Advertisement for orthosearch.org.uk
Results 1 - 20 of 25
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 12 | Pages 1685 - 1689
1 Dec 2010
Gokaraju K Miles J Parratt MTR Blunn GW Pollock RC Skinner JA Cannon SR Briggs TWR

We have reviewed five adult patients treated with endoprosthetic reconstruction of the proximal radius following resection of non-traumatic lesions. The patients had a mean age of 33.4 years (20 to 60) at the time of surgery and the mean follow-up was 7.6 years (0.8 to 16).

Following surgery, all elbows were clinically stable and there was 100% survivorship of the prosthesis. Evaluation of function was assessed clinically and by the Mayo Elbow Performance Score, achieving a mean of 86% (70 to 100).

Results at medium-term follow-up are encouraging with regards to elbow stability, implant survivorship and functional outcome.


The Bone & Joint Journal
Vol. 106-B, Issue 9 | Pages 1000 - 1007
1 Sep 2024
Gong T Lu M Sheng H Li Z Zhou Y Luo Y Min L Tu C

Aims

Endoprosthetic reconstruction following distal femur tumour resection has been widely advocated. In this paper, we present the design of an uncemented endoprosthesis system featuring a short, curved stem, with the goal of enhancing long-term survivorship and functional outcomes.

Methods

This study involved patients who underwent implantation of an uncemented distal femoral endoprosthesis with a short and curved stem between 2014 and 2019. Functional outcomes were assessed using the 1993 version of the Musculoskeletal Tumour Society (MSTS-93) score. Additionally, we quantified five types of complications and assessed osseointegration radiologically. The survivorship of the endoprosthesis was evaluated according to two endpoints. A total of 134 patients with a median age of 26 years (IQR 16 to 41) were included in our study. The median follow-up time was 61 months (IQR 56 to 76), and the median functional MSTS-93 was 83% (IQR 73 to 91) postoperatively.


The Bone & Joint Journal
Vol. 104-B, Issue 6 | Pages 747 - 757
1 Jun 2022
Liang H Yang Y Guo W Yan L Tang X Li D Qu H Zang J Du Z

Aims

The aim of this study was to investigate the feasibility of application of a 3D-printed megaprosthesis with hemiarthroplasty design for defects of the distal humerus or proximal ulna following tumour resection.

Methods

From June 2018 to January 2020, 13 patients with aggressive or malignant tumours involving the distal humerus (n = 8) or proximal ulna (n = 5) were treated by en bloc resection and reconstruction with a 3D-printed megaprosthesis with hemiarthroplasty, designed in our centre. In this paper, we summarize the baseline and operative data, oncological outcome, complication profiles, and functional status of these patients.


The Bone & Joint Journal
Vol. 103-B, Issue 8 | Pages 1405 - 1413
1 Aug 2021
Ogura K Fujiwara T Morris CD Boland PJ Healey JH

Aims

Rotating-hinge knee prostheses are commonly used to reconstruct the distal femur after resection of a tumour, despite the projected long-term burden of reoperation due to complications. Few studies have examined the factors that influence their failure and none, to our knowledge, have used competing risk models to do so. The purpose of this study was to determine the risk factors for failure of a rotating-hinge knee distal femoral arthroplasty using the Fine-Gray competing risk model.

Methods

We retrospectively reviewed 209 consecutive patients who, between 1991 and 2016, had undergone resection of the distal femur for tumour and reconstruction using a rotating-hinge knee prosthesis. The study endpoint was failure of the prosthesis, defined as removal of the femoral component, the tibial component, or the bone-implant fixation; major revision (exchange of the femoral component, tibial component, or the bone-implant fixation); or amputation.


The Bone & Joint Journal
Vol. 103-B, Issue 2 | Pages 382 - 390
1 Feb 2021
Wang H Tang X Ji T Yan T Yang R Guo W

Aims

There is an increased risk of dislocation of the hip after the resection of a periacetabular tumour and endoprosthetic reconstruction of the defect in the hemipelvis. The aim of this study was to determine the rate and timing of dislocation and to identify its risk factors.

Methods

To determine the dislocation rate, we conducted a retrospective single-institution study of 441 patients with a periacetabular tumour who had undergone a standard modular hemipelvic endoprosthetic reconstruction between 2003 and 2019. After excluding ineligible patients, 420 patients were enrolled. Patient-specific, resection-specific, and reconstruction-specific variables were studied using univariate and multivariate analyses.


The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 5 | Pages 714 - 718
1 Jul 2004
Bickels J Kollender Y Merimsky O Isaakov J Petyan-Brand R Meller I

We report our experience with a new technique for cryosurgical ablation of bone tumours which allows accurate determination of the temperature and freezing time within a cavity of any geometrical shape. Between 1997 and 2000, 58 patients diagnosed with 13 malignant and 45 aggressive benign bone tumours underwent argon-based cryoablation. This technique includes removal of the tumour by curettage and filling the cavity with a gel medium into which metal probes are inserted. Argon gas is delivered through the metal probes and both time and temperature are computer-controlled. After formal reconstruction, all patients were followed for more than two years. None had skin necrosis, infection, neurapraxia or thromboembolic complication. Fractures occurred in two patients (3.4%) and the tumour recurred in two patients (3.4%)


The Bone & Joint Journal
Vol. 102-B, Issue 6 | Pages 779 - 787
1 Jun 2020
Gupta S Griffin AM Gundle K Kafchinski L Zarnett O Ferguson PC Wunder J

Aims

Iliac wing (Type I) and iliosacral (Type I/IV) pelvic resections for a primary bone tumour create a large segmental defect in the pelvic ring. The management of this defect is controversial as the surgeon may choose to reconstruct it or not. When no reconstruction is undertaken, the residual ilium collapses back onto the remaining sacrum forming an iliosacral pseudarthrosis. The aim of this study was to evaluate the long-term oncological outcome, complications, and functional outcome after pelvic resection without reconstruction.

Methods

Between 1989 and 2015, 32 patients underwent a Type I or Type I/IV pelvic resection without reconstruction for a primary bone tumour. There were 21 men and 11 women with a mean age of 35 years (15 to 85). The most common diagnosis was chondrosarcoma (50%, n = 16). Local recurrence-free, metastasis-free, and overall survival were assessed using the Kaplan-Meier method. Patient function was evaluated using the Musculoskeletal Tumour Society (MSTS) and Toronto Extremity Salvage Score (TESS).


The Bone & Joint Journal
Vol. 101-B, Issue 12 | Pages 1557 - 1562
1 Dec 2019
Tillman R Tsuda Y Puthiya Veettil M Young PS Sree D Fujiwara T Abudu A

Aims

The aim of this study was to present the long-term surgical outcomes, complications, implant survival, and causes of implant failure in patients treated with the modified Harrington procedure using antegrade large diameter pins.

Patients and Methods

A cohort of 50 consecutive patients who underwent the modified Harrington procedure for periacetabular metastasis or haematological malignancy between January 1996 and April 2018 were studied. The median follow-up time for all survivors was 3.2 years (interquartile range 0.9 to 7.6 years).


The Bone & Joint Journal
Vol. 100-B, Issue 12 | Pages 1640 - 1646
1 Dec 2018
Medellin MR Fujiwara T Clark R Jeys LM

Aims

The aim of this study was to describe, analyze, and compare the survival, functional outcome, and complications of minimally invasive (MI) and non-invasive (NI) lengthening total femoral prostheses.

Patients and Methods

A total of 24 lengthening total femoral prostheses, 11 MI and 13 NI, were implanted between 1991 and 2016. The characteristics, complications, and functional results were recorded. There were ten female patients and ten male patients. Their mean age at the time of surgery was 11 years (2 to 41). The mean follow-up was 13.2 years (seven months to 29.25 years). A survival analysis was performed, and the failures were classified according to the Modified Henderson System.


The Bone & Joint Journal
Vol. 99-B, Issue 2 | Pages 276 - 282
1 Feb 2017
Mumith A Coathup M Chimutengwende-Gordon M Aston W Briggs T Blunn G

Aims

Massive endoprostheses rely on extra-cortical bone bridging (ECBB) to enhance fixation. The aim of this study was to investigate the role of selective laser sintered (SLS) porous collars in augmenting the osseointegration of these prostheses.

Materials and Methods

The two novel designs of porous SLS collars, one with small pores (Ø700 μm, SP) and one with large pores (Ø1500 μm, LP), were compared in an ovine tibial diaphyseal model. Osseointegration of these collars was compared with that of a clinically used solid, grooved design (G). At six months post-operatively, the ovine tibias were retrieved and underwent radiological and histological analysis.


The Bone & Joint Journal
Vol. 99-B, Issue 3 | Pages 401 - 408
1 Mar 2017
Kang S Lee JS Park J Park S

Aims

Children treated for osteosarcoma around the knee often have a substantial leg-length discrepancy at skeletal maturity. The aim of this study was to investigate the results of staged skeletal reconstruction after a leg lengthening procedure using an external fixator in these patients.

Patients and Methods

We reviewed 11 patients who underwent staged reconstruction with either an arthroplasty (n = 6) or an arthrodesis (n = 5). A control group of 11 patients who had undergone wide excision and concurrent reconstruction with an arthroplasty were matched for gender, location, and size of tumour. We investigated the change in leg-length discrepancy, function as assessed by the Musculoskeletal Tumor Society Scale (MSTS) score and complications.


The Bone & Joint Journal
Vol. 95-B, Issue 10 | Pages 1410 - 1416
1 Oct 2013
Gebert C Wessling M Gosheger G Aach M Streitbürger A Henrichs MP Dirksen U Hardes J

To date, all surgical techniques used for reconstruction of the pelvic ring following supra-acetabular tumour resection produce high complication rates. We evaluated the clinical, oncological and functional outcomes of a cohort of 35 patients (15 men and 20 women), including 21 Ewing’s sarcomas, six chondrosarcomas, three sarcomas not otherwise specified, one osteosarcoma, two osseous malignant fibrous histiocytomas, one synovial cell sarcoma and one metastasis. The mean age of the patients was 31 years (8 to 79) and the latest follow-up was carried out at a mean of 46 months (1.9 to 139.5) post-operatively.

We undertook a functional reconstruction of the pelvic ring using polyaxial screws and titanium rods. In 31 patients (89%) the construct was encased in antibiotic-impregnated polymethylmethacrylate. Preservation of the extremities was possible for all patients. The survival rate at three years was 93.9% (95% confidence interval (CI) 77.9 to 98.4), at five years it was 82.4% (95% CI 57.6 to 93.4). For the 21 patients with Ewing’s sarcoma it was 95.2% (95% CI 70.7 to 99.3) and 81.5% (95% CI 52.0 to 93.8), respectively. Wound healing problems were observed in eight patients, deep infection in five and clinically asymptomatic breakage of the screws in six. The five-year implant survival was 93.3% (95% CI 57.8 to 95.7). Patients were mobilised at a mean of 3.5 weeks (1 to 7) post-operatively. A post-operative neurological defect occurred in 12 patients. The mean Musculoskeletal Tumor Society score at last available follow-up was 21.2 (10 to 27).

This reconstruction technique is characterised by simple and oncologically appropriate applicability, achieving high primary stability that allows early mobilisation, good functional results and relatively low complication rates.

Cite this article: Bone Joint J 2013;95-B:1410–16.


The Bone & Joint Journal
Vol. 96-B, Issue 5 | Pages 677 - 683
1 May 2014
Greenberg A Berenstein Weyel T Sosna J Applbaum J Peyser A

Osteoid osteoma is treated primarily by radiofrequency (RF) ablation. However, there is little information about the distribution of heat in bone during the procedure and its safety. We constructed a model of osteoid osteoma to assess the distribution of heat in bone and to define the margins of safety for ablation. Cavities were drilled in cadaver bovine bones and filled with a liver homogenate to simulate the tumour matrix. Temperature-sensing probes were placed in the bone in a radial fashion away from the cavities. RF ablation was performed 107 times in tumours < 10 mm in diameter (72 of which were in cortical bone, 35 in cancellous bone), and 41 times in cortical bone with models > 10 mm in diameter. Significantly higher temperatures were found in cancellous bone than in cortical bone (p <  0.05). For lesions up to 10 mm in diameter, in both bone types, the temperature varied directly with the size of the tumour (p < 0.05), and inversely with the distance from it. Tumours of > 10 mm in diameter showed a trend similar to those of smaller lesions. No temperature rise was seen beyond 12 mm from the edge of a cortical tumour of any size. Formulae were developed to predict the expected temperature in the bone during ablation.

Cite this article: Bone Joint J 2014; 96-B:677–83


The Bone & Joint Journal
Vol. 95-B, Issue 7 | Pages 988 - 992
1 Jul 2013
Glehr M Leithner A Friesenbichler J Goessler W Avian A Andreou D Maurer-Ertl W Windhager R Tunn P

The aims of this study were to evaluate the incidence of local argyria in patients with silver-coated megaprostheses and to identify a possible association between argyria and elevated levels of silver both locally and in the blood. Between 2004 and 2011, 32 megaprostheses with silver coatings were implanted in 20 female and 12 male patients following revision arthroplasty for infection or resection of a malignant tumour, and the levels of silver locally in drains and seromas and in the blood were determined. The mean age of the patients was 46 years (10 to 81); one patient died in the immediate post-operative period and was excluded.

Seven patients (23%) developed local argyria after a median of 25.7 months (interquartile range 2 to 44.5). Patients with and without local argyria had comparable levels of silver in the blood and aspiration fluids. The length of the implant did not influence the development of local argyria. Patients with clinical evidence of local argyria had no neurological symptoms and no evidence of renal or hepatic failure. Thus, we conclude that the short-term surveillance of blood silver levels in these patients is not required.

Cite this article: Bone Joint J 2013;95-B:988–92.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 7 | Pages 999 - 1005
1 Jul 2010
Akiyama T Clark JCM Miki Y Choong PFM

Internal hemipelvectomy is a standard treatment for malignant tumours of the pelvis. Reconstruction using a non-vascularised fibular graft is relatively straightforward compared to other techniques. We describe the surgical and functional outcomes for a series of ten patients who underwent an internal hemipelvectomy (type I or I/IV) with reconstruction by a non-vascularised fibular graft between 1996 and 2009. A key prerequisite for this procedure was a preserved sciatic notch, confirmed pre-operatively on MRI.

Graft-host union was achieved in all patients with a single fibular graft, and in the lower graft where two grafts had been used. The mean time to union was 7.3 months (3 to 12). The upper graft did not unite in four of six cases where two grafts had been used. Seven patients were eventually able to walk without a stick. The mean post-operative Musculoskeletal Tumour Society score was 75.4% (16.7 to 96.7). There were no cases of deep post-operative infection. The mean pelvic shortening was 0.9 cm (0.2 to 3.4). Recurrent tumour occurred in three cases, and death from tumour-related disease occured in one.

Patients who need an internal hemipelvectomy will do well if their pelvic ring is reconstructed with a non-vascularised fibular graft. The complication rate is low, and they attain a good functional outcome.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 8 | Pages 1084 - 1089
1 Aug 2008
Guo W Ji T Yang R Tang X Yang Y

In developing countries locally-made low-cost prostheses are mainly used in limb-salvage surgery to alleviate the economic burden.

We retrospectively collected data on 104 patients treated by limb-salvage surgery between July 1997 and July 2005. We used a locally-designed and fabricated stainless-steel endoprosthesis in each case. Oncological and functional outcomes were evaluated at a mean follow-up of 47 months (12 to 118).

A total of 73 patients (70.2%) were free from disease, nine (8.7%) were alive with disease, 19 (18.2%) had died from their disease and three (2.9%) from unrelated causes. According to the Musculoskeletal Tumor Society scoring system, the mean functional score was 76.3% (SD 17.8). The five-year survival for the implant was 70.5%. There were nine cases (8.7%) of infection, seven early and two late, seven (6.7%) of breakage of the prosthesis, three (2.9%) of aseptic loosening and two (1.9%) of failure of the polyethylene bushing. Multivariate analysis showed that a proximal tibial prosthesis and a resection length of 14 cm or more were significant negative prognostic factors.

