The aim of this study was to report the incidence of implant-related complications, further operations, and their influence on the outcome in a series of patients who underwent primary reverse total shoulder arthroplasty (RTSA). The prospectively collected clinical and radiological data of 797 patients who underwent 854 primary RTSAs between January 2005 and August 2018 were analyzed. The hypothesis was that the presence of complications would adversely affect the outcome. Further procedures were defined as all necessary operations, including reoperations without change of components, and partial or total revisions. The clinical outcome was evaluated using the absolute and relative Constant Scores (aCS, rCS), the Subjective Shoulder Value (SSV) scores, range of motion, and pain.Aims
Methods
The reasons for failure of a hemirthroplasty (HA) when used to
treat a proximal humeral fracture include displaced or necrotic
tuberosities, insufficient metaphyseal bone-stock, and rotator cuff
tears. Reverse total shoulder arthroplasty (rTSA) is often the only
remaining form of treatment in these patients. The aim of this study
was to evaluate the clinical outcome after conversions from a failed
HA to rTSA. A total of 35 patients, in whom a HA, as treatment for a fracture
of the proximal humerus, had failed, underwent conversion to a rTSA.
A total of 28 were available for follow-up at a mean of 61 months
(37 to 91), having been initially reviewed at a mean of 20 months
(12 to 36) postoperatively. Having a convertible design, the humeral
stem could be preserved in nine patients. The stem was removed in
the other 19 patients and a conventional rTSA was implanted. At
final follow-up, patients were assessed using the American Shoulder
and Elbow Surgeons (ASES) score, the Constant Score, and plain radiographs.Aims
Material and Methods
The LockDown device (previously called Surgilig)
is a braided polyester mesh which is mostly used to reconstruct the
dislocated acromioclavicular joint. More than 11 000 have been implanted
worldwide. Little is known about the tissue reaction to the device
nor to its wear products when implanted in an extra-articular site
in humans. This is of importance as an adverse immunological reaction
could result in osteolysis or damage to the local tissues, thereby affecting
the longevity of the implant. We analysed the histology of five LockDown implants retrieved
from five patients over the last seven years by one of the senior
authors. Routine analysis was carried out in all five cases and
immunohistochemistry in one. The LockDown device acts as a scaffold for connective tissue
which forms an investing fibrous pseudoligament. The immunological
response at the histological level seems favourable with a limited
histiocytic and giant cell response to micron-sized wear particles.
The connective tissue envelope around the implant is less organised
than a native ligament. Cite this article: