Advertisement for orthosearch.org.uk
Results 1 - 7 of 7
Results per page:
The Bone & Joint Journal
Vol. 95-B, Issue 8 | Pages 1035 - 1039
1 Aug 2013
Ebreo D Bell PJ Arshad H Donell ST Toms A Nolan JF

Metal artefact reduction (MAR) MRI is now widely considered to be the standard for imaging metal-on-metal (MoM) hip implants. The Medicines and Healthcare Products Regulatory Agency (MHRA) has recommended cross-sectional imaging for all patients with symptomatic MoM bearings. This paper describes the natural history of MoM disease in a 28 mm MoM total hip replacement (THR) using MAR MRI. Inclusion criteria were patients with MoM THRs who had not been revised and had at least two serial MAR MRI scans. All examinations were reported by an experienced observer and classified as A (normal), B (infection) or C1–C3 (mild, moderate, severe MoM-related abnormalities). Between 2002 and 2011 a total of 239 MRIs were performed on 80 patients (two to four scans per THR); 63 initial MRIs (61%) were normal. On subsequent MRIs, six initially normal scans (9.5%) showed progression to a disease state; 15 (15%) of 103 THRs with sequential scans demonstrated worsening disease on subsequent imaging.

Most patients with a MoM THR who do not undergo early revision have normal MRI scans. Late progression (from normal to abnormal, or from mild to more severe MoM disease) is not common and takes place over several years.

Cite this article: Bone Joint J 2013;95-B:1035–9.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 1 | Pages 10 - 15
1 Jan 2012
Ollivere B Wimhurst JA M. Clark I Donell ST

The most frequent cause of failure after total hip replacement in all reported arthroplasty registries is peri-prosthetic osteolysis. Osteolysis is an active biological process initiated in response to wear debris. The eventual response to this process is the activation of macrophages and loss of bone.

Activation of macrophages initiates a complex biological cascade resulting in the final common pathway of an increase in osteolytic activity. The biological initiators, mechanisms for and regulation of this process are beginning to be understood. This article explores current concepts in the causes of, and underlying biological mechanism resulting in peri-prosthetic osteolysis, reviewing the current basic science and clinical literature surrounding the topic.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 11 | Pages 1501 - 1508
1 Nov 2010
Donell ST Darrah C Nolan JF Wimhurst J Toms A Barker THW Case CP Tucker JK

Metal-on-metal total hip replacement has been targeted at younger patients with anticipated long-term survival, but the effect of the production of metal ions is a concern because of their possible toxicity to cells. We have reviewed the results of the use of the Ultima hybrid metal-on-metal total hip replacement, with a cemented polished tapered femoral component with a 28 mm diameter and a cobalt-chrome (CoCr) modular head, articulating with a 28 mm CoCr acetabular bearing surface secured in a titanium alloy uncemented shell.

Between 1997 and 2004, 545 patients with 652 affected hips underwent replacement using this system. Up to 31 January 2008, 90 (13.8%) hips in 82 patients had been revised. Pain was the sole reason for revision in 44 hips (48.9%) of which 35 had normal plain radiographs. Peri-prosthetic fractures occurred in 17 hips (18.9%) with early dislocation in three (3.3%) and late dislocation in 16 (17.8%). Infection was found in nine hips (10.0%).

At operation, a range of changes was noted including cavities containing cloudy fluid under pressure, necrotic soft tissues with avulsed tendons and denuded osteonecrotic upper femora. Corrosion was frequently observed on the retrieved cemented part of the femoral component. Typically, the peri-operative findings confirmed those found on pre-operative metal artefact reduction sequence MRI and histological examination showed severe necrosis.

Metal artefact reduction sequence MRI proved to be useful when investigating these patients with pain in the absence of adverse plain radiological features.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 10 | Pages 1305 - 1309
1 Oct 2009
Chan WCW Musonda P Cooper AS Glasgow MMS Donell ST Walton NP

We retrospectively studied the major complications occurring after one- and two-stage bilateral unicompartmental knee replacements (UKR). Between 1999 and 2008, 911 patients underwent 1150 UKRs through a minimally invasive approach in our unit. Of these, 159 patients (318 UKRs) had one-stage and 80 patients (160 UKRs) had two-stage bilateral UKRs.

The bilateral UKR groups were comparable in age and American Society of Anaesthesiology grade, but more women were in the two-stage group (p = 0.019). Mechanical thromboprophylaxis was used in all cases. Major complications were recorded as death, pulmonary embolus, proximal deep-vein thrombosis and adverse cardiac events within 30 days of surgery. No statistical differences between the groups were found regarding the operating surgeon, the tourniquet time or minor complications except for distal deep-vein thrombosis. The anaesthetic times were longer for the two-stage group (p = 0.0001).

Major complications were seen in 13 patients (8.2%) with one-stage operations but none were encountered in the two-stage group (p = 0.005). Distal deep-vein thrombosis was more frequent in the two-stage group (p = 0.036).

Because of the significantly higher risk of major complications associated with one-stage bilateral UKR we advocate caution before undertaking such a procedure.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 1 | Pages 69 - 77
1 Jan 2006
Costa ML MacMillan K Halliday D Chester R Shepstone L Robinson AHN Donell ST

We performed two independent, randomised, controlled trials in order to assess the potential benefits of immediate weight-bearing mobilisation after rupture of the tendo Achillis. The first trial, on operatively-treated patients showed an improved functional outcome for patients mobilised fully weight-bearing after surgical repair. Two cases of re-rupture in the treatment group suggested that careful patient selection is required as patients need to follow a structured rehabilitation regimen. The second trial, on conservatively-treated patients, provided no evidence of a functional benefit from immediate weight-bearing mobilisation. However, the practical advantages of immediate weight-bearing did not predispose the patients to a higher complication rate. In particular, there was no evidence of tendon lengthening or a higher re-rupture rate. We would advocate immediate weight-bearing mobilisation for the rehabilitation of all patients with rupture of the tendo Achillis.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 8 | Pages 1085 - 1088
1 Aug 2005
Costa ML Kay D Donell ST

One of the factors that influence the outcome after rupture of the tendo Achillis is abnormality of gait. We prospectively assessed 14 patients and 15 normal control subjects using an in-shoe plantar pressure measurement system. There was a significant reduction in peak mean forefoot pressure in the early period of rehabilitation (p < 0.001). There was a concomitant rise in heel pressure on the injured side (p = 0.05). However, there was no difference in cadence, as determined by the duration of the terminal stance and pre-swing phases as a proportion of total stance. The forefoot pressure deficit in the group with tendon ruptures was smaller when assessed six months after the injury but was still significant (p = 0.029). Pedobarographic assessment confirms that there are marked abnormalities within the gait cycle. Rehabilitation programmes which address these abnormalities may improve outcome.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 4 | Pages 458 - 462
1 Apr 2005
Walton NP Darrah C Shepstone L Donell ST Phillips H

We prospectively studied 217 patients who underwent 234 Elite Plus total hip arthroplasties. At a mean of 6.4 (SD 0.7) years post-operatively, 39 patients had died and 22 were either lost to follow-up or had no radiographs available. Clinical (Oxford hip score) and radiological assessments were performed on 156 patients (168 hip arthroplasties) who had a mean age of 67.7 (SD 9.7) years at operation.

In the assessed group, 26 of 159 (16.4%) of femoral stems which had not already been revised and 19 of 159 (11.9%) of acetabular cups were definitely loose. In total, 52 of 168 (31%) of hips had either been revised or had definite evidence of loosening of a component. We could not establish any relationship between clinical and radiological outcomes.

Despite the fact that the clinical outcome and rate of revision for the Elite Plus appeared to meet international standards, our findings give us cause for concern. We believe that joint registries should include radiological surveillance in order to provide reliable information about medium-term outcomes for hip prostheses.