The high revision rates of the DePuy Articular Surface Replacement (ASR) and the DePuy ASR XL (the total hip arthroplasty (THA) version) have led to questions over the viability of metal-on-metal (MoM) hip joints. Some designs of MoM hip joint do, however, have reasonable mid-term performance when implanted in appropriate patients. Investigations into the reasons for implant failure are important to offer help with the choice of implants and direction for future implant designs. One way to assess the performance of explanted hip prostheses is to measure the wear (in terms of material loss) on the joint surfaces. In this study, a coordinate measuring machine (CMM) was used to measure the wear on five failed cementless Biomet Magnum/ReCap/ Taperloc large head MoM THAs, along with one Biomet ReCap resurfacing joint. Surface roughness measurements were also taken. The reason for revision of these implants was pain and/or adverse reaction to metal debris (ARMD) and/or elevated blood metal ion levels.Objectives
Methods
Metal-on-metal hip resurfacing (MOMHR) is available as an alternative
option for younger, more active patients. There are failure modes
that are unique to MOMHR, which include loosening of the femoral
head and fractures of the femoral neck. Previous studies have speculated
that changes in the vascularity of the femoral head may contribute
to these failure modes. This study compares the survivorship between
the standard posterior approach (SPA) and modified posterior approach
(MPA) in MOMHR. A retrospective clinical outcomes study was performed examining
351 hips (279 male, 72 female) replaced with Birmingham Hip Resurfacing
(BHR, Smith and Nephew, Memphis, Tennessee) in 313 patients with
a pre-operative diagnosis of osteoarthritis. The mean follow-up
period for the SPA group was 2.8 years (0.1 to 6.1) and for the
MPA, 2.2 years (0.03 to 5.2); this difference in follow-up period
was statistically significant (p <
0.01). Survival analysis was
completed using the Kaplan–Meier method. Objectives
Methods