header advert
Results 1 - 2 of 2
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 85-B, Issue 5 | Pages 650 - 654
1 Jul 2003
Clarke MT Lee PTH Villar RN

Metal-on-metal (MOM) is a commonly used bearing notable for its ‘suction fit’ when lubricated. In this study, we examined the capacity for MOM bearings to protect against dislocation after total hip replacement (THR).

We undertook a clinical investigation to compare the rate of dislocation of MOM bearings with those of ceramic-on-polyethylene (COP) bearings and found that one MOM bearing dislocated in a series of 109 hips (0.9%) compared with nine of 145 hips (6.2%) in the COP group (p = 0.02).

We also performed an in vitro investigation comparing the peak forces generated during forced separation of the two bearings of the same dimensions at velocities from 1 to 50 cm/s. This revealed that the MOM bearing generated significant resistance to separation at all velocities (maximum mean 24 N), whereas the COP did not (maximum mean 1.9 N, p < 0.001). We conclude that MOM bearings are more stable to dislocation than COP bearings as a result of the interfacial forces provided by a thin, lubricating fluid.


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 4 | Pages 611 - 613
1 Jul 1998
Clarke MT Green JS Harper WM Gregg PJ

A total of 110 total knee replacements (TKRs) was randomised to receive either a cemented or an uncemented prosthesis. Postoperative venography at five to seven days was used to compare the prevalence, site and size of deep-vein thrombosis (DVT). We also compared the findings with those of postoperative venography in a group of patients with cemented total hip replacements (THRs).

The total prevalence of DVT was significantly greater after uncemented (81%) than after cemented TKR (55%). Both knee groups had a significantly higher prevalence of DVT than in cemented hip replacements (32%). We found no difference in the proportion with proximal DVT in the three groups (14%, 15% and 16%). The median length of the thrombi was significantly greater after cemented (26.5 cm) than after uncemented TKR (11 cm) or after cemented THR (7 cm). This difference was mainly due to greater lengths of distal rather than proximal thrombi.

We conclude that the use of cement may affect the formation of DVT after joint replacement, but does not appear to lead to an increased incidence.