Vol. 2, Issue 11 | Pages 921 - 925
9 Nov 2021
Limberg AK Wyles CC Taunton MJ Hanssen AD Pagnano MW Abdel MP

Aims

Varus-valgus constrained (VVC) devices are typically used in revision settings, often with stems to mitigate the risk of aseptic loosening. However, in at least one system, the VVC insert is compatible with the primary posterior-stabilized (PS) femoral component, which may be an option in complex primary situations. We sought to determine the implant survivorship, radiological and clinical outcomes, and complications when this VVC insert was coupled with a PS femur without stems in complex primary total knee arthroplasties (TKAs).

Methods

Through our institution’s total joint registry, we identified 113 primary TKAs (103 patients) performed between 2007 and 2017 in which a VVC insert was coupled with a standard cemented PS femur without stems. Mean age was 68 years (SD 10), mean BMI was 32 kg/m2 (SD 7), and 59 patients (50%) were male. Mean follow-up was four years (2 to 10).


Bone & Joint Open
Vol. 3, Issue 9 | Pages 710 - 715
5 Sep 2022
Khan SK Tyas B Shenfine A Jameson SS Inman DS Muller SD Reed MR

Aims

Despite multiple trials and case series on hip hemiarthroplasty designs, guidance is still lacking on which implant to use. One particularly deficient area is long-term outcomes. We present over 1,000 consecutive cemented Thompson’s hemiarthroplasties over a ten-year period, recording all accessible patient and implant outcomes.

Methods

Patient identifiers for a consecutive cohort treated between 1 January 2003 and 31 December 2011 were linked to radiographs, surgical notes, clinic letters, and mortality data from a national dataset. This allowed charting of their postoperative course, complications, readmissions, returns to theatre, revisions, and deaths. We also identified all postoperative attendances at the Emergency and Outpatient Departments, and recorded any subsequent skeletal injuries.


Bone & Joint Open
Vol. 3, Issue 2 | Pages 107 - 113
1 Feb 2022
Brunt ACC Gillespie M Holland G Brenkel I Walmsley P

Aims

Periprosthetic joint infection (PJI) occurs in approximately 1% to 2% of total knee arthroplasties (TKA) presenting multiple challenges, such as difficulty in diagnosis, technical complexity, and financial costs. Two-stage exchange is the gold standard for treating PJI but emerging evidence suggests 'two-in-one' single-stage revision as an alternative, delivering comparable outcomes, reduced morbidity, and cost-effectiveness. This study investigates five-year results of modified single-stage revision for treatment of PJI following TKA with bone loss.

Methods

Patients were identified from prospective data on all TKA patients with PJI following the primary procedure. Inclusion criteria were: revision for PJI with bone loss requiring reconstruction, and a minimum five years’ follow-up. Patients were followed up for recurrent infection and assessment of function. Tools used to assess function were Oxford Knee Score (OKS) and American Knee Society Score (AKSS).


Bone & Joint Open
Vol. 2, Issue 11 | Pages 1004 - 1016
26 Nov 2021
Wight CM Whyne CM Bogoch ER Zdero R Chapman RM van Citters DW Walsh WR Schemitsch E

Aims

This study investigates head-neck taper corrosion with varying head size in a novel hip simulator instrumented to measure corrosion related electrical activity under torsional loads.

Methods

In all, six 28 mm and six 36 mm titanium stem-cobalt chrome head pairs with polyethylene sockets were tested in a novel instrumented hip simulator. Samples were tested using simulated gait data with incremental increasing loads to determine corrosion onset load and electrochemical activity. Half of each head size group were then cycled with simulated gait and the other half with gait compression only. Damage was measured by area and maximum linear wear depth.