Recent guidance recommends the use of a well-proven
cemented femoral stem for hemiarthroplasty in the management of
fractures of the femoral neck, and the Exeter Trauma Stem (ETS)
has been suggested as an example of such an implant. The design
of this stem was based on the well-proven Exeter Total Hip Replacement
stem (ETHRS). This study assessed the surface finish of the ETS
in comparison with the ETHRS. Two ETSs and two ETHRSs were examined
using a profilometer with a precision of 1 nm and compared with
an explanted Exeter Matt stem. The mean roughness average (RA) of
the ETSs was approximately ten times higher than that of the ETHRSs (0.235 μm
(0.095 to 0.452) Cite this article:
We investigated the capacity of patient warming
devices to disrupt the ultra-clean airflow system. We compared the effects
of two patient warming technologies, forced-air and conductive fabric,
on operating theatre ventilation during simulated hip replacement
and lumbar spinal procedures using a mannequin as a patient. Infection
data were reviewed to determine whether joint infection rates were
associated with the type of patient warming device that was used. Neutral-buoyancy detergent bubbles were released adjacent to
the mannequin’s head and at floor level to assess the movement of
non-sterile air into the clean airflow over the surgical site. During
simulated hip replacement, bubble counts over the surgical site
were greater for forced-air than for conductive fabric warming when
the anaesthesia/surgery drape was laid down (p = 0.010) and at half-height
(p <
0.001). For lumbar surgery, forced-air warming generated
convection currents that mobilised floor air into the surgical site
area. Conductive fabric warming had no such effect. A significant increase in deep joint infection, as demonstrated
by an elevated infection odds ratio (3.8, p = 0.024), was identified
during a period when forced-air warming was used compared to a period
when conductive fabric warming was used. Air-free warming is, therefore,
recommended over forced-air warming for orthopaedic procedures.
Increased femoral head size may reduce dislocation rates following total hip replacement. The National Joint Registry for England and Wales has highlighted a statistically significant increase in the use of femoral heads ≥ 36 mm in diameter from 5% in 2005 to 26% in 2009, together with an increase in the use of the posterior approach. The aim of this study was to determine whether rates of dislocation have fallen over the same period. National data for England for 247 546 procedures were analysed in order to determine trends in the rate of dislocation at three, six, 12 and 18 months after operation during this time. The 18-month revision rates were also examined. Between 2005 and 2009 there were significant decreases in cumulative dislocations at three months (1.12% to 0.86%), six months (1.25% to 0.96%) and 12 months (1.42% to 1.11%) (all p <
0.001), and at 18 months (1.56% to 1.31%) for the period 2005 to 2008 (p <
0.001). The 18-month revision rates did not significantly change during the study period (1.26% to 1.39%, odds ratio 1.10 (95% confidence interval 0.98 to 1.24), p = 0.118). There was no evidence of changes in the coding of dislocations during this time. These data have revealed a significant reduction in dislocations associated with the use of large femoral head sizes, with no change in the 18-month revision rate.
Rivaroxaban has been recommended for routine use as a thromboprophylactic agent in patients undergoing lower-limb arthroplasty. However, trials supporting its use have not fully evaluated the risks of wound complications. This study of 1048 total hip/knee replacements records the rates of return to theatre and infection before and after the change from a low molecular weight heparin (tinzaparin) to rivaroxaban as the agent of chemical thromboprophylaxis in patients undergoing lower-limb arthroplasty. During a period of 13 months, 489 consecutive patients undergoing lower-limb arthroplasty received tinzaparin and the next 559 consecutive patients received rivaroxaban as thromboprophylaxis. Nine patients in the control (tinzaparin) group (1.8%, 95% confidence interval 0.9 to 3.5) returned to theatre with wound complications within 30 days, compared with 22 patients in the rivaroxaban group (3.94%, 95% confidence interval 2.6 to 5.9). This increase was statistically significant (p = 0.046). The proportion of patients who returned to theatre and became infected remained similar (p = 0.10). Our study demonstrates the need for further randomised controlled clinical trials to be conducted to assess the safety and efficacy of rivaroxaban in clinical practice, focusing on the surgical complications as well as the potential prevention of venous thromboembolism.
There are many methods for analysing wear volume in failed polyethylene acetabular components. We compared a radiological technique with three recognised We tested 18 ultra-high-molecular-weight polyethylene acetabular components revised for wear and aseptic loosening, of which 13 had pre-revision radiographs, from which the wear volume was calculated based upon the linear wear. We used a shadowgraph technique on silicone casts of all of the retrievals and a coordinate measuring method on the components directly. For these techniques, the wear vector was calculated for each component and the wear volume extrapolated using mathematical equations. The volumetric wear was also measured directly using a fluid-displacement method. The results of each technique were compared. The series had high wear volumes (mean 1385 mm3; 730 to 1850) and high wear rates (mean 205 mm3/year; 92 to 363). There were wide variations in the measurements of wear volume between the radiological and the other techniques. Radiograph-derived wear volume correlated poorly with that of the fluid-displacement method, co-ordinate measuring method and shadowgraph methods, becoming less accurate as the wear increased. The mean overestimation in radiological wear volume was 47.7% of the fluid-displacement method wear volume. Fluid-displacement method, coordinate measuring method and shadowgraph determinations of wear volume were all better than that of the radiograph-derived linear measurements since they took into account the direction of wear. However, only radiological techniques can be used Interpretation of radiological measurements of acetabular wear must be done judiciously in the clinical setting.
Fatigue fractures which originate at stress-concentrating voids located at the implant-cement interface are a potential cause of septic loosening of cemented femoral components. Heating of the component to 44°C is known to reduce the porosity of the cement-prosthesis interface. The temperature of the cement-bone interface was recorded intra-operatively as 32.3°C. A simulated femoral model was devised to study the effect of heating of the component on the implant-cement interface. Heating of the implant and vacuum mixing have a synergistic effect on the porosity of the implant-cement interface, and heating also reverses the gradients of microhardness in the mantle. Heating of the implant also reduces porosity at the interface depending on the temperature. A minimum difference in temperature between the implant and the bone of 3°C was required to produce this effect. The optimal difference was 7°C, representing a balance between maximal reduction of porosity and an increased risk of thermal injury. Using contemporary cementing techniques, heating the implant to 40°C is recommended to produce an optimum effect.
The ABG I cementless hip prosthesis has demonstrated unacceptably high rates of wear and osteolysis in our patients. We performed a retrospective study of 97 hips implanted between 1992 and 1998. Radiographic analysis revealed high rates of wear of the polyethylene liner with marked periacetabular osteolysis. Clinical examination indicated that many of these patients were initially asymptomatic. Wear-related problems have required ten hips to be revised and a furher 13 are awaiting revision. This gives a failure rate of 24% at a mean follow-up of 69 months. Contributing factors are likely to include poor wear characteristics of the polyethylene liners which were gamma irradiated in air, and increased wear debris caused by a poor fit of the polyethylene liner within the shell. We believe that all ABG I implants should be immediately reviewed and remain under careful, long-term follow-up.
We undertook a prospective, randomised study of 135 total knee arthroplasties to determine the most accurate and reliable technique for alignment of the tibial prosthesis. Tibial resection was guided by either intramedullary or extramedullary alignment jigs. Of the 135 knees, standardised postoperative radiographs suitable for assessment were available in 100. Correct tibial alignment was found in 85% of the intramedullary group compared with 65% of the extramedullary group (p = 0.019). We conclude that intramedullary guides are superior to extramedullary instruments for alignment of the tibial prosthesis.