Endoprosthetic reconstruction following distal femur tumour resection has been widely advocated. In this paper, we present the design of an uncemented endoprosthesis system featuring a short, curved stem, with the goal of enhancing long-term survivorship and functional outcomes. This study involved patients who underwent implantation of an uncemented distal femoral endoprosthesis with a short and curved stem between 2014 and 2019. Functional outcomes were assessed using the 1993 version of the Musculoskeletal Tumour Society (MSTS-93) score. Additionally, we quantified five types of complications and assessed osseointegration radiologically. The survivorship of the endoprosthesis was evaluated according to two endpoints. A total of 134 patients with a median age of 26 years (IQR 16 to 41) were included in our study. The median follow-up time was 61 months (IQR 56 to 76), and the median functional MSTS-93 was 83% (IQR 73 to 91) postoperatively.Aims
Methods
Aseptic loosening is a major cause of failure in cemented endoprosthetic reconstructions. This paper presents the long-term outcomes of a custom-designed cross-pin fixation construct designed to minimize rotational stress and subsequent aseptic loosening in selected patients. The paper will also examine the long-term survivorship and modes of failure when using this technique. A review of 658 consecutive, prospectively collected cemented endoprosthetic reconstructions for oncological diagnoses at a single centre between 1980 and 2017 was performed. A total of 51 patients were identified with 56 endoprosthetic implants with cross-pin fixation, 21 of which were implanted following primary resection of tumour. Locations included distal femoral (n = 36), proximal femoral (n = 7), intercalary (n = 6), proximal humeral (n = 3), proximal tibial (n = 3), and distal humeral (n = 1).Aims
Patients and Methods
Massive endoprostheses rely on extra-cortical bone bridging (ECBB)
to enhance fixation. The aim of this study was to investigate the
role of selective laser sintered (SLS) porous collars in augmenting
the osseointegration of these prostheses. The two novel designs of porous SLS collars, one with small pores
(Ø700 μm, SP) and one with large pores (Ø1500 μm, LP), were compared
in an ovine tibial diaphyseal model. Osseointegration of these collars
was compared with that of a clinically used solid, grooved design
(G). At six months post-operatively, the ovine tibias were retrieved and
underwent radiological and histological analysis.Aims
Materials and Methods
We have reviewed five adult patients treated with endoprosthetic reconstruction of the proximal radius following resection of non-traumatic lesions. The patients had a mean age of 33.4 years (20 to 60) at the time of surgery and the mean follow-up was 7.6 years (0.8 to 16). Following surgery, all elbows were clinically stable and there was 100% survivorship of the prosthesis. Evaluation of function was assessed clinically and by the Mayo Elbow Performance Score, achieving a mean of 86% (70 to 100). Results at medium-term follow-up are encouraging with regards to elbow stability, implant survivorship and functional outcome.
We used a knee-sparing distal femoral endoprosthesis in young patients with malignant bone tumours of the distal femur in whom it was possible to resect the tumour and to preserve the distal femoral condyles. The proximal shaft of the endoprosthesis had a coated hydroxyapatite collar, while the distal end had hydroxyapatite-coated extracortical plates to secure it to the small residual femoral condylar fragment. We reviewed the preliminary results of this endoprosthesis in eight patients with primary bone tumours of the distal femur. Their mean age at surgery was 17.years (14 to 21). The mean follow-up was 24 months (20 to 31). At final follow-up the mean flexion at the knee was 102° (20° to 120°) and the mean Musculoskeletal Tumour Society score was 80% (57% to 96.7%). There was excellent osteointegration at the prosthesis-proximal bone interface with formation of new bone around the hydroxyapatite collar. The prosthesis allowed preservation of the knee and achieved a good functional result. Formation of new bone and remodelling at the interface make the implant more secure. Further follow-up is required to determine the long-term structural integrity of the prosthesis.