Advertisement for orthosearch.org.uk
Results 1 - 2 of 2
Results per page:
Applied filters
The Bone & Joint Journal
Dates
Year From

Year To
The Journal of Bone & Joint Surgery British Volume
Vol. 76-B, Issue 5 | Pages 713 - 716
1 Sep 1994
Bischoff U Freeman M Smith D Tuke M Gregson P

We studied the wear generated by motion between polished and shot-blasted titanium-alloy (Ti-6Al-4V) or cobalt-chrome alloy (Co-Cr) surfaces and cortical bone in vitro. Semicircular sections of human proximal femoral cortex were reamed to fit metal cylinders of each alloy. The cylinders were then fitted in the bone, loaded and rotated in physiological saline. Ti-alloy resulted in more wear both of the bone and of the metal than did Co-Cr alloy. Metal wear was reduced and bone wear was increased by shot-blasting, a procedure which introduces surface residual stresses and roughens the metal surface. We conclude that when there is gross motion between a metal implant and bone, Ti-alloy is likely to generate more wear debris than Co-Cr alloy. The least wear both of bone and of metal was produced by polished Co-Cr.


The Journal of Bone & Joint Surgery British Volume
Vol. 68-B, Issue 4 | Pages 566 - 569
1 Aug 1986
Roper B Tuke M O'Riordan S Bulstrode C

Sixty unconstrained elbow replacements of a new design have been followed prospectively for three to nine years. Review showed that 50% had excellent relief of pain and return of function, 27% had had major complications requiring removal or revision of the prosthesis and 23% had minor complications which marred the result. Further research in this field seems worthwhile.