Advertisement for orthosearch.org.uk
Results 1 - 6 of 6
Results per page:
Bone & Joint Open
Vol. 2, Issue 8 | Pages 599 - 610
1 Aug 2021
Hothi H Bergiers S Henckel J Iliadis AD Goodier WD Wright J Skinner J Calder P Hart AJ

Aims

The aim of this study was to present the first retrieval analysis findings of PRECICE STRYDE intermedullary nails removed from patients, providing useful information in the post-market surveillance of these recently introduced devices.

Methods

We collected ten nails removed from six patients, together with patient clinical data and plain radiograph imaging. We performed macro- and microscopic analysis of all surfaces and graded the presence of corrosion using validated semiquantitative scoring methods. We determined the elemental composition of surface debris using energy dispersive x-ray spectroscopy (EDS) and used metrology analysis to characterize the surface adjacent to the extendable junctions.


Bone & Joint Open
Vol. 2, Issue 10 | Pages 825 - 833
8 Oct 2021
Dailey HL Schwarzenberg P Webb, III EB Boran SAM Guerin S Harty JA

Aims. The study objective was to prospectively assess clinical outcomes for a pilot cohort of tibial shaft fractures treated with a new tibial nailing system that produces controlled axial interfragmentary micromotion. The hypothesis was that axial micromotion enhances fracture healing compared to static interlocking. Methods. Patients were treated in a single level I trauma centre over a 2.5-year period. Group allocation was not randomized; both the micromotion nail and standard-of-care static locking nails (control group) were commercially available and selected at the discretion of the treating surgeons. Injury risk levels were quantified using the Nonunion Risk Determination (NURD) score. Radiological healing was assessed until 24 weeks or clinical union. Low-dose CT scans were acquired at 12 weeks and virtual mechanical testing was performed to objectively assess structural bone healing. Results. A total of 37 micromotion patients and 46 control patients were evaluated. There were no significant differences between groups in terms of age, sex, the proportion of open fractures, or NURD score. There were no nonunions (0%) in the micromotion group versus five (11%) in the control group. The proportion of fractures united was significantly higher in the micromotion group compared to control at 12 weeks (54% vs 30% united; p = 0.043), 18 weeks (81% vs 59%; p = 0.034), and 24 weeks (97% vs 74%; p = 0.005). Structural bone healing scores as assessed by CT scans tended to be higher with micromotion compared to control and this difference reached significance in patients who had biological comorbidities such as smoking. Conclusion. In this pilot study, micromotion fixation was associated with improved healing compared to standard tibial nailing. Further prospective clinical studies will be needed to assess the strength and generalizability of any potential benefits of micromotion fixation. Cite this article: Bone Jt Open 2021;2(10):825–833


Bone & Joint Open
Vol. 2, Issue 1 | Pages 22 - 32
4 Jan 2021
Sprague S Heels-Ansdell D Bzovsky S Zdero R Bhandari M Swiontkowski M Tornetta P Sanders D Schemitsch E

Aims. Using tibial shaft fracture participants from a large, multicentre randomized controlled trial, we investigated if patient and surgical factors were associated with health-related quality of life (HRQoL) at one year post-surgery. Methods. The Study to Prospectively Evaluate Reamed Intramedullary Nails in Patients with Tibial Fractures (SPRINT) trial examined adults with an open or closed tibial shaft fracture who were treated with either reamed or unreamed intramedullary nails. HRQoL was assessed at hospital discharge (for pre-injury level) and at 12 months post-fracture using the Short Musculoskeletal Functional Assessment (SMFA) Dysfunction, SMFA Bother, 36-Item Short Form 36 (SF-36) Physical, and SF-36 Mental Component scores. We used multiple linear regression analysis to determine if baseline and surgical factors, as well as post-intervention procedures within one year of fracture, were associated with these HRQoL outcomes. Significance was set at p < 0.01. We hypothesize that, irrespective of the four measures used, prognosis is guided by both modifiable and non-modifiable factors and that patients do not return to their pre-injury level of function, nor HRQoL. Results. For patient and surgical factors, only pre-injury quality of life and isolated fracture showed a statistical effect on all four HRQoL outcomes, while high-energy injury mechanism, smoking, and race or ethnicity, demonstrated statistical significance for three of the four HRQoL outcomes. Patients who did not require reoperation in response to infection, the need for bone grafts, and/or the need for implant exchanges had statistically superior HRQoL outcomes than those who did require intervention within one year after initial tibial fracture nailing. Conclusion. We identified several baseline patient factors, surgical factors, and post-intervention procedures within one year after intramedullary nailing of a tibial shaft fracture that may influence a patient’s HRQoL. Cite this article: Bone Jt Open 2021;2(1):22–32


Bone & Joint Open
Vol. 3, Issue 8 | Pages 648 - 655
1 Aug 2022
Yeung CM Bhashyam AR Groot OQ Merchan N Newman ET Raskin KA Lozano-Calderón SA

Aims

Due to their radiolucency and favourable mechanical properties, carbon fibre nails may be a preferable alternative to titanium nails for oncology patients. We aim to compare the surgical characteristics and short-term results of patients who underwent intramedullary fixation with either a titanium or carbon fibre nail for pathological long-bone fracture.

Methods

This single tertiary-institutional, retrospectively matched case-control study included 72 patients who underwent prophylactic or therapeutic fixation for pathological fracture of the humerus, femur, or tibia with either a titanium (control group, n = 36) or carbon fibre (case group, n = 36) intramedullary nail between 2016 to 2020. Patients were excluded if intramedullary fixation was combined with any other surgical procedure/fixation method. Outcomes included operating time, blood loss, fluoroscopic time, and complications. Fisher’s exact test and Mann-Whitney U test were used for categorical and continuous outcomes, respectively.


Bone & Joint Open
Vol. 3, Issue 5 | Pages 359 - 366
1 May 2022
Sadekar V Watts AT Moulder E Souroullas P Hadland Y Barron E Muir R Sharma HK

Aims

The timing of when to remove a circular frame is crucial; early removal results in refracture or deformity, while late removal increases the patient morbidity and delay in return to work. This study was designed to assess the effectiveness of a staged reloading protocol. We report the incidence of mechanical failure following both single-stage and two stage reloading protocols and analyze the associated risk factors.

Methods

We identified consecutive patients from our departmental database. Both trauma and elective cases were included, of all ages, frame types, and pathologies who underwent circular frame treatment. Our protocol is either a single-stage or two-stage process implemented by defunctioning the frame, in order to progressively increase the weightbearing load through the bone, and promote full loading prior to frame removal. Before progression, through the process we monitor patients for any increase in pain and assess radiographs for deformity or refracture.


Bone & Joint Open
Vol. 1, Issue 6 | Pages 229 - 235
9 Jun 2020
Lazizi M Marusza CJ Sexton SA Middleton RG

Aims

Elective surgery has been severely curtailed as a result of the COVID-19 pandemic. There is little evidence to guide surgeons in assessing what processes should be put in place to restart elective surgery safely in a time of endemic COVID-19 in the community.

Methods

We used data from a stand-alone hospital admitting and operating on 91 trauma patients. All patients were screened on admission and 100% of patients have been followed-up after discharge to assess outcome.