcome.Aims
Methods
The purpose of this study was to determine the association between prior sleeve gastrectomy in patients undergoing primary total hip and knee arthroplasty, and 90-day complications, incidence of revision arthroplasty, and patient-reported outcome scores at final follow-up. This is a retrospective, single-centre analysis. Patients undergoing primary hip or knee arthroplasty with a prior sleeve gastrectomy were eligible for inclusion (n = 80 patients). A morbidly obese control group was established from the same institutional registry using a 1:2 match, for cases:controls with arthroplasty based on propensity score using age, sex, pre-sleeve gastrectomy BMI, Current Procedural Terminology code to identify anatomical location, and presurgical haemoglobin A1C. Outcomes included 90-day complications, incidence of revision arthroplasty, and patient-reported outcome scores at final follow-up. Multivariable logistic regressions evaluated associations of underlying preoperative demographic and treatment characteristics with outcomes.Aims
Methods
The August 2024 Foot & Ankle Roundup360 looks at: ESWT versus surgery for fifth metatarsal stress fractures; Minimally invasive surgery versus open fusion for hallux rigidus; Diabetes and infection risk in total ankle arthroplasty; Is proximal medial gastrocnemius recession useful for managing chronic plantar fasciitis?; Fuse the great toe in the young!; Conservative surgery for diabetic foot osteomyelitis; Mental health and outcome following foot and ankle surgery.
The purpose of this study was to compare reoperation and revision rates of double plating (DP), single plating using a lateral locking plate (SP), or distal femoral arthroplasty (DFA) for the treatment of periprosthetic distal femur fractures (PDFFs). All patients with PDFF primarily treated with DP, SP, or DFA between 2008 and 2022 at a university teaching hospital were included in this retrospective cohort study. The primary outcome was revision surgery for failure following DP, SP, or DFA. Secondary outcome measures included any reoperation, length of hospital stay, and mortality. All basic demographic and relevant implant and injury details were collected. Radiological analysis included fracture classification and evaluation of metaphyseal and medial comminution.Aims
Methods
This study aims to assess the relationship between history of pseudotumour formation secondary to metal-on-metal (MoM) implants and periprosthetic joint infection (PJI) rate, as well as establish ESR and CRP thresholds that are suggestive of infection in these patients. We hypothesized that patients with a pseudotumour were at increased risk of infection. A total of 1,171 total hip arthroplasty (THA) patients with MoM articulations from August 2000 to March 2014 were retrospectively identified. Of those, 328 patients underwent metal artefact reduction sequence MRI and had minimum two years’ clinical follow-up, and met our inclusion criteria. Data collected included demographic details, surgical indication, laterality, implants used, history of pseudotumour, and their corresponding preoperative ESR (mm/hr) and CRP (mg/dl) levels. Multivariate logistic regression modelling was used to evaluate PJI and history of pseudotumour, and receiver operating characteristic curves were created to assess the diagnostic capabilities of ESR and CRP to determine the presence of infection in patients undergoing revision surgery.Aims
Methods
There is little information in the literature about the use of dual-mobility (DM) bearings in preventing re-dislocation in revision total hip arthroplasty (THA). The aim of this study was to compare the use of DM bearings, standard bearings, and constrained liners in revision THA for recurrent dislocation, and to identify risk factors for re-dislocation. We reviewed 86 consecutive revision THAs performed for dislocation between August 2012 and July 2019. A total of 38 revisions (44.2%) involved a DM bearing, while 39 (45.3%) and nine (10.5%) involved a standard bearing and a constrained liner, respectively. Rates of re-dislocation, re-revision for dislocation, and overall re-revision were compared. Radiographs were assessed for the positioning of the acetabular component, the restoration of the centre of rotation, leg length, and offset. Risk factors for re-dislocation were determined by Cox regression analysis. The modified Harris Hip Scores (mHHSs) were recorded. The mean age of the patients at the time of revision was 70 years (43 to 88); 54 were female (62.8%). The mean follow-up was 5.0 years (2.0 to 8.75).Aims
Methods
There are limited long-term studies reporting on outcomes of the Zimmer Modular Revision (ZMR) stem, and concerns remain regarding failure. Our primary aim was to determine long-term survival free from all-cause revision and stem-related failure for this modular revision stem in revision total hip arthroplasty (THA). Secondary aims included evaluating radiological and functional outcomes. We retrospectively identified all patients in our institutional database who underwent revision THA using the ZMR system from January 2000 to December 2007. We included 106 patients (108 hips) with a mean follow-up of 14.5 years (2.3 to 22.3). Mean patient age was 69.2 years (37.0 to 89.4), and 51.9% were female (n = 55). Indications for index revision included aseptic loosening (73.1%), infection (16.7%), fracture (9.3%), and stem fracture (0.9%). Kaplan-Meier analysis was used to determine the all-cause and stem-related failure revision-free survival. At most recent follow-up, Oxford Hip Scores (OHS) were collected, and radiological stem stability was determined using the Engh classification.Aims
Methods
Robotic-assisted unicompartmental knee arthroplasty (R-UKA) has been proposed as an approach to improve the results of the conventional manual UKA (C-UKA). The aim of this meta-analysis was to analyze the studies comparing R-UKA and C-UKA in terms of clinical outcomes, radiological results, operating time, complications, and revisions. The literature search was conducted on three databases (PubMed, Cochrane, and Web of Science) on 20 February 2024 according to the guidelines for Preferred Reporting Items for Systematic Reviews and Meta-Analysis (PRISMA). Inclusion criteria were comparative studies, written in the English language, with no time limitations, on the comparison of R-UKA and C-UKA. The quality of each article was assessed using the Downs and Black Checklist for Measuring Quality.Aims
Methods
Dislocation remains a leading cause of failure following revision total hip arthroplasty (THA). While dual-mobility (DM) bearings have been shown to mitigate this risk, options are limited when retaining or implanting an uncemented shell without modular DM options. In these circumstances, a monoblock DM cup, designed for cementing, can be cemented into an uncemented acetabular shell. The goal of this study was to describe the implant survival, complications, and radiological outcomes of this construct. We identified 64 patients (65 hips) who had a single-design cemented DM cup cemented into an uncemented acetabular shell during revision THA between 2018 and 2020 at our institution. Cups were cemented into either uncemented cups designed for liner cementing (n = 48; 74%) or retained (n = 17; 26%) acetabular components. Median outer head diameter was 42 mm. Mean age was 69 years (SD 11), mean BMI was 32 kg/m2 (SD 8), and 52% (n = 34) were female. Survival was assessed using Kaplan-Meier methods. Mean follow-up was two years (SD 0.97).Aims
Methods
Patients with a high comorbidity burden (HCB) can achieve similar improvements in quality of life compared with low-risk patients, but greater morbidity may deter surgeons from operating on these patients. Whether surgeon volume influences total hip arthroplasty (THA) outcomes in HCB patients has not been investigated. This study aimed to compare complication rates and implant survivorship in HCB patients operated on by high-volume (HV) and non-HV THA surgeons. Patients with Charlson Comorbidity Index ≥ 5 and American Society of Anesthesiologists grade of III or IV, undergoing primary elective THA between January 2013 and December 2021, were retrospectively reviewed. Patients were separated into groups based on whether they were operated on by a HV surgeon (defined as the top 25% of surgeons at our institution by number of primary THAs per year) or a non-HV surgeon. Groups were propensity-matched 1:1 to control for demographic variables. A total of 1,134 patients were included in the matched analysis. Between groups, 90-day readmissions and revisions were compared, and Kaplan-Meier analysis was used to evaluate implant survivorship within the follow-up period.Aims
Methods
In recent years, the use of a collared cementless femoral prosthesis has risen in popularity. The design intention of collared components is to transfer some load to the resected femoral calcar and prevent implant subsidence within the cancellous bone of the metaphysis. Conversely, the load transfer for a cemented femoral prosthesis depends on the cement-component and cement-bone interface interaction. The aim of our study was to compare the three most commonly used collared cementless components and the three most commonly used tapered polished cemented components in patients aged ≥ 75 years who have undergone a primary total hip arthroplasty (THA) for osteoarthritis (OA). Data from the Australian Orthopaedic Association National Joint Replacement Registry from 1 September 1999 to 31 December 2022 were analyzed. Collared cementless femoral components and cemented components were identified, and the three most commonly used components in each group were analyzed. We identified a total of 11,278 collared cementless components and 47,835 cemented components. Hazard ratios (HRs) from Cox proportional hazards models, adjusting for age and sex, were obtained to compare the revision rates between the groups.Aims
Methods
The benefit of a dual-mobility acetabular component (DMC) for primary total hip arthroplasties (THAs) is controversial. This study aimed to compare the dislocation and complication rates when using a DMC compared to single-mobility (SM) acetabular component in primary elective THA using data collected at a single centre, and compare the revision rates and survival outcomes in these two groups. Between 2010 and 2019, 2,075 primary THAs using either a cementless DM or SM acetabular component were included. Indications for DMC were patients aged older than 70 years or with high risk of dislocation. All other patients received a SM acetabular component. Exclusion criteria were cemented implants, patients treated for femoral neck fracture, and follow-up of less than one year. In total, 1,940 THAs were analyzed: 1,149 DMC (59.2%) and 791 SM (40.8%). The mean age was 73 years (SD 9.2) in the DMC group and 57 years (SD 12) in the SM group. Complications and revisions have been analyzed retrospectively.Aims
Methods
With up to 40% of patients having patellofemoral joint osteoarthritis (PFJ OA), the two arthroplasty options are to replace solely the patellofemoral joint via patellofemoral arthroplasty (PFA), or the entire knee via total knee arthroplasty (TKA). The aim of this study was to assess postoperative success of second-generation PFAs compared to TKAs for patients treated for PFJ OA using patient-reported outcome measures (PROMs) and domains deemed important by patients following a patient and public involvement meeting. MEDLINE, EMBASE via OVID, CINAHL, and EBSCO were searched from inception to January 2022. Any study addressing surgical treatment of primary patellofemoral joint OA using second generation PFA and TKA in patients aged above 18 years with follow-up data of 30 days were included. Studies relating to OA secondary to trauma were excluded. ROB-2 and ROBINS-I bias tools were used.Aims
Methods
The use of cementless total knee arthroplasty (TKA) components has increased during the past decade. The initial design of cementless metal-backed patellar components had shown high failure rates due to many factors. The aim of this study was to evaluate the clinical results of a second-generation cementless, metal-backed patellar component of a modern design. This was a retrospective review of 707 primary TKAs in 590 patients from a single institution, using a cementless, metal-backed patellar component with a mean follow-up of 6.9 years (2 to 12). A total of 409 TKAs were performed in 338 females and 298 TKAs in 252 males. The mean age of the patients was 63 years (34 to 87) and their mean BMI was 34.3 kg/m2 (18.8 to 64.5). The patients were chosen to undergo a cementless procedure based on age and preoperative radiological and intraoperative bone quality. Outcome was assessed using the Knee Society knee and function scores and range of motion (ROM), complications, and revisions.Aims
Methods
This work aimed at answering the following research questions: 1) What is the rate of mechanical complications, nonunion and infection for head/neck femoral fractures, intertrochanteric fractures, and subtrochanteric fractures in the elderly USA population? and 2) Which factors influence adverse outcomes? Proximal femoral fractures occurred between 1 January 2009 and 31 December 2019 were identified from the Medicare Physician Service Records Data Base. The Kaplan-Meier method with Fine and Gray sub-distribution adaptation was used to determine rates for nonunion, infection, and mechanical complications. Semiparametric Cox regression model was applied incorporating 23 measures as covariates to identify risk factors.Aims
Methods
The Vantage Total Ankle System is a fourth-generation low-profile fixed-bearing implant that has been available since 2016. We aimed to describe our early experience with this implant. This is a single-centre retrospective review of patients who underwent primary total ankle arthroplasty (TAA) with a Vantage implant between November 2017 and February 2020, with a minimum of two years’ follow-up. Four surgeons contributed patients. The primary outcome was reoperation and revision rate of the Vantage implant at two years. Secondary outcomes included radiological alignment, peri-implant complications, and pre- and postoperative patient-reported outcomes.Aims
Methods
This study describes the variation in the annual volumes of revision hip arthroplasty (RHA) undertaken by consultant surgeons nationally, and the rate of accrual of RHA and corresponding primary hip arthroplasty (PHA) volume for new consultants entering practice. National Joint Registry (NJR) data for England, Wales, Northern Ireland, and the Isle of Man were received for 84,816 RHAs and 818,979 PHAs recorded between April 2011 and December 2019. RHA data comprised all revision procedures, including first-time revisions of PHA and any subsequent re-revisions recorded in public and private healthcare organizations. Annual procedure volumes undertaken by the responsible consultant surgeon in the 12 months prior to every index procedure were determined. We identified a cohort of ‘new’ HA consultants who commenced practice from 2012 and describe their rate of accrual of PHA and RHA experience.Aims
Methods
This scoping review aims to identify patient-related factors associated with a poorer outcome following total ankle arthroplasty (TAA). A scoping review was performed using the Preferred Reporting Items for Systematic Reviews and Meta-Analyses guidelines. A computer-based literature search was performed in PubMed, Embase, Cochrane trials, and Web of Science. Two reviewers independently performed title/abstract and full-text screening according to predetermined selection criteria. English-language original research studies reporting patient-related factors associated with a poorer outcome following TAA were included. Outcomes were defined as patient-reported outcome measures (PROMs), perioperative complications, and failure.Aims
Methods
The primary aim of this prospective, multicentre study is to describe the rates of returning to golf following hip, knee, ankle, and shoulder arthroplasty in an active golfing population. Secondary aims will include determining the timing of return to golf, changes in ability, handicap, and mobility, and assessing joint-specific and health-related outcomes following surgery. This is a multicentre, prospective, longitudinal study between the Hospital for Special Surgery, (New York City, New York, USA) and Edinburgh Orthopaedics, Royal Infirmary of Edinburgh, (Edinburgh, UK). Both centres are high-volume arthroplasty centres, specializing in upper and lower limb arthroplasty. Patients undergoing hip, knee, ankle, or shoulder arthroplasty at either centre, and who report being golfers prior to arthroplasty, will be included. Patient-reported outcome measures will be obtained at six weeks, three months, six months, and 12 months. A two-year period of recruitment will be undertaken of arthroplasty patients at both sites.Aims
Methods
This study reports mid-term outcomes after periacetabular osteotomy (PAO) exclusively in a borderline hip dysplasia (BHD) population to provide a contrast to published outcomes for arthroscopic surgery of the hip in BHD. We identified 42 hips in 40 patients treated between January 2009 and January 2016 with BHD defined as a lateral centre-edge angle (LCEA) of ≥ 18° but < 25°. A minimum five-year follow-up was available. Patient-reported outcomes (PROMs) including Tegner score, subjective hip value (SHV), modified Harris Hip Score (mHHS), and Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC) were assessed. The following morphological parameters were evaluated: LCEA, acetabular index (AI), α angle, Tönnis staging, acetabular retroversion, femoral version, femoroepiphyseal acetabular roof index (FEAR), iliocapsularis to rectus femoris ratio (IC/RF), and labral and ligamentum teres (LT) pathology.Aims
Methods
Contemporary outcomes of primary total hip arthroplasties (THAs) with highly cross-linked polyethylene (HXLPE) liners in patients with inflammatory arthritis have not been well studied. This study examined the implant survivorship, complications, radiological results, and clinical outcomes of THA in patients with inflammatory arthritis. We identified 418 hips (350 patients) with a primary diagnosis of inflammatory arthritis who underwent primary THA with HXLPE liners from January 2000 to December 2017. Of these hips, 68% had rheumatoid arthritis (n = 286), 13% ankylosing spondylitis (n = 53), 7% juvenile rheumatoid arthritis (n = 29), 6% psoriatic arthritis (n = 24), 5% systemic lupus erythematosus (n = 23), and 1% scleroderma (n = 3). Mean age was 58 years (SD 14.8), 66.3% were female (n = 277), and mean BMI was 29 kg/m2 (SD 7). Uncemented femoral components were used in 77% of cases (n = 320). Uncemented acetabular components were used in all patients. Competing risk analysis was used accounting for death. Mean follow-up was 4.5 years (2 to 18).Aims
Methods
The aims of this study were to determine if an increasing serum cobalt (Co) and/or chromium (Cr) concentration is correlated with a decreasing Harris Hip Score (HHS) and Hip disability and Osteoarthritis Outcome Score (HOOS) in patients who received the Articular Surface Replacement (ASR) hip resurfacing arthroplasty (HRA), and to evaluate the ten-year revision rate and show if sex, inclination angle, and Co level influenced the revision rate. A total of 62 patients with an ASR-HRA were included and monitored yearly postoperatively. At follow-up, serum Co and Cr levels were measured and the HHS and the HOOS were scored. In addition, preoperative patient and implant variables and the need for revision surgery were recorded. We used a linear mixed model to relate the serum Co and Cr levels to different patient-reported outcome measures (PROMs). For the survival analyses we used the Kaplan-Meier and Cox regression model.Aims
Methods
Aims. Nonagenarians (aged 90 to 99 years) have experienced the fastest percent decile population growth in the USA recently, with a consequent increase in the prevalence of nonagenarians living with joint arthroplasties. As such, the number of revision total hip arthroplasties (THAs) and total knee arthroplasties (TKAs) in nonagenarians is expected to increase. We aimed to determine the mortality rate, implant survivorship, and complications of nonagenarians undergoing aseptic revision THAs and revision TKAs. Methods. Our institutional total joint registry was used to identify 96 nonagenarians who underwent 97 aseptic revisions (78 hips and 19 knees) between 1997 and 2018. The most common indications were aseptic loosening and periprosthetic fracture for both revision THAs and revision TKAs. Mean age at revision was 92 years (90 to 98), mean BMI was 27 kg/m. 2. (16 to 47), and 67% (n = 65) were female. Mean time between primary and revision was 18 years (SD 9). Kaplan-Meier survival was used for patient mortality, and compared to age- and sex-matched control populations. Reoperation risk was assessed using cumulative incidence with death as a competing risk. Mean follow-up was five years. Results. Mortality rates were 9%, 18%, 26%, and 62% at 90 days, one year, two years, and five years, respectively, but similar to control populations. There were 43 surgical complications and five reoperations, resulting in a cumulative incidence of reoperation of 4% at five years. Medical complications were common, with a cumulative incidence of 65% at 90 days.
Loosening of components after total knee arthroplasty (TKA) can be associated with the development of radiolucent lines (RLLs). The aim of this study was to assess the rate of formation of RLLs in the cemented original design of the ATTUNE TKA and their relationship to loosening. A systematic search was undertaken using the Cochrane methodology in three online databases: MEDLINE, Embase, and CINAHL. Studies were screened against predetermined criteria, and data were extracted. Available National Joint Registries in the Network of Orthopaedic Registries of Europe were also screened. A random effects model meta-analysis was undertaken.Aims
Methods
Although the Fitmore Hip Stem has been on the market for almost 15 years, it is still not well documented in randomized controlled trials. This study compares the Fitmore stem with the CementLeSs (CLS) in several different clinical and radiological aspects. The hypothesis is that there will be no difference in outcome between stems. In total, 44 patients with bilateral hip osteoarthritis were recruited from the outpatient clinic at a single tertiary orthopaedic centre. The patients were operated with bilateral one-stage total hip arthroplasty. The most painful hip was randomized to either Fitmore or CLS femoral component; the second hip was operated with the femoral component not used on the first side. Patients were evaluated at three and six months and at one, two, and five years postoperatively with patient-reported outcome measures, radiostereometric analysis, dual-energy X-ray absorptiometry, and conventional radiography. A total of 39 patients attended the follow-up visit at two years (primary outcome) and 35 patients at five years. The primary outcome was which hip the patient considered to have the best function at two years.Aims
Methods
To systematically review the efficacy of split tendon transfer surgery on gait-related outcomes for children and adolescents with cerebral palsy (CP) and spastic equinovarus foot deformity. Five databases (CENTRAL, CINAHL, PubMed, Embase, Web of Science) were systematically screened for studies investigating split tibialis anterior or split tibialis posterior tendon transfer for spastic equinovarus foot deformity, with gait-related outcomes (published pre-September 2022). Study quality and evidence were assessed using the Methodological Index for Non-Randomized Studies, the Risk of Bias In Non-Randomized Studies of Interventions, and the Grading of Recommendations Assessment, Development and Evaluation.Aims
Methods
The Exeter V40 femoral stem is the most implanted stem in the National Joint Registry (NJR) for primary total hip arthroplasty (THA). In 2004, the 44/00/125 stem was released for use in ‘cement-in-cement’ revision cases. It has, however, been used ‘off-label’ as a primary stem when patient anatomy requires a smaller stem with a 44 mm offset. We aimed to investigate survival of this implant in comparison to others in the range when used in primary THAs recorded in the NJR. We analyzed 328,737 primary THAs using the Exeter V40 stem, comprising 34.3% of the 958,869 from the start of the NJR to December 2018. Our exposure was the stem, and the outcome was all-cause construct revision. We stratified analyses into four groups: constructs using the 44/00/125 stem, those using the 44/0/150 stem, those including a 35.5/125 stem, and constructs using any other Exeter V40 stem.Aims
Methods
The aim of this study was to determine the prevalence of depressive and anxiety disorders prior to total hip (THA) and total knee arthroplasty (TKA) and to assess their impact on the rates of any infection, revision, or reoperation. Between January 2000 and March 2019, 21,469 primary and revision arthroplasties (10,011 THAs; 11,458 TKAs), which were undertaken in 15,504 patients at a single academic medical centre, were identified from a 27-county linked electronic medical record (EMR) system. Depressive and anxiety disorders were identified by diagnoses in the EMR or by using a natural language processing program with subsequent validation from review of the medical records. Patients with mental health diagnoses other than anxiety or depression were excluded.Aims
Methods
Periprosthetic joint infection (PJI) in total hip arthroplasty in the elderly may occur but has been subject to limited investigation. This study analyzed infection characteristics, surgical outcomes, and perioperative complications of octogenarians undergoing treatment for PJI in a single university-based institution. We identified 33 patients who underwent treatment for PJIs of the hip between January 2010 and December 2019 using our institutional joint registry. Mean age was 82 years (80 to 90), with 19 females (57%) and a mean BMI of 26 kg/m2 (17 to 41). Mean American Society of Anesthesiologists (ASA) grade was 3 (1 to 4) and mean Charlson Comorbidity Index was 6 (4 to 10). Leading pathogens included coagulase-negative Staphylococci (45%) and Aims
Methods
The optimal choice of management for proximal humerus fractures (PHFs) has been increasingly discussed in the literature, and this work aimed to answer the following questions: 1) what are the incidence rates of PHF in the geriatric population in the USA; 2) what is the mortality rate after PHF in the elderly population, specifically for distinct treatment procedures; and 3) what factors influence the mortality rate? PHFs occurring between 1 January 2009 and 31 December 2019 were identified from the Medicare physician service records. Incidence rates were determined, mortality rates were calculated, and semiparametric Cox regression was applied, incorporating 23 demographic, clinical, and socioeconomic covariates, to compare the mortality risk between treatments.Aims
Methods
Rates of reverse total shoulder arthroplasty (rTSA) continue to grow. Glenoid bone loss and deformity remains a technical challenge to the surgeon and may reduce improvements in patients’ outcomes. However, there is no consensus as to the optimal surgical technique to best reconstruct these patients’ anatomy. This review aims to compare the outcomes of glenoid bone grafting versus augmented glenoid prostheses in the management of glenoid bone loss in primary reverse total shoulder arthroplasty. This systematic review and meta-analysis evaluated study-level data in accordance with the Preferred Reporting Items for Systematic Reviews and Meta-Analyses statement. We performed searches of Medline (Ovid), Embase (Ovid), and PubMed from their dates of inception to January 2022. From included studies, we analyzed data for preoperative and postoperative range of motion (ROM), patient-reported functional outcomes, and complication rates.Aims
Methods
It is not known whether preservation of the capsule of the hip positively affects patient-reported outcome measures (PROMs) in total hip arthroplasty using the direct anterior approach (DAA-THA). A recent randomized controlled trial found no clinically significant difference at one year postoperatively. This study aimed to determine whether preservation of the anterolateral capsule and anatomical closure improve the outcome and revision rate, when compared with resection of the anterolateral capsule, at two years postoperatively. Two consecutive groups of patients whose operations were performed by the senior author were compared. The anterolateral capsule was resected in the first group of 430 patients between January 2012 and December 2014, and preserved and anatomically closed in the second group of 450 patients between July 2015 and December 2017. There were no other technical changes between the two groups. Patient characteristics, the Charlson Comorbidity Index (CCI), and surgical data were collected from our database. PROM questionnaires, consisting of the Oxford Hip Score (OHS) and Core Outcome Measures Index (COMI-Hip), were collected two years postoperatively. Data were analyzed with generalized multiple regression analysis.Aims
Methods
The aim of this study was to explore the relationship between reason for revision total hip arthroplasty (rTHA) and outcomes in terms of patient-reported outcome measures (PROMs). We reviewed a prospective cohort of 647 patients undergoing full or partial rTHA at a single high-volume centre with a minimum of two years’ follow-up. The reasons for revision were classified as: infection; aseptic loosening; dislocation; structural failure; and painful THA for other reasons. PROMs (modified Oxford Hip Score (mOHS), EuroQol five-dimension three-level health questionnaire (EQ-5D-3L) score, and visual analogue scales for pain during rest and activity), complication rates, and failure rates were compared among the groups.Aims
Methods
Registry studies on modified acetabular polyethylene (PE) liner designs are limited. We investigated the influence of standard and modified PE acetabular liner designs on the revision rate for mechanical complications in primary cementless total hip arthroplasty (THA). We analyzed 151,096 primary cementless THAs from the German Arthroplasty Registry (EPRD) between November 2012 and November 2020. Cumulative incidence of revision for mechanical complications for standard and four modified PE liners (lipped, offset, angulated/offset, and angulated) was determined using competing risk analysis at one and seven years. Confounders were investigated with a Cox proportional-hazards model.Aims
Methods
The aim of this study was to investigate the rate of revision for distal femoral arthroplasty (DFA) performed as a primary procedure for native knee fractures using data from the Australian Orthopaedic Association National Joint Arthroplasty Registry (AOANJRR). Data from the AOANJRR were obtained for DFA performed as primary procedures for native knee fractures from 1 September 1999 to 31 December 2020. Pathological fractures and revision for failed internal fixation were excluded. The five prostheses identified were the Global Modular Arthroplasty System, the Modular Arthroplasty System, the Modular Universal Tumour And Revision System, the Orthopaedic Salvage System, and the Segmental System. Patient demographic data (age, sex, and American Society of Anesthesiologists grade) were obtained, where available. Kaplan-Meier estimates of survival were used to determine the rate of revision, and the reasons for revision and mortality data were examined.Aims
Methods
The aim of this study was to explore current use of the Global Fragility Fracture Network (FFN) Minimum Common Dataset (MCD) within established national hip fracture registries, and to propose a revised MCD to enable international benchmarking for hip fracture care. We compared all ten established national hip fracture registries: England, Wales, and Northern Ireland; Scotland; Australia and New Zealand; Republic of Ireland; Germany; the Netherlands; Sweden; Norway; Denmark; and Spain. We tabulated all questions included in each registry, and cross-referenced them against the 32 questions of the MCD dataset. Having identified those questions consistently used in the majority of national audits, and which additional fields were used less commonly, we then used consensus methods to establish a revised MCD.Aims
Methods
Endoprosthetic reconstruction with a distal femoral arthroplasty (DFA) can be used to treat distal femoral bone loss from oncological and non-oncological causes. This study reports the short-term implant survivorship, complications, and risk factors for patients who underwent DFA for non-neoplastic indications. We performed a retrospective review of 75 patients from a single institution who underwent DFA for non-neoplastic indications, including aseptic loosening or mechanical failure of a previous prosthesis (n = 25), periprosthetic joint infection (PJI) (n = 23), and native or periprosthetic distal femur fracture or nonunion (n = 27). Patients with less than 24 months’ follow-up were excluded. We collected patient demographic data, complications, and reoperations. Reoperation for implant failure was used to calculate implant survivorship.Aims
Methods
The aim of this study was to determine the outcome of all primary total hip arthroplasties (THAs) and their subsequent revision procedures in patients aged under 50 years performed at our institution. All 1,049 primary THAs which were undertaken in 860 patients aged under 50 years between 1988 and 2018 in our tertiary care institution were included. We used cemented implants in both primary and revision surgery. Impaction bone grafting was used in patients with acetabular or femoral bone defects. Kaplan-Meier analyses were used to determine the survival of primary and revision THA with the endpoint of revision for any reason, and of revision for aseptic loosening.Aims
Methods
The outcome of repeat septic revision after a failed one-stage exchange for periprosthetic joint infection (PJI) in total knee arthroplasty (TKA) remains unknown. The aim of this study was to report the infection-free and all-cause revision-free survival of repeat septic revision after a failed one-stage exchange, and to determine whether the Musculoskeletal Infection Society (MSIS) stage is associated with subsequent infection-related failure. We retrospectively reviewed all repeat septic revision TKAs which were undertaken after a failed one-stage exchange between 2004 and 2017. A total of 33 repeat septic revisions (29 one-stage and four two-stage) met the inclusion criteria. The mean follow-up from repeat septic revision was 68.2 months (8.0 months to 16.1 years). The proportion of patients who had a subsequent infection-related failure and all-cause revision was reported and Kaplan-Meier survival for these endpoints was determined. Patients were categorized according to the MSIS staging system, and the association with subsequent infection was analyzed.Aims
Methods
Total hip arthroplasties (THAs) are performed by surgeons at various stages in training with varying levels of supervision, but we do not know if this is safe practice with comparable outcomes to consultant-performed THA. Our aim was to examine the association between surgeon grade, the senior supervision of trainees, and the risk of revision following THA. We performed an observational study using National Joint Registry (NJR) data. We included adult patients who underwent primary THA for osteoarthritis, recorded in the NJR between 2003 and 2016. Exposures were operating surgeon grade (consultant or trainee) and whether or not trainees were directly supervised by a scrubbed consultant. Outcomes were all-cause revision and the indication for revision up to ten years. We used methods of survival analysis, adjusted for patient, operation, and healthcare setting factors.Aims
Methods
Periprosthetic joint infection (PJI) occurs in approximately 1% to 2% of total knee arthroplasties (TKA) presenting multiple challenges, such as difficulty in diagnosis, technical complexity, and financial costs. Two-stage exchange is the gold standard for treating PJI but emerging evidence suggests 'two-in-one' single-stage revision as an alternative, delivering comparable outcomes, reduced morbidity, and cost-effectiveness. This study investigates five-year results of modified single-stage revision for treatment of PJI following TKA with bone loss. Patients were identified from prospective data on all TKA patients with PJI following the primary procedure. Inclusion criteria were: revision for PJI with bone loss requiring reconstruction, and a minimum five years’ follow-up. Patients were followed up for recurrent infection and assessment of function. Tools used to assess function were Oxford Knee Score (OKS) and American Knee Society Score (AKSS).Aims
Methods
Large-diameter metal-on-metal (MoM) total hip arthroplasty (THA) has demonstrated unexpected high failure rates and pseudotumour formation. The purpose of this prospective cohort study is to report ten-year results in order to establish revision rate, prevalence of pseudotumour formation, and relation with whole blood cobalt levels. All patients were recalled according to the guidelines of the Dutch Orthopaedic Association. They underwent clinical and radiographical assessments (radiograph and CT scan) of the hip prosthesis and whole blood cobalt ion measurements. Overall, 94 patients (95 hips) fulfilled our requirements for a minimum ten-year follow-up.Aims
Methods
One-stage exchange for periprosthetic joint infection (PJI) in total hip arthroplasty (THA) is gaining popularity. The outcome for a repeat one-stage revision THA after a failed one-stage exchange for infection remains unknown. The aim of this study was to report the infection-free and all-cause revision-free survival of repeat one-stage exchange, and to investigate the association between the Musculoskeletal Infection Society (MSIS) staging system and further infection-related failure. We retrospectively reviewed all repeat one-stage revision THAs performed after failed one-stage exchange THA for infection between January 2008 and December 2016. The final cohort included 32 patients. The mean follow-up after repeat one-stage exchange was 5.3 years (1.2 to 13.0). The patients with a further infection-related failure and/or all-cause revision were reported, and Kaplan-Meier survival for these endpoints determined. Patients were categorized according to the MSIS system, and its association with further infection was analyzed.Aims
Methods
Total hip arthroplasty (THA) with dual-mobility components (DM-THA) has been shown to decrease the risk of dislocation in the setting of a displaced neck of femur fracture compared to conventional single-bearing THA (SB-THA). This study assesses if the clinical benefit of a reduced dislocation rate can justify the incremental cost increase of DM-THA compared to SB-THA. Costs and benefits were established for patients aged 75 to 79 years over a five-year time period in the base case from the Canadian Health Payer’s perspective. One-way and probabilistic sensitivity analysis assessed the robustness of the base case model conclusions.Aims
Methods
Proximal femoral endoprosthetic replacements (PFEPRs) are the most common reconstruction option for osseous defects following primary and metastatic tumour resection. This study aimed to compare the rate of implant failure between PFEPRs with monopolar and bipolar hemiarthroplasties and acetabular arthroplasties, and determine the optimum articulation for revision PFEPRs. This is a retrospective review of 233 patients who underwent PFEPR. The mean age was 54.7 years (SD 18.2), and 99 (42.5%) were male. There were 90 patients with primary bone tumours (38.6%), 122 with metastatic bone disease (52.4%), and 21 with haematological malignancy (9.0%). A total of 128 patients had monopolar (54.9%), 74 had bipolar hemiarthroplasty heads (31.8%), and 31 underwent acetabular arthroplasty (13.3%).Aims
Methods
The aim of our study was to investigate the effect of asymmetric crosslinked polyethylene liner use on the risk of revision of cementless and hybrid total hip arthroplasties (THAs). We undertook a registry study combining the National Joint Registry dataset with polyethylene manufacturing characteristics as supplied by the manufacturers. The primary endpoint was revision for any reason. We performed further analyses on other reasons including instability, aseptic loosening, wear, and liner dissociation. The primary analytic approach was Cox proportional hazard regression.Aims
Methods
Single-stage revision total knee arthroplasty (rTKA) is gaining popularity in treating chronic periprosthetic joint infections (PJIs). We have introduced this approach to our clinical practice and sought to evaluate rates of reinfection and re-revision, along with predictors of failure of both single- and two-stage rTKA for chronic PJI. A retrospective comparative cohort study of all rTKAs for chronic PJI between 1 April 2003 and 31 December 2018 was undertaken using prospective databases. Patients with acute infections were excluded; rTKAs were classified as single-stage, stage 1, or stage 2 of two-stage revision. The primary outcome measure was failure to eradicate or recurrent infection. Variables evaluated for failure by regression analysis included age, BMI, American Society of Anesthesiologists grade, infecting organisms, and the presence of a sinus. Patient survivorship was also compared between the groups.Aims
Methods
We studied the outcomes of hip and knee arthroplasties in a high-volume arthroplasty centre to determine if patients with morbid obesity (BMI ≥ 40 kg/m2) had unacceptably worse outcomes as compared to those with BMI < 40 kg/m2. In a two-year period, 4,711 patients had either total hip arthroplasty (THA; n = 2,370), total knee arthroplasty (TKA; n = 2,109), or unicompartmental knee arthroplasty (UKA; n = 232). Of these patients, 392 (8.3%) had morbid obesity. We compared duration of operation, anaesthetic time, length of stay (LOS), LOS > three days, out of hours attendance, emergency department attendance, readmission to hospital, return to theatre, and venous thromboembolism up to 90 days. Readmission for wound infection was recorded to one year. Oxford scores were recorded preoperatively and at one year postoperatively.Aims
Methods
Use of the direct anterior approach (DAA) for total hip arthroplasty (THA) has increased in recent years due to proposed benefits, including a lower risk of dislocation and improved early functional recovery. This study investigates the dislocation rate in a non-selective, consecutive cohort undergoing THA via the DAA without any exclusion or bias in patient selection based on habitus, deformity, age, sex, or fixation method. We retrospectively reviewed all patients undergoing THA via the DAA between 2011 and 2017 at our institution. Primary outcome was dislocation at minimum two-year follow-up. Patients were stratified by demographic details and risk factors for dislocation, and an in-depth analysis of dislocations was performed.Aims
Methods
Uncemented mobile bearing designs in medial unicompartmental knee arthroplasty (UKA) have seen an increase over the last decade. However, there are a lack of large-scale studies comparing survivorship of these specific designs to commonly used cemented mobile and fixed bearing designs. The aim of this study was to evaluate the survivorship of these designs. A total of 21,610 medial UKAs from 2007 to 2018 were selected from the Dutch Arthroplasty Register. Multivariate Cox regression analyses were used to compare uncemented mobile bearings with cemented mobile and fixed bearings. Adjustments were made for patient and surgical factors, with their interactions being considered. Reasons and type of revision in the first two years after surgery were assessed.Aims
Methods
This is a multicentre, non-inventor, prospective observational study of 503 INFINITY fixed bearing total ankle arthroplasties (TAAs). We report our early experience, complications, and radiological and functional outcomes. Patients were recruited from 11 specialist centres between June 2016 and November 2019. Demographic, radiological, and functional outcome data (Ankle Osteoarthritis Scale, Manchester Oxford Questionnaire, and EuroQol five-dimension five-level score) were collected preoperatively, at six months, one year, and two years. The Canadian Orthopaedic Foot and Ankle Society (COFAS) grading system was used to stratify deformity. Early and late complications and reoperations were recorded as adverse events. Radiographs were assessed for lucencies, cysts, and/or subsidence.Aims
Methods
Unicompartmental knee arthroplasty (UKA) is a bone-preserving treatment option for osteoarthritis localized to a single compartment in the knee. The success of the procedure is sensitive to patient selection and alignment errors. Robotic arm-assisted UKA provides technological assistance to intraoperative bony resection accuracy, which is thought to improve ligament balancing. This paper presents the five-year outcomes of a comparison between manual and robotically assisted UKAs. The trial design was a prospective, randomized, parallel, single-centre study comparing surgical alignment in patients undergoing UKA for the treatment of medial compartment osteoarthritis (ISRCTN77119437). Participants underwent surgery using either robotic arm-assisted surgery or conventional manual instrumentation. The primary outcome measure (surgical accuracy) has previously been reported, and, along with secondary outcomes, were collected at one-, two-, and five-year timepoints. Analysis of five-year results and longitudinal analysis for all timepoints was performed to compare the two groups.Aims
Methods
There is no consensus on the treatment of proximal humeral fractures. Hemiarthroplasty has been widely used in patients when non-surgical treatment is not possible. There is, despite extensive use, limited information about the long-term outcome. Our primary aim was to report ten-year patient-reported outcome after hemiarthroplasty for acute proximal humeral fractures. The secondary aims were to report the cumulative revision rate and risk factors for an inferior patient-reported outcome. We obtained data on 1,371 hemiarthroplasties for acute proximal humeral fractures from the Danish Shoulder Arthroplasty Registry between 2006 and 2010. Of these, 549 patients (40%) were alive and available for follow-up. The Western Ontario Osteoarthritis of the Shoulder (WOOS) questionnaire was sent to all patients at nine to 14 years after primary surgery. Revision rates were calculated using the Kaplan-Meier method. Risk factors for an inferior WOOS score were analyzed using the linear regression model.Aims
Methods
The purpose of this study was to evaluate unexpected positive cultures in total hip arthroplasty (THA) revisions for presumed aseptic loosening, to assess the prevalence of low-grade infection using two definition criteria, and to analyze its impact on implant survival after revision. A total of 274 THA revisions performed for presumed aseptic loosening from 2012 to 2016 were reviewed. In addition to obtaining intraoperative tissue cultures from all patients, synovial and sonication fluid samples of the removed implant were obtained in 215 cases (79%) and 101 cases (37%), respectively. Histopathological analysis was performed in 250 cases (91%). Patients were classified as having low-grade infections according to institutional criteria and Musculoskeletal Infection Society (MSIS) International Consensus Meeting (ICM) 2013 criteria. Low-grade infections according to institutional criteria were treated with targeted antibiotics for six weeks postoperatively. Implant failure was defined as the need for re-revision resulting from periprosthetic joint infection (PJI) and aseptic reasons. The mean follow-up was 68 months (26 to 95).Aims
Methods
Debate remains whether the patella should be resurfaced during total knee replacement (TKR). For non-resurfaced TKRs, we estimated what the revision rate would have been if the patella had been resurfaced, and examined the risk of re-revision following secondary patellar resurfacing. A retrospective observational study of the National Joint Registry (NJR) was performed. All primary TKRs for osteoarthritis alone performed between 1 April 2003 and 31 December 2016 were eligible (n = 842,072). Patellar resurfacing during TKR was performed in 36% (n = 305,844). The primary outcome was all-cause revision surgery. Secondary outcomes were the number of excess all-cause revisions associated with using TKRs without (versus with) patellar resurfacing, and the risk of re-revision after secondary patellar resurfacing.Aims
Methods
This systematic review asked which patterns of complications are associated with the three reverse total shoulder arthroplasty (RTSA) prosthetic designs, as classified by Routman et al, in patients undergoing RTSA for the management of cuff tear arthropathy, massive cuff tear, osteoarthritis, and rheumatoid arthritis. The three implant design philosophies investigated were medial glenoid/medial humerus (MGMH), medial glenoid/lateral humerus (MGLH), and lateral glenoid/medial humerus (LGMH). A systematic review of the literature was performed via a search of MEDLINE and Embase. Two reviewers extracted data on complication occurrence and patient-reported outcome measures (PROMs). Meta-analysis was conducted on the reported proportion of complications, weighted by sample size, and PROMs were pooled using the reported standardized mean difference (SMD). Quality of methodology was assessed using Wylde’s non-summative four-point system. The study was registered with PROSPERO (CRD42020193041).Aims
Methods
Periprosthetic femoral fractures (PPF) are a serious complication of total hip arthroplasty (THA) and are becoming an increasingly common indication for revision arthroplasty with the ageing population. This study aimed to identify potential risk factors for PPF based on an analysis of registry data. Cases recorded with PPF as the primary indication for revision arthroplasty in the German Arthroplasty Registry (Endoprothesenregister Deutschland (EPRD)), as well as those classified as having a PPF according to the International Classification of Diseases (ICD) codes in patients’ insurance records were identified from the complete datasets of 249,639 registered primary hip arthroplasties in the EPRD and included in the analysis.Aims
Methods
Complex joint fractures of the lower extremity are often accompanied by soft-tissue swelling and are associated with prolonged hospitalization and soft-tissue complications. The aim of the study was to evaluate the effect of vascular impulse technology (VIT) on soft-tissue conditioning in comparison with conventional elevation. A total of 100 patients were included in this prospective, randomized, controlled monocentre study allocated to the three subgroups of dislocated ankle fracture (n = 40), pilon fracture (n = 20), and intra-articular calcaneal fracture (n = 40). Patients were randomized to the two study groups in a 1:1 ratio. The effectiveness of VIT (intervention) compared with elevation (control) was analyzed separately for the whole study population and for the three subgroups. The primary endpoint was the time from admission until operability (in days).Aims
Methods
To investigate whether chronic kidney disease (CKD) is associated with the risk of all-cause revision or revision due to a periprosthetic joint infection (PJI) after primary hip or knee arthroplasty. This retrospective cohort study comprised 18,979 consecutive hip and knee arthroplasties from a single high-volume academic hospital. At a median of 5.6 years (interquartile range (IQR) 3.5 to 8.1), all deaths and revisions were counted. To overcome the competing risk of death, competing risk analysis using the cumulative incidence function (CIF) was applied to analyze the association between different stages of CKD and revisions. Confounding factors such as diabetes and BMI were considered using either a stratified CIF or the Fine and Gray model.Aims
Methods
Total hip arthroplasty (THA) patients undergoing or having a prior lumbar spine fusion (LSF) have an increased risk of mechanical complications. The aim of this registry-based, retrospective comparative cohort study is to assess the longer term survival of THA in patients who have undergone a LSF during a 17-year period (2000 to 2017). A registry-based population study was conducted on 679 patients who underwent both THA and LSF surgeries. Patients were identified from the regional arthroplasty data base and cross linked to patients with LSF from the regional hospital discharge database between 2000 and 2017. Demographic data, diagnosis leading to primary THA, primary implant survival, perioperative complications, number and causes of failure, and patients requiring revision arthroplasty were collated and compared. For comparison, data from 67,919 primary THAs performed during the same time time period were also retrieved and analyzed.Aims
Methods
The objective of this systematic review was to describe trapeziectomy outcomes and complications in the context of osteoarthritis of the base of the thumb after a five-year minimum follow-up. Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) guidelines were used to guide study design, and 267 full-text articles were assessed for eligibility. After exclusion criteria application, 22 studies were included, involving 728 patients and 823 trapeziectomies. Outcomes included pre- and postoperative clinical and radiological characteristics. Complications and revisions were recorded.Aims
Methods
The aim of this study was to determine whether fixation, as opposed to revision arthroplasty, can be safely used to treat reducible Vancouver B type fractures in association with a cemented collarless polished tapered femoral stem (the Exeter). This retrospective cohort study assessed 152 operatively managed consecutive unilateral Vancouver B fractures involving Exeter stems; 130 were managed with open reduction and internal fixation (ORIF) and 22 with revision arthroplasty. Mean follow-up was 6.5 years (SD 2.6; 3.2 to 12.1). The primary outcome measure was revision of at least one component. Kaplan–Meier survival analysis was performed. Regression analysis was used to identify risk factors for revision following ORIF. Secondary outcomes included any reoperation, complications, blood transfusion, length of hospital stay, and mortality.Aims
Methods
For many designs of total knee arthroplasty (TKA) it remains unclear whether cemented or uncemented fixation provides optimal long-term survival. The main limitation in most studies is a retrospective or non-comparative study design. The same is true for comparative trials looking only at the survival rate as extensive sample sizes are needed to detect true differences in fixation and durability. Studies using radiostereometric analysis (RSA) techniques have shown to be highly predictive in detecting late occurring aseptic loosening at an early stage. To investigate the difference in predicted long-term survival between cemented, uncemented, and hybrid fixation of TKA, we performed a randomized controlled trial using RSA. A total of 105 patients were randomized into three groups (cemented, uncemented, and hybrid fixation of the ACS Mobile Bearing (ACS MB) knee system, implantcast). RSA examinations were performed on the first day after surgery and at scheduled follow-up visits at three months, six months, one year, and two years postoperatively. Patient-reported outcome measures (PROMs) were obtained preoperatively and after two years follow-up. Patients and follow-up investigators were blinded for the result of randomization.Aims
Methods
Preoperative talar valgus deformity ≥ 15° is considered a contraindication for total ankle arthroplasty (TAA). We compared operative procedures and clinical outcomes of TAA in patients with talar valgus deformity ≥ 15° and < 15°. A matched cohort of patients similar for demographics and components used but differing in preoperative coronal-plane tibiotalar valgus deformity ≥ 15° (valgus, n = 50; 52% male, mean age 65.8 years (SD 10.3), mean body mass index (BMI) 29.4 (SD 5.2)) or < 15° (control, n = 50; 58% male, mean age 65.6 years (SD 9.8), mean BMI 28.7 (SD 4.2)), underwent TAA by one surgeon. Preoperative and postoperative radiographs, Ankle Osteoarthritis Scale (AOS) pain and disability and 36-item Short Form Health Survey (SF-36) version 2 scores were collected prospectively. Ancillary procedures, secondary procedures, and complications were recorded.Aims
Methods
Stiffness is a common complication after total knee arthroplasty (TKA). Pathogenesis is not understood, treatment options are limited, and diagnosis is challenging. The aim of this study was to investigate if MRI can be used to visualize intra-articular scarring in patients with stiff, painful knee arthroplasties. Well-functioning primary TKAs (n = 11), failed non-fibrotic TKAs (n = 5), and patients with a clinical diagnosis of fibrosisAims
Methods
Aims. To report mid- to long-term results of Oxford mobile bearing domed lateral unicompartmental knee arthroplasty (UKA), and determine the effect of potential contraindications on outcome. Methods. A total of 325 consecutive domed lateral UKAs undertaken for the recommended indications were included, and their functional and survival outcomes were assessed. The effects of age, weight, activity, and the presence of full-thickness erosions of cartilage in the patellofemoral joint on outcome were evaluated. Results. Median follow-up was seven years (3 to 14), and mean age at surgery was 65 years (39 to 90). Median Oxford Knee Score (OKS) was 43 (interquartile range (IQR) 37 to 47), with 260 (80%) achieving a good or excellent score (OKS > 34).
Aims. The purpose of this study is to examine six types of bearing surfaces implanted at a single institution over three decades to determine whether the reasons for revision vary among the groups and how long it takes to identify differences in survival. Methods. We considered six cohorts that included a total of 1,707 primary hips done between 1982 and 2010. These included 223 conventional polyethylene sterilized with γ irradiation in air (CPE-GA), 114 conventional polyethylene sterilized with gas plasma (CPE-GP), 116 crosslinked polyethylene (XLPE), 1,083 metal-on-metal (MOM), 90 ceramic-on-ceramic (COC), and 81 surface arthroplasties (SAs). With the exception of the COC, all other groups used cobalt-chromium (CoCr) femoral heads. The mean follow-up was 10 (0.008 to 35) years. Descriptive statistics with revisions per 100 component years (re/100 yr) and survival analysis with revision for any reason as the endpoint were used to compare bearing surfaces. Results. XLPE liners demonstrated a lower cumulative incidence of revision at 15 years compared to the CPE-GA and CPE-GP groups owing to the absence of wear-related revisions (4% for XLPE vs 18%, p = 0.02, and 15%, p = 0.003, respectively).