Our survival rates and Musculoskeletal Tumor Society functional scores are similar to those reported in the literature. Although longer follow-up is needed to confirm our results, we believe that a low-cost custom-made endoprosthesis is a cost-effective and reliable reconstructive option for limb salvage in developing countries.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 2 | Pages 265 - 269
1 Feb 2012
Hwang N Grimer RJ Carter SR Tillman RM Abudu A Jeys LM

We reviewed our initial seven-year experience with a non-invasive extendible prosthesis in 34 children with primary bone tumours. The distal femur was replaced in 25 cases, total femur in five, proximal femur in one and proximal tibia in three. The mean follow-up was 44 months (15 to 86) and 27 patients (79%) remain alive. The prostheses were lengthened by an electromagnetic induction mechanism in an outpatient setting and a mean extension of 32 mm (4 to 80) was achieved without anaesthesia. There were lengthening complications in two children: failed lengthening in one and the formation of scar tissue in the other. Deep infection developed in six patients (18%) and local recurrence in three. A total of 11 patients required further surgery to the leg. Amputation was necessary in five patients (20%) and a two-stage revision in another. There were no cases of loosening, but two patients had implant breakage and required revision. The mean Musculoskeletal Tumor Society functional score was 85% (60% to 100%) at last known follow-up. These early results demonstrate that the non-invasive extendible prosthesis allows successful lengthening without surgical intervention, but the high incidence of infection is a cause for concern.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 3 | Pages 401 - 404
1 Mar 2009
Chandrasekar CR Grimer RJ Carter SR Tillman RM Abudu A Jeys LM

We undertook a cemental unipolar proximal femoral endoprosthetic replacement in 131 patients with a mean age of 50 years (2 to 84). Primary malignant tumours were present in 54 patients and 67 had metastatic disease. In addition, eight patients had either lymphoma or myeloma and two had non-oncological disorders. The mean follow-up was 27 months (0 to 180). An acetabular revision was required later in 14 patients, 12 of whom had been under the age of 21 years at the time of insertion of their original prosthesis. The risk of acetabular revision in patients over 21 years of age was 8% at five years compared with 36% in those aged under 21 years. All the unipolar hips in this younger age group required revision within 11 years of the initial operation.

We conclude that unipolar replacement should not be used in younger patients and should be avoided in patients with a life expectancy of more than five years.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 3 | Pages 371 - 376
1 Mar 2008
Falkinstein Y Ahlmann ER Menendez LR

We report our early experience with a new peri-acetabular reconstruction endoprosthesis used for pelvic reconstruction after tumour resection. The outcome of 21 patients who underwent limb salvage following type II pelvic resection and reconstruction using the peri-acetabular reconstruction prosthesis between 2000 and 2006 was retrospectively reviewed. This prosthesis was designed to use the remaining part of the ilium to support a horizontally placed acetabular component secured with internal fixation and bone cement. Into this device a constrained acetabular liner is positioned which is articulates with a conventional femoral component to which a modular extension and modular head are attached. The mean follow-up was 20.5 months (1 to 77). The most common complications were deep infection, superficial wound infections, and dislocation. The mean musculoskeletal tumor society functional outcome score for the survivors was 20.1(11 to 27).

We recommend the use of the peri-acetabular reconstruction prosthesis for reconstruction of large defects after type II pelvic resection, as this design has a greater inherent stability over other available prostheses.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 9 | Pages 1222 - 1227
1 Sep 2008
Jaiswal PK Aston WJS Grimer RJ Abudu A Carter S Blunn G Briggs TWR Cannon S

We treated 98 patients with peri-acetabular tumours by resection and reconstruction with a custom-made pelvic endoprosthesis. The overall survival of the patients was 67% at five years, 54% at ten years and 51% at 30 years. One or more complications occurred in 58.1% of patients (54), of which infection was the most common, affecting 30% (28 patients). The rate of local recurrence was 31% (29 patients) after a mean follow-up of 71 months (11 to 147). Dislocation occurred in 20% of patients (19). Before 1996 the rate was 40.5% (17 patients) but this was reduced to 3.9% (two patients) with the introduction of a larger femoral head. There were six cases of palsy of the femoral nerve with recovery in only two. Revision or excision arthroplasty was performed in 23.7% of patients (22), principally for uncontrolled infection or aseptic loosening. Higher rates of death, infection and revision occurred in men.

This method of treatment is still associated with high morbidity. Patients should be carefully selected and informed of this pre-operatively.