We report an independent prospective review of the first 230 Birmingham hip resurfacings in 212 patients at a mean follow-up of five years (4 to 6). Two patients, one with a loose acetabular component and the other with suspected avascular necrosis of the femoral head, underwent revision. There were two deaths from unrelated causes and one patient was lost to follow-up. The survivorship with the worst-case scenario was 97.8% (95% confidence interval 95.8 to 99.5). The mean Harris hip score improved significantly (paired On radiological review at five years, one patient had a progressive lucent line around the acetabular component and six had progressive lucent lines around the femoral component. A total of 18 femoral components (8%) had migrated into varus and those with lucent lines present migrated a mean of 3.8° (1.02° to 6.54°) more than the rest. Superolateral notching of the femoral neck and reactive sclerosis at the tip of the peg of the femoral component were associated with the presence of lucent lines (chi-squared test, p <
0.05), but not with migration of the femoral component, and are of unknown significance. Our results with the Birmingham hip resurfacing continue to be satisfactory at a mean follow-up of five years.
Narrowing of the femoral neck after resurfacing arthroplasty of the hip has been described previously in both cemented and uncemented hip resurfacing. The natural history of narrowing of the femoral neck is unknown. We retrospectively measured the diameter of the femoral neck in a series of 163 Birmingham hip resurfacings in 163 patients up to a maximum of six years after operation to determine the extent and progression of narrowing. There were 105 men and 58 women with a mean age of 52 years (18 to 82). At a mean follow-up of five years, the mean Harris hip score was 94.8 (47 to 100) and the mean flexion of the hip 112.5° (80° to 160°). There was some narrowing of the femoral neck in 77% (125) of the patients reviewed, and in 27.6% (45) the narrowing exceeded 10% of the diameter of the neck. A multiple logistic regression analysis showed a significant association (chi-squared test (derived from logistic regression) p = 0.01) of narrowing with female gender and a valgus femoral neck/shaft angle. There was no significant association between the range of movement, position or size of the component or radiological lucent lines and narrowing of the neck (chi-squared test; p = 0.10 (flexion), p = 0.08 (size of femoral component), p = 0.09 (size of acetabular component), p = 0.71 (femoral component angulation), p = 0.99 (lucent lines)). There was no significant difference between the diameter of the neck at a mean of three years (2.5 to 3.5) and that at five years (4.5 to 5.5), indicating that any change in the diameter of the neck had stabilised by three years (sign rank test, p = 0.60). We conclude that narrowing of the femoral neck which is found with the Birmingham hip resurfacing arthroplasty is in most cases associated with no adverse clinical or radiological outcome up to a maximum of six years after the initial operation.