It remains controversial whether patellofemoral joint pathology is a contraindication to lateral unicompartmental knee arthroplasty (UKA). This study aimed to evaluate the effect of preoperative radiological degenerative changes and alignment on patient-reported outcome scores (PROMs) after lateral UKA. Secondarily, the influence of lateral UKA on the alignment of the patellofemoral joint was studied. A consecutive series of patients who underwent robotic arm-assisted fixed-bearing lateral UKA with at least two-year follow-up were retrospectively reviewed. Radiological evaluation was conducted to obtain a Kellgren Lawrence (KL) grade, an Altman score, and alignment measurements for each knee. Postoperative PROMs were assessed using the Kujala (Anterior Knee Pain Scale) score, Knee Injury and Osteoarthritis Outcome Score Joint Replacement (KOOS JR), and satisfaction levels.Aims
Methods
Two-stage exchange arthroplasty is the most common definitive treatment for prosthetic joint infection (PJI) in the USA. Complications that occur during treatment are often not considered. The purpose of this study was to analyze complications in patients undergoing two-stage exchange for infected total knee arthroplasty (TKA) and determine when they occur. We analyzed all patients that underwent two-stage exchange arthroplasty for treatment of PJI of the knee from January 2010 to December 2018 at a single institution. We categorized complications as medical versus surgical. The intervals for complications were divided into: interstage; early post-reimplantation (three months); and late post-reimplantation (three months to minimum one year). Minimum follow-up was one year. In total, 134 patients underwent a first stage of a two-stage exchange. There were 69 males and 65 females with an mean age at first stage surgery of 67 years (37 to 89). Success was based on the new Musculoskeletal Infection Society (MSIS) definition of success reporting.Aims
Methods
Aseptic loosening of the tibial component is a frequent cause of failure in primary total knee arthroplasty (TKA). Management options include an isolated tibial revision or full component revision. A full component revision is frequently selected by surgeons unfamiliar with the existing implant or who simply wish to “start again”. This option adds morbidity compared with an isolated tibial revision. While isolated tibial revision has a lower morbidity, it is technically more challenging due to difficulties with exposure and maintaining prosthetic stability. This study was designed to compare these two reconstructive options. Patients undergoing revision TKA for isolated aseptic tibial loosening between 2012 and 2017 were identified. Those with revision implants or revised for infection, instability, osteolysis, or femoral component loosening were excluded. A total of 164 patients were included; 88 had an isolated tibial revision and 76 had revision of both components despite only having a loose tibial component. The demographics and clinical and radiological outcomes were recorded.Aims
Methods
The aim of this study was to determine the impact of the severity of anaemia on postoperative complications following total hip arthroplasty (THA) and total knee arthroplasty (TKA). A retrospective cohort study was conducted using the American College of Surgeons National Quality Improvement Program (ACS-NSQIP) database. All patients who underwent primary TKA or THA between January 2012 and December 2017 were identified and stratified based upon hematocrit level. In this analysis, we defined anaemia as packed cell volume (Hct) < 36% for women and < 39% for men, and further stratified anaemia as mild anaemia (Hct 33% to 36% for women, Hct 33% to 39% for men), and moderate to severe (Hct < 33% for both men and women). Univariate and multivariate analyses were used to evaluate the incidence of multiple adverse events within 30 days of arthroplasty.Aims
Methods
There has been a significant reduction in unicompartmental knee arthroplasty (UKA) procedures recorded in Australia. This follows several national joint registry studies documenting high UKA revision rates when compared to total knee arthroplasty (TKA). With the recent introduction of robotically assisted UKA procedures, it is hoped that outcomes improve. This study examines the cumulative revision rate of UKA procedures implanted with a newly introduced robotic system and compares the results to one of the best performing non-robotically assisted UKA prostheses, as well as all other non-robotically assisted UKA procedures. Data from the Australian Orthopaedic Association National Joint Arthroplasty Registry (AOANJRR) for all UKA procedures performed for osteoarthritis (OA) between 2015 and 2018 were analyzed. Procedures using the Restoris MCK UKA prosthesis implanted using the Mako Robotic-Arm Assisted System were compared to non-robotically assisted Zimmer Unicompartmental High Flex Knee System (ZUK) UKA, a commonly used UKA with previously reported good outcomes and to all other non-robotically assisted UKA procedures using Cox proportional hazard ratios (HRs) and Kaplan-Meier estimates of survivorship.Aim
Methods
Although good clinical outcomes have been reported for monolithic tapered, fluted, titanium stems (TFTS), early results showed high rates of subsidence. Advances in stem design may mitigate these concerns. This study reports on the use of a current monolithic TFTS for a variety of indications. A multi-institutional retrospective study of all consecutive total hip arthroplasty (THA) and revision total hip arthroplasty (rTHA) patients who received the monolithic TFTS was conducted. Surgery was performed by eight fellowship-trained arthroplasty surgeons at four institutions. A total of 157 hips in 153 patients at a mean follow-up of 11.6 months (SD7.8) were included. Mean patient age at the time of surgery was 67.4 years (SD 13.3) and mean body mass index (BMI) was 28.9 kg/m2 (SD 6.5). Outcomes included intraoperative complications, one-year all-cause re-revisions, and subsidence at postoperative time intervals (two weeks, six weeks, six months, nine months, and one year).Aims
Methods
Limited evidence is available on mid-term outcomes of robotic-arm assisted (RA) partial knee arthroplasty (PKA). Therefore, the purpose of this study was to evaluate mid-term survivorship, modes of failure, and patient-reported outcomes of RA PKA. A retrospective review of patients who underwent RA PKA between June 2007 and August 2016 was performed. Patients received a fixed-bearing medial or lateral unicompartmental knee arthroplasty (UKA), patellofemoral arthroplasty (PFA), or bicompartmental knee arthroplasty (BiKA; PFA plus medial UKA). All patients completed a questionnaire regarding revision surgery, reoperations, and level of satisfaction. Knee Injury and Osteoarthritis Outcome Scores (KOOS) were assessed using the KOOS for Joint Replacement Junior survey.Aims
Methods
Unicompartmental knee arthroplasty (UKA) is an alternative to total knee arthroplasty with isolated medial or lateral compartment osteoarthritis. However, polyethylene wear can significantly reduce the lifespan of UKA. Different bearing designs and materials for UKA have been developed to change the rate of polyethylene wear. Therefore, the objective of this study is to investigate the effect of insert conformity and material on the predicted wear in mobile-bearing UKA using a previously developed computational wear method. Two different designs were tested with the same femoral component under identical kinematic input: anatomy mimetic design (AMD) and conforming design inserts with different conformity levels. The insert materials were standard or crosslinked ultra-high-molecular-weight polyethylene (UHMWPE). We evaluated the contact pressure, contact area, wear rate, wear depth, and volumetric wear under gait cycle loading conditions.Objectives
Methods
Knee osteonecrosis in advanced stages may lead to joint degeneration. Total knee arthroplasty (TKA) for osteonecrosis has traditionally been associated with suboptimal results. We analyzed outcomes of contemporary TKAs for osteonecrosis, with particular emphasis on: survivorship free from aseptic loosening, any revision, and any reoperation plus the clinical outcomes, complications, and radiological results. In total, 156 patients undergoing 167 primary TKAs performed for osteonecrosis between 2004 and 2014 at a single institution were reviewed. The mean age at index TKA was 61 years (14 to 93) and the mean body mass index (BMI) was 30 kg/m2 (18 to 51) The mean follow-up was six years (2 to 12). A total of 110 TKAs (66%) were performed for primary osteonecrosis and 57 TKAs (34%) for secondary osteonecrosis. Overall, 15 TKAs (9%) had tibial stems, while 12 TKAs (7%) had femoral stems. Posterior-stabilized designs were used in 147 TKAs (88%) of TKAs. Bivariate Cox regression analysis was conducted to identify risk factors for revision and reoperation.Aims
Patients and Methods
Aseptic loosening is a major cause of failure in cemented endoprosthetic reconstructions. This paper presents the long-term outcomes of a custom-designed cross-pin fixation construct designed to minimize rotational stress and subsequent aseptic loosening in selected patients. The paper will also examine the long-term survivorship and modes of failure when using this technique. A review of 658 consecutive, prospectively collected cemented endoprosthetic reconstructions for oncological diagnoses at a single centre between 1980 and 2017 was performed. A total of 51 patients were identified with 56 endoprosthetic implants with cross-pin fixation, 21 of which were implanted following primary resection of tumour. Locations included distal femoral (n = 36), proximal femoral (n = 7), intercalary (n = 6), proximal humeral (n = 3), proximal tibial (n = 3), and distal humeral (n = 1).Aims
Patients and Methods
There is little literature about total knee arthroplasty (TKA) after distal femoral osteotomy (DFO). Consequently, the purpose of this study was to analyze the outcomes of TKA after DFO, with particular emphasis on: survivorship free from aseptic loosening, revision, or any re-operation; complications; radiological results; and clinical outcome. We retrospectively reviewed 29 patients (17 women, 12 men) from our total joint registry who had undergone 31 cemented TKAs after a DFO between 2000 and 2012. Their mean age at TKA was 51 years (22 to 76) and their mean body mass index 32 kg/m2 (20 to 45). The mean time between DFO and TKA was ten years (2 to 20). The mean follow-up from TKA was ten years (2 to 16). The prostheses were posterior-stabilized in 77%, varus-valgus constraint (VVC) in 13%, and cruciate-retaining in 10%. While no patient had metaphyseal fixation (e.g. cones or sleeves), 16% needed a femoral stem.Aims
Patients and Methods
Aims. Custom flange acetabular components (CFACs) are a patient-specific option for addressing large acetabular defects at revision total hip arthroplasty (THA), but patient and implant characteristics that affect survivorship remain unknown. This study aimed to identify patient and design factors related to survivorship. Patients and Methods. A retrospective review of 91 patients who underwent revision THA using 96 CFACs was undertaken, comparing features between radiologically failed and successful cases. Patient characteristics (demographic, clinical, and radiological) and implant features (design characteristics and intraoperative features) were collected. There were 74 women and 22 men; their mean age was 62 years (31 to 85). The mean follow-up was 24.9 months (. sd. 27.6; 0 to 116). Two sets of statistical analyses were performed: 1) univariate analyses (Pearson’s chi-squared and independent-samples Student’s t-tests) for each feature; and 2) bivariable logistic regressions using features identified from a random forest analysis. Results. Radiological failure and revision rates were 23% and 12.5%, respectively.
The purpose of this study was to determine the functional outcome and implant survivorship of mobile-bearing total ankle arthroplasty (TAA) performed by a single surgeon. We reviewed 205 consecutive patients (210 ankles) who had undergone mobile-bearing TAA (205 patients) for osteoarthritis of the ankle between January 2005 and December 2015. Their mean follow-up was 6.4 years (2.0 to 13.4). Functional outcome was assessed using the Ankle Osteoarthritis Scale, American Orthopaedic Foot and Ankle Society (AOFAS) ankle-hindfoot score, 36-Item Short-Form Health Survey (SF-36) score, visual analogue scale, and range of movement. Implant survivorship and complications were also evaluated.Aims
Patients and Methods
Surface replacement arthroplasty (SRA), compared with traditional total hip arthroplasty (THA), is more expensive and carries unique concern related to metal ions production and hypersensitivity. Additionally, SRA is a more demanding procedure with a decreased margin for error compared with THA. To justify its use, SRA must demonstrate comparable component survival and some clinical advantages. We therefore performed a systematic literature review to investigate the differences in complication rates, patient-reported outcomes, stress shielding, and hip biomechanics between SRA and THA. A systematic review of the literature was completed using MEDLINE and EMBASE search engines. Inclusion criteria were level I to level III articles that reported clinical outcomes following primary SRA compared with THA. An initial search yielded 2503 potential articles for inclusion. Exclusion criteria included review articles, level IV or level V evidence, less than one year’s follow-up, and previously reported data. In total, 27 articles with 4182 patients were available to analyze.Aims
Materials and Methods
Patients with a failed reverse shoulder arthroplasty (RSA) have limited salvage options. The aim of this study was to determine the outcome of revision RSA when used as a salvage procedure for a failed primary RSA. We reviewed all revision RSAs performed for a failed primary RSA between 2006 and 2012, excluding patients with a follow-up of less than two years. A total of 27 revision RSAs were included in the study. The mean age of the patients at the time of revision was 70 years (58 to 82). Of the 27 patients, 14 (52% were female). The mean follow-up was 4.4 years (2 to 10).Aims
Patients and Methods
This study reports on the medium- to long-term implant survivorship
and patient-reported outcomes for the Avon patellofemoral joint
(PFJ) arthroplasty. A total of 558 Avon PFJ arthroplasties in 431 patients, with
minimum two-year follow-up, were identified from a prospective database.
Patient-reported outcomes and implant survivorship were analyzed,
with follow-up of up to 18 years.Aims
Patients and Methods
This study aimed to evaluate implant survival of reverse hybrid
total hip arthroplasty (THA) at medium-term follow-up. A consecutive series of 1082 THAs in 982 patients with mean follow-up
of 7.9 years (5 to 11.3) is presented. Mean age was 69.2 years (21
to 94). Of these, 194 (17.9%) were in patients under 60 years, 663
(61.3%) in female patients and 348 (32.2%) performed by a trainee.
Head size was 28 mm in 953 hips (88.1%) or 32 mm in 129 hips (11.9%).
Survival analysis was performed and subgroups compared using log
rank tests.Aims
Patients and Methods
To report our experience with trunnion corrosion following metal-on-polyethylene
total hip arthroplasty, in particular to report the spectrum of
presentation and determine the mean time to presentation. We report the presenting symptoms and signs, intraoperative findings,
and early results and complications of operative treatment in nine
patients with a mean age of 74 years (60 to 86). The onset of symptoms
was at a mean of seven years (3 to 18) after index surgery.Aims
Patients and Methods
Instability remains a challenging problem in both primary and
revision total hip arthroplasty (THA). Dual mobility components
confer increased stability, but there are concerns about the unique
complications associated with these designs, as well as the long-term
survivorship. We performed a systematic review of all English language articles
dealing with dual mobility THAs published between 2007 and 2016
in the MEDLINE and Embase electronic databases. A total of 54 articles
met inclusion criteria for the final analysis of primary and revision
dual mobility THAs and dual mobility THAs used in the treatment
of fractures of the femoral neck. We analysed the survivorship and
rates of aseptic loosening and of intraprosthetic and extra-articular
dislocation.Aims
Materials and Methods
Little is known about the relative outcomes of revision of unicompartmental
knee arthroplasty (UKA) and high tibial osteotomy (HTO) to total
knee arthroplasty (TKA). The aim of this study is to compare the
outcomes of revision surgery for the two procedures in terms of
complications, re-revision and patient-reported outcome measures (PROMs)
at a minimum of two years follow-up. This study was a retrospective review of data from an institutional
arthroplasty registry for cases performed between 2001 and 2014.
A total of 292 patients were identified, of which 217 had a revision
of HTO to TKA, and 75 had revision of UKA to TKA. While mean follow-up
was longer for the HTO group compared with the UKA group, patient
demographics (age, body mass index and Charlson co-morbidity index)
and PROMs (Short Form-36, Oxford Knee Score, Knee Society Score,
both objective and functional) were similar in the two groups prior
to revision surgery. Outcomes included the rate of complications
and
re-operation, PROMS and patient-reported satisfaction at six months
and two years post-operatively. We also compared the duration of
surgery and the need for revision implants in the two groups. Aims
Patients and Methods
To evaluate the outcomes of cemented total hip arthroplasty (THA)
following a fracture of the acetabulum, with evaluation of risk
factors and comparison with a patient group with no history of fracture. Between 1992 and 2016, 49 patients (33 male) with mean age of
57 years (25 to 87) underwent cemented THA at a mean of 6.5 years
(0.1 to 25) following acetabular fracture. A total of 38 had undergone
surgical fixation and 11 had been treated non-operatively; 13 patients
died at a mean of 10.2 years after THA (0.6 to 19). Patients were
assessed pre-operatively, at one year and at final follow-up (mean
9.1 years, 0.5 to 23) using the Oxford Hip Score (OHS). Implant
survivorship was assessed. An age and gender-matched cohort of THAs
performed for non-traumatic osteoarthritis (OA) or avascular necrosis
(AVN) (n = 98) were used to compare complications and patient-reported outcome
measures (PROMs).Aims
Patients and Methods
We used the National Joint Registry for England, Wales, Northern Ireland and the Isle of Man (NJR) to investigate the risk of revision due to prosthetic joint infection (PJI) for patients undergoing primary and revision hip arthroplasty, the changes in risk over time, and the overall burden created by PJI. We analysed revision total hip arthroplasties (THAs) performed due to a diagnosis of PJI and the linked index procedures recorded in the NJR between 2003 and 2014. The cohort analysed consisted of 623 253 index primary hip arthroplasties, 63 222 index revision hip arthroplasties and 7585 revision THAs performed due to a diagnosis of PJI. The prevalence, cumulative incidence functions and the burden of PJI (total procedures) were calculated. Overall linear trends were investigated with log-linear regression.Objectives
Methods
Aims. Femoral stems with exchangeable (modular) necks were introduced
to offer surgeons an increased choice when determining the version,
offset and length of the femoral neck during total hip arthroplasty
(THA). It was hoped that this would improve outcomes and reduce
complications, particularly dislocation. In 2010, the Australian Orthopaedic
Association National Joint Replacement Registry (AOANJRR) first
reported an increased rate of revision after primary THA using femoral
stems with an exchangeable neck. The aim of this study was to provide
a more comprehensive up-to-date analysis of primary THA using femoral
stems with exchangeable and fixed necks. Materials and Methods. The data included all primary THA procedures performed for osteoarthritis
(OA), reported to the AOANJRR between 01 September 1999 and 31 December
2014. There were 9289 femoral stems with an exchangeable neck and
253 165 femoral stems with a fixed neck. The characteristics of
the patients and prostheses including the bearing surface and stem/neck
metal combinations were examined using Cox proportional hazard ratios
(HRs) and Kaplan-Meier estimates of survivorship. . Results. It was found that prostheses with an exchangeable neck had a
higher rate of revision and this was evident regardless of the bearing
surface or the size of the femoral head. Exchangeable neck prostheses
with a titanium stem and a cobalt-chromium neck had a significantly
higher rate of revision compared with titanium stem/titanium neck
combinations (HR 1.83, 95% confidence interval 1.49 to 2.23, p <
0.001).
The number of revision total knee arthroplasties (TKA) that are
performed is expected to increase. However, previous reports of
the causes of failure after TKA are limited in that they report
the causes at specific institutions, which are often dependent on
referral patterns. Our aim was to report the most common indications
for re-operations and revisions in a large series of posterior-stabilised
TKAs undertaken at a single institution, excluding referrals from
elsewhere, which may bias the causes of failure. A total of 5098 TKAs which were undertaken between 2000 and 2012
were included in the study. Re-operations, revisions with modular
component exchange, and revisions with non-modular component replacement
or removal were identified from the medical records. The mean follow-up
was five years (two to 12).Aims
Patients and Methods