Objectives. The objective of this study was to compare the early migration
characteristics and functional outcome of the Triathlon cemented
knee prosthesis with its predecessor, the Duracon cemented knee
prosthesis (both Stryker). Methods. A total 60 patients were prospectively randomised and tibial
component migration was measured by radiostereometric analysis (RSA)
at three months, one year and two years; clinical outcome was measured
by the American Knee Society score and the Knee Osteoarthritis and
Injury Outcome Score. Results. There were no statistically significant differences in rotation
or translation around or along the three coordinal axes, or in the
maximum total point motion (MTPM) during the two-year follow-up. Conclusions. The Triathlon cemented knee prosthesis has similar early stability
and is likely to perform at least as well as the Duracon cemented
knee prosthesis over the
The mobile bearing Oxford unicompartmental knee arthroplasty (OUKA) is recommended to be performed with the leg in the hanging leg (HL) position, and the thigh placed in a stirrup. This comparative cadaveric study assesses implant positioning and intraoperative kinematics of OUKA implanted either in the HL position or in the supine leg (SL) position. A total of 16 fresh-frozen knees in eight human cadavers, without macroscopic anatomical defects, were selected. The knees from each cadaver were randomized to have the OUKA implanted in the HL or SL position.Aims
Methods
The aim of this study was to examine whether tourniquet use can improve perioperative blood loss, early function recovery, and pain after primary total knee arthroplasty (TKA) in the setting of multiple-dose intravenous tranexamic acid. This was a prospective, randomized clinical trial including 180 patients undergoing TKA with multiple doses of intravenous tranexamic acid. One group was treated with a tourniquet during the entire procedure, the second group received a tourniquet during cementing, and the third group did not receive a tourniquet. All patients received the same protocol of intravenous tranexamic acid (20 mg/kg) before skin incision, and three and six hours later (10 mg/kg). The primary outcome measure was perioperative blood loss. Secondary outcome measures were creatine kinase (CK), CRP, interleukin-6 (IL-6), visual analogue scale (VAS) pain score, limb swelling ratio, quadriceps strength, straight leg raising, range of motion (ROM), American Knee Society Score (KSS), and adverse events.Aims
Methods
The aim of this study was to determine the polyethylene wear rate of Phase 3 Oxford Unicompartmental Knee Replacement bearings and to investigate the effects of resin type and manufacturing process. A total of 63 patients with at least ten years’ follow-up with three bearing types (1900 resin machined, 1050 resin machined, and 1050 resin moulded) were recruited. Patients underwent full weight-bearing model-based radiostereometric analysis to determine the bearing thickness. The linear wear rate was estimated from the change in thickness divided by the duration of implantation.Objectives
Methods
The lack of effective treatment for cartilage defects has prompted investigations using tissue engineering techniques for their regeneration and repair. The success of tissue-engineered repair of cartilage may depend on the rapid and efficient adhesion of transplanted cells to a scaffold. Our aim in this study was to repair full-thickness defects in articular cartilage in the weight-bearing area of a porcine model, and to investigate whether the CD44 monoclonal antibody biotin-avidin (CBA) binding technique could provide satisfactory tissue-engineered cartilage. Cartilage defects were created in the load-bearing region of the lateral femoral condyle of mini-type pigs. The defects were repaired with traditional tissue-engineered cartilage, tissue-engineered cartilage constructed with the biotin-avidin (BA) technique, tissue-engineered cartilage constructed with the CBA technique and with autologous cartilage. The biomechanical properties, Western blot assay, histological findings and immunohistochemical staining were explored.Objectives
Methods
Subtotal or total meniscectomy in the medial or lateral compartment
of the knee results in a high risk of future osteoarthritis. Meniscal
allograft transplantation has been performed for over thirty years
with the scientifically plausible hypothesis that it functions in
a similar way to a native meniscus. It is thought that a meniscal
allograft transplant has a chondroprotective effect, reducing symptoms
and the long-term risk of osteoarthritis. However, this hypothesis has
never been tested in a high-quality study on human participants.
This study aims to address this shortfall by performing a pilot
randomised controlled trial within the context of a comprehensive
cohort study design. Patients will be randomised to receive either meniscal transplant
or a non-operative, personalised knee therapy program. MRIs will
be performed every four months for one year. The primary endpoint
is the mean change in cartilage volume in the weight-bearing area
of the knee at one year post intervention. Secondary outcome measures
include the mean change in cartilage thickness, T2 maps, patient-reported
outcome measures, health economics assessment and complications.Objectives
Methods
We are entering a new era with governmental bodies
taking an increasingly guiding role, gaining control of registries,
demanding direct access with release of open public information
for quality comparisons between hospitals. This review is written
by physicians and scientists who have worked with the Swedish Knee
Arthroplasty Register (SKAR) periodically since it began. It reviews
the history of the register and describes the methods used and lessons
learned. Cite this article:
Lower limb muscle power is thought to influence outcome following
total knee replacement (TKR). Post-operative deficits in muscle
strength are commonly reported, although not explained. We hypothesised
that post-operative recovery of lower limb muscle power would be
influenced by the number of satellite cells in the quadriceps muscle at
time of surgery. Biopsies were obtained from 29 patients undergoing TKR. Power
output was assessed pre-operatively and at six and 26 weeks post-operatively
with a Leg Extensor Power Rig and data were scaled for body weight.
Satellite cell content was assessed in two separate analyses, the
first cohort (n = 18) using immunohistochemistry and the second
(n = 11) by a new quantitative polymerase chain reaction (q-PCR)
protocol for Pax-7 (generic satellite cell marker) and Neural Cell
Adhesion Molecule (NCAM; marker of activated cells).Objectives
Methods