Between 1982 and 1997, 403 consecutive patients (522 elbows) with rheumatoid arthritis underwent Souter-Strathclyde total elbow replacement. By the end of 2007, there had been 66 revisions for aseptic loosening in 60 patients. The mean time of follow-up was 10.6 years (0 to 25) The survival rates at five-, ten, 15 and 19 years were 96% (95%, confidence interval (CI) 95 to 98), 89% (95% CI 86 to 92), 83% (95% CI 78 to 87), and 77% (95% CI 69 to 85), respectively. The small and medium-sized short-stemmed primary humeral components had a 5.6-fold and 3.6-fold risk of revision for aseptic loosening respectively, compared to the medium-sized long-stemmed component. The small and medium-sized all-polyethylene ulnar components had respectively a 28.2-fold and 8.4-fold risk of revision for aseptic loosening, compared to the metal-backed ulnar components. The use of retentive ulnar components was not associated with an increased risk of aseptic loosening compared to non-retentive implants.
We present the results of 525 primary Souter elbow arthroplasties undertaken in 406 patients between 1982 and 1997. There were 372 women and 34 men with a mean age of 57 years; 119 patients had a bilateral procedure. The elbows were affected by chronic inflammatory disease, usually rheumatoid arthritis, which had been present for a mean of 24.7 years (2 to 70). In about 30% the joints were grossly destroyed with significant loss of bone. In 179 elbows the ulnar components were metal-backed and retentive; in the remaining 346, with better bone stock, non-retentive, all-polyethylene prostheses were used. Because of complications, 108 further operations were required in 82 patients. During the early years the incidence of complications was higher. Dislocation was the indication for 30 further procedures in 26 patients. Thirty patients underwent 33 revision procedures for aseptic loosening, 12 had 29 operations because of deep infection, two for superficial infection, and 14 further operations were done for other reasons. The cumulative rate of success, without aseptic loosening, five and ten years after surgery, was 96% and 85%, respectively.