We evaluated the effects of a serum protein coating on
High-intensity narrow-spectrum (HINS) light is
a novel violet-blue light inactivation technology which kills bacteria through
a photodynamic process, and has been shown to have bactericidal
activity against a wide range of species. Specimens from patients
with infected hip and knee arthroplasties were collected over a
one-year period (1 May 2009 to 30 April 2010). A range of these
microbial isolates were tested for sensitivity to HINS-light. During
testing, suspensions of the pathogens were exposed to increasing
doses of HINS-light (of 123mW/cm2 irradiance). Non-light exposed
control samples were also used. The samples were then plated onto
agar plates and incubated at 37°C for 24 hours before enumeration.
Complete inactivation (greater than 4-log10 reduction)
was achieved for all of the isolates. The typical inactivation curve
showed a slow initial reaction followed by a rapid period of inactivation.
The doses of HINS-light required ranged between 118 and 2214 J/cm2.
Gram-positive bacteria were generally found to be more susceptible
than Gram-negative. As HINS-light uses visible wavelengths, it can be safely used
in the presence of patients and staff. This unique feature could
lead to its possible use in the prevention of infection during surgery
and post-operative dressing changes. Cite this article: